Date Initiated by Firm |
July 01, 2004 |
Date Posted |
July 20, 2004 |
Recall Status1 |
Terminated 3 on February 22, 2006 |
Recall Number |
Z-1052-04 |
Recall Event ID |
29479 |
PMA Number |
P030025 |
Product Classification |
Coronary Drug-Eluting Stent - Product Code NIQ
|
Product |
TAXUS Express Paclitaxel-Eluting MONORAIL Coronary Stent System (3.5 x 32mm).
On August 5, 2004 an additional 38 lots were added to this recall, the following TAXUS Express2 Paclitaxel-Eluting Coronary Stent System Part Numbers correspond to the lot numbers listed at the end of the code information below. US part# H7493897024250 H7493897032270 H7493897020300 H7493897020350 H7493897032300 H7493897020350 H7493897024300 H7493897020300 H7493897020300 H7493897032350 H7493897020250 H7493897020270 H7493897024300 H7493897032350 H7493897032270 H7493897020350 H7493897024300 H7493897024250 H7493897024250 H7493897020300 H7493897032350 H7493897032350 H7493897024350 H7493897024250 H7493897024250 H7493897016300 H7493897032350 H7493897024350 H7493897016350 H7493897032300 H7493897016350 H7493897016350 H7493897024350 H7493897016250 H7493897016250 H7493897024300 H7493897024350 H7493897016250 OUS part# H7493800132270 H7493800132270 H7493800132350 H7493800132270 H7493800124350 |
Code Information |
2.25 mm U.S Distributed - 5726202 6425946 2.5 mm U.S Distributed - 5477926 6068099 6068253 6068254 6068255 6068256 6068728 6080205 6068097 6068257 6089076 6089077 6089078 6089079 6090735 6090736 6090737 6090772 6090773 6090774 6090775 6096845 6096847 6096848 6096849 6096850 6097162 6097571 6097573 5922837 6067850 6068100 6068101 6068162 6068252 6068729 6096801 6096842 6096843 6097570 6068098 6127643 6127911 6127912 6127913 6127914 6127918 6127919 6127920 6127921 6128082 6137930 6137932 6138144 6090771 6127645 6127646 6127910 6127915 6127916 6127917 6137926 6137927 6137928 6137929 6138142 6138143 6138145 6067852 6090770 6096846 6096851 6097164 6127647 6215687 6219522 6219828 6220191 6221396 6222937 6226390 6211538 6219953 6224031 6227153 6227595 6263487 6268176 6271813 6272275 6272489 6285158 6289753 6294773 6298392 6328349 6334370 6336268 6332430 6334861 6337078 6335227 6336395 6336584 6337085 6337169 6338384 6341248 6343746 6345608 6346309 6347434 6349873 6352496 6340012 6341609 6343753 6350109 6350803 6350804 6354617 6354857 6352487 6354321 6356146 6358289 6367306 6406591 6420365 6424726 6425117 6425679 6430286 6434917 6440527 6443189 6443708 6491398 6491755 6496654 6500077 6500493 6504635 6506390 6509266 6501051 6503684 6508987 6509275 6509499 6513651 6517766 6524586 6513689 6518413 6521016 6522868 6524901 6527042 6527170 6529133 6531403 6062710 6063137 6063139 6063454 6063456 6063457 6063752 6063755 6063759 6068733 6069001 6069205 6070210 6070217 6070644 6071692 6071697 6080104 6080107 6080119 6071695 6071696 6080081 6068732 6068999 6069207 6070212 6070214 6071693 6080121 6081088 6081089 6081092 6087902 6087903 6087905 6088342 6088343 6088345 6104741 6104763 6105547 6102098 6102100 6105532 6105534 6105542 6105543 6105545 6063138 6063141 6063453 6081091 6087904 6087906 6088301 6104762 6104764 6105533 6068730 6069206 6070211 6070213 6071694 6088344 6105157 6105546 6063758 6069203 6102096 6105544 6182057 6188086 6183275 6192726 6193570 6202474 6206312 6206318 6207480 6190829 6193027 6194497 6197487 6203688 6200769 6201701 6210973 6204771 6206180 6207566 6209505 6211129 6219532 6223461 6226310 6227703 6228403 6254673 6255430 6229540 6233362 6238271 6246020 6237601 6237910 6240836 6242462 6253671 6263889 6256244 6259357 6261235 6263427 6270190 6273883 6276698 6281408 6293469 6315632 6316255 6319064 6311247 6316123 6317341 6337197 6340841 6341954 6346431 6337194 6337195 6342016 6345444 6345621 6346321 6350196 6359472 6363167 6367308 6370185 6375439 6360860 6367876 6373464 6377126 6377546 6381981 6391419 6395897 6380820 6381894 6382749 6404318 6404740 6405007 6406451 6406740 6406956 6408571 6411193 6400608 6405020 6406318 6409330 6406599 6406917 6407122 6422583 6411200 6414127 6424369 6425118 6425733 6426147 6429680 6423110 6425647 6426157 6430284 6430298 6434726 6432082 6434729 6438986 6442131 6445409 6449085 6450065 6441216 6443567 6446843 6448700 6449016 6504711 6508782 6509344 6513691 6517622 6520461 6522737 6526845 6529627 6062538 6066829 6066830 6066831 6066832 6066836 6067673 6067719 6067720 6072003 6072018 6072569 6072809 6072810 6073002 6073006 6076474 6076477 6062533 6062537 6067674 6072483 6073545 6066833 6067672 6072017 6072482 6073007 6073009 6073550 6076475 6076479 6062535 6072571 6073547 6094790 6095165 6115463 6115469 6118819 6118821 6119001 6119374 6130041 6130083 6066838 6072361 6094787 6095164 6106147 6115468 6119378 6072808 6073011 6094788 6094791 6115464 6115467 6119166 6119375 6130040 6138103 6138104 6066834 6066835 6067718 6076473 6094789 6094792 6095163 6095166 6095167 6118816 6119102 6119376 6130085 6138001 6138102 6159913 6161628 6163401 6163724 6167160 6169241 6171341 6066837 6072570 6073012 6106148 6106149 6106150 6106151 6106248 6115465 6137999 6138000 6138972 6138973 6154661 6071901 6072019 6159356 6160781 6167910 6170269 6170964 6176166 6178014 6178060 6179071 6020266 6020268 6020270 6020273 6020274 6020468 6020472 6020525 6020528 6020536 6020540 6020582 6020585 6020587 6020592 6020596 6020601 6020602 6020605 6020606 6020612 6020613 6020614 6020618 6020619 6020620 6020621 6020646 6020648 6020652 6020653 6178493 6179288 6180215 6180830 6181149 6181481 6181941 6182619 6181269 6181823 6193023 6201962 6190379 6196395 6199497 6201465 6202412 6204229 6205414 6206080 6201966 6205616 6196989 6199730 6204612 6215685 6217153 6220132 6221016 6227861 6232359 6233245 6241875 6246032 6249284 6249687 6250802 6251246 6252186 6255089 6258866 6249215 6251856 6254666 6253176 6258862 6259498 6262313 6263749 6263901 6268825 6272413 6281524 6291200 6299414 6299869 6303217 6310712 6312227 6294094 6299371 6301100 6303214 6306879 6319705 6324003 6328037 6328866 6315752 6319901 6329072 6337384 6341539 6342039 6345335 6349515 6352714 6357669 6357801 6406313 6389426 6403961 6429245 6431253 6447421 6448463 6448468 6448694 6448696 6448857 6450691 6451225 6453422 6448697 6448868 6452364 6455655 6451435 6453641 6457296 6458448 6469578 6470205 6464818 6467214 6468584 6469579 6470079 6468189 6470573 6470666 6471371 6472558 6475827 6465583 6479265 6487615 6489089 6489088 6495427 6500494 6496666 6503220 6504624 6518566 6529633 6530081 6530494 6509865 6517074 6518401 6520348 6523579 6179166 6073831 6074548 6077012 6077013 6077017 6078101 6078350 6079006 6079007 6314880 6315304 6073829 6074544 6074545 6074547 6074931 6074933 6077014 6077016 6079005 6104456 6104849 6105304 6105308 6074932 6074935 6077015 6078355 6079009 6105310 6074543 6074936 6078346 6105307 6105309 6104847 6104850 6104851 6104852 6105159 6105311 6074934 6104455 6104848 6119103 6119120 6119121 6119514 6119516 6119517 6123055 6123293 6123443 6123444 6123445 6131667 6105158 6105305 6119513 6119515 6123290 6131665 6131666 6131670 6134235 6134238 6134240 6138278 6157302 6165847 6119117 6123050 6123052 6123053 6123291 6123294 6136373 6119118 6123054 6131669 6138277 6167310 6167731 6119119 6123292 6131668 6134232 6138279 6157574 6158945 6123051 6134233 6136374 6138276 6138280 6159300 6074546 6178654 6181828 6182612 6171430 6171662 6173391 6174426 6176979 6179072 6179426 6183444 6187635 6182181 6182614 6182894 6196803 6199824 6178954 6178960 6178961 6179162 6179163 6179168 6179174 6179182 6219723 6220505 6223620 6227593 6221313 6223148 6224776 6226649 6227322 6241318 6228598 6229603 6233895 6245502 6252030 6258672 6259235 6266116 6267449 6267903 6270707 6271351 6272415 6271352 6272289 6277080 6274243 6276423 6277568 6282796 6289381 6276898 6277131 6277502 6278150 6279106 6280858 6281845 6285600 6287286 6290284 6291982 6294333 6294093 6300562 6302821 6305563 6307926 6322949 6331705 6319385 6320364 6324443 6340830 6363936 6376945 6378916 6382626 6383115 6383118 6385290 6386859 6387508 6391441 6402590 6403676 6391933 6395611 6404305 6406228 6420639 6425645 6439805 6446964 6457284 6460729 6464712 6461772 6467211 6480092 6483619 6487919 6488000 6489091 6478164 6483227 6485701 6488001 6489163 6492173 6496608 6504788 6506741 6507891 6508151 6508794 6509382 6509395 6529145 6530618 6531065 6531406 6531552 6532426 6536586 6522736 6525733 6529756 6530079 6530496 6531000 6531404 6531555 6524592 6527052 6529939 6534485 6539043 6525736 6530167 6531819 6536592 6536884 6536893 6545544 6548017 6080672 6086278 6086279 6087703 6080669 6086276 6087124 6087125 6087702 6087704 6104280 6104544 6105683 6105685 6105762 6105845 6087126 6087127 6104281 6104542 6104543 6104545 6105758 6105844 6107527 6108006 6080671 6086277 6105421 6105684 6105759 6105760 6105847 6107287 6107528 6108009 6105842 6105846 6105962 6105963 6107289 6107526 6107529 6107286 6107288 6105419 6105420 6105460 6105682 6105757 6158018 6167732 6170974 6162374 6163829 6105461 6105761 6163281 6171824 6172037 6175127 6175624 6167319 6167728 6170177 6176971 6177280 6178016 6178861 6171690 6171917 6175422 6175627 6176338 6178011 6176171 6176962 6177737 6179798 6180795 6181022 6181166 6181167 6181524 6181589 6182334 6182615 6182651 6183146 6183326 6183327 6187627 6192642 6193725 6204770 6205044 6206118 6206181 6176684 6179194 6179198 6179199 6179201 6179224 6179228 6179232 6179233 6179234 6179235 6179237 6211127 6211138 6211977 6213192 6214935 6217901 6218635 6219239 6229480 6230271 6231907 6232800 6233029 6237722 6238111 6254006 6247150 6251257 6263422 6263896 6272036 6272404 6276174 6285387 6294879 6305425 6308294 6316754 6317115 6317338 6318141 6319196 6322942 6323736 6309631 6310056 6312580 6316133 6317353 6319857 6323735 6325701 6327220 6317834 6321714 6323590 6324230 6327744 6330737 6346100 6328163 6334862 6335177 6368319 6372207 6371750 6372710 6395514 6396080 6380157 6425387 6425966 6441207 6438997 6443564 6443566 6444584 6447056 6447724 6447789 6448538 6446945 6449707 6452681 6456397 6452884 6457292 6460256 6460954 6461519 6464812 6466455 6468379 6468778 6499971 6500489 6503682 6507663 6508152 6511161 6513790 6521423 6509696 6513650 6513944 6514165 6516499 