| Class 2 Device Recall enGen Laboratory Automation System | |
Date Initiated by Firm | September 07, 2016 |
Create Date | October 22, 2016 |
Recall Status1 |
Terminated 3 on June 15, 2018 |
Recall Number | Z-0300-2017 |
Recall Event ID |
75167 |
510(K)Number | K063144 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
|
Product | enGen Laboratory Automation System (Product Code ENGEN) configured with Thermo Scientific Centrifuge Module and TCAutomation Software Version 3.6.1 and Below; IVD. --- Thermo Scientific Centrifuge Module has Product Code
952040-EG 6844097.
The VITROS 5,1 FS Chemistry System with enGen Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods). |
Code Information |
Software Versions 3.6.1 and below; Affected US Serial Numbers: 5GV6W2J,GBX395J, J2QKG25J, 2Z3YV1J, 3RHC23J, 4HX7S4J, 54FC01J, 5J4B45J, 736BM4J, 7KSC15J, J86PWC5J, 953209, 953221, 953266, 953270, B3ZHP2J, CXL941J, JFBX395J, FCM073J, FXCBG3J, JG7BFB5J, GWXQ52J, HJ5K84J, HPKG25J -- Affected Foreign Serial Numbers: DWXO52J, GWJZY4J, J16XQH4J, BGV6W2J, 17QP94J, 4SHPF4J, 1K5K84J, 37QP34J, H5PWC5J, J2V6LW4J, F6MXR3J, JT6LW4J, N0551107, F1150698, 46XQH4J, 2S5OP0J, CRKM82, J41BCY4J, J56T205J, BGV6W2J, 7YCF43J, 96XQHAJ, C1JNZ5J, C3JTP4J, D3JTP4J, 86T205J, 18BFB5J, 96PWC5J, 1LSSF5J, 4drj44j, 8W7WG3J, J953222, J953244, JCT20Q3J, HWJZY4J, 38769917923. |
Recalling Firm/ Manufacturer |
Ortho-Clinical Diagnostics 100 Indigo Creek Dr Rochester NY 14626-5101
|
For Additional Information Contact | Ms. Jennifer Paine 908-218-8776 |
Manufacturer Reason for Recall | Ortho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fishers centrifuge module. It is possible for mis-associated assay results to be reported from the laboratory prior to the identification of the Cross Check error on the enGen Laboratory Automation System. Assay results associated with the wrong patient and reported out of the laboratory could lead to inappropriate intervention with the potential for serious injury to the patient. |
FDA Determined Cause 2 | Software design |
Action | The firm, Ortho Clinical Diagnostics, sent an "URGENT PRODUCT CORRECTION NOTIFICATION" Letter (Ref. CL2016-183, dated 9/7/2016) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all enGen customers with centrifuge modules to inform them of the issue and advise them to check samples associated with cross check errors as instructed in the operator manual, and perform the preventative maintenance and cleaning procedures for that centrifuge module as provided in the maintenance manual and user guide. The customers were also instructed to Complete and return the Confirmation of Receipt form by 16 September 2016 via Fax:
1.888.557.3759 or 1.585.453.4110 or email:RA-OCDUS-CONFIRMAD@ORTHOCLINICALDIAGNOSTICS.COM.
For questions regarding this notification, please call the Ortho Care Solutions Center any time at 1-800-421-3311. Foreign affiliates were informed by e-mail on 9/7/2016 of the issue and instructed to notify their consignees of the issue and appropriate actions to take. |
Quantity in Commerce | 60 (US: 24; Foreign: 36) |
Distribution | Worldwide Distribution-(Nationwide) and foreign distribution to the following countries: Belgium, Bermuda, Canada, Chile, Denmark, France, Italy, Mexico, Norway, Portugal, Sweden, Taiwan and United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|