• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Biomet Inc., Biomet Sports Medicine, and Biomet

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Biomet Inc., Biomet Sports Medicine, and Biomet see related information
Date Initiated by Firm April 01, 2016
Date Posted May 26, 2016
Recall Status1 Terminated 3 on May 23, 2017
Recall Number Z-1823-2016
Recall Event ID 74038
510(K)Number K992367  
Product Classification Screw, fixation, bone - Product Code HWC
Product Various trauma and sports medicine instruments and implants.
Screw, fixation, bone.
Code Information Part #'s HAA60-11030 HAA60-13030 HAA60-15040 HAA60-16040 HAA60-18040  Lot #'s 016150 016210 165130 016230 133890 181600 133970 134000 181620 
Recalling Firm/
Manufacturer
Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact Audrey Daenzer
574-372-1570
Manufacturer Reason
for Recall
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an infection.
FDA Determined
Cause 2
Equipment maintenance
Action Zimmer Biomet sent an Urgent Medical Device Recall Notice dated April 1, 2016, to all affected customers. Customers were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Distribution Worldwide Distribution - US (Nationwide) and globally to Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecaudor, The Netherlands, Japan, Korea, Malaysia, Mexico, Peru.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = HWC and Original Applicant = ELECTRO-BIOLOGY, INC.
-
-