|
Class 1 Device Recall HeartMate 3 Left Ventricular Assist System |
|
Date Initiated by Firm |
April 05, 2018 |
Date Posted |
May 17, 2018 |
Recall Status1 |
Terminated 3 on October 15, 2021 |
Recall Number |
Z-1774-2018 |
Recall Event ID |
79893 |
PMA Number |
P160054 |
Product Classification |
Ventricular (assist) bypass - Product Code DSQ
|
Product |
Thoratec HeartMate 3 LVAS Implant Kit, Rx Only, Catalog # 106524US, 106524, 10652INT - LVAS KIT, HM 3
Indicated for providing short-term hemodynamic support in patients with advanced refractory left ventricular heart failure.
|
Code Information |
United States: UDI: 00813024013297, Serial Numbers: MLP 000976, MLP 000989, MLP 000991, MLP 000999, MLP 001002, MLP 002020, MLP 002041, MLP 002048, MLP 002049, MLP 002056, MLP 002057, MLP 002058, MLP 002059, MLP 002066, MLP 002069, MLP 002070, MLP 002071, MLP 002072, MLP 002074, MLP 002075, MLP 002076, MLP 002080, MLP 002082, MLP 002084, MLP 002087, MLP 002088, MLP 002090, MLP 002091, MLP 002093, MLP 002096, MLP 002117, MLP 002118, MLP 002119, MLP 002120, MLP 002121, MLP 002123, MLP 002124, MLP 002125, MLP 002126, MLP 002127, MLP 002128, MLP 002130, MLP 002132, MLP 002133, MLP 002134, MLP 002135, MLP 002136, MLP 002137, MLP 002141, MLP 002144, MLP 002152, MLP 002155, MLP 002156, MLP 002157, MLP 002158, MLP 002159, MLP 002160, MLP 002162, MLP 002163, MLP 002161, MLP 002164, MLP 002165, MLP 002188, MLP 002189, MLP 002191, MLP 002192, MLP 002197, MLP 002198, MLP 002200, MLP 002201, MLP 002203, MLP 002204, MLP 002205, MLP 002206, MLP 002207, MLP 002208, MLP 002209, MLP 002211, MLP 002213, MLP 002214, MLP 002216, MLP 002217, MLP 002218, MLP 002219, MLP 002220, MLP 002221, MLP 002222, MLP 002223, MLP 002224, MLP 002226, MLP 002227, MLP 002228, MLP 002229, MLP 002236, MLP 002237, MLP 002238, MLP 002239, MLP 002240, MLP 002241, MLP 002243, MLP 002244, MLP 002245, MLP 002246, MLP 002247, MLP 002248, MLP 002250, MLP 002251, MLP 002254, MLP 002253, MLP 002255, MLP 002257, MLP 002258, MLP 002259, MLP 002261, MLP 002263, MLP 002264, MLP 002266, MLP 002267, MLP 002268, MLP 002269, MLP 002271, MLP 002272, MLP 002273, MLP 002274, MLP 002275, MLP 002276, MLP 002277, MLP 002278, MLP 002280, MLP 002281, MLP 002282, MLP 002283, MLP 002285, MLP 002286, MLP 002287, MLP 002288, MLP 002289, MLP 002290, MLP 002291, MLP 002292, MLP 002293, MLP 002294, MLP 002295, MLP 002300, MLP 002303, MLP 002304, MLP 002306, MLP 002307, MLP 002308, MLP 002309, MLP 002310, MLP 002311, MLP 002312, MLP 002314, MLP 002315, MLP 002316, MLP 002317, MLP 002318, MLP 002319, MLP 002320, MLP 002321, MLP 002324, MLP 002327, MLP 002329, MLP 002330, MLP 002331, MLP 002334, MLP 002336, MLP 002338, MLP 002337, MLP 002340, MLP 002341, MLP 002342, MLP 002343, MLP 002344, MLP 002345, MLP 002346, MLP 002347, MLP 002348, MLP 002349, MLP 002350, MLP 002352, MLP 002354, MLP 002355, MLP 002356, MLP 002357, MLP 002359, MLP 002361, MLP 002362, MLP 002363, MLP 002365, MLP 002366, MLP 002421, MLP 002448, MLP 002449, MLP 002450, MLP 002451, MLP 002452, MLP 002454, MLP 002455, MLP 002456, MLP 002457, MLP 002470, MLP 002471, MLP 002472, MLP 002473, MLP 002474, MLP 002475, MLP 002476, MLP 002477, MLP 002478, MLP 002479, MLP 002480, MLP 002481, MLP 002482, MLP 002483, MLP 002484, MLP 002485, MLP 002486, MLP 002487, MLP 002488, MLP 002509, MLP 002510, MLP 002511, MLP 002512, MLP 002513, MLP 002514, MLP 002516, MLP 002518, MLP 002519, MLP 002520, MLP 002522, MLP 002523, MLP 002524, MLP 002525, MLP 002526, MLP 002529, MLP 002530, MLP 002532, MLP 002531, MLP 002533, MLP 002534, MLP 002536, MLP 002537, MLP 002535, MLP 002538, MLP 002539, MLP 002540, MLP 002541, MLP 002542, MLP 002543, MLP 002544, MLP 002566, MLP 002579, MLP 002580, MLP 002581, MLP 002583, MLP 002584, MLP 002587, MLP 002589, MLP 002590, MLP 002591, MLP 002593, MLP 002596, MLP 002597, MLP 002598, MLP 002600, MLP 002603, MLP 002607, MLP 002611, MLP 002613, MLP 002614, MLP 002616, MLP 002617, MLP 002621, MLP 002624, MLP 002628, MLP 002629, MLP 002632, MLP 002633, MLP 002634, MLP 002635, MLP 002636, MLP 002637, MLP 002638, MLP 002640, MLP 002641, MLP 002643, MLP 002646, MLP 002647, MLP 002648, MLP 002649, MLP 002650, MLP 002651, MLP 002652, MLP 002653, MLP 002654, MLP 002655, MLP 002656, MLP 002658, MLP 002659, MLP 002660, MLP 002661, MLP 002663, MLP 002665, MLP 002666, MLP 002805, MLP 002806, MLP 002807, MLP 002808, MLP 002809, MLP 002810, MLP 002811, MLP 002812, MLP 002813, MLP 002815, MLP 002816, MLP 002817, MLP 002827, MLP 002828, MLP 002829, MLP 002831, MLP 002833, MLP 002834, MLP 002837, MLP 002838, MLP 002840, MLP 002841, MLP 002842, MLP 002843, MLP 002844, MLP 002845, MLP 002847, MLP 002848, MLP 002849, MLP 002850, MLP 002851, MLP 002853, MLP 002854, MLP 002855, MLP 002857, MLP 002858, MLP 002860, MLP 002861, MLP 002863, MLP 002952, MLP 002953, MLP 002954, MLP 002955, MLP 002956, MLP 002957, MLP 002958, MLP 002963, MLP 003024, MLP 003025, MLP 003026, MLP 003028, MLP 003032, MLP 003033, MLP 003039, MLP 003043, MLP 003044, MLP 003050, MLP 003053, MLP 003054, MLP 003057, MLP 003058, MLP 003064, MLP 003065, MLP 003067, MLP 003068, MLP 003069, MLP 003070, MLP 003071, MLP 003072, MLP 003074, MLP 003075, MLP 003077, MLP 003078, MLP 003085, MLP 003086, MLP 003089, MLP 003095, MLP 003096, MLP 003097, MLP 003098, MLP 003099, MLP 003101, MLP 003102, MLP 003105, MLP 003107, MLP 003109, MLP 003110, MLP 003111, MLP 003114, MLP 003122, MLP 003125, MLP 003130, MLP 003131, MLP 003132, MLP 003135, MLP 003139, MLP 003140, MLP 003146, MLP 003148, MLP 003151, MLP 003153, MLP 003163, MLP 003164, MLP 003176, MLP 003177, MLP 003178, MLP 003180, MLP 003183, MLP 003184, MLP 003185, MLP 003186, MLP 003187, MLP 003188, MLP 003189, MLP 003190, MLP 003193, MLP 003237, MLP 003238, MLP 003239, MLP 003240, MLP 003241, MLP 003242, MLP 003243, MLP 003245, MLP 003246, MLP 003247, MLP 003251, MLP 003255, MLP 003256, MLP 003257, MLP 003259, MLP 003260, MLP 003261, MLP 003262, MLP 003263, MLP 003266, MLP 003273, MLP 003274, MLP 003275, MLP 003277, MLP 003278, MLP 003283, MLP 003285, MLP 003286, MLP 003290, MLP 003291, MLP 003293, MLP 003294, MLP 003296, MLP 003298, MLP 003299, MLP 003300, MLP 003302, MLP 003304, MLP 003307, MLP 003308, MLP 003309, MLP 003310, MLP 003311, MLP 003312, MLP 003314, MLP 003315, MLP 003316, MLP 003317, MLP 003330, MLP 003331, MLP 003332, MLP 003333, MLP 003335, MLP 003339, MLP 003340, MLP 003341, MLP 003342, MLP 003343, MLP 003346, MLP 003347, MLP 003348, MLP 003349, MLP 003351, MLP 003352, MLP 003353, MLP 003355, MLP 003358, MLP 003359, MLP 003360, MLP 003362, MLP 003363, MLP 003365, MLP 003367, MLP 003368, MLP 003369, MLP 003370, MLP 003371, MLP 003372, MLP 003373, MLP 003375, MLP 003376, MLP 003377, MLP 003378, MLP 003380, MLP 003381, MLP 003382, MLP 003383, MLP 003384, MLP 003385, MLP 003386, MLP 003387, MLP 003389, MLP 003390, MLP 003391, MLP 003392, MLP 003393, MLP 003394, MLP 003437, MLP 003438, MLP 003439, MLP 003440, MLP 003388, MLP 003441, MLP 003442, MLP 003443, MLP 003446, MLP 003447, MLP 003448, MLP 003455, MLP 003456, MLP 003457, MLP 003458, MLP 003459, MLP 003460, MLP 003461, MLP 003462, MLP 003463, MLP 003464, MLP 003466, MLP 003467, MLP 003468, MLP 003469, MLP 003470, MLP 003471, MLP 003472, MLP 003473, MLP 003474, MLP 003475, MLP 003477, MLP 003478, MLP 003479, MLP 003480, MLP 003481, MLP 003482, MLP 003483, MLP 003484, MLP 003525, MLP 003570, MLP 003572, MLP 003573, MLP 003574, MLP 003576, MLP 003577, MLP 003578, MLP 003579, MLP 003580, MLP 003581, MLP 003582, MLP 003583, MLP 003584, MLP 003586, MLP 003587, MLP 003588, MLP 003589, MLP 003590, MLP 003592, MLP 003593, MLP 003594, MLP 003613, MLP 003614, MLP 003615, MLP 003616, MLP 003617, MLP 003618, MLP 003619, MLP 003620, MLP 003621, MLP 003571, MLP 003622, MLP 003623, MLP 003625, MLP 003626, MLP 003627, MLP 003628, MLP 003629, MLP 003655, MLP 003656, MLP 003657, MLP 003659, MLP 003661, MLP 003663, MLP 003683, MLP 003685, MLP 003687, MLP 003689, MLP 003691, MLP 003692, MLP 003702, MLP 003703, MLP 003704, MLP 003705, MLP 003736, MLP 003737, MLP 003739, MLP 003740, MLP 003741, MLP 003742, MLP 003744, MLP 003745, MLP 003746, MLP 003654, MLP 003826, MLP 003827, MLP 003828, MLP 003829, MLP 003830, MLP 003831, MLP 003832, MLP 003833, MLP 003858, MLP 003859, MLP 003860, MLP 003861, MLP 003738, MLP 003862, MLP 003863, MLP 003864, MLP 003865, MLP 003866, MLP 003867, MLP 003748, MLP 004050, MLP 004093, MLP 004096, MLP 004099, MLP 004100, MLP 004103, MLP 004104, MLP 004108, MLP 004109, MLP 004112, MLP 004115, MLP 004118, MLP 004119, MLP 004124, MLP 004125, MLP 004127, MLP 004128, MLP 004129, MLP 004130, MLP 004133, MLP 004143, MLP 004153, MLP 004154, MLP 004197, MLP 004210, MLP 004237, MLP 004238, MLP 004247, MLP 004249, MLP 004257, MLP 004260, MLP 004267, MLP 004290, MLP 004291, MLP 004292, MLP 004293, MLP 004296, MLP 004298, MLP 004309, MLP 004316, MLP 004317, MLP 