• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall VISIONIST CRTP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall VISIONIST CRTP see related information
Date Initiated by Firm June 03, 2021
Date Posted July 02, 2021
Recall Status1 Open3, Classified
Recall Number Z-2003-2021
Recall Event ID 88087
PMA Number P030005 
Product Classification Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) - Product Code NKE
Product VISIONIST CRT-P pacemaker, Models:
a) U225
b) U226
c) U228
Code Information a) U225, Serial Numbers: 707340, 100094, 100142, 100152, 100160, 100184, 100309, 100364, 100807, 100880, 100908, 100913, 700015, 700197, 700469, 700559, 700563, 700576, 700586, 700598, 700613, 700655, 700681, 700682, 700722, 700725, 700782, 700824, 700895, 700942, 700961, 700998, 701017, 701030, 701045, 701094, 701118, 701127, 701212, 701230, 701264, 701336, 701339, 701343, 701407, 701477, 701504, 702448, 702462, 702471, 702654, 702673, 702685, 702809, 703137, 703696, 703697, 703784, 703827, 703875, 703883, 703998, 707324, 707337, 707338, 707347, 707348, 707398, 707430, 707477, 707502, 707503, 707505, 707510, 707511, 707512, 707522, 707524, 707538, 707546, 707566, 707587, 707590, 707597, 100016, 100021, 100023, 100048, 100053, 100059, 100060, 100061, 100064, 100065, 100070, 100072, 100073, 100075, 100077, 100082, 100087, 100088, 100089, 100090, 100093, 100097, 100098, 100100, 100101, 100103, 100105, 100107, 100108, 100109, 100113, 100114, 100115, 100116, 100117, 100118, 100120, 100121, 100122, 100123, 100125, 100126, 100128, 100129, 100130, 100131, 100134, 100135, 100137, 100138, 100140, 100143, 100144, 100145, 100147, 100151, 100153, 100154, 100155, 100156, 100161, 100168, 100169, 100170, 100172, 100173, 100174, 100176, 100181, 100183, 100185, 100186, 100187, 100190, 100191, 100195, 100196, 100197, 100198, 100233, 100238, 100241, 100252, 100266, 100269, 100271, 100273, 100276, 100278, 100282, 100289, 100293, 100296, 100304, 100312, 100313, 100316, 100388, 100395, 100398, 100424, 100428, 100471, 100543, 100545, 100579, 100581, 100595, 100602, 100604, 100610, 100616, 100627, 100712, 100717, 100720, 100734, 100747, 100787, 100800, 100804, 100806, 100810, 100819, 100827, 100828, 100832, 100834, 100835, 100845, 100849, 100850, 100851, 100852, 100853, 100856, 100858, 100865, 100874, 100879, 100887, 100888, 100889, 100890, 100892, 100894, 100895, 100900, 100901, 100902, 100905, 100906, 100907, 100909, 100915, 100916, 100917, 100919, 100921, 100922, 100923, 100924, 100926, 100928, 100929, 100930, 100931, 100932, 100933, 100934, 100935, 700002, 700006, 700013, 700016, 700021, 700024, 700025, 700027, 700035, 700039, 700040, 700043, 700044, 700047, 700048, 700052, 700056, 700060, 700063, 700064, 700067, 700071, 700075, 700077, 700080, 700082, 700083, 700084, 700085, 700088, 700089, 700092, 700094, 700099, 700100, 700102, 700103, 700104, 700105, 700108, 700109, 700110, 700115, 700121, 700125, 700128, 700132, 700134, 700138, 700142, 700145, 700146, 700148, 700151, 700152, 700153, 700155, 700156, 700157, 700159, 700160, 700165, 700170, 700180, 700188, 700189, 700198, 700202, 700203, 700205, 700206, 700209, 700210, 700211, 700214, 700215, 700216, 700217, 700221, 700222, 700225, 700246, 700249, 700268, 700271, 700274, 700278, 700279, 700281, 700285, 700288, 700291, 700292, 700389, 700425, 700427, 700428, 700429, 700430, 700431, 700432, 700433, 700434, 700435, 700446, 700447, 700448, 700449, 700450, 700451, 700452, 700453, 700454, 700455, 700457, 700458, 700459, 700461, 700462, 700463, 700464, 700465, 700467, 700468, 700470, 700471, 700472, 700473, 700474, 700475, 700476, 700477, 700478, 700479, 700480, 700481, 700482, 700483, 700484, 700485, 700486, 700487, 700490, 700491, 700492, 700493, 700494, 700495, 700496, 700497, 700498, 700499, 700500, 700503, 700504, 700506, 700508, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700522, 700523, 700524, 700525, 700526, 700527, 700528, 700529, 700530, 700531, 700532, 700533, 700534, 700535, 700536, 700537, 700538, 700540, 700541, 700542, 700543, 700544, 700545, 700546, 700547, 700548, 700549, 700550, 700551, 700553, 700555, 700558, 700560, 700561, 700562, 700564, 700566, 700567, 700568, 700570, 700571, 700572, 700573, 700574, 700575, 700577, 700578, 700579, 700580, 700581, 700583, 700584, 700585, 700587, 700588, 700590, 700591, 700592, 700593, 700594, 700595, 700596, 700597, 700599, 700600, 700603, 700604, 700605, 700606, 700607, 700608, 700609, 700610, 700611, 700612, 700614, 700615, 700616, 700617, 700620, 700622, 700623, 700624, 700625, 700626, 700627, 700628, 700629, 700631, 700632, 700633, 700635, 700636, 700637, 700639, 700640, 700641, 700642, 700643, 700644, 700645, 700647, 700648, 700651, 700652, 700653, 700654, 700658, 700659, 700660, 700661, 700663, 700664, 700665, 700666, 700667, 700668, 700670, 700671, 700672, 700673, 700674, 700676, 700677, 700679, 700680, 700683, 700684, 700686, 700687, 700688, 700689, 700690, 700691, 700693, 700694, 700695, 700696, 700698, 700699, 700704, 700705, 700710, 700711, 700712, 700714, 700715, 700716, 700717, 700718, 700719, 700720, 700723, 700724, 700727, 700728, 700729, 700730, 700731, 700732, 700733, 700734, 700736, 700737, 700739, 700740, 700741, 700742, 700743, 700745, 700748, 700749, 700750, 700751, 700752, 700753, 700754, 700755, 700756, 700757, 700758, 700760, 700761, 700764, 700766, 700767, 700768, 700770, 700772, 700773, 700774, 700775, 700776, 700778, 700779, 700780, 700781, 700783, 700785, 700788, 700789, 700790, 700791, 700793, 700794, 700795, 700797, 700799, 700803, 700804, 700807, 700808, 700809, 700810, 700812, 700813, 700815, 700819, 700821, 700822, 700825, 700826, 700827, 700849, 700855, 700857, 700870, 700871, 700872, 700877, 700881, 700885, 700886, 700887, 700888, 700889, 700890, 700891, 700892, 700893, 700894, 700896, 700897, 700899, 700900, 700901, 700902, 700903, 700904, 700905, 700906, 700907, 700908, 700909, 700910, 700913, 700914, 700919, 700920, 700921, 700922, 700923, 700924, 700925, 700926, 700927, 700929, 700933, 700934, 700935, 700936, 700937, 700938, 700939, 700940, 700941, 700945, 700947, 