Class 3 Device Recall Baxter Meridian

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 3 Device Recall Baxter Meridian see related information
Date Initiated by Firm November 15, 2002
Date Posted March 05, 2003
Recall Status1 Terminated 3 on September 19, 2003
Recall Number Z-0605-03
Recall Event ID 25079
510(K)Number K961700  
Product Classification Dialyzer, High Permeability With Or Without Sealed Dialysate System - Product Code KDI
Product Baxter Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, McGaw Park, IL 60085

The instrument is part of a high permeability hemodialysis system which consists of a controlled dialysate delivery system that incorporates an ultrafiltration controller to prevent excessive loss of water from the patient''s blood, an extracorporeal blood set, and a high permeability dialyzer. The standard features of the Meridian instrument incluide a high blood flow rate capacity (for shortened hemodialysis treatment time), automatic ultrafiltration control, standard and variable bicarbonate and sodium capabilities anbd automated chemical disinfection. The Meridian will operate in either the bicarbonate or acetate mode for concentrates.
Code Information product codes 5M5576, 5M5576R, serial numbers 200000 through 202864
Recalling Firm/
Manufacturer
Baxter Healthcare Corp.
Rt 120 & Wilson Rd
Round Lake IL 60073
For Additional Information Contact Center for One Baxter
800-422-9837
Manufacturer Reason
for Recall
Meridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure.
FDA Determined
Cause 2
Other
Action Baxter issued a Safety Alert letter dated 11/15/02 to the direct accounts informing them of the potential for incomplete disinfection of the Meridian instrument during the Renalin/Formalin disinfection cycle due to electomagnetic interference (EMI). The accounts were informed that they will be receiving a radio frequency interference (RFI) filter kit and installation instructions to eliminate EMI by 2/28/03. The accounts were given instructions to follow to ensure that the Renalin/Formalin disinfection cycle has been completed prior to use of the machine.
Quantity in Commerce 2864 units
Distribution Nationwide, Mexico, China, South Korea, Hong Kong
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.