004340, MLP 004349, MLP 004355, MLP 004384, MLP 004444, MLP 004473, MLP 004479, MLP 004480, MLP 005190, MLP 006482, MLP 006705, MLP 006712, MLP 009424, MLP 004097, MLP 004110, MLP 004117, MLP 004131, MLP 004137, MLP 004139, MLP 004140, MLP 004142, MLP 004144, MLP 004145, MLP 004148, MLP 004150, MLP 004151, MLP 004155, MLP 004156, MLP 004157, MLP 004158, MLP 004159, MLP 004161, MLP 004162, MLP 004165, MLP 004166, MLP 004170, MLP 004171, MLP 004172, MLP 004174, MLP 004179, MLP 004185, MLP 004194, MLP 004200, MLP 004208, MLP 004212, MLP 004213, MLP 004216, MLP 004220, MLP 004221, MLP 004227, MLP 004228, MLP 004233, MLP 004234, MLP 004242, MLP 004246, MLP 004253, MLP 004254, MLP 004256, MLP 004259, MLP 004268, MLP 004279, MLP 004281, MLP 004284, MLP 004285, MLP 004289, MLP 004303, MLP 004304, MLP 004305, MLP 004310, MLP 004311, MLP 004346, MLP 004366, MLP 004369, MLP 004378, MLP 004385, MLP 004386, MLP 004391, MLP 004282, MLP 004405, MLP 004414, MLP 004430, MLP 004454, MLP 004469, MLP 004470, MLP 004486, MLP 004489, MLP 004490, MLP 004491, MLP 004494, MLP 004505, MLP 004514, MLP 004524, MLP 004525, MLP 004532, MLP 004545, MLP 004554, MLP 004584, MLP 004585, MLP 004586, MLP 004587, MLP 004588, MLP 004589, MLP 004590, MLP 004591, MLP 004667, MLP 004668, MLP 004670, MLP 004671, MLP 004672, MLP 004675, MLP 004676, MLP 004685, MLP 004686, MLP 004689, MLP 004695, MLP 004697, MLP 004698, MLP 004699, MLP 004700, MLP 004701, MLP 004702, MLP 004703, MLP 004705, MLP 004707, MLP 004797, MLP 004798, MLP 004799, MLP 004802, MLP 004803. MLP 004804, MLP 004805, MLP 004806, MLP 004864, MLP 004865, MLP 004866, MLP 004867, MLP 004868, MLP 004869, MLP 004870, MLP 004871, MLP 004872, MLP 004874, MLP 004875, MLP 004876, MLP 004877, MLP 004878, MLP 004879, MLP 004880, MLP 004881, MLP 004882, MLP 004884, MLP 004885, MLP 004886, MLP 004887, MLP 004888, MLP 004889, MLP 004890, MLP 004891, MLP 004892, MLP 004893, MLP 004903, MLP 004807, MLP 004954, MLP 004956, MLP 004957, MLP 004958, MLP 004959, MLP 004960, MLP 004961, MLP 004962, MLP 004963, MLP 004964, MLP 004965, MLP 004966, MLP 004967, MLP 004968, MLP 004969, MLP 004970, MLP 004971, MLP 004972, MLP 004973, MLP 004983, MLP 004994, MLP 004995, MLP 004996, MLP 004997, MLP 004998, MLP 005000, MLP 005002, MLP 005004, MLP 005005, MLP 005006, MLP 005007, MLP 005009, MLP 005010, MLP 005012, MLP 005013, MLP 005014, MLP 005025, MLP 005026, MLP 005028, MLP 005030, MLP 005032, MLP 005033, MLP 005034, MLP 005035, MLP 005036, MLP 005037, MLP 005048, MLP 005049, MLP 005050, MLP 005051, MLP 005052, MLP 005053, MLP 005054, MLP 005055, MLP 005056, MLP 005057, MLP 005078, MLP 005079, MLP 005080, MLP 005081, MLP 005083, MLP 005086, MLP 005089, MLP 005090, MLP 005091, MLP 005114, MLP 005001, MLP 005115, MLP 005116, MLP 005117, MLP 005120, MLP 005121, MLP 005122, MLP 005123, MLP 005124, MLP 005139, MLP 005140, MLP 005141, MLP 005142, MLP 005143, MLP 005144, MLP 005145, MLP 005146, MLP 005147, MLP 005148, MLP 005177, MLP 005178, MLP 005179, MLP 005180, MLP 005181, MLP 005182, MLP 005183, MLP 005184, MLP 005185, MLP 005186, MLP 005187, MLP 005188, MLP 005189, MLP 005191, MLP 005192, MLP 005193, MLP 005194, MLP 005195, MLP 005196, MLP 005208, MLP 005209, MLP 005210, MLP 005211, MLP 005212, MLP 005213, MLP 005214, MLP 005215, MLP 005216, MLP 005219, MLP 005220, MLP 005221, MLP 005222, MLP 005223, MLP 005224, MLP 005225, MLP 005226, MLP 005227, MLP 005228, MLP 005258, MLP 005259, MLP 005261, MLP 005262, MLP 005263, MLP 005264, MLP 005266, MLP 005268, MLP 005290, MLP 005291, MLP 005292, MLP 005293, MLP 005294, MLP 005295, MLP 005296, MLP 005297, MLP 005298, MLP 005299, MLP 005320, MLP 005321, MLP 005322, MLP 005323, MLP 005324, MLP 005326, MLP 005327, MLP 005328, MLP 005329, MLP 005330, MLP 005332, MLP 005333, MLP 005334, MLP 005335, MLP 005337, MLP 005339, MLP 005365, MLP 005385, MLP 005389, MLP 005390, MLP 005391, MLP 005392, MLP 005393, MLP 005394, MLP 005395, MLP 005396, MLP 005398, MLP 005399, MLP 005400, MLP 005260, MLP 005401, MLP 005402, MLP 005403, MLP 005404, MLP 005405, MLP 005406, MLP 005407, MLP 005408, MLP 005409, MLP 005410, MLP 005411, MLP 005412, MLP 005413, MLP 005414, MLP 005415, MLP 005416, MLP 005420, MLP 005426, MLP 005438, MLP 005440, MLP 005441, MLP 005442, MLP 005443, MLP 005444, MLP 005447, MLP 005454, MLP 005460, MLP 005462, MLP 005463, MLP 005465, MLP 005468, MLP 005471, MLP 005479, MLP 005481, MLP 005482, MLP 005483, MLP 005484, MLP 005485, MLP 005488, MLP 005489, MLP 005490, MLP 005492, MLP 005494, MLP 005495, MLP 005497, MLP 005498, MLP 005499, MLP 005502, MLP 005503, MLP 005512, MLP 005513, MLP 005514, MLP 005516, MLP 005528, MLP 005529, MLP 005530, MLP 005531, MLP 005532, MLP 005533, MLP 005534, MLP 005539, MLP 005540, MLP 005541, MLP 005543, MLP 005544, MLP 005545, MLP 005551, MLP 005552, MLP 005553, MLP 005554, MLP 005555, MLP 005559, MLP 005560, MLP 005561, MLP 005574, MLP 005575, MLP 005576, MLP 005578, MLP 005579, MLP 005591, MLP 005592, MLP 005593, MLP 005594, MLP 005595, MLP 005597, MLP 005600, MLP 005601, MLP 005602, MLP 005604, MLP 005605, MLP 005606, MLP 005608, MLP 005610, MLP 005613, MLP 005616, MLP 005617, MLP 005618, MLP 005619, MLP 005459, MLP 005620, MLP 005621, MLP 005622, MLP 005478, MLP 005636, MLP 005638, MLP 005639, MLP 005640, MLP 005641, MLP 005642, MLP 005648, MLP 005650, MLP 005651, MLP 005653, MLP 005654, MLP 005655, MLP 005660, MLP 005661, MLP 005663, MLP 005668, MLP 005680, MLP 005684, MLP 005685, MLP 005686, MLP 005687, MLP 005695, MLP 005696, MLP 005697, MLP 005698, MLP 005700, MLP 005701, MLP 005703, MLP 005704, MLP 005706, MLP 005707, MLP 005708, MLP 005709, MLP 005728, MLP 005729, MLP 005730, MLP 005731, MLP 005742, MLP 005745, MLP 005746, MLP 005747, MLP 005748, MLP 005749, MLP 005750, MLP 005751, MLP 005752, MLP 005754, MLP 005761, MLP 005766, MLP 005767, MLP 005769, MLP 005773, MLP 005774, MLP 005779, MLP 005791, MLP 005793, MLP 005794, MLP 005795, MLP 005803, MLP 005804, MLP 005805, MLP 005806, MLP 005807, MLP 005808, MLP 005809, MLP 005810, MLP 005811, MLP 005812, MLP 005813, MLP 005814, MLP 005816, MLP 005818, MLP 005820, MLP 005821, MLP 005822, MLP 005823, MLP 005824, MLP 005825, MLP 005826, MLP 005828, MLP 005830, MLP 005831, MLP 005833, MLP 005835, MLP 005836, MLP 005838, MLP 005839, MLP 005840, MLP 005841, MLP 005852, MLP 005853, MLP 005854, MLP 005699, MLP 005871, MLP 005873, MLP 005874, MLP 005875, MLP 005877, MLP 005878, MLP 005879, MLP 005900, MLP 005901, MLP 005902, MLP 005924, MLP 005926, MLP 005927, MLP 005928, MLP 005930, MLP 005931, MLP 005932, MLP 005933, MLP 005934, MLP 005792, MLP 005798, MLP 005946, MLP 005947, MLP 005948, MLP 005949, MLP 005950, MLP 005951, MLP 005962, MLP 005965, MLP 005815, MLP 005967, MLP 005968, MLP 006002, MLP 006003, MLP 006004, MLP 006005, MLP 006006, MLP 006007, MLP 006035, MLP 006036, MLP 006037, MLP 006040, MLP 006041, MLP 006042, MLP 006044, MLP 006046, MLP 006047, MLP 006048, MLP 006050, MLP 006051, MLP 006054, MLP 006055, MLP 006056, MLP 006067, MLP 006069, MLP 006073, MLP 006081, MLP 006090, MLP 006096, MLP 006100, MLP 006122, MLP 006125, MLP 006126, MLP 006127, MLP 006128, MLP 006186, MLP 006187, MLP 006189, MLP 006190, MLP 006191, MLP 006192, MLP 006193, MLP 006194, MLP 006195, MLP 006196, MLP 006197, MLP 006198, MLP 006200, MLP 006203, MLP 006204, MLP 005964, MLP 006205, MLP 006206, MLP 006207, MLP 006213, MLP 006215, MLP 006216, MLP 006218, MLP 006219, MLP 006221, MLP 006222, MLP 006223, MLP 006225, MLP 006226, MLP 006236, MLP 006237, MLP 006238, MLP 006239, MLP 006240, MLP 006244, MLP 006245, MLP 006246, MLP 006247, MLP 006248, MLP 006249, MLP 006250, MLP 006293, MLP 006294, MLP 006295, MLP 006298, MLP 006300, MLP 006301, MLP 006302, MLP 006303, MLP 006304, MLP 006305, MLP 006359, MLP 006045, MLP 006365, MLP 006053, MLP 006063, MLP 006386, MLP 006387, MLP 006388, MLP 006389, MLP 006390, MLP 006095, MLP 006407, MLP 006408, MLP 006409, MLP 006410, MLP 006411, MLP 006412, MLP 006413, MLP 006414, MLP 006415, MLP 006417, MLP 006422, MLP 006423, MLP 006424, MLP 006425, MLP 006430, MLP 006431, MLP 006202, MLP 006433, MLP 006434, MLP 006435, MLP 006437, MLP 006449, MLP 006450, MLP 006451, MLP 006452, MLP 006454, MLP 006460, MLP 006461, MLP 006463, MLP 006464, MLP 006466, MLP 006467, MLP 006468, MLP 006243, MLP 006469, MLP 006470, MLP 006471, MLP 006477, MLP 006479, MLP 006480, MLP 006481, MLP 006483, MLP 006484, MLP 006485, MLP 006486, MLP 006487, MLP 006488, MLP 006489, MLP 006490, MLP 006493, MLP 006518, MLP 006519, MLP 006520, MLP 006522, MLP 006533, MLP 006534, MLP 006535, MLP 006536, MLP 006537, MLP 006538, MLP 006539, MLP 006540, MLP 006541, MLP 006542, MLP 006553, MLP 006554, MLP 006555, MLP 006557, MLP 006558, MLP 006570, MLP 006571, MLP 006572, MLP 006577, MLP 006578, MLP 006579, MLP 006580, MLP 006581, MLP 006582, MLP 006583, MLP 006584, MLP 006585, MLP 006586, MLP 006587, MLP 006588, MLP 