700948, 700949, 700950, 700951, 700953, 700954, 700955, 700957, 700958, 700959, 700962, 700963, 700964, 700966, 700967, 700968, 700969, 700970, 700971, 700972, 700973, 700974, 700975, 700976, 700977, 700979, 700980, 700981, 700983, 700986, 700987, 700988, 700989, 700990, 700991, 700992, 700993, 700994, 700995, 700996, 700997, 701000, 701001, 701004, 701006, 701007, 701009, 701010, 701011, 701013, 701016, 701018, 701019, 701020, 701021, 701022, 701025, 701026, 701027, 701028, 701031, 701035, 701037, 701038, 701039, 701040, 701042, 701043, 701044, 701046, 701047, 701048, 701049, 701051, 701052, 701053, 701054, 701057, 701059, 701061, 701063, 701064, 701066, 701068, 701069, 701070, 701071, 701073, 701074, 701075, 701076, 701078, 701079, 701080, 701083, 701085, 701086, 701087, 701088, 701089, 701090, 701092, 701093, 701095, 701096, 701097, 701098, 701099, 701100, 701101, 701106, 701107, 701108, 701109, 701111, 701112, 701113, 701114, 701116, 701119, 701120, 701121, 701122, 701123, 701124, 701126, 701128, 701129, 701130, 701131, 701132, 701133, 701134, 701135, 701138, 701139, 701140, 701141, 701142, 701143, 701145, 701146, 701148, 701149, 701151, 701152, 701154, 701155, 701158, 701159, 701161, 701162, 701163, 701164, 701166, 701167, 701169, 701170, 701172, 701173, 701174, 701175, 701176, 701177, 701178, 701179, 701180, 701181, 701182, 701183, 701184, 701186, 701187, 701188, 701189, 701190, 701211, 701213, 701214, 701215, 701216, 701217, 701220, 701221, 701225, 701226, 701227, 701228, 701233, 701234, 701235, 701237, 701238, 701239, 701242, 701244, 701245, 701246, 701247, 701248, 701253, 701255, 701257, 701261, 701262, 701263, 701265, 701267, 701268, 701269, 701272, 701274, 701275, 701277, 701286, 701290, 701292, 701293, 701294, 701295, 701298, 701319, 701322, 701324, 701326, 701327, 701328, 701329, 701330, 701331, 701335, 701337, 701340, 701344, 701346, 701347, 701349, 701350, 701353, 701354, 701359, 701360, 701361, 701362, 701363, 701365, 701366, 701367, 701368, 701369, 701370, 701371, 701372, 701373, 701374, 701375, 701376, 701377, 701378, 701379, 701381, 701382, 701383, 701385, 701386, 701387, 701389, 701390, 701395, 701396, 701397, 701398, 701399, 701400, 701401, 701402, 701403, 701404, 701405, 701406, 701408, 701409, 701410, 701412, 701413, 701415, 701417, 701418, 701419, 701420, 701421, 701423, 701425, 701426, 701427, 701428, 701429, 701430, 701431, 701433, 701435, 701437, 701438, 701439, 701440, 701441, 701443, 701444, 701445, 701446, 701448, 701450, 701452, 701453, 701454, 701455, 701456, 701457, 701458, 701459, 701460, 701461, 701462, 701463, 701464, 701465, 701466, 701467, 701468, 701469, 701470, 701471, 701473, 701474, 701475, 701476, 701478, 701479, 701480, 701481, 701482, 701483, 701484, 701487, 701488, 701489, 701490, 701491, 701492, 701493, 701494, 701495, 701496, 701497, 701499, 701500, 701501, 701502, 701503, 701505, 701506, 701507, 701508, 701509, 701510, 701511, 701512, 701513, 701515, 701516, 701518, 701521, 701522, 701523, 701524, 701525, 701526, 701527, 701528, 701530, 701532, 701533, 701534, 701535, 701536, 701537, 701538, 701540, 701541, 701542, 701543, 701544, 701546, 701547, 701548, 701549, 701550, 701551, 701552, 701553, 701554, 701555, 701556, 701557, 701558, 702382, 702397, 702399, 702401, 702415, 702421, 702437, 702442, 702443, 702444, 702454, 702463, 702464, 702465, 702477, 702483, 702501, 702509, 702510, 702514, 702521, 702522, 702523, 702527, 702531, 702536, 702545, 702546, 702547, 702549, 702552, 702555, 702557, 702558, 702561, 702562, 702566, 702570, 702573, 702576, 702579, 702580, 702581, 702582, 702584, 702586, 702590, 702591, 702593, 702595, 702600, 702601, 702603, 702604, 702606, 702610, 702615, 702617, 702618, 702623, 702624, 702626, 702631, 702632, 702636, 702637, 702641, 702643, 702651, 702653, 702655, 702657, 702658, 702660, 702662, 702664, 702666, 702667, 702668, 702669, 702672, 702678, 702682, 702689, 702690, 702698, 702699, 702700, 702701, 702703, 702705, 702706, 702708, 702709, 702711, 702712, 702713, 702714, 702717, 702719, 702721, 702722, 702723, 702725, 702726, 702727, 702728, 702730, 702733, 702735, 702738, 702745, 702756, 702770, 702773, 702779, 702791, 702794, 702796, 702797, 702799, 702831, 702834, 702835, 702837, 702838, 702841, 702843, 702844, 702846, 702847, 702848, 702849, 702850, 702853, 702854, 702857, 702860, 702861, 702862, 702864, 702865, 702867, 702869, 702870, 702871, 702873, 702874, 702875, 702878, 702882, 702883, 702884, 702885, 702886, 702887, 702888, 702890, 702962, 702963, 702965, 702969, 702975, 702979, 702980, 702981, 702983, 702988, 702989, 702994, 702997, 703007, 703035, 703036, 703049, 703053, 703058, 703111, 703115, 703124, 703125, 703263, 703269, 703271, 703273, 703275, 703276, 703563, 703564, 703565, 703566, 703567, 703568, 703569, 703570, 703571, 703615, 703618, 703619, 703620, 703662, 703663, 703664, 703666, 703667, 703668, 703669, 703682, 703685, 703686, 703688, 703694, 703695, 703698, 703699, 703700, 703703, 703704, 703707, 703710, 703719, 703722, 703723, 703729, 703730, 703732, 703739, 703740, 703741, 703747, 703750, 703755, 703765, 703778, 703779, 703790, 703795, 703808, 703810, 703823, 703826, 703830, 703842, 703846, 703847, 703849, 703850, 703858, 703869, 703871, 703872, 703873, 703876, 703879, 703880, 703881, 703882, 703886, 704033, 704034, 704040, 704046, 704049, 704053, 704057, 704062, 704063, 704065, 704066, 704068, 704070, 704079, 704081, 704082, 704083, 704084, 704086, 705127, 705129, 705131, 705133, 705135, 705138, 705139, 707293, 707294, 707298, 707316, 707326, 707327, 707329, 707336, 707339, 707342, 707344, 707346, 707349, 707354, 707356, 707362, 707366, 707378, 707380, 707382, 707386, 707388, 707390, 707392, 707393, 707395, 707396, 707397, 707400, 707401, 707403, 707404, 707405, 707406, 707407, 707408, 707411, 707425, 707426, 707427, 707429, 707431, 707433, 707434, 707436, 707437, 707439, 707440, 707441, 707453, 707463, 707464, 707465, 707466, 707467, 707469, 707474, 707482, 707487, 707489, 707492, 707501, 707506, 707507, 707509, 707515, 707516, 707521, 707523, 707525, 707533, 707534, 707539, 707542, 707543, 