6517726 6518353 6519988 6520453 6526166 6531777 6531786 6531812 6531816 6540958 6536838 6539113 6543291 2.75 mm U.S Distributed - 6062532 6066570 6066571 6066572 6066573 6066574 6066575 6066576 6066577 6066578 6067856 6067858 6067859 6067860 6080700 6091400 6091401 6091469 6091473 6091697 6067857 6080701 6091394 6091398 6091399 6091463 6091464 6091467 6091468 6091470 6127652 6127653 6128211 6128215 6128220 6128221 6128262 6136139 6136140 6136141 6136362 6136364 6127648 6127649 6127650 6128213 6128218 6128219 6136363 6166797 6170583 6066569 6091395 6127651 6218719 6220180 6222767 6224279 6227426 6224027 6228934 6244894 6244895 6268621 6285480 6293468 6327292 6328019 6336994 6331431 6334192 6335702 6336391 6337168 6331477 6334742 6336828 6343745 6358564 6414135 6438991 6503529 6507127 6514143 6525420 6062018 6062020 6062021 6062184 6065427 6065618 6065619 6068990 6065426 6065617 6068991 6068994 6070126 6080355 6084630 6065310 6080357 6084628 6068988 6068989 6084631 6084632 6084634 6098144 6098477 6098478 6084623 6084625 6084626 6098143 6098474 6098479 6098850 6068992 6070124 6070129 6084629 6098475 6098476 6098851 6068993 6068995 6070128 6098145 6098849 6182112 6182645 6183774 6202628 6206829 6187841 6197324 6200146 6205613 6201977 6204589 6206832 6206828 6220018 6224762 6228537 6249846 6251346 6229569 6233260 6240647 6248138 6252664 6256256 6255632 6268163 6273022 6280850 6312228 6312507 6312030 6317112 6346809 6339033 6347700 6348683 6382115 6413666 6424739 6437831 6428346 6448367 6503836 6505102 6507148 6508042 6548718 6549717 6062336 6065714 6065715 6067651 6062335 6065717 6065719 6067652 6074561 6073841 6074558 6062334 6065713 6065720 6067653 6073608 6073609 6073610 6073839 6074560 6099439 6099442 6099446 6106154 6106155 6106591 6106596 6113807 6113808 6113809 6113810 6129701 6129782 6129799 6133118 6133427 6133428 6133865 6065716 6073840 6074722 6099443 6106153 6106592 6106593 6106595 6099441 6106157 6106594 6113805 6133110 6133866 6133867 6129380 6129381 6129422 6129426 6131043 6131047 6133109 6133111 6133120 6133425 6099440 6133864 6163797 6170275 6073605 6106156 6113806 6129423 6129424 6129802 6129803 6133112 6133426 6167308 6170622 6171128 6177822 6178974 6179427 6179931 6180470 6180479 6181826 6182240 6182892 6129783 6182618 6182899 6190943 6200096 6199644 6201205 6205413 6195065 6196659 6199210 6199732 6201745 6203685 6206063 6213488 6232704 6241344 6243742 6228288 6230913 6237592 6250445 6250444 6254888 6249442 6250817 6256627 6264544 6272562 6293031 6294323 6304740 6308262 6332201 6333273 6317396 6324002 6327382 6331336 6337086 6339515 6356626 6360010 6384344 6405145 6412760 6425732 6447055 6448374 6448916 6456335 6456388 6459353 6461233 6461780 6456113 6460877 6466133 6468804 6469730 6474119 6465380 6465584 6469723 6466036 6467497 6468730 6468740 6468745 6470074 6470193 6470403 6474376 6474826 6477179 6478331 6478970 6483223 6489232 6491377 6499100 6491757 6495565 6522250 6524111 6517457 6077425 6077427 6077428 6077429 6078089 6078093 6078094 6080006 6080109 6077424 6077426 6078087 6078091 6080008 6080009 6080111 6103214 6103215 6103267 6104088 6104204 6104205 6104207 6106251 6106252 6079629 6078085 6078090 6079627 6079628 6080005 6103212 6104086 6104087 6104089 6104206 6078092 6103213 6103216 6103264 6103266 6103268 6103263 6106253 6120751 6120755 6128714 6129039 6129041 6129184 6129687 6103265 6106250 6119970 6119971 6128784 6129685 6135821 6119972 6120752 6122418 6128716 6128717 6129040 6129182 6129198 6129686 6135802 6135810 6120754 6128713 6128719 6128782 6129199 6129201 6129262 6135804 6166937 6167737 6171736 6106249 6120742 6120753 6122550 6128715 6128720 6128721 6129183 6129200 6119969 6128718 6128783 6128785 6135814 6135816 6163387 6178977 6181264 6182115 6170839 6171691 6171966 6172157 6174422 6176967 6177532 6177718 6177827 6182911 6184081 6191025 6196390 6215370 6226684 6232966 6237257 6243759 6233246 6252022 6253425 6255009 6258662 6259704 6267527 6267616 6272240 6272277 6272565 6281361 6276594 6283157 6285703 6289063 6289610 6290850 6289886 6298989 6301583 6307700 6295682 6307327 6321938 6323738 6328351 6336685 6337390 6335158 6337113 6381655 6382888 6425739 6426725 6448699 6458289 6453633 6467494 6491371 6491615 6480505 6491320 6494213 6509780 6531809 6534328 6530498 6531826 6528519 6528894 6082489 6086275 6082485 6082491 6086274 6104829 6104834 6082492 6104830 6104833 6106142 6107717 6107720 6082486 6082490 6104832 6106144 6106145 6107715 6107716 6107718 6108538 6106133 6106136 6108527 6108528 6108530 6108531 6108534 6134444 6134445 6134446 6104831 6106146 6106139 6107712 6108535 6108541 6171925 6158944 6106134 6106141 6108662 6172501 6176010 6167725 6171738 6176241 6180078 6175705 6177738 6180466 6182313 6182910 6205421 6215683 6215684 6216737 6230691 6233244 6240642 6247869 6251861 6259711 6270990 6272286 6273024 6286322 6295522 6296383 6299166 6308276 6321924 6327996 6331704 6336189 6373719 6442930 6448275 6450510 6455155 6457290 6464432 6465515 6460632 6509824 6517305 6512800 6523448 6524832 6078658 6078783 6078896 6101283 6078659 6078660 6078755 6078760 6078796 6078887 6078894 6100593 6101241 6100595 6101282 6101286 6101292 6104551 6104554 6106194 6106195 6100596 6100597 6104552 6104553 6104556 6106196 6106197 6104555 6118362 6118557 6118558 6118678 6127024 6118679 6127026 6127028 6127098 6118341 6118369 6118372 6127027 6127101 6127163 6101293 6118340 6118559 6127025 6127099 6127100 6194933 6196799 6197267 6197497 6206307 6207140 6241232 6248239 6251303 6260132 6262938 6267568 6270712 6270989 6272290 6295144 6297418 6299442 6303218 6307543 6315354 6327629 6331584 6332019 6336017 6354789 6357923 6358746 6358888 6359570 6361142 6376855 6377140 6395926 6402389 6429879 6434918 6438995 6438792 6443094 6443410 6443411 6443681 6444712 6455861 6461078 6465311 6483334 6504626 6507109 6508850 6509281 6509635 6531497 6537009 6090874 6090875 6090876 6090877 6090878 6090879 6090880 6105453 6105454 6105455 6105456 6105457 6105458 6105459 6115295 6117908 6117909 6118268 6118271 6118272 6118273 6118275 6115292 6115294 6115502 6118267 6159444 6161942 6115293 6115503 6115504 6118270 6118274 6023183 6023207 6023208 6117910 6182055 6182178 6206403 6209969 6214094 6206177 6206405 6209338 6218128 6234371 6238263 6233363 6237731 6249451 6249508 6255091 6258677 6242472 6248436 6264262 6251244 6255784 6266934 6281481 6282463 6276593 6277574 6282591 6307997 6315527 6316127 6333624 6334981 6336398 6346685 6421329 6403865 6406913 6423561 6424594 6427790 6439512 6440724 6442671 6448462 6461091 6465389 6470940 6475251 6492764 6500162 6502601 6508560 6509092 6508848 6509268 6509636 6509677 6514305 3.0 mm U.S Distributed - 5695814 5695967 5695975 5696041 6060777 6064351 6064352 6064387 5887312 5887317 5887335 5887570 5887571 5887574 5887575 5887623 5887627 5887806 5887807 5887808 5887810 5887813 5887815 5887881 5887887 5887893 5888169 5888170 5888172 5888175 5888228 5888236 5888291 5888463 5888464 5888465 5888466 5888475 5888477 5888541 5888542 5888544 6067903 6067905 6067906 6067907 6067908 6067910 6067911 6067913 6080284 6080298 6080299 6084137 6080297 6080300 6080301 6080342 6084133 6084136 6095664 6095666 6096184 6096434 6099922 6099927 5887883 5887886 5887892 5888174 5888546 5888549 6067861 6067902 6067904 6067909 6067912 6067914 6084134 6084135 6087705 6095665 6095667 6096182 6096183 6099917 6099920 6099928 5887336 5887573 5887889 5888177 5888467 5888479 5888545 5970919 5971133 5971137 5971138 5971139 5971195 5971196 5971197 5971199 5971743 5971744 5971745 5971746 5971747 5971748 5971749 5971750 5971974 5971977 5972032 5972123 5972124 5972131 5972135 5972136 5972140 5972225 5972226 5972228 5972230 5972231 5972233 5972239 5972329 5972332 6127755 6127756 6127759 6127760 6128134 6128205 6128206 6129473 6129476 6137081 6137082 6099925 6127758 6128207 6128208 6129474 6129475 6137084 6137085 6169963 6177877 6168199 6176978 6127754 6127757 6137083 6181030 6181190 6182648 6185981 6184294 6192033 6228330 6223931 6227596 6228896 6229165 6224140 6228284 6229035 6273220 6283341 6294215 6303707 6294213 6295156 6299168 6326030 6335161 6336585 6336662 6337767 6334969 6336775 6339521 6342727 6344524 6341088 6345330 6346316 6347992 6351976 6353617 6354476 6354675 6357758 6357848 6367944 6368308 6372206 6372708 6380073 6386615 6392546 6406916 6411054 6422166 6424177 6406915 6406945 6410337 6411348 6421328 6425684 6425390 6428536 6430723 6433000 6441673 6444022 6446874 6447054 6447613 6447846 6496415 6502210 6504259 6500864 6504796 6505073 6505718 6507520 6508045 6508551 6508928 6512782 6518559 6521163 6517076 6528334 6530072 6530275 6530664 6530990 6531602 6062192 6062194 6062196 6062197 6065431 6065529 6066042 6066043 6066614 6068527 6080423 6080426 6065528 6065560 6066044 6077211 6077212 6077214 6077215 6080427 6080798 6081448 6084879 6085549 6065312 6065314 6080799 6081367 6081375 6087816 6087818 6087820 6087821 6088066 6089325 6065429 6065561 6068529 6080424 6080425 6080801 6080842 6085546 6085548 6087720 6087807 6088068 6089326 6089327 6089328 6099355 6102901 6103102 6104697 6104699 6104700 6104701 6105522 6084878 6085547 6087721 6087806 6087815 6087817 6087819 6088067 6089324 6098520 6098521 6098522 6099352 6102897 6102900 6103103 6068533 6077213 6081366 6087803 6098524 6099353 6102899 6171521 6172780 6081372 6098523 6098525 6099354 6099356 6099357 6183328 6186942 6183157 6183329 6187631 6203261 6205418 6192343 6198629 6202491 6205546 6202683 6206277 6211879 6214375 6217144 6224005 6227172 6227709 6228618 6224210 6227098 6229409 6236713 6242251 6232187 6232888 6239090 6245817 6255088 6261911 6263946 6253168 6257866 6259356 6268019 6271904 6273457 6276687 6279952 6290311 6294646 6315716 6317113 6317662 6307995 6316295 6337199 6342944 6345615 6350884 6352343 6354186 6354188 6341612 6342477 6350170 6350805 6350882 6351636 6353839 6354730 6354073 6356855 6358436 6358885 6358887 6359053 6359111 6362132 6363923 6363962 6364260 6367303 6368130 6369189 6358892 6359358 6364721 6364314 6368310 6370650 6371963 6372406 6372707 6375337 6376872 6379349 6382698 6382896 6383236 6381886 6382754 6383591 6386637 6389916 