006590, MLP 006591, MLP 006595, MLP 006432, MLP 006596, MLP 006436, MLP 006627, MLP 006628, MLP 006629, MLP 006630, MLP 006631, MLP 006632, MLP 006633, MLP 006634, MLP 006635, MLP 006636, MLP 006657, MLP 006658, MLP 006659, MLP 006660, MLP 006662, MLP 006663, MLP 006670, MLP 006671, MLP 006672, MLP 006673, MLP 006674, MLP 006675, MLP 006676, MLP 006677, MLP 006678, MLP 006688, MLP 006690, MLP 006695, MLP 006706, MLP 006708, MLP 006709, MLP 006710, MLP 006711, MLP 006714, MLP 006715, MLP 006716, MLP 006556, MLP 006739, MLP 006742, MLP 006756, MLP 006764, MLP 006765, MLP 006768, MLP 006592, MLP 006773, MLP 006774, MLP 006776, MLP 006777, MLP 006800, MLP 006801, MLP 006811, MLP 006812, MLP 006824, MLP 006825, MLP 006826, MLP 006827, MLP 006828, MLP 006829, MLP 006830, MLP 006831, MLP 006832, MLP 006833, MLP 006661, MLP 006844, MLP 006845, MLP 006847, MLP 006848, MLP 006849, MLP 006850, MLP 006851, MLP 006852, MLP 006864, MLP 006865, MLP 006866, MLP 006867, MLP 006870, MLP 006875, MLP 006876, MLP 006687, MLP 006689, MLP 006904, MLP 006905, MLP 006707, MLP 007249, MLP 007250, MLP 006766, MLP 006767, MLP 006778, MLP 007379, MLP 007380, MLP 007381, MLP 007382, MLP 007383, MLP 007384, MLP 007385, MLP 006846, MLP 007386, MLP 007387, MLP 007388, MLP 007389, MLP 007390, MLP 007391, MLP 007392, MLP 006853, MLP 007393, MLP 007394, MLP 007395, MLP 007396, MLP 007398, MLP 007400, MLP 007401, MLP 007402, MLP 007403, MLP 007406, MLP 007407, MLP 007408, MLP 007410, MLP 007414, MLP 007415, MLP 007417, MLP 007419, MLP 007420, MLP 007421, MLP 006906, MLP 007422, MLP 007428, MLP 007429, MLP 007431, MLP 007432, MLP 007433, MLP 007435, MLP 007436, MLP 007437, MLP 007438, MLP 007439, MLP 007440, MLP 007441, MLP 007442, MLP 007443, MLP 007444, MLP 007446, MLP 007448, MLP 007449, MLP 007451, MLP 007452, MLP 007453, MLP 007454, MLP 007455, MLP 007458, MLP 007459, MLP 007460, MLP 007461, MLP 007462, MLP 007463, MLP 007464, MLP 007465, MLP 007467, MLP 007469, MLP 007470, MLP 007474, MLP 007476, MLP 007477, MLP 007479, MLP 007480, MLP 007501, MLP 007503, MLP 007504, MLP 007506, MLP 007507, MLP 007508, MLP 007509, MLP 007511, MLP 007536, MLP 007537, MLP 007538, MLP 007539, MLP 007540, MLP 007541, MLP 007542, MLP 007543, MLP 007545, MLP 007547, MLP 007548,
MLP 007550, MLP 007551, MLP 007552, MLP 007553, MLP 007583, MLP 007584, MLP 007585, MLP 007586, MLP 007587, MLP 007588, MLP 007589, MLP 007591, MLP 007602, MLP 007603, MLP 007605, MLP 007608, MLP 007610, MLP 007611, MLP 007622, MLP 007623, MLP 007624, MLP 007625, MLP 007626, MLP 007627, MLP 007628, MLP 007629, MLP 007631, MLP 007632, MLP 007634, MLP 007636, MLP 007638, MLP 007639, MLP 007640, MLP 007642, MLP 007655, MLP 007656, MLP 007657, MLP 007658, MLP 007659, MLP 007660, MLP 007661, MLP 007662, MLP 007664, MLP 007399, MLP 007665, MLP 007667, MLP 007668, MLP 007669, MLP 007672, MLP 007405, MLP 007673, MLP 007674, MLP 007675, MLP 007676, MLP 007679, MLP 007681, MLP 007411, MLP 007682, MLP 007683, MLP 007686, MLP 007689, MLP 007690, MLP 007418, MLP 007701, MLP 007702, MLP 007703, MLP 007704, MLP 007430, MLP 007705, MLP 007707, MLP 007708, MLP 007711, MLP 007434, MLP 007712, MLP 007713, MLP 007716, MLP 007717, MLP 007718, MLP 007719, MLP 007720, MLP 007445, MLP 007447, MLP 007466, MLP 007468, MLP 007753, MLP 007754, MLP 007755, MLP 007756, MLP 007473, MLP 007757, MLP 007758, MLP 007475, MLP 007759, MLP 007760, MLP 007765, MLP 007767, MLP 007481, MLP 007768, MLP 007482, MLP 007769, MLP 007770, MLP 007771, MLP 007772, MLP 007773, MLP 007774, MLP 007775, MLP 007778, MLP 007779, MLP 007792, MLP 007793, MLP 007795, MLP 007796, MLP 007798, MLP 007799, MLP 007800, MLP 007801, MLP 007806, MLP 007510, MLP 007807, MLP 007808, MLP 007811, MLP 007822, MLP 007823, MLP 007825, MLP 007826, MLP 007827, MLP 007828, MLP 007829, MLP 007830, MLP 007832, MLP 007833, MLP 007534, MLP 007834, MLP 007835, MLP 007836, MLP 007838, MLP 007839, MLP 007840, MLP 007544, MLP 007546, MLP 007549, MLP 007582, MLP 007870, MLP 007871, MLP 007872, MLP 007590, MLP 007874, MLP 007875, MLP 007876, MLP 007877, MLP 007879, MLP 007881, MLP 007882, MLP 007885, MLP 007887, MLP 007888, MLP 007889, MLP 007890, MLP 007604, MLP 007606, MLP 007607, MLP 007902, MLP 007904, MLP 007905, MLP 007907, MLP 007909, MLP 007910, MLP 007911, MLP 007912, MLP 007913, MLP 007915, MLP 007916, MLP 007920, MLP 007923, MLP 007630, MLP 007926, MLP 007927, MLP 007928, MLP 007929, MLP 007635, MLP 007930, MLP 007931, MLP 007637, MLP 007932, MLP 007933, MLP 007936, MLP 007937, MLP 007940, MLP 007643, MLP 007941, MLP 007944, MLP 007945, MLP 007990, MLP 008001, MLP 007663, MLP 008024, MLP 007670, MLP 008026, MLP 007671, MLP 008027, MLP 008029, MLP 008030, MLP 008032, MLP 007677, MLP 007678, MLP 007680, MLP 008045, MLP 008046, MLP 008047, MLP 007685, MLP 008048, MLP 008049, MLP 007687, MLP 008051, MLP 007688, MLP 008052, MLP 008053, MLP 008068, MLP 008069, MLP 008070, MLP 008071, MLP 008074, MLP 008076, MLP 008080, MLP 007706, MLP 008082, MLP 008083, MLP 008084, MLP 007709, MLP 008085, MLP 007710, MLP 008086, MLP 008088, MLP 008089, MLP 008090, MLP 007715, MLP 008091, MLP 008105, MLP 008106, MLP 008108, MLP 008110, MLP 008112, MLP 008114, MLP 008116, MLP 008118, MLP 008119, MLP 008120, MLP 008126, MLP 008128, MLP 008130, MLP 008131, MLP 008135, MLP 008137, MLP 008139, MLP 008141, MLP 008159, MLP 008162, MLP 008163, MLP 008166, MLP 008168, MLP 008169, MLP 007761, MLP 007762, MLP 007764, MLP 008191, MLP 008192, MLP 008193, MLP 008194, MLP 008195, MLP 008197, MLP 007776, MLP 008198, MLP 007777, MLP 008199, MLP 008200, MLP 008201, MLP 007780, MLP 008202, MLP 007781, MLP 008204, MLP 008206, MLP 008207, MLP 008208, MLP 008210, MLP 008215, MLP 008216, MLP 008217, MLP 008218, MLP 007794, MLP 008219, MLP 008220, MLP 008221, MLP 008223, MLP 008224, MLP 007802, MLP 007803, MLP 008245, MLP 008248, MLP 008249, MLP 008251, MLP 008252, MLP 008253, MLP 008254, MLP 008255, MLP 008258, MLP 008259, MLP 008260, MLP 008261, MLP 008265, MLP 008266, MLP 008267, MLP 008268, MLP 008270, MLP 008271, MLP 008272, MLP 008274, MLP 008275, MLP 008276, MLP 008278, MLP 008279, MLP 008280, MLP 007841, MLP 008281, MLP 008283, MLP 008284, MLP 008295, MLP 008296, MLP 008298, MLP 008299, MLP 008300, MLP 008301, MLP 008306, MLP 008307, MLP 008308, MLP 008315, MLP 008327, MLP 008328, MLP 008329, MLP 008330, MLP 008332, MLP 008333, MLP 008334, MLP 008338, MLP 008339, MLP 007880, MLP 008343, MLP 008345, MLP 008347, MLP 007883, MLP 008349, MLP 008350, MLP 007886, MLP 008352, MLP 008353, MLP 008356, MLP 008357, MLP 008361, MLP 008362, MLP 008363, MLP 008364, MLP 008365, MLP 008367, MLP 008369, MLP 007897, MLP 007899, MLP 007903, MLP 008381, MLP 008383, MLP 008384, MLP 007908, MLP 008385, MLP 008386, MLP 008387, MLP 008388, MLP 008389, MLP 008390, MLP 007914, MLP 008391, MLP 008392, MLP 008393, MLP 008394, MLP 008398, MLP 008399, MLP 007921, MLP 008401, MLP 007922, MLP 008402, MLP 008403, MLP 008404, MLP 008416, MLP 008417, MLP 008418, MLP 008419, MLP 008421, MLP 008422, MLP 008423, MLP 008424, MLP 008425, MLP 007934, MLP 008428, MLP 007935, MLP 008430, MLP 007938, MLP 007939, MLP 008443, MLP 007942, MLP 008444, MLP 007943, MLP 008445, MLP 008446, MLP 008447, MLP 007946, MLP 007983, MLP 007986, MLP 007987, MLP 007989, MLP 007999, MLP 008000, MLP 008002, MLP 008005, MLP 008006, MLP 008008, MLP 008009, MLP 008010, MLP 008013, MLP 008022, MLP 008023, MLP 008025, MLP 008028, MLP 008031, MLP 008034, MLP 008054, MLP 008066, MLP 008073, MLP 008077, MLP 008078, MLP 008079, MLP 008092, MLP 008093, MLP 008104, MLP 008107, MLP 008115, MLP 008117, MLP 008132, MLP 008133, MLP 008134, MLP 008138, MLP 008161, MLP 008164, MLP 008165, MLP 008167, MLP 008196, MLP 008203, MLP 008205, MLP 008209, MLP 008222, MLP 008246, MLP 008247, MLP 008250, MLP 008257, MLP 008262, MLP 008263, MLP 008264, MLP 008269, MLP 008273, MLP 008277, MLP 008282, MLP 008302, MLP 008309, MLP 008310, MLP 008311, MLP 008312, MLP 008314, MLP 008331, MLP 008336, MLP 008340, MLP 008341, MLP 008342, MLP 008344, MLP 008346, MLP 008348, MLP 008351, MLP 008359, MLP 008360, MLP 008366, MLP 008368, MLP 008380, MLP 008382, MLP 008396, MLP 008397, MLP 008400, MLP 008405, MLP 008420, MLP 008426, MLP 008429, MLP 008441, MLP 008442, MLP 008448, MLP 008449, MLP 008450, MLP 008461, MLP 008462, MLP 008463, MLP 008464, MLP 008465, MLP 008466, MLP 008467, MLP 008468, MLP 008469, MLP 008470, MLP 008501, MLP 008502, MLP 008503, MLP 008504, MLP 008505, MLP 008506, MLP 008507, MLP 008508, MLP 008509, MLP 008510, MLP 008511, MLP 008512, MLP 008513, MLP 008515, MLP 008516, MLP 008517, MLP 008518, MLP 008519, MLP 008520, MLP 008521, MLP 008522, MLP 008523, MLP 008526, MLP 008531, MLP 008532, MLP 008534, MLP 008535, MLP 008538, MLP 008539, MLP 008540, MLP 008541, MLP 008542, MLP 008549, MLP 008550, MLP 008551, MLP 008552, MLP 008553, MLP 008554, MLP 008557, MLP 008558, MLP 