707544, 707545, 707548, 707550, 707551, 707553, 707554, 707558, 707559, 707560, 707562, 707567, 707568, 707569, 707570, 707572, 707574, 707576, 707577, 707578, 707579, 707580, 707581, 707583, 707584, 707586, 707589, 707591, 707593, 707594, 707596, 707599, 707600, 707602, 707604, 707605, 707606, 707607, 707608, 707609, 707610, 707613, 707615, 707617, 707619, 707620 b) U226, Serial Numbers: 100044, 100069, 100116, 700077, 700093, 700140, 700163, 700220, 700306, 700323, 700331, 100006, 100009, 100011, 100012, 100013, 100014, 100015, 100016, 100017, 100018, 100019, 100021, 100022, 100023, 100024, 100025, 100026, 100027, 100028, 100029, 100030, 100032, 100034, 100035, 100038, 100039, 100042, 100045, 100046, 100048, 100049, 100050, 100052, 100055, 100058, 100059, 100060, 100061, 100062, 100063, 100066, 100067, 100068, 100070, 100072, 100073, 100074, 100075, 100076, 100077, 100081, 100082, 100085, 100087, 100089, 100090, 100091, 100092, 100094, 100095, 100096, 100098, 100100, 100101, 100102, 100105, 100106, 100107, 100108, 100109, 100111, 100112, 100113, 100117, 100118, 100120, 100123, 100124, 100126, 100127, 100128, 100129, 100132, 100133, 100134, 100135, 100136, 700002, 700003, 700004, 700006, 700007, 700008, 700009, 700010, 700011, 700012, 700013, 700014, 700015, 700018, 700019, 700020, 700021, 700023, 700024, 700025, 700026, 700030, 700031, 700032, 700033, 700034, 700035, 700036, 700037, 700039, 700040, 700042, 700043, 700044, 700045, 700048, 700050, 700051, 700052, 700053, 700054, 700056, 700057, 700058, 700059, 700060, 700061, 700063, 700065, 700066, 700067, 700068, 700071, 700072, 700074, 700075, 700076, 700079, 700081, 700082, 700083, 700084, 700087, 700088, 700089, 700090, 700091, 700092, 700094, 700095, 700096, 700098, 700099, 700101, 700102, 700103, 700106, 700107, 700108, 700109, 700110, 700111, 700113, 700114, 700115, 700116, 700118, 700120, 700121, 700122, 700123, 700124, 700125, 700126, 700132, 700135, 700137, 700139, 700141, 700142, 700143, 700145, 700146, 700147, 700149, 700151, 700153, 700154, 700155, 700156, 700159, 700164, 700167, 700168, 700169, 700170, 700171, 700172, 700183, 700184, 700186, 700187, 700191, 700192, 700193, 700194, 700198, 700199, 700201, 700205, 700207, 700208, 700209, 700210, 700211, 700212, 700213, 700214, 700215, 700216, 700218, 700219, 700222, 700225, 700226, 700227, 700229, 700230, 700231, 700232, 700236, 700237, 700239, 700240, 700241, 700245, 700246, 700247, 700249, 700265, 700309, 700310, 700311, 700312, 700313, 700314, 700316, 700318, 700319, 700321, 700325, 700326, 700327, 700328, 700329 c) U228, Serial Numbers: 702970, 726570, 100020, 100024, 100048, 100074, 100080, 100127, 100129, 100211, 100225, 100230, 100255, 100299, 100496, 101262, 101351, 101386, 101531, 101620, 101741, 101787, 101800, 101883, 101896, 101910, 101982, 102125, 102224, 102298, 700031, 700063, 700124, 700151, 700220, 700234, 700238, 700294, 700403, 700704, 700931, 701037, 701243, 701353, 701496, 701866, 701912, 701959, 702057, 702434, 702597, 702602, 702651, 702839, 703191, 703259, 703318, 703362, 703363, 703962, 703972, 703995, 704020, 704106, 704161, 704579, 704639, 704925, 704982, 705083, 705235, 705390, 705454, 705487, 705575, 705700, 705734, 705739, 705811, 706030, 706035, 706251, 706572, 706585, 706653, 706784, 707069, 709016, 709198, 709260, 709300, 709337, 709982, 710064, 710069, 710090, 710869, 710888, 711240, 711260, 711899, 712093, 712156, 712749, 712858, 712859, 712872, 712885, 712894, 712903, 712908, 713152, 713173, 713246, 713360, 713372, 713398, 713458, 713472, 713513, 713515, 713529, 713537, 713580, 713656, 713734, 713838, 713840, 713886, 713925, 714277, 714284, 714288, 714301, 714353, 714433, 714438, 714448, 714453, 714684, 714696, 714763, 714791, 714794, 714795, 714799, 714815, 715582, 718328, 726154, 726191, 726232, 726331, 726368, 726369, 726381, 726418, 726439, 726448, 726454, 726455, 726457, 726462, 726472, 726482, 726501, 726502, 726519, 726536, 726588, 726663, 726695, 726780, 726797, 726932, 727077, 100004, 100006, 100009, 100010, 100011, 100012, 100015, 100016, 100018, 100019, 100022, 100031, 100032, 100033, 100034, 100035, 100036, 100037, 100039, 100041, 100042, 100045, 100046, 100047, 100049, 100050, 100051, 100053, 100055, 100057, 100059, 100061, 100062, 100064, 100065, 100067, 100069, 100070, 100071, 100073, 100075, 100076, 100078, 100079, 100082, 100083, 100084, 100085, 100086, 100088, 100089, 100090, 100091, 100093, 100099, 100102, 100104, 100105, 100106, 100108, 100109, 100111, 100117, 100118, 100120, 100125, 100128, 100130, 100131, 100132, 100133, 100134, 100135, 100136, 100137, 100143, 100145, 100148, 100149, 100150, 100153, 100154, 100155, 100157, 100158, 100159, 100160, 100161, 100162, 100163, 100164, 100165, 100166, 100167, 100168, 100169, 100170, 100172, 100173, 100174, 100175, 100176, 100178, 100179, 100180, 100183, 100185, 100186, 100188, 100190, 100191, 100193, 100194, 100195, 100196, 100198, 100199, 100206, 100207, 100214, 100218, 100219, 100220, 100221, 100223, 100224, 100227, 100228, 100229, 100231, 100233, 100234, 100235, 100237, 100238, 100239, 100241, 100242, 100243, 100244, 100248, 100249, 100250, 100252, 100254, 100259, 100261, 100263, 100264, 100265, 100268, 100269, 100270, 100271, 100273, 100276, 100278, 100279, 100280, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100290, 100291, 100292, 100294, 100296, 100300, 100301, 100303, 100304, 100305, 100306, 100307, 100309, 100310, 100311, 100312, 100313, 100314, 100315, 100316, 100318, 100319, 100320, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100332, 100333, 100334, 100335, 100336, 100337, 100342, 100343, 100344, 100345, 100349, 100352, 100353, 100354, 100356, 100358, 100366, 100368, 100370, 100374, 100382, 100395, 100396, 100406, 100412, 100421, 100423, 100424, 100425, 100426, 100429, 100431, 100432, 100434, 100439, 100442, 100443, 100446, 100448, 100460, 100462, 100464, 100465, 100472, 100498, 100508, 100515, 100518, 100528, 100534, 100543, 100546, 100548, 100558, 100578, 100587, 100598, 100604, 100606, 100612, 100622, 100623, 100815, 100937, 100976, 100986, 101029, 