6392015 6382752 6386769 6386913 6405870 6410659 6399309 6402400 6403987 6405144 6411368 6423915 6410941 6420749 6421322 6421831 6423913 6424271 6425384 6425542 6425689 6425699 6426108 6429980 6433830 6434723 6438796 6439736 6424436 6425200 6429315 6430297 6425695 6425967 6429419 6430418 6434522 6435335 6439465 6440735 6444253 6447060 6448396 6448698 6452070 6453310 6442659 6446160 6447059 6448395 6449901 6509493 6509864 6514050 5439807 5439813 5440870 5444444 5444454 5478733 5574205 5711075 5712353 5712355 5712641 5890712 5890713 5890714 5890715 5891006 5891007 5891008 5891009 5891010 5891011 5891013 5891150 5891327 5891329 5891332 5891335 5891340 5891421 5892634 5892638 5892639 5892640 5892862 5892863 5892865 5892985 5892987 5892988 5892990 5892991 5892997 5892998 5892999 6062017 6062182 6062185 6064305 6065315 6065611 6069383 6069385 6071897 6072811 6062019 6065541 6065602 6065603 6065612 6070138 6070208 6071048 6071049 6071691 6063671 6064241 6069401 6071053 6071688 6071783 6072353 6072566 6072567 6072907 6072908 6063670 6063673 6064306 6064308 6070141 6070206 6070207 6071784 6071896 6072812 6072905 6072906 6111422 6111424 6111434 6112901 6127037 6127039 6128678 6128769 6129037 6129255 6129406 6137864 6063663 6065316 6065540 6071687 6071689 6072352 6072568 6111428 6111429 6112892 6112895 6112896 6112899 6113023 6127035 6127307 6069389 6069391 6070204 6072355 6112894 6112898 6112900 6113022 6127034 6127306 6127312 6128762 6128763 6128766 6129038 6129252 6129253 6129402 6129403 6071690 6071782 6111427 6111432 6111433 6128671 6128673 6128765 6128767 6129034 6129035 6129251 6129254 6137862 6160359 6167909 6170397 6171129 6171733 6069390 6071686 6072813 6111426 6111430 6112897 6127040 6127308 6128674 6128677 6128679 6128680 6128768 6129033 6129036 6129256 6129405 6129407 6137741 6137863 6071898 6159914 6160957 6163784 6169784 6171585 5891012 5892637 5974148 5974149 5974150 5974152 5974856 5974924 5974925 5974926 5974932 5974933 5975057 5975058 5975060 5975141 5975142 5975143 5975198 5975199 5975200 5975582 5975583 5975813 5975831 5977073 5977074 5977079 5977081 5977262 5977263 5977756 5978448 5978450 6175716 6176980 6177278 5974860 5975838 5975840 5975890 5975891 5977072 5977755 6178203 6181268 6181968 6181990 6182053 6129404 6181531 6182116 6182610 6182643 6182649 6182891 6187112 6188458 6193022 6191533 6192347 6197327 6201640 6204238 6187622 6188396 6188926 6192398 6196389 6197273 6201203 6202808 6202812 6204290 6204772 6206062 6206077 6214378 6231149 6236570 6237336 6241084 6241992 6227858 6245054 6250307 6250451 6256418 6249290 6251251 6255011 6260798 6262934 6270995 6271696 6277337 6283435 6303865 6307561 6311713 6292699 6293117 6295025 6311706 6314629 6372404 6376104 6317678 6323195 6327291 6331207 6316284 6317663 6319704 6337389 6346319 6332092 6353387 6354533 6357231 6357876 6358745 6354733 6358680 6358898 6359116 6355742 6356781 6357662 6357785 6358027 6358527 6358754 6359046 6359047 6359465 6359939 6362963 6364027 6368137 6376586 6376735 6379846 6379847 6384966 6386641 6387363 6388094 6400523 6400785 6386541 6386777 6391082 6391941 6395805 6396138 6398113 6400605 6395480 6396231 6397855 6400799 6402819 6403392 6404086 6405581 6406156 6406862 6406943 6411045 6430282 6430539 6435140 6450430 6452475 6453781 6459047 6453322 6461674 6470639 6473141 6474880 6476207 6468780 6469999 6470402 6468920 6469576 6469577 6470063 6473974 6474839 6469306 6470613 6473595 6474160 6480286 6483824 6488826 6491369 6491099 6491613 6492108 6494077 6496999 6500487 6521262 6514157 6073216 6073217 6074338 6075048 6074336 6074337 6075047 6075050 6075053 6075058 6077700 6077790 6077791 6088287 6088288 6088290 6104878 6104879 6104998 6105558 6105559 6106050 6106051 6106059 6106504 6073219 6081508 6081509 6081511 6081512 6088286 6105555 6105557 6077788 6077789 6088282 6088283 6088285 6088289 6092749 6092750 6092751 6104735 6104736 6104880 6105000 6105554 6106052 6106056 6106057 6106584 6119713 6122558 6104732 6104733 6104737 6104738 6104876 6104997 6106054 6106505 6088291 6119706 6119711 6121937 6122709 6123106 6129822 6129826 6130438 6133115 6133117 6133795 6073214 6073218 6104875 6105001 6121933 6123105 6129820 6133430 6133431 6133432 6133797 6138281 6138342 6138343 6138344 6138370 6138371 6155417 6105556 6106058 6106585 6106586 6106587 6119708 6121935 6122554 6123056 6129813 6130431 6130435 6133429 6133434 6138047 6106509 6119709 6121936 6122705 6122708 6129818 6133794 6138048 6138050 6138051 6138372 6138373 6138375 6167317 6170063 6170833 6171249 6106506 6106507 6106508 6129824 6133113 6133116 6133433 6138049 6138376 6155472 6156248 6159451 6073215 6119703 6133119 6133796 6133798 6133799 6155416 6156558 6158145 6159915 6160947 6168225 6133114 6138046 6181830 6106053 6171689 6174741 6176722 6178747 6179801 6183144 6183835 6186233 6187840 6188633 6182121 6184287 6194799 6192808 6193028 6197323 6200898 6201359 6224226 6226309 6223518 6224781 6227147 6233242 6238637 6239961 6241352 6228736 6231325 6237048 6251506 6254785 6256173 6260265 6266121 6267692 6267689 6269227 6270993 6272209 6269942 6271902 6272287 6273584 6277062 6280570 6275262 6281458 6287546 6289757 6291588 6284794 6289986 6292664 6293046 6286205 6292871 6299415 6298825 6301395 6319633 6328001 6317323 6322944 6324014 6324342 6334955 6335157 6334678 6337933 6342030 6359044 6359050 6363337 6359054 6359158 6363249 6364037 6371264 6377720 6372633 6376870 6379337 6380249 6382213 6387502 6389436 6391925 6396947 6399882 6400147 6405150 6421158 6422969 6423911 6424731 6425120 6425397 6447198 6448434 6441205 6447610 6452896 6453317 6454595 6465315 6465586 6457597 6461313 6464441 6468558 6468738 6486407 6482720 6486402 6491779 6495760 6500198 6504836 6507421 6509398 6509498 6512871 6513440 6536965 6541062 6526981 6527829 6529346 6529745 6531001 6531407 6526766 6527991 6530270 6531556 6531558 6532282 6540642 6530269 6531479 6531561 6531804 6534339 6549070 6549884 6076594 6076595 6076596 6076597 6076598 6076690 6076691 6076692 6076693 6076694
6076900 6076901 6077002 6077003 6077004 6077007 6103505 6103508 6103277 6103280 6103506 6105752 6103276 6103281 6105776 6107706 6103507 6103509 6105748 6105749 6105750 6105778 6106515 6106589 6107159 6107709 6105754 6105779 6105781 6106511 6106512 6106514 6107158 6107538 6110592 6110943 6105751 6105753 6106516 6103278 6103279 6105777 6106588 6106590 6107705 6110589 6110590 6110591 6110593 6110944 6110945 6110947 6156374 6167735 6170534 6170829 6174723 6175128 6151112 6155474 6156244 6160646 6105755 6105780 6106513 6107534 6107707 6110588 6110946 6110948 6163831 6171687 6173383 6174014 6163800 6164117 6167583 6170073 6175433 6175784 6176723 6177540 6178204 6171940 6175796 6176964 6177828 6179425 6180019 6180305 6180471 6181938 6182616 6182642 6182650 6182921 6188400 6188843 6191291 6193029 6193564 6196795 6206836 6210977 6213923 6214717 6215686 6216467 6219395 6220513 6229005 6234083 6244788 6257254 6271544 6274038 6292514 6294216 6294644 6294992 6312236 6312327 6314815 6316527 6316772 6317339 6317394 6318519 6319394 6322943 6317340 6317674 6319565 6338834 6341313 6346307 6349412 6351871 6333497 6336666 6367635 6369889 6372595 6375165 6400090 6400305 6403984 6424937 6440353 6444992 6445138 6445319 6446158 6446850 6447193 6447618 6447851 6448533 6448905 6451235 6453311 6448782 6456391 6459877 6465388 6467893 6459968 6461625 6461637 6463480 6464439 6465078 6465387 6468746 6469310 6469674 6509996 6518564 6522247 6512049 6516750 6521832 6522875 6535540 6545000 6090866 6090868 6090871 6091045 6091046 6091048 6091203 6091204 6090867 6090869 6090870 6091043 6091044 6091047 6091207 6091208 6091205 6101380 6104725 6104726 6104727 6104728 6104730 6105524 6106402 6106403 6106650 6101375 6104729 6105525 6105526 6106160 6106161 6106185 6106187 6106254 6106255 6106257 6106361 6106404 6106651 6106653 6101376 6105527 6105528 6105529 6106186 6106652 6119662 6120476 6120502 6124185 6124748 6124750 6124751 6126052 6101381 6106256 6119518 6119521 6119670 6119671 6120480 6124121 6124184 6126051 6126053 6126139 6126548 6126550 6101482 6119664 6119666 6120481 6124183 6124747 6126002 6126138 6119519 6125977 6181295 6181824 6182111 6196800 6205042 6206085 6206621 6208402 6227590 6228926 6237462 6237851 6242855 6246618 6237333 6239950 6248740 6250988 6256419 6259493 6273586 6278153 6281195 6294865 6299971 6300717 6303873 6308270 6312678 6313142 6315541 6316498 6315331 6318899 6322453 6327676 6334364 6337577 6355544 6358038 6368134 6370199 6378044 6379896 6381644 6382006 6382126 6388669 6390707 6391462 6392773 6396078 6398390 6400234 6400238 6427258 6430417 6432753 6438583 6440117 6434040 6434538 6436191 6439806 6440062 6441211 6441668 6443081 6443680 6446127 6454066 6457297 6461603 6466183 6477784 6481036 6531813 6532285 6539897 6088346 6088347 6088724 6088725 6088726 5918942 5918944 5918949 5918953 5918958 5918959 5918960 5918985 5918986 5918989 5918991 5918993 5918996 5918999 5919041 5919042 5919046 5919048 5919050 5919053 5919054 5919055 5919060 5919125 6088348 6088349 6088727 6100494 6100488 6100489 6105737 6105861 6106042 6106049 5918946 5918948 5918981 5919052 5919121 6105731 6105732 6105736 6106044 6106045 6106046 6114315 6114319 6114896 6114898 6124182 6125104 6126055 6126056 6126478 6126669 6126670 6105735 6106047 6114314 6114318 6114649 6114650 6114897 6126477 6159909 6160950 6105733 6105734 6105860 6106043 6106048 6114316 6114317 6114652 6114653 6114654 6125965 6126210 6126487 6126546 6178306 5979140 5979141 5979263 5979264 5979708 5979709 5979710 5979808 5979817 5979820 5979821 5979854 5979855 5979856 5979857 5979858 5980360 5980361 5980610 5980611 5980787 5980788 5980790 5980791 5980792 5980793 5980794 6100491 6114651 6124981 6125102 6125103 6125963 6125967 6126054 6180464 6182179 6210818 6211135 6214088 6206079 6206271 6206826 6206838 6209264 6214522 6207750 6210301 6211973 6213640 6215688 6227704 6229006 6241663 6246162 6232312 6235743 6237341 6238717 6241873 6248504 6249449 6259702 6241600 6241877 6248736 6261543 6263294 6263998 6268166 6246314 6248985 6250443 6250453 6256638 6264801 6267728 6273023 6275276 6278490 6280563 6293424 6303874 