008559, MLP 008560, MLP 008561, MLP 008562, MLP 008568, MLP 008570, MLP 008574, MLP 008575, MLP 008577, MLP 008578, MLP 008579, MLP 008580, MLP 008581, MLP 008582, MLP 008583, MLP 008584, MLP 008585, MLP 008586, MLP 008587, MLP 008588, MLP 008589, MLP 008590, MLP 008591, MLP 008592, MLP 008593, MLP 008594, MLP 008605, MLP 008606, MLP 008607, MLP 008608, MLP 008609, MLP 008610, MLP 008611, MLP 008612, MLP 008613, MLP 008614, MLP 008615, MLP 008616, MLP 008617, MLP 008619, MLP 008620, MLP 008621, MLP 008623, MLP 008624, MLP 008645, MLP 008646, MLP 008647, MLP 008648, MLP 008649, MLP 008650, MLP 008651, MLP 008652, MLP 008653, MLP 008654, MLP 008655, MLP 008656, MLP 008657, MLP 008658, MLP 008659, MLP 008660, MLP 008661, MLP 008662, MLP 008663, MLP 008664, MLP 008676, MLP 008678, MLP 008679, MLP 008680, MLP 008681, MLP 008684, MLP 008685, MLP 008763, MLP 008764, MLP 008765, MLP 008766, MLP 008767, MLP 008768, MLP 008769, MLP 008770, MLP 008771, MLP 008772, MLP 008773, MLP 008774, MLP 008775, MLP 008776, MLP 008777, MLP 008778, MLP 008779, MLP 008780, MLP 008781, MLP 008782, MLP 008794, MLP 008795, MLP 008796, MLP 008797, MLP 008798, MLP 008799, MLP 008800, MLP 008801, MLP 008803, MLP 008804, MLP 008805, MLP 008806, MLP 008808, MLP 008810, MLP 008811, MLP 008812, MLP 008813, MLP 008814, MLP 008815, MLP 008816, MLP 008817, MLP 008818, MLP 008819, MLP 008820, MLP 008821, MLP 008822, MLP 008823, MLP 008824, MLP 008825, MLP 008826, MLP 008827, MLP 008828, MLP 008829, MLP 008830, MLP 008831, MLP 008832, MLP 008833, MLP 008834, MLP 008835, MLP 008836, MLP 008837, MLP 008839, MLP 008840, MLP 008841, MLP 008842, MLP 008843, MLP 008844, MLP 008845, MLP 008846, MLP 008847, MLP 008868, MLP 008869, MLP 008870, MLP 008871, MLP 008872, MLP 008873, MLP 008874, MLP 008875, MLP 008876, MLP 008877, MLP 008878, MLP 008879, MLP 008880, MLP 008881, MLP 008882, MLP 008883, MLP 008884, MLP 008885, MLP 008887, MLP 008888, MLP 008889, MLP 008890, MLP 008891, MLP 008892, MLP 008893, MLP 008894, MLP 008895, MLP 008896, MLP 008897, MLP 008898, MLP 008899, MLP 008900, MLP 008901, MLP 008902, MLP 008903, MLP 008904, MLP 008905, MLP 008906, MLP 008907, MLP 008908, MLP 008909, MLP 008910, MLP 008911, MLP 008912, MLP 008913, MLP 008914, MLP 008915, MLP 008916, MLP 008917, MLP 008918, MLP 008919, MLP 008920, MLP 008921, MLP 008922, MLP 008923, MLP 008924, MLP 008925, MLP 008927, MLP 008928, MLP 008929, MLP 008930, MLP 008931, MLP 008932, MLP 008933, MLP 008934, MLP 008935, MLP 008936, MLP 008937, MLP 008939, MLP 008960, MLP 008961, MLP 008962, MLP 008963, MLP 008964, MLP 008965, MLP 008966, MLP 008967, MLP 008968, MLP 008969, MLP 008970, MLP 008971, MLP 008972, MLP 008973, MLP 008974, MLP 008975, MLP 008976, MLP 008977, MLP 008978, MLP 008979, MLP 008980, MLP 008981, MLP 008982, MLP 008983, MLP 008984, MLP 008985, MLP 008986, MLP 008987, MLP 008988, MLP 008989, MLP 008990, MLP 008991, MLP 008992, MLP 008993, MLP 008994, MLP 008995, MLP 008996, MLP 008997, MLP 008998, MLP 009000,
MLP 009001, MLP 009002, MLP 009003, MLP 009004, MLP 009005, MLP 009006, MLP 009007, MLP 009008, MLP 009009, MLP 009010, MLP 009011, MLP 009012, MLP 009013, MLP 009032, MLP 009043, MLP 009044, MLP 009045, MLP 009046, MLP 009047, MLP 009048, MLP 009049, MLP 009050, MLP 009051, MLP 009052, MLP 009053, MLP 009054, MLP 009060, MLP 009061, MLP 009062, MLP 009063, MLP 009064, MLP 009066, MLP 009067, MLP 009068, MLP 009069, MLP 009070, MLP 009071, MLP 009072, MLP 009073, MLP 009074, MLP 009075, MLP 009076, MLP 009077, MLP 009078, MLP 009079, MLP 009080, MLP 009081, MLP 009082, MLP 009084, MLP 009086, MLP 009087, MLP 009088, MLP 009089, MLP 009090, MLP 009091, MLP 009092, MLP 009093, MLP 009094, MLP 009095, MLP 009096, MLP 009097, MLP 009102, MLP 009105, MLP 009108, MLP 009109, MLP 009110, MLP 009111, MLP 009122, MLP 009123, MLP 009124, MLP 009125, MLP 009126, MLP 009127, MLP 009128, MLP 009129, MLP 009130, MLP 009131, MLP 009132, MLP 009133, MLP 009134, MLP 009135, MLP 009136, MLP 009137, MLP 009152, MLP 009153, MLP 009155, MLP 009219, MLP 009227, MLP 009228, MLP 009259, MLP 009261, MLP 009262, MLP 009265, MLP 009266, MLP 009268, MLP 009269, MLP 009270, MLP 009272, MLP 009273, MLP 009274, MLP 009275, MLP 009276, MLP 009277, MLP 009278, MLP 009279, MLP 009280, MLP 009282, MLP 009283, MLP 009284, MLP 009285, MLP 009286, MLP 009287, MLP 009288, MLP 009289, MLP 009291, MLP 009292, MLP 009294, MLP 009305, MLP 009308, MLP 009309, MLP 009310, MLP 009311, MLP 009312, MLP 009313, MLP 009329, MLP 009330, MLP 009331, MLP 009332, MLP 009333, MLP 009334, MLP 009335, MLP 009336, MLP 009337, MLP 009338, MLP 009339, MLP 009340, MLP 009341, MLP 009342, MLP 009343, MLP 009354, MLP 009355, MLP 009356, MLP 009357, MLP 009358, MLP 009359, MLP 009360, MLP 009361, MLP 009364, MLP 009365, MLP 009367, MLP 009368, MLP 009369, MLP 009370, MLP 009389, MLP 009390, MLP 009391, MLP 009393, MLP 009395, MLP 009396, MLP 009398, MLP 009399, MLP 009401, MLP 009403, MLP 009408, MLP 009412, MLP 009413, MLP 009414, MLP 009416, MLP 009417, MLP 009418, MLP 009419, MLP 009422, MLP 009423, MLP 009425, MLP 009426, MLP 009427, MLP 009429, MLP 009433, MLP 009436, MLP 009437, MLP 009438, MLP 009439, MLP 009440, MLP 009442, MLP 009443, MLP 009444, MLP 009446, MLP 009447, MLP 009448, MLP 009449, MLP 009450, MLP 009451, MLP 009452, MLP 009453, MLP 009454, MLP 009455, MLP 009456, MLP 009478, MLP 009479, MLP 009480, MLP 009481, MLP 009482, MLP 009483, MLP 009484, MLP 009486, MLP 009488, MLP 009491, MLP 009492, MLP 009506, MLP 009547, MLP 009551, MLP 009568, MLP 009595, MLP 009596, MLP 009597, MLP 009613, MLP 009614, MLP 009616, MLP 009630, MLP 009637, MLP 009642, MLP 009647, MLP 009648, MLP 009649, MLP 009650, MLP 009651, MLP 009652, MLP 009654, MLP 009655, MLP 009656, MLP 009712, MLP 009715, MLP 009716, MLP 009718, MLP 009732, MLP 009733, MLP 009734, MLP 009735, MLP 009736, MLP 009737, MLP 009740, MLP 009741, MLP 009742, MLP 009743, MLP 009744, MLP 009747, MLP 009748, MLP 009749, MLP 009750, MLP 009751, MLP 009752, MLP 009753, MLP 009754, MLP 009755, MLP 009756, MLP 009757, MLP 009758, MLP 009760, MLP 009761, MLP 009762, MLP 009763, MLP 009764, MLP 009765, MLP 009766, MLP 009767, MLP 009768, MLP 009769, MLP 009771, MLP 009772, MLP 009774, MLP 009775, MLP 009776, MLP 009777, MLP 009778, MLP 009779, MLP 009780, MLP 009781, MLP 009782, MLP 009783, MLP 009785, MLP 009793, MLP 009795, MLP 009796, MLP 009797, MLP 009798, MLP 009799, MLP 009802, MLP 009803, MLP 009804, MLP 009807, MLP 009812, MLP 009813, MLP 009814, MLP 009816, MLP 009818, MLP 009819, Serial Numbers Outside of the U.S. MLP 008324, MLP 002570, MLP 002574, MLP 002582, MLP 002601, MLP 002606, MLP 002681, MLP 002683, MLP 002820, MLP 002826, MLP 005163, MLP 002918, MLP 002930, MLP 003020, MLP 007366, MLP 003172, MLP 003182, MLP 003166, MLP 003198, MLP 003404, MLP 003494, MLP 003538, MLP 007368, MLP 007369, MLP 003609, MLP 007370, MLP 003752, MLP 003763, MLP 003715, MLP 003810, MLP 003813, MLP 003888, MLP 004169, MLP 007371, MLP 007373, MLP 003891, MLP 003893, MLP 003899, MLP 003921, MLP 003988, MLP 004052, MLP 004053, MLP 004065, MLP 004066, MLP 004067, MLP 004076, MLP 004107, MLP 007374, MLP 004272, MLP 004351, MLP 004361, MLP 004364, MLP 004373, MLP 004375, MLP 004379, MLP 004269, MLP 004408, MLP 004442, MLP 004455, MLP 004464, MLP 007489, MLP 004515, MLP 004550, MLP 004573, MLP 004630, MLP 004634, MLP 008212, MLP 004750, MLP 004769, MLP 008495, MLP 004816, MLP 004856, MLP 008788, MLP 008863, MLP 008867, MLP 004912, MLP 004914, MLP 008942, MLP 008946, MLP 004945, MLP 004946, MLP 009024, MLP 004975, MLP 004989, MLP 004990, MLP 004992, MLP 005003, MLP 009229, MLP 005016, MLP 005018, MLP 009230, MLP 005038, MLP 009231, MLP 009477, MLP 005112, MLP 005162, MLP 005167, MLP 005203, MLP 005234, MLP 005310, MLP 005740, MLP 005920, MLP 005983, MLP 006077, MLP 006109, MLP 006498, MLP 006724, MLP 006786, MLP 006809, MLP 006821, MLP 006839, MLP 007299, MLP 007318, MLP 007348, MLP 007367, MLP 007518, MLP 007526, MLP 007530, MLP 007723, MLP 007952, MLP 008015, MLP 008789, MLP 008855, MLP 008857, MLP 008858, MLP 009464, MLP 002437, MLP 007490, MLP 007532, MLP 002768, MLP 002777, MLP 002778, MLP 002786, MLP 002787, MLP 002791, MLP 002916, MLP 002922, MLP 002926, MLP 002928, MLP 002929, MLP 002931, MLP 002932, MLP 002938, MLP 002940, MLP 003002, MLP 003006, MLP 003009, MLP 003034, MLP 003138, MLP 003159, MLP 003212, MLP 003250, MLP 003264, MLP 003305, MLP 003306, MLP 003412, MLP 003414, MLP 003415, MLP 003418, MLP 003497, MLP 003514, MLP 003517, MLP 003521, MLP 003537, MLP 003545, MLP 003600, MLP 003611, MLP 003722, MLP 003723, MLP 003724, MLP 003753, MLP 003758, MLP 003766, MLP 003772, MLP 003774, MLP 003775, MLP 003780, MLP 003782, MLP 003735, MLP 003834, MLP 003836, MLP 003837, MLP 003990, MLP 004055, MLP 004070, MLP 004102, MLP 004163, MLP 004335, MLP 004343, MLP 004347, MLP 004367, MLP 