Class 2 Device Recall Planer BT37 Incubator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Planer BT37 Incubator see related information
Date Initiated by Firm August 12, 2022
Date Posted September 28, 2022
Recall Status1 Open3, Classified
Recall Number Z-1822-2022
Recall Event ID 90782
510(K)Number K121566  
Product Classification Accessory, assisted reproduction - Product Code MQG
Product Humidifier bottle and syringe filter kit (6 count of each) for the BT37
Mark I Benchtop Incubator (AY102295).

The Mark I model is no longer produced, but the legacy humidifier
bottle component design can also be used with the BT37 Mark II
Benchtop Incubator.
Code Information Lot Numbers 05593V291121 18-1192 1812-68 18-1268 G000575 G003811 G003893 G004517 G004874 G005109 G005476 G005723 E160415 Expiration date range: 27 Jun 2022 to 07 Dec 2024
Recalling Firm/
Manufacturer
CooperSurgical, Inc.
95 Corporate Dr
Trumbull CT 06611-1350
For Additional Information Contact Customer Service
203-601-5200 Ext. 03300
Manufacturer Reason
for Recall
There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.
FDA Determined
Cause 2
Under Investigation by firm
Action On August 12, 2022, the firm notified customers via an Urgent Medical Device Safety Notice. Customers were instructed to quarantine affected product in their inventory. Once customers complete and return the Acknowledgement and Receipt Form attached to the recall letter, the firm will arrange for the affected product to be returned at no cost. Customers will receive credit for returned affected product. CooperSurgical may be reached at 203-601-5200 ext. 03300 or recall@coopersurgical.com.
Quantity in Commerce 5,182 kits
Distribution Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS MA MI MN MO NJ NY OH OR SC TX VA VT WA WI WV Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database 510(K)s with Product Code = MQG and Original Applicant = PLANER PLC
-
-