|
Class 2 Device Recall EPIQ 5 Ultrasound System iwth Pediatric Cardiology option |
|
Date Initiated by Firm |
April 10, 2015 |
Date Posted |
May 05, 2015 |
Recall Status1 |
Terminated 3 on February 29, 2016 |
Recall Number |
Z-1578-2015 |
Recall Event ID |
71084 |
510(K)Number |
K132304
|
Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
Product |
EPIQ 5 Ultrasound System with Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W. Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3) 453561800581 (1.3.1) 453561805181 (1.3.2)
|
Code Information |
*********************SYSTEM CODES IN THE US*********************************** US114C0108, US114C0206, US114C0210, US114C0239, US114C0240, US114C0241, US114C0271, US114C0272, US114C0273, US114C0286, US114C0321, US114C0322, US114C0331, US114C0332, US114C0333, US114C0334, US114C0444, US114C0534, US114C0605, US114C0674, US115C0140, US115C0382, US115C0468, US214C0106, US214C0143, US214C0144, US214C0192, US214C0534, US214C0535, US214C0536, US215C0071, US215C0394, US215C0536, US215C0774, US215C0839, US215C0843, US215C0844, US215C1010, US215C1049, US215C1050, US314C0055, US314C0073, US314C0235, US314C0296, US314C0297, US314C0322, US314C0444, US314C0446, US314C0557, US314C0613, US314C0746, US314C0835, US315C0055, US414C0014, US414C0035, US414C0069, US414C0158, US414C0376, US414C0377, US414C0616, US514C0044, US514C0045, US514C0046, US514C0084, US514C0085, US514C0149, US514C0175, US514C0338, US514C0373, US514C0374, US514C0375, US514C0473, US514C0474, US514C0639, US514C0657, US514C0668, US514C0669, US514C0670, US514C0671, US514C0684, US514C0741, US514C0742, US614C0154, US614C0155, US614C0156, US614C0157, US614C0288, US614C0289, US614C0294, US614C0333, US614C0419, US614C0463, US614C0464, US614C0466, US614C0467, US614C0468, US614C0487, US614C0488, US614C0489, US614C0490, US614C0491, US614C0493, US614C0494, US614C0495, US614C0496, US614C0500, US614C0501, US614C0502, US614C0503, US614C0506, US614C0624, US614C0628, US614C0700, US714C0097, US714C0164, US714C0370, US714C0508, US814C0025, US814C0137, US814C0155, US814C0157, US914C0076, US914C0190, US914C0245, US914C0246, US914C0263, US914C0264, US914C0265, US914C0266, US914C0267, US914C0268, US914C0793, US914C0998, USD13B0662, USD13C0018, USD13C0070, USD13C0071, USD13C0072, USD13C0073, USD13C0234, USD13C0235, USD13C0237, USD13C0255, USD13C0256, USD13C0258, USD13C0291, USD13C0365, USD13C0439, USD13C0440, USD13C0441, USD13C0442, USD13C0443, USD13C0522, USD13C0523, USD13C0524, USD13C0525, USD13C0526, USD13C0527, USD13C0569, USD13C0570, USD13C0571, USD13C0583, USD13C0584, USD13C0585, USD13C0614, USD13C0615, USD13C0616, USD13C0617, USD13C0618, USD13C0619, USD13C0620, USD13C0621, USD13C0642, USD13C0668, USD13C0670, USD13C0671, USD13C0672, USD13C0678, USD13C0698, USD13C0707, USD13C0708, USD13C0710, USD13C0711, USD13C0712, USD13C0718, USD13C0719, USD13C0720, USD14C0167, USD14C0168, USD14C0370, USD14C0371, USD14C0372, USD14C0373, USD14C0374, USD14C0597, USD14C0603, USD14C0604, USD14C0605, USD14C0758, USD14C0759, USD14C0760, USD14C0886, USD14C0887, USD14C0888, USD14C0889, USD14C1024, USD14C1025, USN13B0039, USN13C0054, USN13C0182, USN13C0183, USN13C0184, USN13C0185, USN13C0281, USN13C0282, USN13C0283, USN13C0284, USN13C0285, USN13C0286, USN13C0362, USN13C0363, USN13C0379, USN13C0380, USN13C0504, USN13C0574, USN13C0575, USN13C0626, USN14C0038, USN14C0189, USN14C0298, USN14C0402, USN14C0780, USN14C1044, USO13C0350, USO13C0352, USO13C0353, USO13C0354, USO14C0056, USO14C0126, USO14C0127, USO14C0156, USO14C0157, USO14C0158, USO14C0159, USO14C0160, and USO14C0188. *****************************SYSTEM CODES IN