004388, MLP 004394, MLP 004401, MLP 004402, MLP 004403, MLP 004406, MLP 004428, MLP 004431, MLP 004433, MLP 004566, MLP 004569, MLP 004570, MLP 004576, MLP 004607, MLP 004608, MLP 004609, MLP 004610, MLP 004612, MLP 004613, MLP 004616, MLP 004618, MLP 004619, MLP 004626, MLP 004638, MLP 004644, MLP 004739, MLP 004740, MLP 004741, MLP 004744, MLP 004746, MLP 004748, MLP 004749, MLP 004925, MLP 005062, MLP 005103, MLP 005125, MLP 005126, MLP 005133, MLP 005161, MLP 005106, MLP 005165, MLP 005168, MLP 005170, MLP 005172, MLP 005173, MLP 006031, MLP 006598, MLP 007531, MLP 007697, MLP 005474, MLP 007745, MLP 005504, MLP 005511, MLP 005582, MLP 005583, MLP 005584, MLP 007790, MLP 005629, MLP 005631, MLP 005633, MLP 005634, MLP 005635, MLP 005645, MLP 007971, MLP 007978, MLP 008669, MLP 008671, MLP 008672, MLP 008783, MLP 005712, MLP 005714, MLP 005715, MLP 005716, MLP 008785, MLP 008852, MLP 005756, MLP 005757, MLP 005759, MLP 005882, MLP 005886, MLP 005891, MLP 005894, MLP 005895, MLP 008866, MLP 008943, MLP 006026, MLP 006030, MLP 006032, MLP 008944, MLP 006057, MLP 006059, MLP 008945, MLP 008947, MLP 009021, MLP 009022, MLP 006230, MLP 009217, MLP 009563, MLP 006457, MLP 009658, MLP 006473, MLP 006474, MLP 006517, MLP 006526, MLP 006531, MLP 009725, MLP 006532, MLP 009726, MLP 009728, MLP 006600, MLP 006601, MLP 006606, MLP 009888, MLP 006650, MLP 006651, MLP 006653, MLP 006654, MLP 006665, MLP 009954, MLP 009958, MLP 010086, MLP 006728, MLP 006731, MLP 006732, MLP 006740, MLP 006745, MLP 006747, MLP 006750, MLP 006753, MLP 006757, MLP 006814, MLP 006815, MLP 006816, MLP 006823, MLP 007692, MLP 007693, MLP 007698, MLP 007699, MLP 007721, MLP 007722, MLP 007724, MLP 007728, MLP 007736, MLP 007741, MLP 007742, MLP 007744, MLP 007782, MLP 007789, MLP 007817, MLP 008438, MLP 008472, MLP 008478, MLP 008481, MLP 008482, MLP 008483, MLP 008500, MLP 008670, MLP 008673, MLP 008950, MLP 008951, MLP 008957, MLP 002199, MLP 002428, MLP 002443, MLP 002623, MLP 002691, MLP 002704, MLP 002705, MLP 002749, MLP 003008, MLP 003121, MLP 003144, MLP 003284, MLP 003501, MLP 003534, MLP 003716, MLP 003764, MLP 003769, MLP 003846, MLP 004126, MLP 004206, MLP 004209, MLP 004345, MLP 004360, MLP 004542, MLP 004621, MLP 004650, MLP 004664, MLP 004735, MLP 005064, MLP 005066, MLP 005073, MLP 005135, MLP 005316, MLP 005369, MLP 005691, MLP 005762, MLP 005994, MLP 006112, MLP 006567, MLP 006599, MLP 006667, MLP 006769, MLP 006788, MLP 007267, MLP 007286, MLP 007495, MLP 007850, MLP 007972, MLP 008379, MLP 009348, MLP 007314, MLP 007649, MLP 007961, MLP 008371, MLP 008374, MLP 008665, MLP 009020, MLP 009235, MLP 010085, MLP 006475, MLP 009462, MLP 003052, MLP 003403, MLP 003486, MLP 003489, MLP 003498, MLP 003546, MLP 003942, MLP 003873, MLP 003874, MLP 003927, MLP 004074, MLP 004204, MLP 004372, MLP 004425, MLP 004445, MLP 004488, MLP 004536, MLP 004596, MLP 004657, MLP 004662, MLP 004732, MLP 004852, MLP 004935, MLP 004943, MLP 005059, MLP 005067, MLP 005070, MLP 005092, MLP 005099, MLP 005314, MLP 005657, MLP 005915, MLP 005921, MLP 006439, MLP 006444, MLP 006501, MLP 006748, MLP 006834, MLP 007236, MLP 005507, MLP 006021, MLP 006683, MLP 007727, MLP 007321, MLP 007787, MLP 007355, MLP 007820, MLP 007846, MLP 008491, MLP 008638, MLP 008674, MLP 007500, MLP 008183, MLP 009377, MLP 007700, MLP 007740, MLP 007748, MLP 007859, MLP 008017, MLP 008039, MLP 008101, MLP 008213, MLP 008322, MLP 008631, MLP 008786, MLP 008864, MLP 009115, MLP 010214, MLP 010016, MLP 003976, MLP 004001, MLP 004003, MLP 004071, MLP 004458, MLP 004475, MLP 004752, MLP 004756, MLP 004781, MLP 005071, MLP 005282, MLP 005431, MLP 006058, MLP 006270, MLP 006835, MLP 006836, MLP 007353, MLP 007618, MLP 007653, MLP 008226, MLP 007593, MLP 008493, MLP 008955, MLP 007892, MLP 007976, MLP 007982, MLP 009027, MLP 009098, MLP 005061, MLP 005434, MLP 007955, MLP 005042, MLP 005207, MLP 005312, MLP 005909, MLP 005985, MLP 007492, MLP 007595, MLP 007621, MLP 007981, MLP 008094, MLP 008435, MLP 008451, MLP 008452, MLP 008497, MLP 008602, MLP 007644 MLP 010076, MLP 010078, MLP 007737, MLP 007739, MLP 008146, MLP 008439, MLP 009963, MLP 010145, MLP 010254, MLP 002489, MLP 002517, MLP 002668, MLP 003289, MLP 003419, MLP 003696, MLP 003877, MLP 003879, MLP 003923, MLP 004002, MLP 004261, MLP 004426, MLP 004547, MLP 004723, MLP 004724, MLP 005383, MLP 005960, MLP 006022, MLP 006442, MLP 006694, MLP 006789, MLP 007274, MLP 003161, MLP 006691, MLP 007285, MLP 008152, MLP 009435, MLP 007259, MLP 007272, MLP 007344, MLP 009117, MLP 009215, MLP 008634, MLP 002701, MLP 002914, MLP 002923, MLP 002924, MLP 002927, MLP 002947, MLP 003732, MLP 003840, MLP 003733, MLP 003838, MLP 003842, MLP 003844, MLP 004306, MLP 004357, MLP 004363, MLP 004383, MLP 004793, MLP 004794, MLP 005796, MLP 005923, MLP 005940, MLP 005944, MLP 005945, MLP 005952, MLP 006260, MLP 006261, MLP 006264, MLP 006274, MLP 006376, MLP 006391, MLP 007255, MLP 007650, MLP 005943, MLP 007851, MLP 006232, MLP 006368, MLP 007856, MLP 007957, MLP 008020, MLP 007954, MLP 007966, MLP 007967, MLP 007968, MLP 007973, MLP 008041, MLP 008145, MLP 008148, MLP 008042, MLP 008293, MLP 009950, MLP 010074, MLP 010075, MLP 008291, MLP 008143, MLP 008184, MLP 008214, MLP 005846, MLP 006530, MLP 006543, MLP 005246, MLP 005308, MLP 005317, MLP 005451, MLP 002682, MLP 002821, MLP 003407, MLP 003824, MLP 004063, MLP 004084, MLP 004180, MLP 004187, MLP 004365, MLP 004371 MLP 004811, MLP 004896, MLP 004926, MLP 004927, MLP 004931, MLP 004933, MLP 005176, MLP 005199, MLP 005271, MLP 005760, MLP 005799, MLP 005801, MLP 005884, MLP 005989, MLP 005995, MLP 006015, MLP 006105, MLP 006235, MLP 006385, MLP 006399, MLP 006497, MLP 006559, MLP 006743, MLP 006837, MLP 007294, MLP 007305, MLP 007616, MLP 007646, MLP 007893, MLP 007894, MLP 007951, MLP 008097, MLP 008320, MLP 008431, MLP 003900, MLP 008437, MLP 009029, MLP 009218, MLP 009434, MLP 009458, MLP 008632, MLP 008848, MLP 009239, MLP 009243, MLP 009254, MLP 009376, MLP 009463, MLP 010144, MLP 002595, MLP 002604, MLP 002619, MLP 002693, MLP 002732, MLP 002784, MLP 002935, MLP 002939, MLP 003112, MLP 003116, MLP 003452, MLP 003503, MLP 003504, MLP 003147, MLP 003913, MLP 003938, MLP 003973, MLP 003980, MLP 003981, MLP 003984, MLP 003996, MLP 004062, MLP 004080, MLP 004094, MLP 004225, MLP 004356, MLP 004377, MLP 004438, MLP 004449, MLP 004484, MLP 004501, MLP 004512, MLP 004709, MLP 004713, MLP 004720, MLP 004721, MLP 004722, MLP 004734, MLP 004736, MLP 004742, MLP 004812, MLP 004851, MLP 004902, MLP 004922, MLP 004953, MLP 004988, MLP 005044, MLP 005045, MLP 005109, MLP 005134, MLP 005159, MLP 005171, MLP 005386, MLP 005388, MLP 005589, MLP 005624, MLP 005628, MLP 005689, MLP 005802, MLP 005844, MLP 005869, MLP 005907, MLP 006382, MLP 006392, MLP 006395, MLP 006402, MLP 006549, MLP 006613, MLP 006626, MLP 006666, MLP 006692, MLP 006723, MLP 006758, MLP 006763, MLP 006787, MLP 006796, MLP 006797, MLP 006818, MLP 006822, MLP 007243, MLP 007284, MLP 007360, MLP 007486, MLP 007496, MLP 007580, MLP 007600, MLP 007730, MLP 007733, MLP 007735, MLP 007842, MLP 007844, MLP 007849, MLP 007861, MLP 007949, MLP 007956, MLP 008211, MLP 002915, MLP 007300, MLP 007845, MLP 007891, MLP 008433, MLP 008456, MLP 008473, MLP 008479, MLP 008480, MLP 008492, MLP 008603, MLP 009621, MLP 009730, MLP 008598, MLP 002003, MLP 002010, MLP 002016, MLP 002018, MLP 002024, MLP 002026, MLP 002032, MLP 002037, MLP 002039, MLP 002095, MLP 002097, MLP 002098, MLP 002100, MLP 002101, MLP 002103, MLP 002104, MLP 002105, MLP 002106, MLP 002110, MLP 002111, MLP 002113, MLP 002114, MLP 002115, MLP 002116, MLP 002122, MLP 002143, MLP 002145, MLP 002146, MLP 002147, MLP 002149, MLP 002150, MLP 002154, MLP 002190, MLP 002194, MLP 002195, MLP 002196, MLP 002210, MLP 002230, MLP 002322, MLP 002323, MLP 002328, MLP 002608, MLP 002689, MLP 002735, MLP 002736, MLP 002759, MLP 002765, MLP 002767, MLP 002771, MLP 002776, MLP 002917, MLP 002945, MLP 002950, MLP 003411, MLP 003734, MLP 003916, MLP 004073, MLP 004245, MLP 004248, MLP 004287, MLP 004302, MLP 004330, MLP 004333, MLP 004358, MLP 004460, MLP 004467, MLP 004478, MLP 004533, MLP 004557, MLP 004560, MLP 004563, MLP 004976, MLP 004980, MLP 004982, MLP 004984, MLP 004985, MLP 005065, MLP 005424, MLP 005656, MLP 005658, MLP 005973, MLP 005998, MLP 006000, MLP 006114, MLP 006117, MLP 006639, MLP 006640, MLP 006643, MLP 006644, MLP 006911, MLP 006912, MLP 007491, MLP 007525, MLP 007592, MLP 007725, MLP 007726, MLP 007857, MLP 007858, MLP 007960, MLP 007964, MLP 008036, MLP 008037, MLP 008225, MLP 008316, MLP 008317, MLP 004544, MLP 004653, MLP 005022, MLP 005040, MLP 005984, MLP 006400, MLP 006645, MLP 006717, MLP 009211, MLP 009233, MLP 009559, MLP 009622, MLP 009886, MLP 010081, MLP 010087, MLP 009465, MLP 009557, MLP 002602, MLP 002620, MLP 002625, MLP 002731, MLP 002750, MLP 002752, MLP 