• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall see related information
Date Initiated by Firm August 24, 2021
Create Date September 22, 2021
Recall Status1 Open3, Classified
Recall Number Z-2505-2021
Recall Event ID 88549
510(K)Number K834352  
Product Classification Catheter, steerable - Product Code DRA
Product DLP Left Heart Vent Catheters, 18 FR. Sterile, Rx Only
Code Information Model/Reference Number 12118, Lot Number (GTIN Number): 2018080452 (20643169086702), 2018080452 (00673978462950), 2018090435 (20643169086702), 2018090435 (00673978462950), 2018100525 (20643169086702), 2018100525 (00673978462950), 2018111024 (20643169086702), 2018111024 (00673978462950), 2018111421 (20643169086702), 2018111421 (00673978462950), 2019020369 (20643169086702), 2019020369 (00673978462950), 2019030426 (20643169086702), 2019030426 (00673978462950), 2019031168 (20643169086702), 2019031168 (00673978462950), 2019041190 (20643169086702), 2019041190 (00673978462950), 2019060424 (20643169086702), 2019060424 (00673978462950), 2019070714 (20643169086702), 2019070714 (00673978462950), 2019070715 (20643169086702), 2019070715 (00673978462950), 2019080686 (20643169086702), 2019080686 (00673978462950), 2019081292 (20643169086702), 2019081292 (00673978462950), 2019090394 (20643169086702), 2019090394 (00673978462950), 2019091081 (20643169086702), 2019091081 (00673978462950), 2019100159 (20643169086702), 2019100159 (00673978462950), 2019101446 (20643169086702), 2019101446 (00673978462950), 2019111051 (20643169086702), 2019111051 (00673978462950), 2019120418 (20643169086702), 2019120418 (00673978462950), 2020010051 (20643169086702), 2020010051 (00673978462950), 2020010446 (20643169086702), 2020010446 (00673978462950), 2020020094 (20643169086702), 2020020094 (00673978462950), 2020020764 (20643169086702), 2020020764 (00673978462950), 2020021113 (20643169086702), 2020021113 (00673978462950), 2020040403 (20643169086702), 2020040403 (00673978462950), 2020041125 (20643169086702), 2020041125 (00673978462950), 2020041431 (20643169086702), 2020041431 (00673978462950), 2020050353 (20643169086702), 2020050353 (00673978462950), 2020050919 (20643169086702), 2020050919 (00673978462950), 2020060370 (20643169086702), 2020060370 (00673978462950), 2020060673 (20643169086702), 2020060673 (00673978462950), 2020070374 (20643169086702), 2020070374 (00673978462950), 2020080075 (20643169086702), 2020080075 (00673978462950), 2020120620 (20643169086702), 2020120620 (00673978462950), 2021010035 (20643169086702), 2021010035 (00673978462950), 201906C177 (20643169086702), 201906C177 (00673978462950), 2017090238 (00673978462950), 2018080450 (20643169086702), 2018080451 (20643169086702), 2018090065 (20643169086702), 2018091021 (20643169086702), 2018100523 (20643169086702), 2018100524 (20643169086702), 2018120893 (20643169086702), 2019010457 (20643169086702), 2019010765 (20643169086702), 2019011214 (20643169086702), 2019020007 (20643169086702), 2019030064 (20643169086702), 2019030065 (20643169086702), 2019030721 (20643169086702), 2019031169 (20643169086702), 2019040844 (20643169086702), 2019040845 (20643169086702), 2019040846 (20643169086702), 2019041191 (20643169086702), 2019050167 (20643169086702), 2019050577 (20643169086702), 2019050846 (20643169086702), 2019051127 (20643169086702), 2019060219 (20643169086702), 2019060220 (20643169086702), 2019060687 (20643169086702), 2019060985 (20643169086702), 2019060986 (20643169086702), 2019070613 (20643169086702), 2019071115 (20643169086702), 2019081002 (20643169086702), 2019090072 (20643169086702), 2019090393 (20643169086702), 2019091025 (20643169086702), 2019100418 (20643169086702), 2019101095 (20643169086702), 2019110084 (20643169086702), 2019111050 (20643169086702), 2019120927 (20643169086702), 2020010052 (20643169086702), 2020011621 (20643169086702), 2020020472 (20643169086702), 2020030065 (20643169086702), 2020050354 (20643169086702), 2020050700 (20643169086702), 2020060067 (20643169086702), 2020060068 (20643169086702), 2020060964 (20643169086702), 2020060965 (20643169086702), 2020070055 (20643169086702), 2020070655 (20643169086702), 2020070969 (20643169086702), 2020071273 (20643169086702), 2020081034 (20643169086702), 2020090632 (20643169086702), 2020110762 (20643169086702), 2020120079 (20643169086702), 2020120621 (20643169086702), 2021010553 (20643169086702), 2021011168 (20643169086702), 2021011169 (20643169086702), 2021011170 (20643169086702), 2021031116 (20643169086702), 2021031308 (20643169086702), 2021031309 (20643169086702), 2021040093 (20643169086702), 2021040871 (20643169086702), 2021041149 (20643169086702), 2021050635 (20643169086702), 201901C051 (20643169086702), 201903C128 (20643169086702), 201905C013 (20643169086702), 201906C176 (20643169086702), 202011C198 (20643169086702), 202106C012 (20643169086702), 202106C013 (20643169086702), 202106C014 (20643169086702), 202106C020 (20643169086702), 202106C021 (20643169086702). Expiration date range: 09/30/2020 - 05/19/2024
Recalling Firm/
Manufacturer
Medtronic Perfusion Systems
7611 Northland Dr N
Brooklyn Park MN 55428-1088
For Additional Information Contact
763-391-9000
Manufacturer Reason
for Recall
Potential for a wire protrusion through the left heart vent catheter tip
FDA Determined
Cause 2
Process control
Action On about 08/24/2021, Medtronic notified customers via 2-day tracked mail letter titled, "URGENT MEDICAL DEVICE RECALL DLP Left Heart Vent Catheters." The letter instructed customers identify and quarantine all affected DLP Left heart Vent Catheters, return all unused affected product in inventory to Medtronic by contacting Medtronic Customer Service at 1-800-854-3570 and to contact your local Medtronic Representative can assist you in the return of this product, and complete and return the provided Customer Confirmation Form and email to rs.cfqfca@medtronic.com. Additional instructions included to notify anyone within your organization that should be aware or to any organization where the potentially affected devices have been transferred.
Quantity in Commerce 48,672 units
Distribution Nationwide distribution to AL, CA, CO, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI. International distribution to Albania, Algeria, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, Colombia, Costa Rica, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Latvia, Lebanon, Libya, Luxembourg, Macedonia, Malaysia, Mauritius, Mexico, Morocco, Myanmar, Nepal, New Zealand, Nigeria, Norway, Pakistan, Philippines, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Viet Nam, Yemen.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = DRA and Original Applicant = DLP, INC.
-
-