CANADA******************** US115C0274, US115C0466, US115C0467, US115C0494, US214C0486, US215C0034, US215C0036, US215C0347, US414C0111, US414C0112, US414C0113, US414C0114, US414C0477, US414C0478, US514C0339, US614C0688, US614C0854, US914C0179, US914C0180, USN14C0237, and USO14C1093. ****************************SYSTEM CODES IN OTHER COUNTRIES************* US114C0077, US114C0078, US114C0079, US114C0080, US114C0105, US114C0106, US114C0109, US114C0190, US114C0191, US114C0284, US114C0285, US114C0361, US114C0362, US114C0365, US114C0378, US114C0405, US114C0406, US114C0407, US114C0443, US114C0505, US114C0531, US114C0533, US114C0672, US114C0673, US114C0675, US114C0730, US114C0751, US115C0018, US115C0055, US115C0097, US115C0098, US115C0099, US115C0215, US115C0275, US115C0383, US115C0449, US115C0450, US115C0463, US115C0564, US115C0566, US115C0573, US115C0574, US115C0616, US115C0647, US115C0648, US115C0649, US115C0650, US214C0017, US214C0021, US214C0022, US214C0044, US214C0045, US214C0085, US214C0086, US214C0087, US214C0140, US214C0177, US214C0178, US214C0179, US214C0250, US214C0416, US214C0417, US214C0418, US214C0419, US214C0420, US214C0421, US214C0422, US214C0423, US214C0424, US214C0425, US214C0477, US214C0478, US214C0479, US214C0481, US214C0482, US214C0484, US214C0485, US215C0031, US215C0130, US215C0131, US215C0174, US215C0175, US215C0177, US215C0344, US215C0345, US215C0382, US215C0383, US215C0384, US215C0429, US215C0430, US215C0431, US215C0432, US215C0433, US215C0434, US215C0468, US215C0511, US215C0529, US215C0530, US215C0531, US215C0569, US215C0570, US215C0574, US215C0575, US215C0576, US215C0577, US215C0578, US215C0621, US215C0622, US215C0623, US215C0764, US215C0765, US215C0766, US215C0840, US215C0841, US215C0842, US215C0912, US215C0913, US215C0914, US215C0917, US215C0918, US215C0922, US215C0923, US215C0924, US215C0925, US215C0950, US215C1044, US215C1045, US314C0054, US314C0067, US314C0068, US314C0069, US314C0070, US314C0071, US314C0072, US314C0100, US314C0101, US314C0102, US314C0103, US314C0104, US314C0167, US314C0199, US314C0200, US314C0201, US314C0230, US314C0231, US314C0232, US314C0233, US314C0234, US314C0298, US314C0299, US314C0300, US314C0321, US314C0728, US314C0729, US314C0730, US314C0742, US314C0743, US314C0744, US314C0745, US314C0775, US314C0776, US314C0777, US314C0778, US314C0779, US314C0798, US314C0799, US314C0800, US314C0801, US314C0834, US315C0049, US315C0054, US414C0012, US414C0013, US414C0037, US414C0038, US414C0066, US414C0067, US414C0068, US414C0074, US414C0075, US414C0076, US414C0077, US414C0078, US414C0109, US414C0110, US414C0142, US414C0157, US414C0159, US414C0231, US414C0307, US414C0384, US414C0395, US414C0519, US414C0614, US414C0647, US513C0412, US513C0539, US514C0024, US514C0081, US514C0108, US514C0115, US514C0116, US514C0272, US514C0273, US514C0274, US514C0275, US514C0299, US514C0472, US514C0535, US514C0536, US514C0537, US514C0538, US514C0539, US514C0540, US514C0541, US514C0542, US514C0543, US514C0544, US514C0545, US514C0546, US514C0560, US514C0636, US514C0637, US514C0638, US514C0718, US514C0732, US614C0024, US614C0189, US614C0192, US614C0383, US614C0418, US614C0441, US614C0497, US614C0498, US614C0745, US714C0071, US714C0077, US714C0078, US714C0079, US714C0080, US714C0081, US714C0095, US714C0096, US714C0121, US714C0135, US714C0167, US714C0170, US714C0199, US714C0254, US714C0255, US714C0291, US714C0304, US714C0439, US714C0461, US714C0464, US714C0505, US714C0506, US714C0548, US714C0736, US814C0023, US814C0024, US814C0244, US814C0311, US814C0423, US814C0424, US814C0425, US814C0426, US814C0427, US814C0428, US814C0429, US814C0430, US814C0458, US814C0461, US814C0462, US814C0468, US814C0469, US814C0508, US814C0510, US814C0626, US814C0627, US914C0017, US914C0018, US914C0019, US914C0