002754, MLP 003143, MLP 003410, MLP 003792, MLP 003806, MLP 003725, MLP 003729, MLP 003730, MLP 004380, MLP 003931, MLP 003974, MLP 003978, MLP 003994, MLP 004060, MLP 004089, MLP 004092, MLP 004116, MLP 004178, MLP 004188, MLP 004191, MLP 004192, MLP 004195, MLP 004196, MLP 004198, MLP 004199, MLP 004205, MLP 004214, MLP 004240, MLP 004319, MLP 004325, MLP 004328, MLP 004334, MLP 004362, MLP 004374, MLP 004437, MLP 004468, MLP 004574, MLP 004599, MLP 004600, MLP 004614, MLP 004660, MLP 004661, MLP 004712, MLP 004718, MLP 004787, MLP 004904, MLP 004906, MLP 004907, MLP 004909, MLP 005041, MLP 005046, MLP 005058, MLP 005063, MLP 005068, MLP 005137, MLP 005149, MLP 005151, MLP 005153, MLP 005204, MLP 005233, MLP 005235, MLP 005244, MLP 005370, MLP 005371, MLP 005376, MLP 005423, MLP 005452, MLP 005522, MLP 005523, MLP 005727, MLP 005787, MLP 005788, MLP 005789, MLP 005790, MLP 005867, MLP 005868, MLP 005880, MLP 005881, MLP 005885, MLP 006210, MLP 006284, MLP 006370, MLP 006373, MLP 006377, MLP 006381, MLP 006456, MLP 006476, MLP 006529, MLP 006551, MLP 006573, MLP 006576, MLP 006604, MLP 006608, MLP 006646, MLP 006668, MLP 006669, MLP 006760, MLP 006762, MLP 007263, MLP 007269, MLP 007271, MLP 007273, MLP 007278, MLP 007291, MLP 007349, MLP 007354, MLP 007356, MLP 007524, MLP 008058, MLP 008064, MLP 008067, MLP 008099, MLP 007372, MLP 007979, MLP 007980, MLP 008063, MLP 008144, MLP 009552, MLP 009553, MLP 009554, MLP 009555, MLP 009575, MLP 009618, MLP 009724, MLP 009727, MLP 004255, MLP 004659, MLP 004747, MLP 004745, MLP 005047, MLP 005060, MLP 005509, MLP 005519, MLP 005521, MLP 005526, MLP 005538, MLP 005572, MLP 006722, MLP 006746, MLP 008536, MLP 002568, MLP 002569, MLP 002645, MLP 002697, MLP 002700, MLP 002739, MLP 002744, MLP 002773, MLP 002779, MLP 002792, MLP 002862, MLP 002865, MLP 002866, MLP 002868, MLP 002874, MLP 003011, MLP 003023, MLP 003036, MLP 003045, MLP 003119, MLP 003157, MLP 003158, MLP 003156, MLP 003181, MLP 003444, MLP 003449, MLP 003450, MLP 003405, MLP 003533, MLP 003719, MLP 003720, MLP 003605, MLP 003712, MLP 003778, MLP 003789, MLP 003801, MLP 003815, MLP 003819, MLP 003845, MLP 003849, MLP 003854, MLP 003895, MLP 003940, MLP 004054, MLP 004061, MLP 004138, MLP 004146, MLP 004167, MLP 004186, MLP 004229, MLP 004232, MLP 004263, MLP 004320, MLP 004332, MLP 004341, MLP 004342, MLP 004344, MLP 004389, MLP 004393, MLP 004396, MLP 004398, MLP 004410, MLP 004481, MLP 004495, MLP 004496, MLP 004519, MLP 004535, MLP 004578, MLP 004631, MLP 004633, MLP 004714, MLP 004719, MLP 004766, MLP 004786, MLP 004790, MLP 004809, MLP 004820, MLP 004858, MLP 004895, MLP 004897, MLP 004905, MLP 004977, MLP 004979, MLP 004981, MLP 004986, MLP 005108, MLP 005111, MLP 005206, MLP 005270, MLP 005300, MLP 005309, MLP 005318, MLP 005367, MLP 005428, MLP 005458, MLP 005565, MLP 005587, MLP 005694, MLP 005710, MLP 005713, MLP 005726, MLP 005738, MLP 005908, MLP 005922, MLP 005981, MLP 005986, MLP 006009, MLP 006013, MLP 006019, MLP 006020, MLP 006060, MLP 006211, MLP 006251, MLP 006252, MLP 006257, MLP 006378, MLP 006445, MLP 006495, MLP 006500, MLP 006502, MLP 006609, MLP 006618, MLP 006649, MLP 006681, MLP 006693, MLP 006703, MLP 006772, MLP 006877, MLP 007240, MLP 007242, MLP 007275, MLP 007320, MLP 007483, MLP 007514, MLP 007519, MLP 007596, MLP 007597, MLP 007619, MLP 007731, MLP 007864, MLP 008016, MLP 008181, MLP 008185, MLP 008496, MLP 008596, MLP 008601, MLP 008629, MLP 008639, MLP 008644, MLP 006785, MLP 007694, MLP 008666, MLP 008851, MLP 008956, MLP 009237, MLP 009346, MLP 010084, MLP 004629, MLP 006614, MLP 007617, MLP 008952, MLP 008436, MLP 008489, MLP 009118, MLP 009015, MLP 009722, MLP 009147, MLP 009378, MLP 009721, MLP 005286, MLP 006066, MLP 007651, MLP 008190, MLP 008325, MLP 009242, MLP 009249, MLP 009574, MLP 009576, MLP 009946, MLP 009983, MLP 010092, MLP 010216, MLP 002772, MLP 002925, MLP 003016, MLP 003019, MLP 003021, MLP 003022, MLP 003030, MLP 003059, MLP 003093, MLP 003136, MLP 003141, MLP 003785, MLP 003788, MLP 003790, MLP 003791, MLP 003794, MLP 003795, MLP 003797, MLP 003798, MLP 003805, MLP 003807, MLP 003808, MLP 003809, MLP 004324, MLP 004948, MLP 004950, MLP 004993, MLP 005020, MLP 005024, MLP 005166, MLP 005175, MLP 005200, MLP 005281, MLP 005283, MLP 005284, MLP 005375, MLP 005524, MLP 005536, MLP 005567, MLP 005569, MLP 005647, MLP 005755, MLP 005855, MLP 005887, MLP 005889, MLP 005892, MLP 005988, MLP 006012, MLP 006017, MLP 006018, MLP 006494, MLP 006805, MLP 006879, MLP 006885, MLP 007283, MLP 007292, MLP 007293, MLP 007295, MLP 007302, MLP 007303, MLP 007323, MLP 007334, MLP 007336, MLP 007853, MLP 008038, MLP 008292, MLP 008232, MLP 009469, MLP 009474, MLP 009562, MLP 009580, MLP 009583, MLP 009951, MLP 009955, MLP 009964, MLP 009965, MLP 010080, MLP 003409, MLP 003761, MLP 003762, MLP 004069, MLP 004086, MLP 004087, MLP 004422, MLP 003406, MLP 005077, MLP 005096, MLP 005644, MLP 005723, MLP 005734, MLP 005897, MLP 007746, MLP 008040, MLP 007963, MLP 008454, MLP 008494, MLP 007317, MLP 005898, MLP 007529, MLP 006383, MLP 006367, MLP 000975, MLP 000997, MLP 000998, MLP 002000, MLP 002001, MLP 002013, MLP 002019, MLP 002022, MLP 002023, MLP 002025, MLP 002027, MLP 002031, MLP 002044, MLP 002046, MLP 002047, MLP 002052, MLP 002060, MLP 002061, MLP 002062, MLP 002063, MLP 002067, MLP 002092, MLP 002138, MLP 002139, MLP 002186, MLP 002212, MLP 002262, MLP 002313, MLP 002325, MLP 002326, MLP 002332, MLP 002333, MLP 002335, MLP 002351, MLP 002423, MLP 002425, MLP 002426, MLP 002429, MLP 002430, MLP 002433, MLP 002436, MLP 002438, MLP 002440, MLP 002441, MLP 002444, MLP 002445, MLP 002446, MLP 002458, MLP 002459, MLP 002460, MLP 002461, MLP 002463, MLP 002464, MLP 002465, MLP 002466, MLP 002515, MLP 002527, MLP 002528, MLP 002567, MLP 002572, MLP 002573, MLP 002575, MLP 002576, MLP 002577, MLP 002599, MLP 002605, MLP 002615, MLP 002622, MLP 002626, MLP 002627, MLP 002630, MLP 002631, MLP 002642, MLP 002667, MLP 002670, MLP 002672, MLP 002674, MLP 002678, MLP 002685, MLP 002687, MLP 002688, MLP 002690, MLP 002696, MLP 002699, MLP 002698, MLP 002702, MLP 002703, MLP 002706, MLP 002730, MLP 002733, MLP 002734, MLP 002741, MLP 002743, MLP 002738, MLP 002746, MLP 002747, MLP 002755, MLP 002757, MLP 002760, MLP 002761, MLP 002763, MLP 002764, MLP 002766, MLP 002762, MLP 002769, MLP 002770, MLP 002774, MLP 002775, MLP 002781, MLP 002782, MLP 002783, MLP 002785, MLP 002789, MLP 002793, MLP 002794, MLP 002818, MLP 002819, MLP 002822, MLP 002823, MLP 002824, MLP 002869, MLP 002871, MLP 002872, MLP 002867, MLP 002876, MLP 002913, MLP 002919, MLP 002941, MLP 002942, MLP 002943, MLP 002944, MLP 002936, MLP 002946, MLP 002948, MLP 002949, MLP 002951, MLP 002964, MLP 003001, MLP 003003, MLP 003005, MLP 003010, MLP 003012, MLP 003013, MLP 003014, MLP 003015, MLP 003029, MLP 003031, MLP 003035, MLP 003041, MLP 003042, MLP 003046, MLP 003047, MLP 003038, MLP 003048, MLP 003049, MLP 003051, MLP 003055, MLP 003061, MLP 003062, MLP 003063, MLP 003066, MLP 003076, MLP 003079, MLP 003080, MLP 003081, MLP 003084, MLP 003087, MLP 003090, MLP 003091, MLP 003088, MLP 003103, MLP 003092, MLP 003106, MLP 003094, MLP 003108, MLP 003115, MLP 003117, MLP 003120, MLP 003123, MLP 003113, MLP 003126, MLP 003127, MLP 003133, MLP 003134, MLP 003137, MLP 003142, MLP 003149, MLP 003150, MLP 003154, MLP 003155, MLP 003165, MLP 003152, MLP 003170, MLP 003171, MLP 003174, MLP 003175, MLP 003160, MLP 003168, MLP 003173, MLP 003192, MLP 003265, MLP 003272, MLP 003249, MLP 003287, MLP 003313, MLP 003301, MLP 003400, MLP 003401, MLP 003399, MLP 003451, MLP 003454, MLP 003417, MLP 003420, MLP 003421, MLP 003423, MLP 003445, MLP 003485, MLP 003453, MLP 003487, MLP 003488, MLP 003492, MLP 003496, MLP 003499, MLP 003502, MLP 003505, MLP 003507, MLP 003508, MLP 003510, MLP 003511, MLP 003512, MLP 003513 MLP 003515, MLP 003523, MLP 003524, MLP 003526, MLP 003528, MLP 003490, MLP 003529, MLP 003491, MLP 003531, MLP 003532, MLP 003535, MLP 003536, MLP 003539, MLP 003540, MLP 003500, MLP 003541, MLP 003542, MLP 003543, MLP 003547, MLP 003548, MLP 003506, MLP 003552, MLP 003585, MLP 003527, MLP 003530, MLP 003595, MLP 003599, MLP 003601, MLP 003544, MLP 003670, MLP 003701, MLP 003603, MLP 003607, MLP 003750, MLP 003754, MLP 003767, MLP 003768, MLP 003776, MLP 003779, MLP 003781, MLP 003786, MLP 003707, MLP 003708, MLP 003711, MLP 003796, MLP 003713, MLP 003718, MLP 003721, MLP 003822, MLP 003727, MLP 003728, MLP 003731, MLP 003749, MLP 003751, MLP 003756, MLP 003757, MLP 003760, MLP 003765, MLP 003773, MLP 003787, MLP 003793, MLP 003800, MLP 003802, MLP 003812, MLP 003817, MLP 003818, MLP 003820, MLP 003821, MLP 003823, MLP 003839, MLP 003841, MLP 003843, MLP 003848, MLP 003850, MLP 003853, MLP 004409, MLP 004432, MLP 003855, MLP 004606, MLP 004777, MLP 005692, MLP 005863, MLP 003868, MLP 005905, MLP 003869, MLP 003872, MLP 