6310338 6316767 6331707 6334483 6336664 6336691 6336693 6337116 6337119 6340021 6342174 6345600 6346426 6350410 6350184 6355854 6359379 6361626 6363119 6363614 6364617 6367315 6372205 6382127 6382756 6386105 6386772 6421099 6423258 6424423 6387787 6390864 6391440 6395818 6402811 6404041 6404876 6405151 6406505 6406601 6406914 6412903 6421086 6424215 6425947 6426242 6427949 6430287 6429973 6431249 6434161 6437681 6438419 6443408 6447850 6448871 6449459 6452603 6454079 6456918 6465587 6469782 6470628 6475650 6479263 6474268 6478973 6494428 6504177 6509274 6509541 6509941 6512106 6513399 6513942 6514795 6531815 3.5 mm U.S Distributed - 6062186 6062188 6062189 6062190 6066580 6066581 6066823 6066825 6066828 6067678 6067679 6067680 6066826 6066827 6067681 6091822 6091823 6091826 6091828 6091830 6091831 6091833 6092129 6092130 6092133 6092134 6092135 6092136 6092137 6066579 6091824 6091825 6091827 6092131 6092132 6127468 6127471 6127472 6127473 6127474 6127475 6127477 6127478 6127479 6127480 6127942 6128126 6128127 6130676 6130766 6137266 6137267 6137268 6137735 6137736 6127469 6127470 6127476 6127481 6137073 6137074 6137075 6137076 6137077 6137078 6137269 6137270 6137271 6066822 6220693 6222776 6224276 6227280 6228286 6228684 6229895 6267693 6272210 6290845 6293023 6294083 6331658 6334910 6336587 6337581 6332441 6335705 6336665 6336692 6337120 6337379 6341611 6341948 6351022 6353179 6354478 6355467 6338401 6342272 6342643 6347270 6354424 6356904 6357115 6357796 6354102 6356157 6358040 6358677 6366358 6410650 6421087 6421535 6425642 6423331 6424220 6425385 6425678 6426149 6434304 6435562 6438985 6445036 6446736 6502720 6509393 6509400 6509492 6509868 6514207 6528467 6531498 6062524 6062525 6062712 6063042 6063044 6063045 6063381 6064132 6070645 6070646 6081447 6062528 6063377 6068088 6068280 6070461 6071684 6080896 6080947 6063442 6068382 6071682 6071683 6071685 6080898 6081217 6081453 6086518 6086519 6086902 6093525 6093533 6098519 6064130 6064134 6080944 6081215 6081216 6086520 6093526 6093528 6093529 6098036 6098516 6098518 6098557 6062527 6068090 6070643 6080897 6080943 6081218 6081452 6093524 6098038 6098556 6063001 6068089 6068092 6068093 6068281 6068383 6086906 6098514 6098515 6098517 6098558 6171259 6172404 6172923 6062711 6063380 6068091 6098037 6182646 6182896 6183770 6188629 6190707 6191527 6205048 6211140 6192722 6205548 6224335 6226896 6228287 6252441 6232732 6234367 6229227 6242533 6260122 6267131 6255096 6264274 6270709 6272484 6272787 6285585 6286727 6289626 6285266 6292698 6316773 6317395 6314996 6350387 6340824 6341779 6344096 6345609 6346092 6359210 6362844 6367231 6364258 6368123 6363769 6368305 6371555 6372410 6384123 6381892 6382817 6386261 6380657 6383639 6389919 6391348 6403756 6400855 6405159 6406944 6407039 6404731 6405143 6406503 6420378 6421083 6421373 6423542 6425386 6426119 6430296 6424462 6424596 6425399 6425736 6425969 6426257 6431601 6432269 6425968 6438723 6447725 6444031 6447786 6513168 6518366 6522871 6526312 6527181 6062078 6062079 6062081 6062702 6062713 6062714 6062777 6062779 6063191 6067954 6067957 6067958 6067959 6068085 6073018 6062080 6062776 6063189 6063190 6067955 6073016 6094978 6095178 6062703 6068082 6068083 6068084 6068277 6073017 6094590 6094593 6094595 6094979 6094980 6095179 6110918 6110919 6110962 6111513 6111515 6111713 6113370 6128708 6128712 6128916 6129028 6129031 6129193 6068086 6094589 6094591 6095180 6095181 6068275 6068276 6073013 6073015 6073020 6095362 6110917 6111516 6129032 6129191 6129196 6130677 6094592 6094977 6110920 6111707 6128709 6128710 6128919 6129030 6129195 6094981 6161129 6163474 6167304 6170972 6171584 6067956 6068087 6110921 6111518 6111709 6111716 6128915 6128917 6128921 6129029 6129192 6129194 6161621 6163830 6166789 6169110 6170460 6171685 6176730 6178856 6179795 6180213 6128918 6178494 6179679 6181165 6182180 6182611 6182647 6182318 6182621 6199283 6192819 6203027 6205615 6197172 6200220 6201747 6202818 6204240 6197494 6199175 6201690 6205556 6206266 6206320 6206179 6216250 6238270 6228593 6233882 6237052 6242523 6251255 6255427 6259496 6266458 6268173 6272038 6271345 6276428 6280575 6292587 6307859 6289473 6312031 6319950 6319982 6323741 6326237 6322701 6328180 6343205 6357783 6358541 6359052 6349411 6354027 6361802 6358895 6371561 6371772 6381129 6381853 6387004 6391382 6405152 6406314 6400475 6402150 6404084 6403780 6404479 6426114 6426309 6432743 6446152 6446852 6448524 6457285 6450719 6454234 6457295 6452470 6452791 6463808 6470080 6466852 6467209 6470679 6469312 6469330 6469681 6469722 6470000 6470001 6470075 6469095 6470322 6470397 6470433 6471856 6478820 6488391 6489164 6496662 6501046 6492106 6495087 6497504 6501045 6503524 6504832 6518565 6522869 6527823 6528949 6530738 6509494 6511815 6513272 6071785 6071895 6072020 6072182 6072347 6072348 6072349 6072576 6076481 6076582 6076583 6076601 6076682 6077008 6077009 6071893 6071894 6072575 6072577 6072805 6076599 6098861 6099199 6104773 6105027 6105315 6105316 6105381 6071787 6072021 6072806 6072807 6076600 6105034 6105035 6076480 6099197 6099198 6104767 6104771 6105033 6105038 6122417 6098859 6099342 6105380 6121036 6122416 6122420 6122548 6129221 6129242 6129243 6130637 6131100 6131101 6131445 6131447 6133437 6133671 6105313 6105379 6122544 6131443 6133791 6137936 6122421 6122547 6129244 6129965 6129967 6129983 6129998 6131098 6131099 6131448 6133440 6133441 6133672 6137933 6137935 6138096 6138098 6071786 6104772 6105314 6121035 6121037 6122545 6122546 6130635 6130639 6130640 6131444 6133439 6133674 6133675 6133790 6137934 6138097 6164115 6170647 6171688 6121038 6133673 6133792 6133793 6137937 6122542 6129220 6133438 6138099 6098858 6098860 6099022 6099023 6099196 6099200 6181969 6171133 6171251 6172532 6177539 6178061 6183145 6183277 6197392 6203032 6216369 6224854 6227584 6223504 6225417 6229460 6237847 6228897 6233367 6235754 6249690 6250804 6253024 6255639 6266453 6272409 6272488 6267699 6282277 6277063 6278835 6285591 6274668 6277076 6280209 6280573 6281447 6283154 6285275 6287123 6284251 6289957 6293395 6294869 6299420 6303199 6319076 6322951 6327414 6319532 6337466 6359550 6367599 6362012 6366092 6367305 6368135 6376182 6379325 6381449 6381708 6382750 6386369 6395362 6396197 6398121 6400957 6406942 6420635 6421840 6426494 6446724 6451882 6456395 6461679 6463463 6465582 6469332 6469733 6469923 6478162 6478958 6483328 6489165 6478972 6484566 6487071 6488242 6488829 6488833 6501367 6503685 6506514 6507232 6509064 6509394 6510069 6530080 6538476 6526391 6527126 6527987 6529196 6530162 6529411 6530499 6530739 6536903 6531213 6531817 6541070 6544737 6080619 6080620 6080662 6081818 6080667 6081816 6081817 6081819 6095172 6095177 6103113 6103116 6105549 6105552 6080663 6081804 6095173 6095174 6095175 6095176 6103114 6103115 6106696 6080665 6105548 6105550 6106198 6106655 6106659 6106697 6107291 6103118 6106199 6106200 6106654 6106656 6106657 6107292 6109908 6109909 6109910 6132942 6132944 6132945 6132946 6134460 6107290 6109911 6134450 6105551 6105553 6106658 6106695 6106698 6109913 6132901 6134461 6168888 6170391 6172531 6173788 6157315 6159809 6163675 6105843 6109912 6132943 6134662 6134663 6134664 6171725 6166793 6180465 6174733 6103117 6177849 6178057 6181053 6181972 6183147 6184840 6190955 6192035 6192820 6207468 6214530 6214931 6218613 6221279 6223722 6228692 6229553 6233037 6234379 6230086 6235439 6259720 6283151 6291511 6299642 6315780 6336627 6345598 6371996 6374512 6375941 6376856 6376320 6376740 6402592 6442656 6446608 6446873 6448437 6447535 6452885 6453323 6456386 6463468 6464819 6456831 6459493 6466402 6467213 6469925 6502325 6504622 6504793 6508555 6514166 6509602 6512990 6514262 6518562 6531402 6531494 6535828 6541024 6090938 6090939 6090940 6090941 6094540 6094541 6098467 6098469 6098852 6098853 6090777 6090778 6090780 6090937 6094537 6094538 6094539 6098854 6099123 6098468 6098856 6098857 6104561 6104605 6104606 6104607 6104608 6098471 6099429 6106190 6106192 6106193 6104602 6106191 6118762 6118765 6118766 6119017 6119021 6119182 6119707 6126746 6126922 6098470 6098855 6118601 6118764 6119019 6119704 6119710 6126748 6126749 6126752 6126989 6126991 6098466 6119020 6119183 6119705 6119712 6126753 6126754 6126923 6126924 6126925 6126990 6181844 6182117 6196137 6199828 6205420 6206404 6245982 6249505 6252195 6257473 6263060 6264784 6268967 6271552 6272483 6277137 6282275 6294335 6296151 6302703 6311899 6311984 6317033 6317061 6320362 6327081 6329385 6334731 6338198 6362976 6365116 6478645 6479155 6487419 6488116 6504164 6507942 6508214 6508796 6531560 6535041 6538882 5480444 6081520 6081521 6081682 6084427 6084429 6084431 6084435 6092755 6092758 6092759 6099461 6099502 6099504 6099505 6084432 6092757 6105535 6105536 6105537 6105539 6105540 6105541 6115196 6116317 6123611 6123742 6123743 6123862 6123867 6115195 6116304 6116316 6123707 6123868 6157033 6158855 6159917 6115194 6116302 6116303 6116305 6116306 6116315 6116318 6116319 6123447 6123610 6123614 6123708 6123761 6099460 6099503 6123446 6123449 6123709 6123745 6123760 6182110 6182613 6182942 6182231 6183442 6183764 6206644 6206696 6212276 6214790 6206620 6207124 6208824 6210596 6210600 6210979 6211143 6212279 6213674 6215381 6208448 6211133 6213776 6216336 6227095 6229237 6233585 6239417 6229607 6234365 6238264 6241391 6244308 6245449 6246092 6250454 6250659 6265607 6245471 6246038 6247470 6261542 6248669 6249379 6258401 6260655 6272747 6276583 6279251 6281154 6286737 6291064 6302817 6334813 6358540 6358750 6359051 6365192 6379887 6420451 6405328 6406315 6406584 6406598 6410456 6435125 6444591 6446138 6447741 6461768 6469329 6469806 6470204 6470423 6470668 6472231 6477316 6495711 6496446 6498657 6500879 6501047 6508099 6522769 6522862 6525727 6526592 6527061 6529142 6530078 6530500 6531063 6535698 No 4.0 mm lots distributed in U.S. 2.25 mm Distributed OUS. - 5701706 5796837 5796838 5800202 5800204 5701700 5752721 5752725 5752727 5752728 5759895 5759896 5838234 5796835 5796836 5800205 5811243 5814480 5814501 5814502 5796839 5831630 5831631 5831632 5831633 5831634 5878260 5878841 6150382 6150383 6150384 