 Class 2 Device Recall Dimension QuikLYTE Dilution Checksee related information
Date Initiated by FirmFebruary 24, 2015
Date PostedApril 07, 2015
Recall Status1 Terminated 3 on February 01, 2017
Recall NumberZ-1398-2015
Recall Event ID 70627
510(K)NumberK970330 
Product Classification Electrode, ion specific, potassium - Product Code CEM
Product QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry systems with the QuikLYTE module. Intended Use of the QuikLYTE module : The Na+, K+, Cl and TCO2 methods on the Dimension clinical chemistry system with the QuikLYTE module are in vitro diagnostic tests intended for the quantitative measurement of sodium, potassium, chloride and total carbon dioxide in human serum and heparinized plasma. They are also intended for the quantitative measurement of Na+, K+ and Cl- in urine.
Code Information Catalog Number S640, Siemens Material Number (SMN) 10444872, Lot Number 4MD707 
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics, Inc.
500 Gbc Dr Ms 514
PO BOX 6101
Newark DE 19702-2466
For Additional Information ContactCustomer Support
800-441-9250
Manufacturer Reason
for Recall
A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.
FDA Determined
Cause 2
Under Investigation by firm
ActionAn Urgent Medical Device Recall letter, dated February 2015, was provided to all Dimension system customers who received the affected lot to notify them that Siemens confirmed customer complaints of failing dilution checks when using QuikLYTE Dilution Check lot 4MD707. The letter also instructed them to immediately discontinue use and discard all inventory of Dimension QuikLYTE Dilution Check lot 4MD707. Siemens offered a no charge replacement with a non-impacted lot.
Quantity in Commerce5981
DistributionWorldwide Distribution.
Total Product Life CycleTPLC Device Report