006795, MLP 003875, MLP 007238, MLP 003880, MLP 003884, MLP 003885, MLP 003887, MLP 003890, MLP 003894, MLP 003897, MLP 003902, MLP 003905, MLP 003906, MLP 003908, MLP 003909, MLP 003911, MLP 003912, MLP 003914, MLP 003915, MLP 003917, MLP 003919, MLP 003920, MLP 003922, MLP 003924, MLP 003926, MLP 003929, MLP 003930, MLP 003934, MLP 003935, MLP 003936, MLP 003937, MLP 003939, MLP 003971, MLP 003972, MLP 003975, MLP 003977, MLP 003979, MLP 003982, MLP 003983, MLP 003987, MLP 003991, MLP 003992, MLP 003993, MLP 003998, MLP 004000, MLP 004004, MLP 004007, MLP 004008, MLP 004051, MLP 004056, MLP 004057, MLP 004058, MLP 004059, MLP 004072, MLP 004078, MLP 004079, MLP 004083, MLP 004085, MLP 004090, MLP 004106, MLP 004111, MLP 004113, MLP 004120, MLP 004121, MLP 004123, MLP 004132, MLP 004134, MLP 004141, MLP 004164, MLP 004168, MLP 004177, MLP 004182, MLP 004183, MLP 004184, MLP 004189, MLP 004190, MLP 004193, MLP 004201, MLP 004202, MLP 004203, MLP 004207, MLP 004211, MLP 004215, MLP 004217, MLP 004223, MLP 004224, MLP 004226, MLP 004230, MLP 004243, MLP 004244, MLP 004270, MLP 004288, MLP 004295, MLP 004313, MLP 004314, MLP 004315, MLP 004318, MLP 004322, MLP 004323, MLP 004327, MLP 004329, MLP 004331, MLP 004337, MLP 004338, MLP 004339, MLP 004348, MLP 004350, MLP 004352, MLP 004354, MLP 004368, MLP 004381, MLP 004382, MLP 004387, MLP 004390, MLP 004395, MLP 004397, MLP 004404, MLP 004413, MLP 004415, MLP 004417, MLP 004418, MLP 004419, MLP 004420, MLP 004421, MLP 004423, MLP 004424, MLP 004427, MLP 004429, MLP 004434, MLP 004435, MLP 004436, MLP 004439, MLP 004440, MLP 004441, MLP 004448, MLP 004450, MLP 004451, MLP 004452, MLP 004453, MLP 004459, MLP 004461, MLP 004472, MLP 004477, MLP 004487, MLP 004497, MLP 004499, MLP 004503, MLP 004504, MLP 004506, MLP 004508, MLP 004511, MLP 004518, MLP 004521, MLP 004526, MLP 004534, MLP 004537, MLP 004539, MLP 004540, MLP 004551, MLP 004552, MLP 004555, MLP 004556, MLP 004558, MLP 004559, MLP 004562, MLP 004567, MLP 004568, MLP 004571, MLP 004572, MLP 004575, MLP 004579, MLP 004580, MLP 004581, MLP 004582, MLP 004593, MLP 004594, MLP 004595, MLP 004597, MLP 004598, MLP 004601, MLP 004602, MLP 004603, MLP 004604, MLP 004605, MLP 004611, MLP 004615, MLP 004617, MLP 004620, MLP 004622, MLP 004623, MLP 004627, MLP 004628, MLP 004632, MLP 004635, MLP 004636, MLP 004637, MLP 004639, MLP 004641, MLP 004642, MLP 004643, MLP 004645, MLP 004646, MLP 004647, MLP 004648, MLP 004649, MLP 004651, MLP 004652, MLP 004655, MLP 004656, MLP 004663, MLP 004690, MLP 004710, MLP 004711, MLP 004715, MLP 004716, MLP 004717, MLP 004728, MLP 004729, MLP 004730, MLP 004731, MLP 004737, MLP 004738, MLP 004743, MLP 004751, MLP 004753, MLP 004755, MLP 004759, MLP 004760, MLP 004761, MLP 004767, MLP 004768, MLP 004770, MLP 004772, MLP 004773, MLP 004774, MLP 004775, MLP 004776, MLP 004778, MLP 004779, MLP 004780, MLP 004784, MLP 004785, MLP 004789, MLP 004796, MLP 004808, MLP 004810, MLP 004814, MLP 004817, MLP 004821, MLP 004822, MLP 004823, MLP 004849, MLP 004854, MLP 004859, MLP 004860, MLP 004862, MLP 004883, MLP 004898, MLP 004899, MLP 004901, MLP 004910, MLP 004915, MLP 004916, MLP 004917, MLP 004919, MLP 004921, MLP 004924, MLP 004929, MLP 004930, MLP 004932, MLP 004934, MLP 004936, MLP 004938, MLP 004939, MLP 004944, MLP 004947, MLP 004949, MLP 004855, MLP 004951, MLP 004952, MLP 004974, MLP 004978, MLP 005015, MLP 005017, MLP 005019, MLP 005021, MLP 005023, MLP 005039, MLP 005043, MLP 005074, MLP 005075, MLP 005076, MLP 005093, MLP 005094, MLP 005095, MLP 005097, MLP 005098, MLP 005100, MLP 005101, MLP 005102, MLP 005105, MLP 005110, MLP 005113, MLP 005132, MLP 005136, MLP 005138, MLP 005150, MLP 005152, MLP 005154, MLP 005156, MLP 005157, MLP 005158, MLP 005160, MLP 005164, MLP 005174, MLP 005198, MLP 005201, MLP 005205, MLP 005217, MLP 005218, MLP 005230, MLP 005231, MLP 005232, MLP 005236, MLP 005237, MLP 005238, MLP 005239, MLP 005240, MLP 005241, MLP 005242, MLP 005243, MLP 005245, MLP 005250, MLP 005251, MLP 005252, MLP 005253, MLP 005254, MLP 005255, MLP 005256, MLP 005269, MLP 005272, MLP 005273, MLP 005274, MLP 005275, MLP 005276, MLP 005278, MLP 005279, MLP 005280, MLP 005285, MLP 005287, MLP 005288, MLP 005289, MLP 005301, MLP 005302, MLP 005303, MLP 005304, MLP 005305, MLP 005306, MLP 005311, MLP 005315, MLP 005319, MLP 005368, MLP 005372, MLP 005374, MLP 005377, MLP 005378, MLP 005380, MLP 005381, MLP 005384, MLP 005387, MLP 005421, MLP 005427, MLP 005433, MLP 005435, MLP 005439, MLP 005448, MLP 005450, MLP 005456, MLP 005457, MLP 005475, MLP 005505, MLP 005506, MLP 005525, MLP 005535, MLP 005580, MLP 005581, MLP 005585, MLP 005588, MLP 005643, MLP 005659, MLP 005674, MLP 005676, MLP 005678, MLP 005679, MLP 005681, MLP 005690, MLP 005693, MLP 005719, MLP 005721, MLP 005724, MLP 005725, MLP 005733, MLP 005735, MLP 005739, MLP 005741, MLP 005780, MLP 005800, MLP 005843, MLP 005845, MLP 005850, MLP 005856, MLP 005888, MLP 005903, MLP 005904, MLP 005911, MLP 005913, MLP 005914, MLP 005918, MLP 005935, MLP 005936, MLP 005937, MLP 005938, MLP 005939, MLP 005941, MLP 005953, MLP 005957, MLP 005958, MLP 005963, MLP 005969, MLP 005972, MLP 005974, MLP 005975, MLP 005977, MLP 005978, MLP 005979, MLP 005980, MLP 005993, MLP 005996, MLP 006010, MLP 006011, MLP 006016, MLP 006024, MLP 006025, MLP 006027, MLP 006028, MLP 006029, MLP 006033, MLP 006061, MLP 006062, MLP 006065, MLP 006103, MLP 006119, MLP 006120, MLP 006121, MLP 006208, MLP 006209, MLP 006227, MLP 006234, MLP 006256, MLP 006267, MLP 006275, MLP 006282, MLP 006366, MLP 006369, MLP 006371, MLP 006372, MLP 006374, MLP 006379, MLP 006380, MLP 006393, MLP 006401, MLP 006403, MLP 006404, MLP 006405, MLP 006427, MLP 006429, MLP 006438, MLP 006447, MLP 006455, MLP 006458, MLP 006459, MLP 006472, MLP 006496, MLP 006515, MLP 006524, MLP 006525, MLP 006527, MLP 006545, MLP 006547, MLP 006548, MLP 006550, MLP 006552, MLP 006560, MLP 006561, MLP 006562, MLP 006564, MLP 006565, MLP 006566, MLP 006568, MLP 006569, MLP 006574, MLP 006575, MLP 006607, MLP 006611, MLP 006612, MLP 006616, MLP 006622, MLP 006625, MLP 006638, MLP 006641, MLP 006642, MLP 006647, MLP 006648, MLP 006652, MLP 006664, MLP 006684, MLP 006685, MLP 006697, MLP 006698, MLP 006702, MLP 006718, MLP 006720, MLP 006721, MLP 006725, MLP 006726, MLP 006729, MLP 006730, MLP 006733, MLP 006734, MLP 006738, MLP 006741, MLP 006744, MLP 006751, MLP 006754, MLP 006759, MLP 006770, MLP 006780, MLP 006781, MLP 006782, MLP 006790, MLP 006791, MLP 006792, MLP 006794, MLP 006798, MLP 006806, MLP 006807, MLP 006817, MLP 006840, MLP 006842, MLP 006871, MLP 006872, MLP 006878, MLP 006881, MLP 006882, MLP 006883, MLP 006908, MLP 006910, MLP 006913, MLP 006914, MLP 007232, MLP 007233, MLP 007234, MLP 007235, MLP 007239, MLP 007241, MLP 007244, MLP 007245, MLP 007247, MLP 007251, MLP 007252, MLP 007253, MLP 007254, MLP 007256, MLP 007260, MLP 007261, MLP 007262, MLP 007270, MLP 007276, MLP 007277, MLP 007280, MLP 007281, MLP 007282, MLP 007288, MLP 007290, MLP 007297, MLP 007301, MLP 007306, MLP 007308, MLP 007310, MLP 007311, MLP 007312, MLP 007313, MLP 007315, MLP 007319, MLP 007324, MLP 007326, MLP 007327, MLP 007330, MLP 007335, MLP 007337, MLP 007339, MLP 007340, MLP 007341, MLP 007342, MLP 007343, MLP 007350, MLP 007351, MLP 007357, MLP 007358, MLP 007359, MLP 007361, MLP 007362, MLP 007365, MLP 007375, MLP 007376, MLP 007377, MLP 007378, MLP 007488, MLP 007493, MLP 007494, MLP 007497, MLP 007499, MLP 007513, MLP 007515, MLP 007516, MLP 007520, MLP 007522, MLP 007527, MLP 007528, MLP 007581, MLP 007598, MLP 007599, MLP 007601, MLP 007612, MLP 007645, MLP 007785, MLP 007791, MLP 007814, MLP 007815, MLP 007862, MLP 007863, MLP 007868, MLP 007869, MLP 007895, MLP 007906, MLP 007950, MLP 008043, MLP 008044, MLP 008059, MLP 008096, MLP 008098, MLP 008100, MLP 008102, MLP 008142, MLP 008147, MLP 008149, MLP 008151, MLP 008170, MLP 008173, MLP 008174, MLP 008178, MLP 008186, MLP 008187, MLP 008227, MLP 008228, MLP 008231, MLP 008289, MLP 008290, MLP 008319, MLP 008321, MLP 008373, MLP 008432, MLP 008434, MLP 008453, MLP 008484, MLP 008485, MLP 008487, MLP 008488, MLP 008597, MLP 008599, MLP 008625, MLP 008626, MLP 008633, MLP 008641, MLP 008784, MLP 008787, MLP 008850, MLP 008853, MLP 008854, MLP 008859, MLP 008862, MLP 008865, MLP 008941, MLP 008959, MLP 009016, MLP 009018, MLP 009023, MLP 009025, MLP 009026, MLP 009100, MLP 009112, MLP 009113, MLP 009114, MLP 009119, MLP 009142, MLP 009144, MLP 009146, MLP 009149, MLP 009210, MLP 009212, MLP 009241, MLP 009245, MLP 009252, MLP 009253, MLP 009256, MLP 009240, MLP 003970, MLP 004376, MLP 009351, MLP 004762, MLP 004765, MLP 004795, MLP 004819, MLP 004861, MLP 004908, MLP 004928, MLP 004941, MLP 004942, MLP 009471, MLP 005249, MLP 005568, MLP 005858, MLP 006544, MLP 006696, MLP 007363, MLP 007257, MLP 007264, MLP 009560, MLP 007265, MLP 