Date Initiated by Firm |
August 12, 2022 |
Date Posted |
September 28, 2022 |
Recall Status1 |
Open3, Classified |
Recall Number |
Z-1822-2022 |
Recall Event ID |
90782 |
510(K)Number |
K121566
|
Product Classification |
Accessory, assisted reproduction - Product Code MQG
|
Product |
Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295).
The Mark I model is no longer produced, but the legacy humidifier bottle component design can also be used with the BT37 Mark II Benchtop Incubator. |
Code Information |
Lot Numbers
05593V291121
18-1192
1812-68
18-1268
G000575
G003811
G003893
G004517
G004874
G005109
G005476
G005723
E160415
Expiration date range: 27 Jun 2022 to 07 Dec 2024 |
Recalling Firm/ Manufacturer |
CooperSurgical, Inc. 95 Corporate Dr Trumbull CT 06611-1350
|
For Additional Information Contact |
Customer Service 203-601-5200 Ext. 03300
|
Manufacturer Reason for Recall |
There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.
|
FDA Determined Cause 2 |
Under Investigation by firm |
Action |
On August 12, 2022, the firm notified customers via an Urgent Medical Device Safety Notice.
Customers were instructed to quarantine affected product in their inventory. Once customers complete and return the Acknowledgement and Receipt Form attached to the recall letter, the firm will arrange for the affected product to be returned at no cost. Customers will receive credit for returned affected product.
CooperSurgical may be reached at 203-601-5200 ext. 03300 or recall@coopersurgical.com. |
Quantity in Commerce |
5,182 kits |
Distribution |
Domestic distribution to the following states:
AL
AZ
CA
CT
DE
FL
GA
HI
IL
KS
MA
MI
MN
MO
NJ
NY
OH
OR
SC
TX
VA
VT
WA
WI
WV
Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
510(K) Database |
510(K)s with Product Code = MQG and Original Applicant = PLANER PLC
|