Class 2 Device Recall Sterile, CV0479 Open Heart Pk surgical kit

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Sterile, CV0479 Open Heart Pk surgical kitsee related information
Date Initiated by FirmApril 22, 2015
Date PostedOctober 07, 2015
Recall Status1 Terminated 3 on March 28, 2017
Recall NumberZ-0033-2016
Recall Event ID 71904
Product Classification General and plastic surgery - Product Code FSY
Productregard Item Number: 800410, Sterile, CV0479 - Open Heart Pk - surgical kit containing cover light handle soft, RX. The responsible firm on the label is ROi, LLC, St. Louis, MO. Custom procedure tray contains surgical instruments and accessories intended for use during a surgical procedure
Code Information Lot numbers 15134 and 17453.
Recalling Firm/
Manufacturer
Resource Optimization & Innovation Llc
2909 N Neergard Ave
Springfield MO 65803
For Additional Information ContactRobert W. Callahan
314-364-6561
Manufacturer Reason
for Recall
The kits contain an equipment glove component which may contain splits or holes compromising the sterility.
FDA Determined
Cause 2
Process control
ActionROi initiated a recall of the custom kits via e-mail to their customers on April 22, 2015. The e-mail included their recall letter dated April 17, 2015. and the supplier's recall letter dated April 16, 2015. The customer was requested to discard the recalled component from their procedure tray at the time of set up and replace it with sterile stock. Customers were instructed to contact any ROi custom pack solutions manager with any concerns. The recalling firm issued a second e-mail notification on 10/19/2015 to the recall coordinator of the distributor. The notice informed them they had been receiving uncorrected product and reminded them to discard the defective component at the time the kit is set up for surgery. Any future distribution will be labeled with an alert to remove the defective component until the kits contain acceptable light handle covers. For questions regarding this recall call 314-364-6561.
Quantity in Commerce199 kits
DistributionDistribution was made to LA, MS, and TX.