101030, 101031, 101033, 101034, 101036, 101068, 101105, 101132, 101163, 101179, 101215, 101231, 101252, 101253, 101257, 101270, 101286, 101287, 101288, 101315, 101317, 101320, 101323, 101331, 101332, 101339, 101340, 101342, 101346, 101348, 101349, 101352, 101356, 101368, 101379, 101380, 101382, 101383, 101384, 101385, 101388, 101394, 101396, 101397, 101399, 101400, 101405, 101406, 101408, 101409, 101421, 101422, 101423, 101424, 101425, 101439, 101440, 101441, 101443, 101444, 101447, 101448, 101449, 101450, 101452, 101453, 101456, 101458, 101461, 101465, 101466, 101467, 101468, 101471, 101475, 101476, 101477, 101478, 101479, 101483, 101485, 101487, 101490, 101494, 101495, 101497, 101498, 101499, 101500, 101501, 101502, 101503, 101504, 101505, 101506, 101507, 101508, 101511, 101512, 101513, 101515, 101517, 101519, 101520, 101521, 101522, 101523, 101526, 101527, 101529, 101530, 101532, 101533, 101534, 101535, 101537, 101538, 101540, 101543, 101544, 101545, 101547, 101548, 101549, 101550, 101551, 101619, 101622, 101625, 101626, 101627, 101628, 101630, 101632, 101634, 101637, 101638, 101640, 101643, 101645, 101646, 101648, 101649, 101650, 101651, 101653, 101654, 101655, 101657, 101658, 101660, 101661, 101662, 101664, 101666, 101667, 101668, 101669, 101670, 101671, 101672, 101673, 101674, 101678, 101704, 101730, 101731, 101735, 101736, 101737, 101740, 101742, 101743, 101745, 101748, 101750, 101754, 101755, 101756, 101757, 101758, 101759, 101761, 101764, 101765, 101766, 101767, 101768, 101769, 101770, 101771, 101772, 101773, 101780, 101782, 101783, 101784, 101785, 101786, 101790, 101792, 101793, 101794, 101795, 101796, 101797, 101798, 101799, 101803, 101804, 101805, 101806, 101807, 101808, 101809, 101811, 101812, 101815, 101816, 101817, 101818, 101819, 101820, 101821, 101822, 101823, 101824, 101825, 101826, 101828, 101829, 101830, 101831, 101833, 101836, 101837, 101839, 101840, 101843, 101846, 101847, 101850, 101851, 101852, 101855, 101856, 101858, 101859, 101860, 101861, 101862, 101863, 101864, 101865, 101866, 101867, 101869, 101872, 101873, 101874, 101875, 101876, 101877, 101878, 101880, 101882, 101885, 101886, 101887, 101889, 101891, 101893, 101894, 101895, 101897, 101898, 101899, 101900, 101901, 101902, 101903, 101905, 101906, 101908, 101912, 101913, 101914, 101915, 101916, 101918, 101920, 101922, 101923, 101925, 101926, 101927, 101928, 101929, 101931, 101933, 101936, 101937, 101938, 101939, 101940, 101941, 101942, 101943, 101945, 101946, 101947, 101949, 101952, 101953, 101956, 101957, 101958, 101959, 101960, 101961, 101967, 101969, 101971, 101973, 101975, 101976, 101978, 101980, 101984, 101988, 101990, 101991, 101994, 101995, 101996, 101997, 101999, 102000, 102006, 102009, 102020, 102025, 102029, 102030, 102031, 102034, 102041, 102046, 102048, 102056, 102068, 102072, 102075, 102077, 102080, 102083, 102084, 102086, 102094, 102095, 102096, 102098, 102105, 102114, 102115, 102122, 102124, 102130, 102133, 102149, 102153, 102156, 102160, 102162, 102164, 102169, 102173, 102177, 102179, 102181, 102183, 102184, 102185, 102186, 102187, 102188, 102192, 102193, 102194, 102195, 102199, 102200, 102202, 102204, 102205, 102207, 102208, 102209, 102215, 102216, 102217, 102218, 102219, 102220, 102222, 102223, 102227, 102228, 102231, 102232, 102236, 102238, 102239, 102241, 102244, 102250, 102253, 102254, 102256, 102259, 102264, 102266, 102267, 102268, 102269, 102270, 102271, 102272, 102273, 102274, 102275, 102276, 102277, 102278, 102279, 102280, 102281, 102282, 102283, 102285, 102286, 102287, 102288, 102290, 102293, 102294, 102295, 102300, 102302, 102303, 102304, 102305, 102306, 102307, 102308, 102309, 102310, 102313, 102321, 102322, 102323, 102326, 102329, 102330, 102331, 102332, 102333, 102335, 102336, 102338, 102339, 102340, 102341, 102342, 102343, 102344, 102345, 102347, 102348, 102357, 102358, 102360, 102361, 102362, 102364, 102365, 102368, 102369, 102370, 102372, 102373, 102375, 102376, 102377, 102378, 102379, 102381, 102383, 102385, 102386, 102387, 102391, 102392, 102393, 102394, 102395, 102396, 102399, 102402, 102406, 102408, 102409, 102410, 102411, 102412, 102419, 102421, 102422, 102423, 102425, 102426, 102427, 102430, 102432, 102438, 102442, 102454, 102456, 102460, 102479, 102482, 102483, 700001, 700002, 700004, 700005, 700006, 700007, 700008, 700010, 700011, 700012, 700014, 700016, 700017, 700018, 700019, 700020, 700021, 700022, 700023, 700024, 700025, 700026, 700027, 700028, 700029, 700030, 700032, 700033, 700035, 700036, 700037, 700038, 700040, 700041, 700042, 700043, 700046, 700047, 700048, 700049, 700050, 700051, 700052, 700053, 700054, 700055, 700056, 700057, 700058, 700059, 700060, 700061, 700062, 700064, 700067, 700068, 700069, 700072, 700073, 700074, 700075, 700076, 700077, 700078, 700079, 700080, 700081, 700082, 700083, 700085, 700087, 700088, 700089, 700092, 700094, 700095, 700096, 700098, 700100, 700101, 700102, 700104, 700106, 700107, 700108, 700109, 700111, 700113, 700114, 700115, 700116, 700117, 700118, 700119, 700120, 700121, 700122, 700123, 700125, 700127, 700128, 700129, 700131, 700133, 700134, 700135, 700137, 700138, 700139, 700140, 700141, 700142, 700143, 700145, 700146, 700147, 700148, 700149, 700150, 700152, 700153, 700154, 700155, 700156, 700157, 700158, 700161, 700162, 700163, 700164, 700165, 700167, 700170, 700171, 700172, 700173, 700174, 700176, 700178, 700179, 700181, 700182, 700183, 700184, 700186, 700187, 700188, 700189, 700191, 700193, 700196, 700198, 700199, 700200, 700201, 700202, 700204, 700207, 700208, 700209, 700210, 700211, 700214, 700215, 700216, 700218, 700221, 700224, 700225, 700226, 700228, 700229, 700232, 700235, 700236, 700237, 700239, 700240, 700243, 700246, 700247, 700255, 700259, 700263, 700265, 700266, 700269, 700276, 700282, 700284, 700288, 700295, 700297, 700299, 700302, 700303, 700305, 700317, 700333, 700335, 700338, 700342, 700347, 700368, 700375, 700376, 700377, 700382, 700383, 700384, 700385, 700386, 700387, 700393, 700394, 700395, 700397, 700402, 700404, 700405, 700406, 700409, 700411, 700412, 700413, 700414, 700419, 700420, 700422, 700423, 700430, 700431, 