6150385 5881901 5878461 5881902 5881903 5881904 5906425 5883374 5935099 6089972 5935098 5935100 5965396 5965398 5965683 5965689 5935243 5965397 5965523 5965524 5965525 5965684 5965690 6001080 6181225 6181226 6181227 6181228 6181229 6181235 6181236 6181238 6181240 6181242 6181243 6181985 6181986 6181987 6181988 6181989 6182092 6181201 6181224 6181237 6181239 6181241 6182093 6314318 6314319 6314320 6314321 6314563 6314564 6314565 6314566 6314567 6315234 6315235 6493004 6495428 6496616 6499088 6500298 6497646 6500074 6500867 6500868 6501050 6503523 6503835 6505184 6507895 5771748 5779263 5792852 5792853 5792854 5793272 5793273 5838235 5757188 5779272 5771746 5779254 5779260 5779273 5838352 5814659 5814660 5814761 5814762 5845487 5845488 5845489 5846013 5846015 5846016 5846274 5846276 6029265 5845491 5845492 5846014 6029266 5859165 5859820 5875726 5875735 5875656 5895007 5895009 5902501 5902515 5902517 5923704 5923713 6215821 5902502 5902503 5902516 5950151 5950152 5950153 5950154 5950155 5943505 5943512 5979870 5979871 5979872 5979876 5979878 5983277 5983279 6216164 6216168 6216171 6216227 5982406 5982408 5982410 5983278 6004575 6006613 6216172 6216229 5983276 6215819 6215820 6216162 6216169 5979875 6002504 6002505 6002518 6004572 6004573 6004574 6004577 6004578 6006612 5982407 5982411 5982414 6315859 6378710 6050091 6050095 6050097 6393233 6396458 6397331 6399950 6392019 6392461 6392543 6396239 6471222 6473594 6473978 6474849 6478968 6479765 6483017 6483886 6486391 6488235 6486310 6489083 6489168 6489226 6491512 6492166 6493228 6496228 5668228 5682722 5658677 5658679 5658821 5668229 5682725 5794372 5759198 5781130 5781135 5762775 5762776 5762777 5781129 5781134 5794741 5794743 5794999 5794741B1 5794741B2 5811246 5811245 5817809 5817811 5817810 5817812 5817813 5817814 5828800 5828842 5832943 5832956 5828841 5832944 5832957 5832960 5912006 5912009 5922303 5922694 5923732 5923763 5999323 5876061 5876067 5896401 5896402 5896403 5912011 5876067B1 5896403B1 5906424 5922693 5922301 5943517 6029269 5922299 5922300 5922692 5922946 5943518 6400219 5943687 5943688 5943519 5943521 5943685 5943686 5943691 5943693 5964834 6400212 5876062 5943690 5964836 5965943 6402383 6407610 5964833 5964835 5964903 5965942 5965944 5999300 5999322 5999491 5999493 5999630 6000043 6000045 6399192 5965941 5999494 5999638 5999639 5999640 6006857 6029267 6032102 6032103 6032105 6032121 6032243 6215733 6315236 6315237 6315238 6316102 6316103 6316105 6332641 6397596 6402158 6006611 6006679 6006680 6006724 6006856 6032242 6400208 5999629 6400245 5999488 5999632 6000042 6000044 6032104 6396657 6403534 6006681 6006723 6006858 6215725 6215726 6215727 6215729 6215730 6215731 6215732 6215734 6215735 6054669 6054670 6054953 6054955 6055960 6400214 6400216 6054663 6054665 6054666 6054956 6054957 6055956 6056022 6315239 6315861 6316104 6332637 6332639 6332640 6373706 6373714 6374735 6377709 6382131 6382774 6382801 6382894 6383926 6400861 6401951 6403887 6404463 6481663 6483697 6485900 6488760 6489090 6489167 6489228 6489295 6491379 6492144 6492168 6495085 6495996 5645383 5645384 5667790 5671840 5672021 5667788 5667791 5672022 5703133 5682308 5710287 5710288 5710287B1 5786568 5786569 5788137 5797987 5797988 5797995 5801264 5797990 5801265 5801711 5801714 5797991 5801713 5797994 5829059 5846307 5846308 5846310 5846312 5846313 5846314 5832644 5846309 5867585 5867586 5869141 5868899 5868900 5876479 5876564 5883297 5883298 6032441 5906426 5906428 5954825 5954826 5955012 5975138 5978843 5978846 5984180 5954824 5954827 5954828 5955004 5955011 5984538 5947155 5947158 5968023 5968024 5975136 5978845 5979398 5947153 5947154 5947156 5947157 5968021 5968022 5975731 5979397 5984171 5984535 5995568 6029790 5984178 5984536 5984537 5984181 5995574 5995999 5996104 5996107 6007305 5995567 5995569 5995572 5995576 6006933 6006938 6007302 6011290 6011291 6011292 6011704 6011705 6006934 6006935 6006936 6006937 6007304 6011706 6007303 6011293 6011703 6032442 6032445 6405580 6032443 6032444 6032446 6203662 6203664 6203665 6203667 6203668 6212663 6380315 6381813 6382692 6386155 6387315 6387796 6388674 6389686 6389904 6391436 6469998 6473821 6475312 6477740 6478810 6479507 6491508 6491620 6492146 6494205 6495686 6496601 6497621 6500294 5682307 5819842 5693769 5693766 5789308 5789309 5759186 5759187 5762390 5762391 5770625 5770626 5770628 5770625B1 5797069 5797071 5797070 5829164 5836107 5829161 5830413 5830416 5836106 5846372 5852699 5852700 5853824 5853829 5852761 5881801 5883289 5883290 5883291 5883365 5906423 5883364 5937840 5943038 5950560 5950604 5950605 5937840B1 5943101 5943039 5943040 5950606 5959424 5959425 5972042 5972043 5972048 5977744 5971819 5972049 5972050 5977541 5977742 6325167 5972551 6007824 6007827 6007828 6401052 6049293 6049365 6054329 6049291 5972546 6049292 6054328 6212354 6212355 6212495 6212497 6212500 6212501 6216393 6216394 6216395 6216402 6229315 6229317 6229318 6229319 6332825 6007823 6007826 6049290 6049366 6049367 6049368 6054330 6089973 6203670 6203671 6203673 6203674 6203675 6203676 6203677 6203678 6203679 6203680 6203681 6203722 6203723 6212490 6212491 6212492 6212493 6212494 6212496 6212498 6212499 6216396 6216397 6216398 6216399 6216401 6229313 6229314 6229316 6300749 6300752 6300753 6300754 6300755 6300757 6300760 6300761 6300883 6300884 6380010 6380067 6380246 6382005 6382128 6383120 6385890 6386368 6387317 6388088 6391802 6473895 6474862 6476964 6478438 6479002 6481195 6480293 6481419 6483018 6484957 6486707 6488111 6489082 6489227 6489296 6491618 6492171 2.5 mm Distributed OUS - 5647893 5647896 5647901 5700051 5700052 5700053 5700054 5701770 5701771 5701773 5701774 5701777 5701778 5647901B1 5753153 5753155 5753503 5753504 5753915 5753916 5754055 5754365 5761358 5761725 5763733 5754056 5757715 5757716 5757717 5760746 5760747 5761356 5761721 5763735 5763737 5760875 5760876 5761981 5867617 5867618 5867619 5867620 5867734 5867736 5867738 5867997 5867998 5867999 5868000 5867731 5867996 5879059 5879101 5878495 5879057 5879058 5879060 5883378 5883379 5883485 5883486 5883487 5883672 5883673 5884246 5884247 5884248 5884249 5884250 5884254 5884341 5885649 5885655 5900968 5883380 5883671 5885708 5885709 5900969 5900970 5925155 5925159 5925921 5925158 5925373 5925923 5925156 5925370 5925371 5925372 5925922 6002978 6712239 6719359 6719514 6712169 5658492 5658806 5669569 5669570 5669572 5669573 5669574 5669577 5673892 5673893 5673894 5673895 5673896 5673897 5673899 5703996 5705205 5705208 5705752 5705753 5739428 5669573B1 5739428B1 5669571 5705206 5739087 5739088 5739293 5739294 5739427 5739431 5739537 5739539 5739703 5755387 5770634 5770724 5770725 5770940 5771231 5792836 5739087B1 5739427B1 5755385 5755386 5771316 5771667 5782787 5782788 5782789 5782790 5782792 5782898 5783164 5783355 5783356 5783357 5786162 5786163 5786165 5792833 5792837 5794274 5794278 5795110 5795111 5754366 5757713 5770632 5770633 5770723 5771232 5771317 5782786 5782896 5783163 5771044 5783161 5783436 5786099 5786164 5794280 5815081 5815082 5815083 5815084 5815275 5817663 5815445 5846019 5846306 5846481 5846482 5846483 5846018 5846278 5846279 5846303 5846304 5846305 5846376 5846379 5846280 5856126 5856128 5856132 5856284 5856285 5870431 5870879 5870963 5870998 5871000 5871201 5871401 5870437 5870442 5870667 5870668 5870670 5870877 5870878 5870880 5870962 5870999 5871097 5871099 5871100 5871202 5890658 5890659 5890660 5906817 5906818 5906820 5931482 5931483 5931486 5931484 5964041 5967223 5967224 5967225 6000180 6000343 6001107 6001111 5963918 5995993 6000053 6000581 6001108 6010796 6023332 6023333 6023411 6023413 6024156 5995994 5995995 5996650 6000055 5963920 5967407 5967408 5996206 5996207 5996208 5996649 5996651 5996653 6000056 6000179 6000181 6000722 6001109 6002977 6010792 5963919 5996209 6000054 6010791 6010793 6010794 6010795 6010797 6017642 6017646 6017644 6017647 6017648 6023334 6023335 6023409 6023410 6023414 6023721 6023782 6023783 6023784 6024154 6024157 6017643 6023412 5663047 5663051 5663053 5664348 5664349 5668551 5668552 5668699 5668790 5680480 5680482 5680483 5680485 5657525 5657526 5657527 5657528 5658089 5658099 5663048 5663050 5663054 5664350 5680479 5657525B1 5658100 5680486 5713788 5714402 5713785 5714374 5714379 5714403 5714411 5730046 5730047 5764851 5764851B1 5758787 5758788 5758790 5759181 5759182 5764001 5764002 5764222 5764710 5764711 5770938 5772462 5772463 5772464 5813070 5738273 5738478 5758578 5758580 5758721 5758785 5758789 5758800 5759183 5763959 5764223 5764224 5764230 5764233 5765003 5770935 5796721 5796721B1 5796605 5796606 5796608 5796657 5796685 5796688 5796723 5796685B1 5796607 5811248 5811249 5811250 5811445 5811446 5813068 5813069 5813071 5817680 5817803 5811446B1 5817801 5821295 5821296 5821297 5821298 5821299 5821300 5821341 5821345 5821349 5824622 5824623 5824624 5824625 5824626 5828547 5828548 5828549 5832319 5832321 5832322 5835057 5828550 5832286 5832287 5833770 5835054 5835059 5835221 5835224 5835225 5835337 5835338 5835339 5835566 5835573 5838872 5838874 5839703 5839704 5839705 5839970 5840126 5890277 5905139 5928492 5930436 5930958 5931021 5943150 5943156 5839972 5839973 5855392 5855393 5855394 5855396 5855397 5855398 5890278 5890279 5902296 5902297 5902300 5902301 5904551 5904553 5905125 5905961 5905962 5905964 5905965 5905970 5928494 5928593 5902298 5902302 5904554 5928489 5928491 5928586 5930437 5930959 5902302B1 5902299 5905121 5905131 5905134 5905760 5928490 5928582 5928588 5928591 5930438 5930610 5930611 5930612 5930940 5930613 5930960 5943357 5954592 5954595 5955746 5955747 5955946 5955950 5954594 5955948 5955949 5958871 5959139 5962131 5962132 5962134 5965063 5965068 5965069 5931022 5931221 5958876 5983562 5931222 5943151 5954593 5954596 5955945 5958870 5958874 5959140 5963389 5943356 5958872 5958875 5962128 5962129 5963390 5965064 5965066 5987107 5991822 5991826 5943155 5962130 5962133 5963391 5963392 5991224 6027779 5946857 5989207 5989212 5991222 6010097 6010296 6028009 6028847 6028848 6028849 5989304 5991824 6008010 6010297 6028011 6028053 6028056 6028846 6028850 6010099 6010500 6010662 6019018 6019143 6019144 