Date Initiated by Firm |
November 15, 2002 |
Date Posted |
March 05, 2003 |
Recall Status1 |
Terminated 3 on September 19, 2003 |
Recall Number |
Z-0605-03 |
Recall Event ID |
25079 |
510(K)Number |
K961700
|
Product Classification |
Dialyzer, High Permeability With Or Without Sealed Dialysate System - Product Code KDI
|
Product |
Baxter Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, McGaw Park, IL 60085
The instrument is part of a high permeability hemodialysis system which consists of a controlled dialysate delivery system that incorporates an ultrafiltration controller to prevent excessive loss of water from the patient''s blood, an extracorporeal blood set, and a high permeability dialyzer. The standard features of the Meridian instrument incluide a high blood flow rate capacity (for shortened hemodialysis treatment time), automatic ultrafiltration control, standard and variable bicarbonate and sodium capabilities anbd automated chemical disinfection. The Meridian will operate in either the bicarbonate or acetate mode for concentrates. |
Code Information |
product codes 5M5576, 5M5576R, serial numbers 200000 through 202864 |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corp. Rt 120 & Wilson Rd Round Lake IL 60073
|
For Additional Information Contact |
Center for One Baxter 800-422-9837
|
Manufacturer Reason for Recall |
Meridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure.
|
FDA Determined Cause 2 |
Other |
Action |
Baxter issued a Safety Alert letter dated 11/15/02 to the direct accounts informing them of the potential for incomplete disinfection of the Meridian instrument during the Renalin/Formalin disinfection cycle due to electomagnetic interference (EMI). The accounts were informed that they will be receiving a radio frequency interference (RFI) filter kit and installation instructions to eliminate EMI by 2/28/03. The accounts were given instructions to follow to ensure that the Renalin/Formalin disinfection cycle has been completed prior to use of the machine. |
Quantity in Commerce |
2864 units |
Distribution |
Nationwide, Mexico, China, South Korea, Hong Kong |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|