700432, 700437, 700440, 700442, 700444, 700448, 700462, 700463, 700464, 700468, 700469, 700470, 700471, 700472, 700473, 700475, 700476, 700478, 700479, 700482, 700483, 700484, 700485, 700488, 700489, 700492, 700493, 700494, 700496, 700497, 700498, 700499, 700502, 700503, 700505, 700506, 700509, 700510, 700511, 700512, 700513, 700514, 700515, 700516, 700517, 700518, 700698, 700701, 700703, 700710, 700712, 700725, 700812, 700819, 700836, 700838, 700899, 700908, 700912, 700925, 700974, 700990, 701015, 701017, 701033, 701034, 701040, 701043, 701048, 701055, 701060, 701061, 701062, 701065, 701071, 701075, 701077, 701082, 701085, 701086, 701089, 701090, 701093, 701100, 701103, 701104, 701107, 701118, 701120, 701122, 701126, 701128, 701148, 701163, 701165, 701170, 701184, 701191, 701192, 701197, 701201, 701202, 701203, 701205, 701213, 701220, 701222, 701229, 701238, 701239, 701252, 701257, 701265, 701269, 701270, 701293, 701298, 701301, 701305, 701309, 701310, 701311, 701319, 701327, 701346, 701351, 701374, 701375, 701377, 701378, 701382, 701384, 701385, 701427, 701432, 701443, 701444, 701446, 701451, 701454, 701457, 701461, 701475, 701476, 701477, 701487, 701488, 701491, 701494, 701649, 701650, 701653, 701658, 701659, 701671, 701672, 701674, 701677, 701682, 701690, 701691, 701701, 701705, 701706, 701709, 701711, 701719, 701721, 701726, 701730, 701734, 701735, 701737, 701742, 701743, 701748, 701749, 701753, 701757, 701758, 701759, 701769, 701771, 701774, 701776, 701778, 701780, 701783, 701784, 701790, 701794, 701795, 701798, 701799, 701800, 701801, 701803, 701804, 701805, 701806, 701807, 701810, 701813, 701814, 701818, 701822, 701823, 701830, 701835, 701848, 701850, 701858, 701861, 701869, 701872, 701875, 701879, 701882, 701884, 701886, 701889, 701890, 701891, 701896, 701897, 701903, 701918, 701919, 701926, 701928, 701929, 701933, 701935, 701937, 701938, 701946, 701958, 701960, 701961, 701967, 701970, 701972, 701978, 701979, 701990, 701991, 701992, 701993, 701994, 701995, 701996, 701997, 702000, 702004, 702008, 702009, 702012, 702014, 702016, 702018, 702020, 702022, 702023, 702025, 702028, 702034, 702040, 702043, 702047, 702048, 702054, 702061, 702062, 702064, 702065, 702069, 702070, 702073, 702088, 702094, 702098, 702099, 702100, 702102, 702106, 702115, 702116, 702120, 702122, 702123, 702126, 702127, 702128, 702129, 702130, 702133, 702135, 702137, 702138, 702141, 702142, 702146, 702150, 702151, 702157, 702158, 702160, 702162, 702164, 702165, 702169, 702174, 702178, 702181, 702184, 702188, 702189, 702190, 702191, 702192, 702193, 702195, 702197, 702208, 702209, 702212, 702215, 702216, 702217, 702222, 702223, 702224, 702232, 702235, 702247, 702249, 702263, 702267, 702279, 702280, 702300, 702303, 702304, 702305, 702306, 702307, 702308, 702311, 702313, 702323, 702324, 702327, 702330, 702331, 702344, 702353, 702354, 702358, 702367, 702369, 702370, 702374, 702375, 702379, 702380, 702384, 702390, 702394, 702404, 702407, 702417, 702424, 702433, 702436, 702439, 702443, 702444, 702445, 702446, 702456, 702458, 702459, 702463, 702465, 702469, 702470, 702472, 702479, 702480, 702484, 702485, 702487, 702489, 702492, 702511, 702513, 702517, 702518, 702520, 702521, 702524, 702527, 702529, 702531, 702532, 702533, 702535, 702537, 702538, 702555, 702556, 702558, 702563, 702564, 702566, 702570, 702571, 702577, 702578, 702583, 702585, 702591, 702592, 702594, 702604, 702609, 702615, 702617, 702625, 702629, 702631, 702634, 702635, 702636, 702639, 702642, 702643, 702644, 702649, 702654, 702660, 702661, 702664, 702666, 702669, 702670, 702673, 702675, 702678, 702682, 702683, 702687, 702688, 702689, 702692, 702699, 702700, 702704, 702707, 702709, 702710, 702711, 702712, 702720, 702723, 702724, 702725, 702729, 702730, 702734, 702741, 702742, 702749, 702750, 702752, 702753, 702755, 702756, 702758, 702764, 702773, 702774, 702776, 702780, 702783, 702788, 702791, 702794, 702795, 702797, 702799, 702805, 702808, 702811, 702816, 702819, 702820, 702821, 702826, 702835, 702841, 702842, 702843, 702847, 702849, 702850, 702851, 702852, 702856, 702859, 702865, 702867, 702868, 702869, 702870, 702872, 702880, 702890, 702926, 702934, 702943, 702951, 702953, 702964, 702966, 702972, 702977, 702983, 703016, 703020, 703022, 703023, 703036, 703040, 703043, 703045, 703066, 703069, 703071, 703074, 703078, 703081, 703083, 703088, 703089, 703093, 703095, 703097, 703100, 703102, 703103, 703107, 703113, 703121, 703123, 703125, 703127, 703144, 703145, 703149, 703153, 703154, 703155, 703156, 703160, 703163, 703172, 703174, 703178, 703179, 703181, 703183, 703187, 703189, 703190, 703193, 703194, 703197, 703199, 703200, 703203, 703226, 703230, 703233, 703234, 703235, 703236, 703238, 703239, 703240, 703241, 703242, 703243, 703244, 703248, 703251, 703254, 703255, 703257, 703260, 703261, 703263, 703264, 703266, 703267, 703268, 703270, 703271, 703272, 703273, 703280, 703300, 703303, 703306, 703320, 703321, 703322, 703334, 703335, 703336, 703337, 703342, 703343, 703345, 703346, 703348, 703350, 703353, 703354, 703358, 703359, 703360, 703361, 703364, 703366, 703367, 703368, 703369, 703370, 703371, 703375, 703380, 703381, 703382, 703383, 703384, 703386, 703387, 703388, 703389, 703390, 703391, 703392, 703393, 703394, 703395, 703396, 703398, 703559, 703887, 703890, 703891, 703892, 703899, 703901, 703902, 703906, 703907, 703908, 703911, 703912, 703916, 703917, 703918, 703919, 703920, 703922, 703924, 703927, 703929, 703930, 703931, 703935, 703936, 703941, 703942, 703943, 703945, 703948, 703949, 703950, 703952, 703954, 703957, 703959, 703960, 703963, 703964, 703971, 703974, 703975, 703981, 703982, 703983, 703984, 703985, 703986, 703987, 703988, 703989, 703990, 703992, 703993, 703994, 703996, 703998, 704001, 704004, 704006, 704011, 704014, 704015, 704016, 704017, 704019, 704021, 704022, 704023, 704024, 704025, 704027, 704030, 704035, 704036, 704037, 704038, 704039, 704040, 704042, 704045, 704047, 704048, 704050, 704051, 704054, 704056, 704057, 704060, 704062, 704063, 704065, 704067, 704068, 704075, 704076, 704077, 704078, 704079, 704083, 704085, 704087, 704089, 704090, 704092, 