6019348 6019352 6019353 6027778 6028010 6028055 5989208 5989302 5991223 5991225 6008004 6008007 6008009 6008011 6008895 6008901 6010098 6010298 6010501 5987106 5987108 5989220 5989221 5989303 5991226 5991621 5991823 5991825 6008898 6008899 6027777 6019347 6019349 6019350 6028052 6044319 6044320 6044321 6044342 6044343 6048424 6056361 6057326 6057327 6058246 6047369 6047380 6047381 6048423 6050684 6050685 6050686 6050690 6047370 6047482 6048422 6048425 6050687 6050700 6056443 6056444 6056445 6057324 6057325 6057753 6057754 6057755 6057756 6058247 5640142 5643498 5643499 5646049 5646088 5646089 5646090 5646091 5660812 5660814 5660815 5660817 5660813 5660816 5690699 5704188 5704189 5641416 5671717 5671718 5692090 5692172 5704185 5704186 5704193 5692172B1 5661194 5682310 5682314 5704184 5704187 5704191 5704194 5704191B1 5704196 5737175 5737176 5737177 5737484 5737485 5737486 5737627 5737628 5786950 5786951 5788136 5795832 5787188 5795776 5795780 5795821 5795831 5848233 5802193 5803512 5824020 5824062 5824063 5923815 5803513 5824065 5828274 5923902 5827658 5827659 5827660 5827781 5827782 5828269 5828273 5832556 5832558 5832641 5839444 5839445 5840122 5840123 5843089 5843090 6411515 5828270 5828272 5832642 5843799 5843800 5844178 5848226 5848464 5848470 5848471 5852018 5852884 5923896 5848160 5848222 5848224 5848462 5848465 5851839 5852019 5852883 5832557 5848232 5848234 5848400 5851836 5851837 5851838 5852885 5875248 5875289 6411518 5851836B1 5851836B2 5848221 5848225 5851835 5851840 5852017 5869316 5875256 5875260 5875286 5875288 5898060 5898340 5898361 5898362 5898507 5898508 5905613 5906429 5906431 5906432 5906433 5898340B1 5906427 5906430 5907876 5946407 5955013 5955014 5955015 5955016 5955017 5955018 5955019 5963511 5946927 5947008 5963508 5963513 5963515 5963516 5967632 5898059 5946405 5946406 5946851 5946855 5946856 5946925 5946926 5947005 5947007 5963507 5963509 5963510 5963512 5963514 5963517 5963518 5967633 5967634 5982066 5982110 5982112 5983860 5983861 5983929 5983932 5989588 5990388 5982067 5982111 5983922 5989587 5989592 5982068 5983858 5983859 5983930 5983931 5989591 5989593 5990201 5990382 5990383 6007548 6004197 6004198 6004200 6007545 6007546 6004195 6004196 6004199 6004201 6007549 6007499 6027247 6027251 6027252 6027249 6027250 6027670 6027672 6027673 6404586 6027248 6027671 6723982 6728279 6725069 6718475 5658353 5658356 5660983 5660984 5660985 5660986 5660987 5679803 5680546 5679802 5679804 5680544 5695245 5695247 5698369 5698370 5698512 5698514 5660988 5680548 5680549 5709093 5709122 5709125 5709606 5709608 5709609 5698371 5709127 5721313 5721314 5789421 5789434 5912017 5721313B1 5721310 5721311 5721316 5759871 5759872 5759873 5769328 5778843 5778844 5778845 5778932 5778933 5778934 5778935 5779621 5779624 5789320 5789437 5789625 5789626 5789627 5789789 5789790 5789791 5721315 5799019 5799020 5799182 5814319 5814320 5814362 5844519 5829251 5829257 5834205 5834206 5834208 5835858 5835859 5835860 5839626 5842828 5912015 5835860B1 5814361 5829259 5912019 5829260 5834207 5836619 5842827 5842829 5842830 5842840 5842901 5846103 5856135 5861919 5861920 5861921 5861929 5861931 5861934 5861935 5863590 5863592 5874403 5874417 5874483 5861918 5861932 5863591 5869317 5869318 5874301 5874304 5874404 5874416 5874419 6029274 5901079 5901080 5904126 5904131 5904132 5904136 5904302 5904304 5904308 5905170 5905557 5937827 5938771 5938772 5938774 5938776 6029275 5893123 5893124 5901201 6029273 5923903 5937825 5938769 5938770 5938773 5938775 5938777 5942772 5942773 5951010 5951012 5952065 5952655 5952656 5952657 5985812 5937835 5942768 5942769 5942770 5942771 5951002 5951003 5952079 5985810 5950858 5950859 5950860 5951001 5951009 5952080 5977751 5977752 5977754 5978768 5978859 5978769 5978860 5977750 5978772 5985807 5985811 5988312 5988317 5988319 6023089 6040046 6040048 6040346 6040975 6040976 6040981 6041002 6041003 6043057 6040981B1 5985808 5988316 6040044 6040304 6040345 6040977 6040979 6041004 6042037 6042038 6043055 6043058 6043217 5978858 6040007 6042720 6043056 6053594 6053638 6053639 6053640 6023090 6039696 6039701 6040006 6040319 6040978 6040980 6050142 6053595 6053922 6053923 6023087 6023088 6050144 6053596 6053641 6040050 6042027 6042028 6045551 6045552 6045553 6045554 6045555 6045556 6050148 6053597 6062791 6088920 6088921 6089868 6089869 6089870 6089871 6089872 6089873 6089874 2.75 mm Distributed OUS. - 5644654 5699226 5699227 5699230 5705128 5705128B1 5810920 5810941 5810942 5811000 5811010 5705129 5754059 5754082 5755665 5755666 5756318 5756319 5759658 5759659 5759660 5762266 5762268 5762269 5810999 5811001 5811013 5811016 5811141 5831345 5831346 5831347 5831348 5831349 5863731 5863732 5863734 5863736 5863738 5876005 5879044 5879076 5879104 5876001 5879042 5879080 5879103 5879105 5879142 5879074 5879102 5879148 5883246 5883247 5883248 5883250 5883251 5883249 5883252 5931228 5931285 5931286 5931288 5931289 5762770 5767418 5768373 5779556 5783169 5783170 5793351 5793352 5793354 5793724 5793728 5793732 5767417 5767582 5768314 5768315 5779548 5783171 5783172 5768315B1 5767414 5767585 5767594 5779545 5783174 5800261 5800262 5807168 5807801 5808160 5808261 5815773 5815821 5815824 5817393 5817394 5817395 5817396 5817397 5817398 5817399 5817661 5817963 5844518 5815823 5817819 5817964 5817965 5817966 5837232 5837235 5837389 5846669 5846079 5846080 5846495 5846500 5846671 5846361 5846362 5846496 5846670 5872414 5874135 5883191 5883193 5883194 5883195 5883196 5883197 5846667 5872346 5872401 5872410 5872416 5874134 5883190 5883192 5883198 5902630 5902631 5902632 5927036 5927350 5926630 5926631 5926632 5927037 5927349 5927351 5950544 5983280 5988681 5983281 5988303 5990006 6012363 6012958 6018615 5927038 5950545 5950548 5988304 5988802 6002979 6012956 5988241 5988305 5988803 5988805 5988808 5990004 5990005 5990007 5990008 5991968 5991969 6012362 6012364 6012365 6012959 6012960 6018294 6018430 6018661 6018293 6018431 6018616 6018825 6736029 6713753 6740505 6740284 5671832 5680638 5680662 5680667 5691293 5693453 5693569 5693573 5693576 5671831 5671833 5680664 5680665 5680666 5693571 5712704 5712705 5712706 5712707 5712843 5713348 5713349 5713404 5731747 5671833B1 5680664B1 5712703 5712842 5713347 5763723 5730189 5763722 5763782 5763783 5763785 5781487 5782871 5792316 5792622 5781483 5781484 5781485 5789792 5792314 5792315 5792563 5792564 5792565 5792569 5792570 5792580 5792621 5792625 5792626 5794744 5781484B1 5792570B1 5792570B2 5792625B1 5781482 5782869 5782870 5792568 5794618 5794619 5812906 5812915 5812916 5812917 5818370 5818371 5818372 5818373 5818374 5818375 5818376 5818377 5818378 5831147 5833595 5833597 5831148 5833596 5835575 5835576 5835577 5835741 5835742 5835743 5835848 5836114 5917126 5917270 5931309 5931313 5835574 5862275 5862273 5862274 5862274B1 5917127 5917728 5922206 5931311 5935724 5935727 5935728 5935734 5931311B1 5917269 5917271 5917272 5917365 5917727 5922207 5922211 5922214 5931308 5935735 5936086 6029271 5958669 5962138 5962140 6029272 5931310 5935600 5935722 5958672 5958674 5958864 5963396 5983563 5917128 5931312 5958673 5958865 5987558 5962136 5962137 5963394 5987329 5987331 5987333 5987556 5987954 5991963 5991967 5935730 5963393 6024566 5935721 5935725 5936085 5987552 5987560 5987682 5987955 5987956 5987958 6024143 6024490 6024563 6028058 6028059 6024040 6033101 6024484 6024493 5991965 5983564 5987330 5987550 5987551 5987553 5987957 6033100 6033283 6024564 6028057 6023915 6023916 6023917 6023918 6024039 6024142 6024483 6051560 6051623 6752406 6754657 6754498 6746851 6746336 6746840 5666427 5666431 5671798 5673166 5673168 5666428 5666433 5672395 5673167 5673169 5682544 5682546 5700630 5666432 5672387 5700633 5700632 5706751 5706752 5706753 5711049 5727133 5706753B1 5792764 5796172 5796319 5792765 5792766 5792768 5792769 5792770 5792776 5795701 5795702 5795763 5795818 5795819 5795820 5796173 5796176 5796318 5796320 5801650 5801651 5801652 5801653 5827005 5827006 5840124 5840132 5842917 5842918 5842919 5840242 5842920 5836424 5871252 5843081 5880079 5880080 5880181 5880182 5885313 5885315 5885318 5885319 5885660 5885701 5885702 5885703 5917886 5880079B1 5880182B1 5907873 5931353 5931354 5931356 5931357 5931358 5954829 5954830 5954832 5954833 5947290 5967401 5967626 5931355 5947291 5947293 5954831 5967076 5967077 5967625 5988355 5988357 6032638 5967075 5985017 5995334 5947292 5967402 5967403 5967627 5985016 5985019 5985020 5995322 5995325 5995362 6011508 6019912 6011509 6019231 6019232 6019233 6019234 6004389 6004390 6004391 6004392 6004393 6004394 6004395 6011506 6011507 6011713 6011714 6019230 6019910 6019911 6019229 6032640 6032633 6046752 6046754 6046758 6046869 6047991 6047998 6048358 6048383 6048388 6048390 5658020 5658081 5658083 5658084 5679530 5679801 5658083B1 5679528 5679529 5679620 5693304 5693305 5694903 5679620B1 5693307 5694779 5694910 5702529 5702531 5762394 5814157 5786391 5786393 5786394 5786395 5762393 5762395 5972662 5786392 5786396 5799185 5799256 5836688 5799183 5799184 5799257 5799258 5799442 5799443 5799659 5799881 5814156 5821431 5838358 5821291 5821293 5821294 5822209 5822210 5821293B1 5843683 5843684 5843685 5843686 5844177 5874923 5846794 5846798 5874921 5874922 5886777 5886840 5886841 5886942 5887026 5886780 5907875 5929140 5929225 5931291 5937934 5886779 5887025 5887027 5929137 5929226 5931292 5931293 5931294 5931295 5931296 5937933 5951013 5937933B1 5937935 5947731 5947734 5951014 5972664 5947733 5947738 5947739 5947783 5972638 5975302 5972663 5975156 5976509 5976510 5988360 5988361 6030445 6030447 6030448 5975315 5975363 5976508 6030442 6030449 5975365 6043442 6043443 6043444 5972661 6049569 6049570 6049572 6049949 6030340 6030444 6043446 6049568 6049573 6030341 6030450 6030452 6049571 5758356 5758357 5758358 5758562 5758563 5758565 5758566 5758783 5758784 5758792 5758799 5772473 5772474 5758791 5772475 5773082 5773086 5773088 5773089 5799656 5799657 5799658 5799974 5799975 5799976 5800212 5800213 5800214 5800215 5800456 5827165 5827168 5827169 5827170 5842833 5842835 5842839 5867284 5867285 5867370 5867371 5867472 5892738 5892740 5892841 5892843 