| Class 2 Device Recall VITROS 5,1 FS Chemistry System | |
Date Initiated by Firm | December 21, 2015 |
Date Posted | March 08, 2016 |
Recall Status1 |
Terminated 3 on May 30, 2018 |
Recall Number | Z-1075-2016 |
Recall Event ID |
72979 |
510(K)Number | K031924 K063144 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | VITROS 5,1 FS Chemistry System, Catalog Number 6801375, Unique Device Identifier Number 10758750001132; and VITROS 5,1 FS Chemistry System (Refurbished), Catalog Number 6801890, Unique Device Identifier Number 10758750001644; IVD.
Product Usage:
The VITROS 5,1 FS Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.) |
Code Information |
Software Version 2.8 and below; VITROS 5,1 FS Chemistry System Serial Numbers: J34000119 - J34002323; VITROS 5,1 FS Chemistry System (Refurbished) Serial Numbers: 34000133, 34000193, 34000197, 34000212, 34000285, 34000316, 34000318, 34000337, 34000360, 34000420, 34000451, 34000478, 34000495, 34000497, 34000529, 34000538, 34000552, 34000557, 34000590, 34000596, 34000601, 34000631, 34000642, 34000761, 34000799, 34000805, 34000833, 34000851, 34000893, 34000946, 34000947, 34001006, 34001172, 34001218, 34001240, 34001243, 34001248, 34001251, 34001274, 34001297, 34001317, 34001345, 34001378, 34001379, 34001389, 34001390, 34001396, 34001422, 34001472, 34001504, 34001569, 34001572, 34001612, 34001626, 34001633, 34001668, 34001755, 34001809, 34001840, 34001850, 34001875, 34001897, 34000123, 34000143, 34000191, 34000215, 34000219, 34000235, 34000251, 34000256, 34000260, 34000261, 34000319, 34000331, 34000380, 34000382, 34000406, 34000408, 34000412, 34000426, 34000431, 34000454, 34000464, 34000473, 34000501, 34000523, 34000530, 34000541, 34000542, 34000546, 34000548, 34000553, 34000558, 34000576, 34000577, 34000584, 34000585, 34000609, 34000612, 34000615, 34000616, 34000617, 34000618, 34000620, 34000625, 34000637, 34000644, 34000672, 34000744, 34000745, 34000755, 34000759, 34000767, 34000776, 34000793, 34000800, 34000814, 34000831, 34000844, 34000846, 34000852, 34000856, 34000870, 34000876, 34000887, 34000892, 34000902, 34000912, 34000920, 34000938, 34000956, 34000969, 34000976, 34001007, 34001009, 34001044, 34001045, 34001052, 34001071, 34001075, 34001098, 34001099, 34001111, 34001129, 34001154, 34001161, 34001171, 34001182, 34001192, 34001195, 34001201, 34001204, 34001208, 34001212, 34001234, 34001247, 34001254, 34001271, 34001278, 34001284, 34001288, 34001289, 34001293, 34001298, 34001301, 34001302, 34001316, 34001325, 34001337, 34001340, 34001348, 34001367, 34001369, 34001399, 34001400, 34001407, 34001410, 34001416, 34001423, 34001427, 34001429, 34001457, 34001490, 34001518, 34001526, 34001551, 34001573, 34001580, 34001581, 34001582, 34001587, 34001589, 34001594, 34001610, 34001618, 34001623, 34001627, 34001673, 34001688, 34001693, 34001725, 34001738, 34001745, 34001747, 34001790, 34001796, 34001798, 34001810, 34001822, 34001831, 34001833, 34001849, 34001855, 34001859, 34001860, 34001867, 34001871, 34001872, 34001881, 34001905, 34001916, 34001923, 34001995 |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Ms. Jennifer Paine 908-218-8776 |
Manufacturer Reason for Recall | Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall). |
FDA Determined Cause 2 | Error in labeling |
Action | Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION Letter (Ref. CL2015-201, dated December 21, 2015) via FedEx overnight courier and/or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address). The notification letter was sent to all customers who received the affected product. Foreign affiliates were informed by e-mail on 12/20/2015 of the issue and instructed to notify their consignees of the issue and required actions. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Technical Services at 1-800-421-3311. |
Quantity in Commerce | US = 994; Foreign = 1405 |
Distribution | Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|