704093, 704094, 704095, 704096, 704098, 704100, 704101, 704102, 704103, 704104, 704105, 704107, 704108, 704109, 704110, 704113, 704114, 704115, 704117, 704118, 704120, 704121, 704123, 704124, 704127, 704129, 704132, 704133, 704134, 704136, 704138, 704140, 704142, 704143, 704144, 704145, 704146, 704149, 704150, 704151, 704152, 704154, 704156, 704158, 704159, 704165, 704166, 704168, 704171, 704174, 704175, 704176, 704177, 704186, 704187, 704205, 704209, 704222, 704234, 704237, 704260, 704262, 704277, 704292, 704297, 704298, 704301, 704302, 704303, 704304, 704306, 704307, 704309, 704315, 704318, 704321, 704324, 704339, 704345, 704347, 704355, 704358, 704359, 704365, 704366, 704374, 704387, 704388, 704389, 704391, 704398, 704403, 704413, 704417, 704418, 704432, 704444, 704449, 704453, 704461, 704467, 704472, 704473, 704474, 704476, 704480, 704481, 704488, 704493, 704494, 704496, 704505, 704506, 704507, 704508, 704509, 704510, 704518, 704520, 704522, 704528, 704532, 704535, 704541, 704550, 704555, 704558, 704561, 704562, 704571, 704576, 704582, 704583, 704585, 704592, 704594, 704596, 704600, 704606, 704608, 704615, 704616, 704619, 704620, 704623, 704625, 704633, 704634, 704635, 704636, 704637, 704648, 704650, 704667, 704691, 704767, 704771, 704772, 704775, 704782, 704786, 704789, 704790, 704791, 704797, 704802, 704813, 704819, 704830, 704868, 704873, 704878, 704892, 704894, 704905, 704906, 704918, 704921, 704928, 704929, 704931, 704962, 704966, 704969, 704975, 704983, 704999, 705004, 705134, 705229, 705231, 705232, 705233, 705234, 705236, 705237, 705238, 705240, 705241, 705242, 705243, 705244, 705245, 705246, 705247, 705248, 705249, 705250, 705251, 705252, 705253, 705254, 705255, 705264, 705265, 705266, 705273, 705275, 705292, 705293, 705294, 705316, 705318, 705319, 705320, 705340, 705365, 705367, 705369, 705370, 705372, 705374, 705375, 705376, 705377, 705378, 705379, 705381, 705383, 705384, 705385, 705387, 705388, 705389, 705392, 705395, 705397, 705398, 705399, 705400, 705401, 705402, 705404, 705408, 705409, 705410, 705411, 705413, 705414, 705415, 705416, 705418, 705419, 705421, 705425, 705426, 705427, 705428, 705429, 705432, 705433, 705434, 705435, 705437, 705438, 705440, 705442, 705443, 705447, 705448, 705450, 705453, 705455, 705456, 705457, 705459, 705460, 705461, 705462, 705464, 705466, 705467, 705468, 705472, 705473, 705476, 705477, 705478, 705479, 705480, 705481, 705483, 705484, 705485, 705486, 705488, 705489, 705490, 705491, 705493, 705494, 705495, 705496, 705514, 705516, 705517, 705518, 705519, 705523, 705528, 705529, 705530, 705532, 705534, 705535, 705537, 705538, 705539, 705540, 705542, 705543, 705544, 705545, 705546, 705547, 705548, 705549, 705551, 705554, 705555, 705558, 705559, 705560, 705562, 705566, 705568, 705569, 705579, 705583, 705621, 705623, 705624, 705629, 705633, 705637, 705638, 705639, 705641, 705642, 705644, 705645, 705648, 705652, 705653, 705655, 705658, 705660, 705662, 705663, 705667, 705668, 705669, 705670, 705673, 705674, 705676, 705679, 705680, 705683, 705685, 705686, 705687, 705688, 705689, 705690, 705692, 705693, 705694, 705701, 705702, 705703, 705704, 705706, 705708, 705710, 705713, 705715, 705716, 705717, 705718, 705719, 705720, 705723, 705725, 705729, 705735, 705736, 705738, 705741, 705743, 705746, 705749, 705753, 705754, 705755, 705756, 705757, 705758, 705759, 705762, 705763, 705764, 705766, 705767, 705768, 705769, 705770, 705771, 705772, 705775, 705776, 705777, 705779, 705780, 705781, 705782, 705783, 705784, 705785, 705786, 705793, 705798, 705799, 705801, 705803, 705804, 705812, 705818, 705820, 705825, 705826, 705829, 705831, 705832, 705833, 705834, 705836, 705837, 705838, 705839, 705846, 705851, 705852, 705859, 705860, 705861, 705864, 705866, 705867, 705868, 705869, 705878, 705882, 705883, 705885, 705886, 705889, 705896, 705897, 705901, 705903, 705910, 705912, 705914, 705918, 705920, 705927, 705932, 705937, 705942, 705947, 705953, 705955, 705956, 705963, 705964, 705968, 705974, 705981, 705984, 705987, 705995, 705996, 705997, 705999, 706002, 706008, 706009, 706012, 706017, 706019, 706025, 706029, 706039, 706040, 706041, 706043, 706044, 706045, 706046, 706047, 706053, 706058, 706059, 706063, 706066, 706067, 706070, 706087, 706090, 706093, 706113, 706114, 706134, 706143, 706159, 706160, 706173, 706176, 706177, 706184, 706189, 706190, 706193, 706199, 706201, 706203, 706204, 706205, 706229, 706231, 706235, 706247, 706254, 706257, 706258, 706260, 706261, 706263, 706266, 706268, 706272, 706278, 706280, 706285, 706286, 706295, 706298, 706299, 706304, 706305, 706306, 706313, 706315, 706322, 706326, 706327, 706329, 706334, 706335, 706339, 706341, 706343, 706344, 706345, 706346, 706352, 706353, 706363, 706366, 706367, 706368, 706371, 706375, 706377, 706381, 706385, 706386, 706388, 706392, 706393, 706397, 706404, 706405, 706407, 706414, 706418, 706420, 706427, 706430, 706439, 706445, 706448, 706451, 706453, 706462, 706468, 706470, 706472, 706474, 706477, 706483, 706489, 706490, 706494, 706497, 706498, 706500, 706501, 706507, 706510, 706512, 706516, 706535, 706536, 706540, 706542, 706547, 706550, 706551, 706562, 706566, 706567, 706583, 706587, 706598, 706604, 706606, 706608, 706610, 706615, 706616, 706619, 706622, 706623, 706624, 706625, 706628, 706629, 706630, 706631, 706632, 706634, 706635, 706636, 706638, 706641, 706642, 706643, 706645, 706647, 706649, 706651, 706654, 706655, 706657, 706658, 706659, 706661, 706663, 706665, 706667, 706668, 706669, 706670, 706671, 706674, 706675, 706676, 706677, 706678, 706680, 706681, 706684, 706685, 706686, 706688, 706689, 706690, 706691, 706693, 706694, 706696, 706698, 706701, 706704, 706705, 706708, 706709, 706710, 706711, 706712, 706713, 706717, 706718, 706720, 706722, 706727, 706728, 706730, 706731, 706733, 706738, 706740, 706742, 706743, 706744, 706746, 706748, 706749, 706750, 706751, 706752, 706753, 706754, 706755, 706756, 706758, 706761, 706762, 706765, 706766, 706768, 706777, 706778, 706780, 706781, 706782, 706783, 706786, 706787, 706790, 706791, 706795, 706798, 706799, 706800, 706806, 706809, 706810, 706811, 706812, 706814, 