5901216 5930422 5930425 6029277 5866286 5866287 5866288 5866290 5867283 5867300 5867473 5867474 5867582 5867583 5867584 5901218 5897846 5897847 5898517 5901215 5901217 5901219 5930374 5930426 5930638 5930639 5930681 5935091 5897843 5923074 5930640 5962121 6029789 5961798 5961800 5962122 5962123 5962124 6054942 6056784 6054129 6054130 6054132 6054134 6054298 6054917 6054921 6054944 6056691 6056697 6057871 6057872 6057873 6057881 6054131 6056692 6058002 6058628 6300885 6305282 6305283 6305284 6305285 6325180 5672811 5672812 5676206 5676211 5676217 5672810 5676216 5676323 5676326 5676328 5676319 5709313 5709315 5709316 5709319 5709604 5717596 5717597 5717598 5719617 5719619 5720077 5721307 5721308 5806858 5806859 5807583 5807740 5808262 5808343 5808455 5809522 5809913 5809914 5814780 5814885 5815254 5836535 5814885B1 5806857 5806860 5807364 5808274 5808456 5809457 5809458 5809521 5844515 5814881 5814882 5819847 5832086 5832643 5829180 5829241 5829242 5829243 5831767 5831768 5831769 5832087 5836249 5836251 5836254 5836257 5836259 5836302 5921902 5921903 5843084 5855707 5855711 5862367 5875364 5875374 5900863 5903675 5900862 5900864 5900866 5855709 5875365 5875367 5875369 5921598 5900865 5903676 5907877 5907879 5907921 5907923 5921599 5921600 5921599B1 5907878 5946409 5946641 5946642 5946643 5921901 5946410 5946420 6050459 6050461 6050544 6050550 6058405 6058409 6058413 6058221 6058262 6058263 6089488 6089493 6089494 6089495 6089496 6089497 6089875 6089876 6089877 6089878 6089879 6089880 Distributed OUS - 3.0 mm - 5626405 5632869 5643848 5699217 5699222 5699273 5699274 5699275 5699770 5699771 5699772 5700048 5700049 5703995 5705127 5700049B1 5700050 5428044 5810698 5810700 5810808 5810809 5810812 5810823 5645454 5645455 5699223 5705126 5738154 5738157 5738261 5738625 5738626 5721455 5738219 5738220 5738623 5748626 5748627 5748630 5748632 5748635 5751248 5751254 5751259 5751472 5751473 5751909 5751915 5751920 5753149 5753150 5753786 5753787 5760745 5810699 5810810 5810811 5810821 5810822 5810824 5816833 5816834 5816835 5816836 5816838 5816839 5816840 5817488 5817516 5817517 5817520 5817552 5817554 5823543 5721584 5816832 5816837 5817181 5817513 5817514 5817551 5817553 5863725 5863726 5863727 5867426 5867427 5867428 5876567 5879108 5879112 5868087 5879106 5879107 5879109 5879110 5879111 5887056 5887057 5887059 5887068 5887069 5887070 5901745 5909847 5909849 5909850 5909900 5910001 5887060 5910002 5921144 5922281 5922283 5922285 5922960 5923002 5923003 5923019 5939091 5923019B1 5923006 5923062 5933923 5935491 5935492 5935495 5935496 5938849 5938850 5938853 5938854 5938856 5939092 5939094 5970296 5922286 5923018 5933860 5933921 5933924 5935493 5935494 5935497 5935498 5938845 5938846 5938847 5938848 5938852 5939093 5939099 5969381 5982814 5987695 5987696 5987697 5987698 5987699 6179091 6179094 6179095 6179092 6404377 6508843 5645457 5669784 5669785 5669786 5669793 5669800 5669802 5669803 5674990 5674991 5675271 5675272 5675280 5675421 5740620 5740961 5740962 5740967 5675272B1 5669801 5679886 5740619 5740801 5740802 5740804 5740966 5741892 5741893 5741894 5741896 5745079 5745080 5745123 5745125 5745126 5751823 5751827 5751829 5752172 5754368 5754369 5755053 5755056 5755393 5755395 5759199 5759200 5745123B1 5751827B1 5754050 5755135 5781015 5783375 5783379 5789422 5789424 5789440 5789561 5792838 5794366 5794367 5794368 5741895 5745121 5752171 5754051 5755055 5755134 5755138 5755394 5765406 5765408 5765411 5779192 5779193 5781013 5781014 5781016 5781125 5779189 5779190 5779191 5779478 5779479 5779541 5783376 5792840 5792842 5800273 5800274 5800275 5800395 5800396 5800399 5678817 5721588 5735807 5749151 5807098 5816103 5816110 5816112 5816113 5816846 5816854 5816858 5816951 5816958 5816962 5816965 5817073 5817082 5817084 5817215 5817351 5817352 5817353 5817354 5817355 5817356 5818364 5818368 5817083 5817085 5817214 5818170 5818171 5818172 5818309 5818313 5818314 5818361 5818366 5837390 5841789 5841790 5846676 5846678 5847888 5847889 5847947 5818317 5818318 5818367 5846677 5847333 5847334 5847337 5847886 5847887 5847946 5847948 5848908 5857297 5857404 5857405 5860719 5861372 5847332 5848912 5857299 5857401 5866685 5866772 5866790 5882017 5882041 5882042 5882044 5882045 5882048 5882049 5882050 5866686 5866692 5866763 5868091 5882018 5882043 5882047 5882046 5893800 5902627 5902628 5902629 5910097 5910098 5910099 5910100 5910203 5910208 5910209 5910210 5910211 5910212 5910213 5910214 5910215 5910497 5910498 5910499 5910500 5927753 5928063 5928067 5928070 5910207 5927746 5927749 5927751 5927754 5928062 5928064 5928066 5931632 5935088 5950291 5950292 5950444 5950967 5928061 5941756 5950181 5950182 5964050 5964052 5968355 5968497 5968498 5968521 5968676 5968679 5986980 5987218 5987686 6332635 5950159 5950160 5950183 5950293 5962815 5968500 5987326 6011048 6026810 6026811 6026813 6027082 5950294 5962811 5968354 5987003 5987080 5987081 5987216 5987325 5987547 5987549 5987684 5987685 5987692 5968352 5970403 5987002 5987076 5987102 5987213 5987215 5987217 5987328 5987548 6011049 6011051 6011052 6050729 5968351 5968675 5968677 5968678 5982808 5987214 5987327 5987546 5987683 5987693 6011050 6015622 6015703 6015704 6015787 5982818 6011053 6015481 6015623 6015788 6026812 6027061 6027083 6027099 6027100 6027101 6050726 6332636 6380530 6022163 6027084 6049178 6049179 6049180 6049181 6049283 6049284 6049285 6049286 6049577 6049578 6049580 6049764 6049765 6049766 6049951 6049952 6050232 6050234 6050235 6050722 6050724 6050730 6051554 6051555 6051627 6742142 6743175 6742144 6743138 6740705 6742146 6742864 6740709 6740802 6740809 6742686 6742690 6735792 6735794 6736014 6740708 6740800 6740808 6742145 6742452 6742501 6742677 6742865 6735793 6736003 6740699 6740701 6740704 6740796 6740812 5645459 5663664 5665134 5665135 5665136 5683222 5692166 5651110 5651120 5651203 5651205 5651206 5651207 5651209 5651211 5660482 5661733 5661734 5661735 5661736 5663661 5664981 5665352 5663661B1 5663661B2 5665352B1 1000421 5637477 5663662 5664920 5664982 5664983 5665345 5665349 5683221 5692165 5692167 5693013 5695244 5741598 5744741 5744871 5745063 5745975 5745977 5664983B1 5683221B1 5692167B1 5745975B1 1000422 1000423 1000426 5661729 5663663 5664984 5664985 5665132 5665133 5665347 5667016 5702894 5711083 5711232 5740973 5741353 5741488 5744620 5744746 5678501 5760757 5762780 5779897 5785664 5785665 5785724 5785728 5788729 5790025 5785728B1 5721590 5731263 5731752 5731753 5731754 5740974 5740975 5741348 5741354 5741491 5741597 5744747 5744872 5745061 5760758 5762678 5762679 5762680 5762761 5762778 5762779 5763876 5763878 5779893 5779894 5780081 5781968 5781970 5781974 5789799 5791911 5792578 5792704 5780083 5780085 5780087 5780724 5780725 5780726 5780830 5780831 5780832 5781967 5781972 5781973 5788730 5788738 5788739 5789137 5789138 5789800 5792571 5792572 5792577 5792627 5792629 5792703 5792777 5780831B1 5785540 5785723 5785729 5791909 5792075 5792573 5792579 5792628 5792705 5797167 5803935 5804012 5804015 5804717 5797167B1 5790740 5796727 5796731 5797161 5804008 5804010 5804011 5804013 5804718 5804719 5804720 5804804 5810934 5810994 5810996 5810997 5810998 5811050 5811051 1000613 1000614 1000615 1000617 1000619 1000621 1000622 1000625 1000626 5761952 5761956 5771810 5804014 5811048 5811049 5817074 5817189 5817206 5817217 1000615B1 1000620 1000623 1000624 5812506 5817075 5817207 5822818 5823435 5823436 5827109 5827110 5827111 5827112 5828744 5828745 5828746 5828747 5828849 5828850 5828264 5828266 5835632 5835744 5835749 5835758 5835841 5835850 5835851 5835953 5835955 5836110 5836112 5836113 5836119 5836120 5837671 5902321 5902325 5903309 5903363 5909844 5909845 5909846 5911297 5914187 5914197 5921905 5925462 5936633 5938722 5941610 5941617 5837674 5845883 5845884 5845885 5846000 5846141 5846142 5846143 5846145 5846146 5846147 5846148 5846149 5846150 5846151 5848914 5852888 5852895 5853835 5863374 5863529 5863531 5866792 5866795 5866902 5867151 5852889 5852892 5863373 5863375 5863530 5866799 5867150 5867152 5871236 5871237 5872275 5872276 5872277 5872278 5874627 5874628 5884424 5893962 5939567 5874629 5893965 5902322 5902552 5902553 5902554 5903307 5903308 5910003 5914196 5914199 5921867 5921904 5921906 5925466 5925666 5874631 5901747 5902323 5902551 5903304 5903362 5903365 5903367 5909898 5909899 5914188 5914189 5914190 5914200 5925665 5938618 5902324 5902550 5914183 5921146 5925463 5925464 5925664 5925667 5926628 5939566 5939622 5941613 5941614 5947483 5926629 5931627 5936272 5936273 5938619 5938723 5939625 5939628 5941609 5941611 5941612 5947439 5951074 5951387 5951388 5951653 5951804 5953246 5953247 5954458 5954459 5954460 5954542 5954548 5954551 5955426 5955427 5959421 5931681 5947436 5947487 5951076 5954457 5955424 5955429 5956236 5956346 5956348 5959991 5959994 5967072 5936635 5936636 5938620 5939575 5941616 5941618 5941748 5955425 5955428 5956157 5956158 5956234 5956235 5956240 5956347 5959219 5959989 5959992 5988920 5921866 5931684 5939574 5939577 5941615 5947435 5947437 5951077 5951384 5951386 5951651 5951652 5951654 5951655 5953245 5953823 5954547 5956156 5959996 5967622 5951384B1 5951654B1 5936270 5936271 5941758 5941764 5947482 5953248 5953824 5954541 5954543 5959422 5959988 5959993 5959995 5959997 5962594 5967070 5967071 5967073 5967074 5967623 5967624 5967638 5969382 5988592 5988916 5991962 5996689 5997709 5997710 5997711 5941760 5947440 5959990 5963397 5963398 5967069 5991465 5996699 5996871 6010101 6015624 6027964 6028285 6032257 5936634 5938721 5947438 5951073 5951075 5951385 5962807 5967621 5967637 5988917 5991467 6015789 6022940 6023720 6032258 6044166 6044167 5970415 5982680 5988585 5988589 5991469 6010100 6010362 6011514 6023339 6023340 6027963 6028297 6032317 6032725 6032728 6033033 6033091 6033092 6033093 6033296 6033297 5970409 6010282 6010299 6010301 6011515 6011516 6015790 6015791 6015792 6015793 6022903 6022939 6023377 6027962 6028061 6028296 6028298 6032256 6032726 5963399 5982804 5991466 5991781 5996215 5996216 6011517 6043938 6044165 5982810 5988586 5988588 5991463 5991464 5991780 5996213 5996214 5996688 5996700 5996870 