• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 3 Device Recall CDl H/S Cuvette Perfusion Pack or Smart Perfusion Pack

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 3 Device Recall CDl H/S Cuvette Perfusion Pack or Smart Perfusion Pack see related information
Date Initiated by Firm January 19, 2016
Create Date February 25, 2016
Recall Status1 Terminated 3 on October 24, 2016
Recall Number Z-1018-2016
Recall Event ID 73126
510(K)Number K881330  
Product Classification Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWF
Product CDI H/S Cuvette, non-sterile, Manufactured by Terumo Cardiovascular Group, Perfusion Pack or Smart Perfusion Pack.

Product Usage:
Heart/Lung Perfusion Pack is either a customized tubing pack built to user specifications or a stock tubing pack, designed for use during surgery requiring cardiopulmonary bypass. It is not designed for long-term use (greater than six hours). Perfusion Packs labeled as SMART¿ contain components that have an additive to improve blood compatibility. Non-SMART packs may contain some SMART tubing and/or connectors depending on the customer requirements.
Code Information Catalog Number 6922, 1/2 inch -lots TE27, TH06 and TK03 Catalog Number 6923, 3/8 inch - lots TE13, TE18, TG15, TH20 and TK17 b. Model Number(s): 020204612, 020208807, 020210805, 020211806, 020493806, 044002300, 044003200, 044004102, 044004700, 044010200, 044016900, 044017601, 044018700, 044019901, 044027700, 044035101, 075104103, 075104404, 020135802, 088511700, 084103204, 047000500, 046001601, 065255600, 066110500, 075203800, 627265002, 627304502, 084118101, 020277801, 075101900, 627309201, 075104800, 627138002, 627241801, 627305101, 627239301, 044006801, 627347601, 075203501, 627232601, 627123006, 627280901, 088523400, 627282602, 044022001, 627252001, 044020900, 627147304, 627147403, 627153904, 046003400, 627308201, 044013600, 088511801, 084515200, 044009000, 627282001, 044022000, 066109900, 627286901, 075201201, 075202000, 084103204, 088504901, 088508000, 088513300, 088513501, 627133104, 627139306, 627165604, 627166101, 627167301, 627167501, 627167601, 627167701, 627168001, 627168101, 627179903, 627190501, 627190801, 627197802, 627198001, 627198102, 627212201, 627297401, 047002300, 044000600, 627251001, 627254501, 627283501, 627283601, 627321201, 044003901, 088504400, 627198903, 627214003, 044028900, 627231801, 627240802, 627243501, 627243601, 627243701, 627244602, 627244701, 627246002, 627252001, 627256901, 627260701, 627262303, 627266301, 627269502, 627286802, 627286901, 627290001, 627303901, 627304901, 627336301, 627336402, 627343801, 627346601, 627346701, 627349401 Catalog Number 6924, 1/4 inch - lots TE20, TF11, TG01 , TG15, TH 13 and TH27
Recalling Firm/
Manufacturer
Sorin Group USA, Inc.
14401 W 65th Way
Arvada CO 80004-3503
For Additional Information Contact Raymond Reid
303-467-6632
Manufacturer Reason
for Recall
The CDI System 500 monitor displays the "H/S DISCONNECT AT CUVETTE" error message when the CDI HIS Cuvette does not make a proper connection with the CDI H/S Probe. When this occurs, blood parameter values for HCT, Hgb and S02 do not display.
FDA Determined
Cause 2
Component design/selection
Action A Safety Alert letter dated January 13, 2015 via registered mail to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check the communication of the Cuvette prior to initiating Cardiopulmonary Bypass (CPS). The Terumo Safety Alert did not request removal or return of the product. The instructions that Sarin has communicated to the customers is identical to the information from Terumo. Acknowledgement of the letter and understanding of the contents is required for all customer responses.
Quantity in Commerce 6859
Distribution US Nationwide Distribution and Canada.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = DWF and Original Applicant = COBE LABORATORIES, INC.
-
-