706815, 706817, 706818, 706819, 706820, 706822, 706823, 706824, 706825, 706830, 706836, 706837, 706838, 706840, 706842, 706843, 706846, 706848, 706849, 706850, 706851, 706852, 706853, 706854, 706855, 706857, 706859, 706860, 706864, 706866, 706867, 706868, 706870, 706871, 706872, 706874, 706875, 706877, 706879, 706880, 706887, 706888, 706897, 706900, 706901, 706903, 706904, 706905, 706906, 706910, 706911, 706912, 706913, 706914, 706922, 706923, 706924, 706925, 706946, 706947, 706952, 706953, 706954, 706955, 706966, 706972, 706973, 706975, 706978, 706980, 706981, 706983, 706984, 706988, 706989, 706990, 706993, 706995, 706996, 706997, 706998, 707002, 707006, 707012, 707013, 707016, 707017, 707018, 707022, 707028, 707032, 707033, 707034, 707036, 707039, 707042, 707043, 707044, 707045, 707047, 707050, 707051, 707053, 707055, 707059, 707061, 707062, 707063, 707064, 707066, 707067, 707068, 707070, 707073, 707075, 708827, 708832, 708846, 708848, 708864, 708879, 708889, 708892, 708898, 708905, 708908, 708913, 708919, 708920, 708934, 708939, 708943, 708945, 708950, 708952, 708953, 708955, 708958, 708966, 708968, 708969, 708971, 708973, 708974, 708976, 708980, 708981, 708982, 708984, 708986, 708989, 708990, 708991, 708992, 708995, 708996, 708997, 708999, 709000, 709001, 709002, 709003, 709007, 709008, 709010, 709011, 709012, 709017, 709018, 709019, 709021, 709022, 709023, 709027, 709028, 709031, 709032, 709033, 709035, 709036, 709037, 709038, 709039, 709041, 709042, 709043, 709045, 709048, 709049, 709050, 709053, 709054, 709057, 709059, 709063, 709064, 709066, 709067, 709069, 709070, 709072, 709073, 709075, 709076, 709077, 709078, 709079, 709080, 709082, 709083, 709085, 709086, 709087, 709090, 709091, 709092, 709095, 709098, 709099, 709101, 709107, 709109, 709115, 709116, 709117, 709119, 709123, 709125, 709130, 709140, 709146, 709150, 709151, 709157, 709158, 709161, 709162, 709166, 709168, 709169, 709171, 709172, 709173, 709176, 709178, 709180, 709182, 709183, 709184, 709186, 709188, 709193, 709194, 709196, 709197, 709199, 709202, 709204, 709205, 709206, 709207, 709209, 709210, 709212, 709213, 709216, 709218, 709219, 709220, 709221, 709226, 709227, 709228, 709230, 709231, 709232, 709233, 709241, 709248, 709249, 709253, 709255, 709261, 709264, 709265, 709266, 709267, 709268, 709270, 709272, 709274, 709279, 709281, 709284, 709287, 709290, 709292, 709293, 709298, 709306, 709311, 709316, 709321, 709325, 709327, 709328, 709329, 709332, 709335, 709339, 709340, 709341, 709355, 709360, 709361, 709362, 709368, 709369, 709371, 709372, 709373, 709393, 709395, 709397, 709458, 709459, 709461, 709464, 709468, 709474, 709475, 709494, 709498, 709523, 709564, 709565, 709566, 709567, 709568, 709572, 709574, 709578, 709587, 709589, 709593, 709778, 709784, 709786, 709787, 709789, 709791, 709849, 709859, 709924, 709938, 709941, 709943, 709948, 709949, 709950, 709951, 709952, 709953, 709954, 709955, 709957, 709958, 709960, 709961, 709962, 709963, 709969, 709970, 709971, 709972, 709973, 709975, 709978, 709979, 709980, 709983, 710003, 710005, 710006, 710008, 710009, 710021, 710028, 710036, 710037, 710039, 710040, 710042, 710044, 710054, 710055, 710056, 710057, 710058, 710059, 710061, 710066, 710068, 710071, 710072, 710073, 710074, 710075, 710076, 710078, 710079, 710080, 710081, 710083, 710084, 710085, 710089, 710091, 710092, 710093, 710100, 710154, 710164, 710166, 710167, 710171, 710172, 710173, 710244, 710245, 710247, 710248, 710249, 710250, 710251, 710252, 710254, 710258, 710260, 710262, 710265, 710266, 710267, 710269, 710272, 710276, 710278, 710280, 710282, 710285, 710291, 710294, 710296, 710305, 710307, 710311, 710312, 710315, 710319, 710366, 710369, 710370, 710372, 710375, 710381, 710386, 710392, 710415, 710416, 710422, 710429, 710433, 710434, 710440, 710467, 710475, 710483, 710511, 710615, 710754, 710756, 710759, 710766, 710768, 710772, 710774, 710775, 710778, 710780, 710783, 710786, 710787, 710788, 710797, 710799, 710803, 710808, 710811, 710820, 710823, 710826, 710836, 710850, 710851, 710853, 710856, 710857, 710859, 710860, 710867, 710870, 710885, 711119, 711129, 711130, 711134, 711138, 711142, 711143, 711146, 711156, 711158, 711162, 711164, 711178, 711181, 711182, 711184, 711185, 711186, 711201, 711205, 711210, 711211, 711212, 711216, 711223, 711225, 711228, 711229, 711232, 711236, 711237, 711239, 711242, 711246, 711252, 711255, 711258, 711261, 711262, 711263, 711332, 711338, 711340, 711342, 711344, 711345, 711347, 711348, 711349, 711350, 711352, 711356, 711359, 711361, 711382, 711383, 711417, 711421, 711535, 711539, 711540, 711541, 711542, 711543, 711554, 711555, 711559, 711603, 711607, 711608, 711614, 711616, 711618, 711757, 711778, 711779, 711798, 711801, 711802, 711853, 711857, 711858, 711859, 711860, 711862, 711863, 711883, 711893, 711895, 711896, 711901, 711903, 711904, 711906, 711908, 711910, 711915, 711916, 711917, 711919, 711920, 711922, 711926, 711930, 711931, 711933, 711935, 711936, 711937, 711938, 711939, 711940, 711948, 711951, 711952, 711954, 711956, 711957, 711958, 711959, 711960, 711990, 711999, 712002, 712010, 712012, 712014, 712015, 712017, 712018, 712025, 712033, 712039, 712044, 712045, 712062, 712063, 712066, 712067, 712070, 712072, 712078, 712082, 712085, 712086, 712087, 712089, 712092, 712124, 712133, 712135, 712177, 712184, 712186, 712188, 712192, 712193, 712200, 712204, 712206, 712207, 712212, 712213, 712221, 712723, 712758, 712781, 712819, 712839, 712842, 712845, 712848, 712849, 712851, 712852, 712862, 712868, 712874, 712876, 712881, 712884, 712887, 712888, 712897, 712900, 712902, 712904, 712905, 712912, 712914, 712916, 712919, 712921, 712922, 712926, 712930, 712940, 712943, 712945, 712952, 712954, 713050, 713051, 713053, 713063, 713067, 713071, 713081, 713088, 713090, 713091, 713092, 713094, 713096, 713097, 713098, 713100, 713102, 713103, 713106, 713118, 713127, 713136, 713137, 713138, 713141, 713142, 713145, 713151, 713155, 713159, 713160, 713164, 713169, 713172, 713176, 713177, 713180, 713184, 713188, 713189, 713196, 713197, 713198, 713202, 713206, 713208, 713210, 713215, 713226, 713232, 713235, 