Date Initiated by Firm |
November 16, 2015 |
Create Date |
December 24, 2015 |
Recall Status1 |
Terminated 3 on November 09, 2016 |
Recall Number |
Z-0533-2016 |
Recall Event ID |
72651 |
Product Classification |
Enzyme immunoassay, benzodiazipine - Product Code JXM
|
Product |
ProScreen 8 Panel Dip Card w/Adult, Item No. PSDA-8P, PSDA-8MBO300
|
Code Information |
152950 152951 153325 |
Recalling Firm/ Manufacturer |
Ameditech Inc 9940 Mesa Rim Rd San Diego CA 92121-2910
|
For Additional Information Contact |
858-535-1968
|
Manufacturer Reason for Recall |
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
A Medical Device Labeling Correction letter was sent on 11/16/15 to distributors and end users to inform them that Ameditechs Drugs of Abuse Tests have shown reduced reactivity for up to two (2) of the nineteen (19) BZO compounds for which specificity information is provided in the product insert. The letter informs the customers that the particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide. Customers with questions are instructed to contact Ameditech at (844) 616-4453 or email at amt.fieldaction@alere.com. |
Distribution |
Nationwide Distribution. |