007298, MLP 007316, MLP 007620, MLP 007965, MLP 008125, MLP 009619, MLP 008176, MLP 008189, MLP 008490, MLP 008498, MLP 008627, MLP 008628, MLP 008954, MLP 009121, MLP 009143, MLP 009151, MLP 009244, MLP 009379, MLP 009460, MLP 009475, MLP 007289, MLP 004726, MLP 005717, MLP 006231, MLP 006736, MLP 006841, MLP 007325, MLP 007338, MLP 007352, MLP 007485, MLP 007652, MLP 007813, MLP 008072, MLP 008154, MLP 008175, MLP 008179, MLP 008595, MLP 008600, MLP 008630, MLP 008635, MLP 005677, MLP 008637, MLP 008790, MLP 008856, MLP 008958, MLP 006719, MLP 009019, MLP 009099, MLP 009116, MLP 009145, MLP 009148, MLP 009150, MLP 009238, MLP 009246, MLP 009345, MLP 009349, MLP 009375, MLP 009457, MLP 009470, MLP 009476, MLP 007977, MLP 009258, MLP 009374, MLP 009578, MLP 009579, MLP 009620, MLP 009846, MLP 009850, MLP 009949, MLP 010017, MLP 010111, MLP 010139, MLP 010143, MLP 010146, MLP 010148, MLP 009623, MLP 009624, MLP 009947, MLP 005910, MLP 005961, MLP 005971, MLP 006263, MLP 006265, MLP 006637, MLP 006679, MLP 006761, MLP 006783, MLP 006820, MLP 006843, MLP 007296, MLP 007613, MLP 007614, MLP 007647, MLP 007691, MLP 007729, MLP 007732, MLP 007738, MLP 007743, MLP 007747, MLP 007749, MLP 007750, MLP 007786, MLP 007788, MLP 007812, MLP 007819, MLP 007847, MLP 007848, MLP 007860, MLP 007866, MLP 007867, MLP 007896, MLP 007953, MLP 008018, MLP 008057, MLP 008060, MLP 008062, MLP 008124, MLP 008171, MLP 008229, MLP 008230, MLP 008233, MLP 008370, MLP 008377, MLP 008643, MLP 008791, MLP 008940, MLP 008949, MLP 009030, MLP 009031, MLP 009213, MLP 009214, MLP 009247, MLP 009459, MLP 009582, MLP 008286,
MLP 008294, MLP 008378, MLP 009101, MLP 009250, MLP 009257, MLP 006228, MLP 007783, MLP 007854, MLP 008056, MLP 008065, MLP 008372, MLP 008375, MLP 008376, MLP 008440, MLP 008792, MLP 009028, MLP 009232, MLP 009466, MLP 009561, MLP 008948, MLP 009468, MLP 009581, MLP 009845, MLP 003518, MLP 003519, MLP 003520, MLP 003549, MLP 003550, MLP 003596, MLP 003598, MLP 003608, MLP 003606, MLP 003610, MLP 002427, MLP 002462, MLP 002912, MLP 003169, MLP 003398, MLP 003516, MLP 003602, MLP 004561, MLP 005432, MLP 003925, MLP 003989, MLP 004236, MLP 004857, MLP 004873, MLP 004900, MLP 005169, MLP 005537, MLP 005797, MLP 005992, MLP 006605, MLP 006680, MLP 007246, MLP 008636, MLP 009472, MLP 006621, MLP 008486, MLP 000992, MLP 002005, MLP 002007, MLP 002009, MLP 002014, MLP 002015, MLP 002028, MLP 002050, MLP 002055, MLP 002064, MLP 002131, MLP 002142, MLP 002148, MLP 002153, MLP 002193, MLP 002202, MLP 003604, MLP 004006, MLP 005916, MLP 006008, MLP 007695, MLP 009353, MLP 005382, MLP 003783, MLP 003784, MLP 009380, MLP 009381, MLP 006735, MLP 006737, MLP 009017, MLP 009461, MLP 001000, MLP 002038, MLP 002053, MLP 002085, MLP 002102, MLP 002187, MLP 002265, MLP 002301, MLP 002302, MLP 002339, MLP 002468, MLP 002469, MLP 002594, MLP 002846, MLP 006286, MLP 006289, MLP 006291, MLP 006360, MLP 006361, MLP 006363, MLP 006364, MLP 006813, MLP 007248, MLP 007409, MLP 007424, MLP 007425, MLP 007456, MLP 007457, MLP 007763, MLP 007797, MLP 007873, MLP 007917, MLP 007918, MLP 007924, MLP 007925, MLP 008122, MLP 008123, MLP 008256, MLP 009138, MLP 009139, MLP 009140, MLP 009141, MLP 009290, MLP 009306, MLP 009307, MLP 009314, MLP 009372, MLP 009404, MLP 009405, MLP 009406, MLP 009407, MLP 009409, MLP 009410, MLP 009411, MLP 009415, MLP 009420, MLP 009421, MLP 009430, MLP 009599, MLP 009602, MLP 009603, MLP 009627, MLP 009644, MLP 009645, MLP 009646, MLP 009657, MLP 009719, MLP 007413, MLP 007416, MLP 007423, MLP 007684, MLP 008326, MLP 009363, MLP 009392, MLP 009394, MLP 009653, MLP 009201, MLP 009203, MLP 009205, MLP 009206, MLP 009207, MLP 009667, MLP 004326, MLP 003941, MLP 004456, MLP 004457, MLP 004538, MLP 004543, MLP 004549, MLP 004583, MLP 004592, MLP 004684, MLP 004725, MLP 004733, MLP 004813, MLP 004913, MLP 004937, MLP 004940, MLP 005069, MLP 005104, MLP 005247, MLP 005313, MLP 005425, MLP 005430, MLP 005517, MLP 005720, MLP 005782, MLP 005990, MLP 006104, MLP 006277, MLP 006375, MLP 006384, MLP 006880, MLP 007333, MLP 007484, MLP 007654, MLP 007821, MLP 008642, MLP 008849, MLP 008861, MLP 009234, MLP 008182, MLP 008953, MLP 009347, MLP 009729, MLP 009344, MLP 009350, MLP 009352, MLP 010015, MLP 000990, MLP 001001, MLP 002002, MLP 002006, MLP 002029, MLP 002043, MLP 002068, MLP 002260, MLP 002279, MLP 002358, MLP 002422, MLP 002435, MLP 002439, MLP 002442, MLP 002467, MLP 002592, MLP 002675, MLP 002676, MLP 002753, MLP 002870, MLP 002873, MLP 002875, MLP 003004, MLP 003040, MLP 003083, MLP 003104, MLP 003191, MLP 003402, MLP 003522, MLP 003726, MLP 003804, MLP 003825, MLP 003870, MLP 003892, MLP 003896, MLP 003898, MLP 003904, MLP 003910, MLP 003943, MLP 003995, MLP 003997, MLP 003999, MLP 004219, MLP 004312, MLP 004392, MLP 004447, MLP 004462, MLP 004498, MLP 004548, MLP 004625, MLP 004771, MLP 004783, MLP 004791, MLP 004853, MLP 004894, MLP 004987, MLP 005127, MLP 005155, MLP 005197, MLP 005248, MLP 005518, MLP 005520, MLP 005573, MLP 005912, MLP 005917, MLP 005955, MLP 006398, MLP 006426, MLP 006428, MLP 006441, MLP 006446, MLP 006499, MLP 006523, MLP 006563, MLP 006597, MLP 006624, MLP 007258, MLP 007307, MLP 007487, MLP 007523, MLP 007818, MLP 007865, MLP 008014, MLP 008285, MLP 008288, MLP 008455, MLP 008860, MLP 006448, MLP 006516, MLP 006819, MLP 007498, MLP 009236, MLP 004091, MLP 006804, MLP 008667, MLP 009626, MLP 000988, MLP 002083, MLP 002965, MLP 004105, MLP 004400, MLP 005982, MLP 004411, MLP 004412, MLP 004918, MLP 004920, MLP 004923, MLP 005307, MLP 005379, MLP 005453, MLP 005477, MLP 005508, MLP 005860, MLP 005862, MLP 005896, MLP 005959, MLP 005987, MLP 006443, MLP 006656, MLP 006749, MLP 006752, MLP 006755, MLP 007237, MLP 007266, MLP 007268, MLP 007328, MLP 007345, MLP 007347, MLP 007346, MLP 002453, MLP 005954, MLP 005970, MLP 007331, MLP 007329 MLP 008150, |
Recalling Firm/ Manufacturer |
Abbott 6035 Stoneridge Dr Pleasanton CA 94588-3270
|
For Additional Information Contact |
Justin Paquette 651-756-6293
|
Manufacturer Reason for Recall |
Reports of outflow graft twist occlusions. Patients whose devices experience outflow graft occlusions will experience a persistent low flow alarm. Outflow graft occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death.
|
FDA Determined Cause 2 |
Device Design |
Action |
Customers were notified via an Important Medical Device Advisory letter, dated 04/05/2018, that the firm has received reports of outflow graft twist occlusions in the HeartMate 3 (HM3) Left Ventricular Assist System. As a result, patients whose devices experience this issue will experience a persistent low flow alarm. Outflow graft twist occlusions can result in serious adverse events such as hemodynamic compromise, thrombus, and death. Physicians managing patients that will be implanted or already implanted with HM3 devices should follow the recommendations provided in the letter. Additionally, physicians managing patients that exhibit a persistent low flow alarm should determine patient care recommendations based on each unique clinical case. Questions should be directed to customers' local Abbott MCS Clinical Specialist or MCS HeartLine 1-800-456-1477.
The firm issued a press release, dated 05/04/2018, notifying physicians about a field corrective action related to the HeartMate 3 Left Ventricular Assist Device (LVAD). Outflow graft twisting could occur post-implant in the HeartMate 3 LVAD, resulting in a persistent low flow alarm that may signal a potential safety risk to patients, such as low blood flow or clotting. No devices are being recalled from patients or hospitals as a result of the corrective action, and the recommendations made by the firm remain in place for physicians managing patients implanted with HeartMate 3.
On 10/17/2018, the firm mailed an Urgent Medical Device Correction notice to consignees informing them that the Outflow Graft Clip was now available. The Outflow Graft Clip is intended to be used with all new patient implants and may be used on patient cases where an Outflow Graft twist is confirmed and surgical intervention is required. All HM3 implanting centers are required to be trained on the proper attachment of the clip prior to use of the clip. |
Quantity in Commerce |
5,607 |
Distribution |
Worldwide Distribution -- U.S., including the states of MN, PA, IL, WI, AZ, AR, TN, MO, TX, MA, NE, MI, CA, NC, OH, NY, GA, VA, FL, OK, MD, ME, SC, IN, LA, OR, WA, AL, CO, IA, KS, KY, UT, DC, and CT; and, the countries of Australia, Austria, Belgium, Brazil, Brunei, Canada, Cayman Islands, Colombia, Croatia,
Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Kuwait, Lebanon, Malaysia, Montenegro, Netherlands, Poland, Qatar, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia,
Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = DSQ and Original Applicant = Abbott Medical
|
|
|
|