713238, 713239, 713240, 713247, 713253, 713255, 713258, 713263, 713267, 713268, 713269, 713274, 713275, 713277, 713278, 713279, 713289, 713290, 713291, 713292, 713293, 713295, 713296, 713297, 713299, 713301, 713302, 713303, 713306, 713307, 713313, 713315, 713318, 713319, 713320, 713326, 713328, 713329, 713330, 713331, 713332, 713335, 713336, 713338, 713341, 713345, 713348, 713349, 713350, 713351, 713354, 713355, 713356, 713358, 713362, 713363, 713370, 713373, 713375, 713377, 713383, 713387, 713390, 713391, 713395, 713396, 713399, 713400, 713403, 713404, 713405, 713406, 713407, 713408, 713410, 713411, 713414, 713416, 713417, 713418, 713419, 713420, 713421, 713423, 713424, 713425, 713427, 713429, 713431, 713433, 713434, 713435, 713436, 713437, 713438, 713439, 713442, 713443, 713444, 713445, 713446, 713447, 713448, 713450, 713451, 713452, 713453, 713454, 713455, 713456, 713457, 713459, 713462, 713465, 713470, 713471, 713474, 713476, 713477, 713479, 713484, 713487, 713488, 713489, 713493, 713495, 713500, 713503, 713523, 713525, 713527, 713531, 713532, 713533, 713534, 713535, 713542, 713547, 713550, 713554, 713557, 713559, 713563, 713565, 713568, 713569, 713571, 713577, 713578, 713579, 713582, 713591, 713593, 713599, 713603, 713604, 713605, 713611, 713612, 713619, 713622, 713629, 713633, 713639, 713641, 713643, 713644, 713650, 713658, 713663, 713666, 713674, 713675, 713679, 713681, 713682, 713683, 713692, 713696, 713701, 713702, 713703, 713708, 713714, 713717, 713721, 713724, 713728, 713732, 713737, 713739, 713741, 713742, 713743, 713744, 713745, 713747, 713748, 713749, 713751, 713752, 713753, 713754, 713755, 713756, 713760, 713765, 713766, 713767, 713770, 713793, 713801, 713802, 713803, 713806, 713807, 713809, 713813, 713816, 713817, 713832, 713834, 713837, 713839, 713841, 713842, 713843, 713848, 713854, 713855, 713858, 713860, 713870, 713890, 713891, 713892, 713893, 713894, 713895, 713896, 713897, 713899, 713900, 713902, 713904, 713905, 713906, 713909, 713910, 713911, 713912, 713913, 713914, 713918, 713928, 713930, 713931, 713936, 713940, 713941, 713942, 713943, 713945, 713947, 713951, 713952, 714261, 714262, 714263, 714264, 714265, 714267, 714268, 714275, 714279, 714281, 714282, 714283, 714289, 714290, 714305, 714318, 714320, 714322, 714328, 714375, 714405, 714418, 714419, 714422, 714430, 714439, 714444, 714446, 714454, 714461, 714463, 714465, 714467, 714471, 714472, 714482, 714485, 714492, 714498, 714500, 714501, 714507, 714509, 714510, 714566, 714599, 714627, 714645, 714691, 714700, 714705, 714706, 714727, 714760, 714762, 714764, 714766, 714792, 714793, 714796, 714797, 714798, 714813, 714816, 714830, 714836, 714871, 714878, 714879, 714882, 714885, 714887, 714888, 714902, 714908, 718332, 718336, 718337, 718351, 718353, 726088, 726089, 726100, 726102, 726106, 726128, 726130, 726132, 726133, 726138, 726141, 726148, 726149, 726150, 726152, 726157, 726159, 726168, 726169, 726174, 726175, 726176, 726177, 726180, 726181, 726183, 726184, 726193, 726194, 726195, 726198, 726200, 726203, 726210, 726211, 726219, 726223, 726225, 726234, 726235, 726240, 726250, 726253, 726254, 726255, 726256, 726257, 726258, 726260, 726261, 726262, 726264, 726266, 726267, 726268, 726270, 726272, 726277, 726279, 726285, 726286, 726294, 726295, 726318, 726321, 726322, 726325, 726333, 726341, 726342, 726345, 726349, 726350, 726352, 726353, 726354, 726355, 726357, 726358, 726361, 726365, 726370, 726372, 726373, 726374, 726376, 726377, 726378, 726379, 726380, 726382, 726384, 726385, 726387, 726388, 726389, 726391, 726392, 726394, 726398, 726399, 726400, 726401, 726402, 726403, 726404, 726407, 726408, 726409, 726410, 726411, 726412, 726415, 726416, 726417, 726419, 726420, 726421, 726422, 726423, 726424, 726425, 726426, 726427, 726431, 726432, 726433, 726434, 726435, 726437, 726438, 726440, 726441, 726442, 726443, 726446, 726447, 726449, 726450, 726451, 726452, 726453, 726459, 726460, 726461, 726464, 726466, 726474, 726475, 726477, 726478, 726480, 726481, 726484, 726487, 726488, 726490, 726492, 726493, 726498, 726499, 726503, 726504, 726505, 726507, 726508, 726509, 726510, 726513, 726515, 726516, 726518, 726520, 726524, 726526, 726527, 726529, 726531, 726533, 726534, 726538, 726539, 726540, 726545, 726548, 726549, 726552, 726553, 726554, 726556, 726558, 726561, 726562, 726566, 726567, 726568, 726569, 726571, 726572, 726573, 726574, 726575, 726576, 726577, 726578, 726580, 726583, 726584, 726585, 726587, 726591, 726592, 726598, 726599, 726600, 726601, 726602, 726604, 726605, 726606, 726607, 726608, 726610, 726615, 726617, 726627, 726628, 726630, 726638, 726639, 726642, 726643, 726644, 726645, 726655, 726662, 726664, 726666, 726669, 726670, 726672, 726676, 726677, 726703, 726706, 726719, 726721, 726727, 726728, 726731, 726734, 726735, 726742, 726749, 726758, 726759, 726768, 726770, 726775, 726782, 726784, 726786, 726787, 726789, 726792, 726793, 726794, 726806, 726816, 726821, 726834, 726838, 726840, 726844, 726852, 726874, 726889, 726896, 726899, 726900, 726902, 726915, 726927, 726928, 726929, 726931, 726940, 726942, 727058, 727061, 727064, 727066, 727067, 727078, 727079, 727085, 727087, 727214, 727215, 727216
Recalling Firm/
Manufacturer
Boston Scientific Corporation
4100 Hamline Ave N Bldg 3
Saint Paul MN 55112-5700
For Additional Information Contact United States Technical Services
800-227-3422
Manufacturer Reason
for Recall
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.
FDA Determined
Cause 2
Nonconforming Material/Component
Action The firm began notifying its sales representatives on 06/02/2021 by email or hand delivery and followed with notifying the implanting/following physicians and centers on 06/03/2021 by letter. The letter explained the issue and provided recommendations for the physicians. Information is also posted on www.BostonScientific.com/advisory. Patient notification is done by the implanting/following physician.
Quantity in Commerce 5920 units
Distribution Worldwide distribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA Database PMAs with Product Code = NKE and Original Applicant = GUIDANT CORP.
-
-