053, US914C0261, US914C0376, US914C0377, US914C0378, US914C0483, US914C0485, US914C0636, US914C0701, US914C0737, US914C0742, US914C0873, US914C0874, US914C0957, US914C0958, USD13C0031, USD13C0033, USD13C0034, USD13C0035, USD13C0036, USD13C0064, USD13C0065, USD13C0066, USD13C0067, USD13C0068, USD13C0140, USD13C0141, USD13C0142, USD13C0195, USD13C0230, USD13C0231, USD13C0233, USD13C0254, USD13C0257, USD13C0357, USD13C0358, USD13C0359, USD13C0360, USD13C0361, USD13C0362, USD13C0363, USD13C0371, USD13C0372, USD13C0373, USD13C0374, USD13C0375, USD13C0376, USD13C0377, USD13C0478, USD13C0479, USD13C0480, USD13C0481, USD13C0729, USD14C0108, USD14C0109, USD14C0110, USD14C0111, USD14C0114, USD14C0117, USD14C0277, USD14C0278, USD14C0279, USD14C0281, USD14C0284, USD14C0285, USD14C0288, USD14C0289, USD14C0368, USD14C0369, USD14C0400, USD14C0527, USD14C0661, USD14C0662, USD14C0663, USD14C0664, USD14C0669, USD14C0670, USD14C0761, USN13C0014, USN13C0015, USN13C0016, USN13C0017, USN13C0018, USN13C0040, USN13C0041, USN13C0042, USN13C0043, USN13C0044, USN13C0045, USN13C0046, USN13C0047, USN13C0048, USN13C0049, USN13C0050, USN13C0051, USN13C0052, USN13C0053, USN13C0075, USN13C0076, USN13C0077, USN13C0078, USN13C0151, USN13C0152, USN13C0205, USN13C0206, USN13C0277, USN13C0278, USN13C0279, USN13C0280, USN13C0360, USN13C0361, USN13C0393, USN13C0456, USN13C0457, USN13C0460, USN13C0469, USN13C0499, USN13C0501, USN13C0502, USN13C0532, USN13C0573, USN13C0579, USN13C0580, USN13C0593, USN13C0622, USN13C0623, USN13C0624, USN13C0625, USN13C0627, USN13C0652, USN13C0653, USN13C0654, USN13C0655, USN13C0656, USN13C0657, USN13C0658, USN14C0030, USN14C0101, USN14C0102, USN14C0104, USN14C0311, USN14C0365, USN14C0469, USN14C0470, USN14C0601, USN14C0655, USN14C0936, USN14C0937, USN14C0938, USN14C0940, USN14C0941, USN14C1040, USN14C1041, USN14C1042, USO13C0361, USO13C0396, USO13C0397, USO13C0399, USO13C0400, USO13C0401, USO13C0402, USO13C0403, USO13C0453, USO13C0454, USO13C0465, USO13C0470, USO13C0471, USO13C0472, USO13C0473, USO13C0553, USO13C0593, USO13C0629, USO13C0678, USO13C0680, USO13C0681, USO13C0682, USO13C0683, USO13C0685, USO14C0051, USO14C0052, USO14C0053, USO14C0096, USO14C0125, USO14C0185, USO14C0186, USO14C0187, USO14C0478, USO14C0601, USO14C0602, USO14C0603, USO14C0604, USO14C0605, USO14C0606, USO14C0721, USO14C0845, USO14C1091, USO14C1124, USO14C1125, USO14C1126, USO14C1127, and USO14C1180. |
Recalling Firm/ Manufacturer |
Philips Ultrasound, Inc. 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact |
Philips Customer Service 800-722-9377
|
Manufacturer Reason for Recall |
When Epiq 5 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA).
|
FDA Determined Cause 2 |
Software design |
Action |
Philips sent the Urgent-Medical Device Correction-Philips Model EPIQ5 and Ultrasound system letter, dated 2015 APRIL 10, and the Urgent-Medical Device Correction-Philips Model EPIQ7 and Ultrasound system letter, dated 2015 APRIL 10, letters to their consignees.
These systems WITH Pediatric Cardiology option are affected.
Philips issued two Field Change Orders (FCO) to their Field Service Engineer to install upgraded software:
-FCO 79500336 for EPIQ5
-FCO 79500337 for EPIQ7
Customers with questions can call Philips Customer Service at 1-800-722-9377. |
Quantity in Commerce |
751 units total (246 units in the US and 505 units outside the US) |
Distribution |
Worldwide Distribution - US (nationwide and DC) and Internationally to Norway, India, New Zealand, Bahrain, Sweden, Italy, Australia, France, Netherlands, Switzerland, Germany, Austria, and Turkey. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IYN and Original Applicant = PHILIPS ULTRASOUND, INC.
|
|
|
|