5997708 6032259 6032316 6032318 6032319 6033094 5992374 5992377 6015625 6022904 6023719 6028283 6028284 6022069 6022906 6022962 6023091 6023092 6023093 6023094 6023095 6023341 6023378 6023379 6023380 6023381 6023522 6023717 6023718 6023786 6023787 6024029 6024031 6022077 6023785 6043937 6043940 6043941 6044163 6044164 6044168 6044169 6044170 6044171 6044172 6044173 6044174 6044175 6044176 6022079 6045547 6045559 6043939 6044162 6045544 6045545 6045546 6045560 6045783 6045784 6045785 6045786 6046213 6046470 6046478 6052039 6096237 6024033 6046210 6046211 6046215 6052142 6068538 6068539 6068540 6300889 5640928 5645279 5649400 5649662 5649663 5649664 5649828 5649830 5660484 5667955 5668222 5669173 5669174 5672819 5673161 5678022 5649664B1 5673161B1 5645381 5667957 5668221 5668223 5668689 5669165 5669170 5669171 5669172 5672027 5672238 5672820 5673164 5673165 5674683 5678141 5678143 5678503 5672820B1 5668690 5668691 5669164 5669167 5669168 5669169 5672026 5673162 5673163 5678021 5678509 5704131 5704132 5704133 5669167B1 5704133B1 5604292 5668692 5672028 5704140 5704322 5704432 5704433 5720206 5720351 5720356 5720463 5720466 5720473 5720537 5727124 5727125 5727126 5727129 5727127 5727130 5730749 5730751 5730767 5730768 5731749 5731750 5784054 5784055 5784056 5784058 5784060 5784183 5784438 5797305 5784436 5784437 5797301 5797302 5797303 5797304 5797307 5797799 5797800 5797901 5797902 5797903 5801001 5784436B1 5783103 5797904 5800899 5801009 5801011 5801716 5802187 5802382 5823851 5824152 5824158 5824160 5824201 5824207 5781590 5801715 5823850 5823852 5823853 5823859 5823903 5823908 5823910 5823911 5824150 5824153 5824156 5824206 5637501 5678505 5678507 5827646 5827647 5827648 5827650 5827651 5827652 5827653 5827655 5827656 5827657 5839027 5827649 5827654 5839028 5839029 5841794 5851294 5851634 5596614 5839026 5851296 5851636 5852001 5851287 5851635 5851789 5851791 5852005 5852015 5852016 5867379 5867380 5867422 5867424 5867425 5868095 5871186 5883158 5883369 5851788 5852004 5882701 5882703 5883482 5883483 5883484 5883877 5883878 5883879 5883921 5898515 5898516 5899378 5899380 5899441 5899442 5899443 5901742 5901743 5902960 5903065 5903069 5903074 5903075 5903076 5903078 5903079 5910668 5910670 5910671 5899442B1 5901749 5903061 5903063 5903077 5921739 5921741 5921742 5921743 5921744 5921745 5922496 5922497 5922498 5922562 5922757 5922758 5922760 5922821 5937937 5937939 5938011 5938013 5938225 5940271 5940272 5955421 5979385 6411381 5921148 5955158 5955159 5955160 5955439 5955440 5955742 5955743 5955744 6411482 5903068 5910669 5938008 5938159 5941746 5955741 5960843 5960845 5960846 5960847 5960848 5961642 5961752 5961753 5961754 5975197 5979427 5983937 5903068B1 5921740 5921746 5931689 5937938 5938156 5938157 5940264 5940265 5940267 5940270 5941752 5955157 5960849 5962588 5974937 5975194 5975196 5975204 5979280 5979386 5979425 5983923 5983933 5983936 5962817 5975153 5975155 5983935 5992014 6411380 5960842 5960844 5961420 5961641 5970290 5970401 5973171 5974935 5975183 5975184 5975376 5979279 5983925 5983934 5992011 5992013 5992015 5992019 5992020 5992021 5992222 5992223 5992388 5992390 5992391 5999328 5999331 5999368 5999575 5999702 5999703 5999707 5999720 5999816 6002776 6004582 6004600 6004723 6006605 6006673 6006674 5982802 5992012 5999574 6002772 6002774 6002775 6066840 5982816 5983924 6004383 6004385 6004386 6004387 6004388 6004579 6004580 6004581 6004583 6004584 6004594 6004722 6006540 6006602 6030184 5992382 6004382 6004384 6004592 6004593 6004601 6004724 6006541 6006604 6006675 6006676 6006736 6006737 6006738 6001700 6006672 6006678 6030198 6032555 6022071 6028738 6028741 6029799 6029801 6030022 6030028 6030199 6030201 6032552 6081846 6028735 6028736 6028739 6028740 6029800 6030183 6030200 6032551 6032554 6032557 6032558 6064864 6087028 6745113 5660486 5678511 5686733 5686737 5689105 5689108 5689109 5689995 5689999 5690142 5690152 5690231 5690273 5690517 5690518 5689105B1 5631632 5686927 5686928 5686937 5686938 5690025 5690033 5690062 5690077 5690148 5690383 5690521 5695250 5702953 5702954 5718734 5736829 5679888 5686727 5690223 5690232 5718801 5729890 5729895 5729955 5734810 5735241 5736553 5736827 5746635 5746762 5746765 5746767 5746772 5718731 5718736 5721581 5729888 5734816 5734817 5735116 5735120 5735240 5735242 5736555 5736556 5736833 5746631 5746634 5746763 5746764 5746770 5748107 5761359 5735240B1 5746773 5747644 5747645 5748108 5748109 5762274 5762275 5762276 5748108B1 5796324 5796325 5796326 5796510 5796511 5796513 5796697 5796698 5799903 5799904 5799905 5799906 5799907 5790782 5820650 5820652 5820653 5820656 5822813 5830039 5830181 5833079 5833202 5833079B1 5731272 5749154 5788105 5749438 5820649 5820651 5820658 5820741 5820847 5820848 5830033 5830034 5830037 5761958 5830040 5833201 5848007 5848008 5848009 5848492 5848496 5853833 5856386 5856390 5856398 5856481 5848011 5848012 5848152 5848153 5848154 5848157 5848159 5856385 5856388 5856393 5856394 5856397 5856399 5857276 5867578 5867579 5867722 5867723 5868001 5870453 5871050 5848158 5867580 5867725 5868002 5870456 5870719 5870720 5870821 5882705 5883286 5883287 5883288 5871051 5893968 5893969 5897464 5897466 5897963 5898363 5898364 5898510 5921869 5921872 5936799 5936800 5867724 5870454 5870703 5883285 5893967 5897467 5897468 5897469 5897964 5898181 5898182 5898190
5898231 5898234 5898235 5898365 5898509 5898514 5902195 5902231 5902232 5902234 5902233 5921151 5921870 5921871 5931634 5936921 5936923 5943024 5943029 5951301 5951317 5951319 5951928 5951932 5951937 5952759 5952760 5953021 5953401 5953404 5983499 5943029B1 5931638 5936924 5942848 5942849 5942850 5942852 5942853 5943025 5943026 5943027 5943028 5951079 5951080 5951303 5951318 5953402 5959338 5942848B1 5941762 5951078 5951320 5951929 5951934 5953023 5972803 5942851 5943023 5959339 5959340 5962809 5969857 5983582 5970299 5982806 5983494 5983495 5983500 5983501 5983583 5983585 5983586 5983587 5983590 6024323 6025721 6025863 6033289 6033908 6033921 5982812 5983497 5983498 6021890 6022801 6022890 6033290 6033915 5983492 5983584 5983588 5986775 5986866 5992369 5992372 6042712 6042714 6042721 6042782 6043218 6043219 6043324 6043531 6044309 5983493 5983496 6001588 5972804 6001695 6043220 6043323 6043325 6043330 6043532 6043533 6043631 6043632 6043633 6043634 6043635 6048934 6081845 6001697 6001698 6044306 6054672 6021639 6021640 6021641 6021892 6021893 6022073 6022081 6022800 6022882 6022902 6024325 6024846 6025441 6025484 6025862 6025864 6033292 6033295 6033913 6048864 6048865 6049062 6022074 6022075 6043534 6044302 6044303 6044304 6044305 6044307 6044308 6047842 6047845 6048398 6048399 6048400 6048401 6048862 6048863 6048999 6049000 6049001 6024322 6024324 6024845 6024850 6025482 6054671 6064867 6064869 6088908 6088914 6088997 6088998 6089881 6089962 6089963 6089964 6089965 6089966 6089967 6089968 6089969 5484301 5756764 5756765 5756994 5756996 5757284 5757298 5757522 5757523 5757524 5765504 5765508 5765511 5772454 5772461 5773250 5735812 5772460 5773251 5773252 5778686 5778691 5778692 5778694 5778836 5778839 5778840 5778841 5779475 5798011 5798012 5798013 5798016 5798012B1 5784504 5784505 5798006 5798009 5798010 5811254 5811256 5811257 5811699 5811781 5812455 5812456 5811699B1 5762001 5811255 5811700 5812425 5812453 5822212 5822214 5824542 5824543 5824553 5824554 5824560 5824551 5831247 5831248 5831249 5831250 5831251 5831252 5831900 5831961 5841796 5849228 5849229 5849230 5849234 5855229 5855230 5855231 5855232 5855233 6411378 5857278 5898748 5898750 5898757 5899696 5900870 5900871 5901030 5934004 5934614 5898748B1 5868099 5877898 5877899 5877900 5878144 5878250 5878251 5878254 5878492 5878494 5878842 5878843 5878844 5898755 5899694 5893971 5898749 5898756 5898758 5899693 5899695 5900868 5900869 5900872 5902328 5902329 5902330 5934395 5934398 5934612 5934613 5900875 5934001 5934002 5934003 5941750 5943502 5943510 5943537 5934394 5934397 5943501 5943508 5943526 5943536 5943561 5943562 5943969 5943970 5943973 5943976 5962813 6011513 5943539 6001691 6001692 6045460 6045461 6045542 6045543 6045557 6045558 6053163 6053165 6053394 6053396 6053860 6043741 6053162 6053164 6053166 6053167 6053395 6053861 6053902 6053905 6053906 6054291 6054292 6053859 6725200 6729338 6729345 6735282 6735285 6725199 6729340 6729517 6733357 6729336 6732480 6725198 6725306 6728293 6728496 6728503 6728777 6729335 6731881 6725302 6728481 6728998 6729000 6725201 6725301 6725304 6732571 6732635 6733358 5646059 5646081 5646082 5646251 5660488 5671447 5676596 5677011 5677013 5679339 5679805 5679807 5645744 5671448 5676598 5676599 5679282 5679283 5679284 5679338 5679340 5679806 5679808 5679809 5679810 5679811 5679812 5679813 5679814 5679810B1 5626409 5702896 5706430 5706431 5706432 5706434 5706435 5706453 5706454 5706455 5706663 5706665 5706701 5706702 5706703 5727255 5727257 5728439 5734151 5734351 5734352 5734522 5735182 5735183 5735367 5740601 5740606 5734351B1 5724568 5727256 5727258 5727408 5728437 5728438 5734144 5734147 5734150 5802564 5802567 5803063 5803064 5803446 5803447 5803505 5803506 5803518 5803521 5803539 5804134 5804135 5804224 5804225 5804226 5804227 5804228 5804230 5807038 5807739 5807944 5808275 5808341 5808342 5815957 5769448 5769450 5797974 5803449 5804130 5804229 5807039 5807169 5807367 5807039B1 5802565 5807100 5815946 5815949 5828606 5828608 5815946B1 5823438 5828607 5828611 5843095 5912013 5911361 5913163 5913164 5913165 5913168 5913410 5913411 5913412 5913413 5913414 5913551 5913554 5913555 5913556 5914523 5914525 5917769 5917884 5917980 5923556 5846261 5846262 5846263 5846264 5855579 5855580 5855581 5855582 5857279 5857280 5859014 5862156 5862157 5862158 5862203 5859013 5865259 5868204 5871247 5881905 5881907 5882489 5882491 5882493 5882494 5882495 5881814 5881815 5899838 5911363 5862160 5862205 5862207 5865248 5868201 5871244 5871245 5871246 5871248 5871249 5912012 5882492 5900964 5901751 5913167 5913553 5914288 5914291 5914292 5914300 5914521 5917768 5917770 5917775 5917777 5917885 5918101 5918102 5918200 5918562 5919242 5923673 5923677 5928322 5881816 5882490 5882496 5882707 5899837 5899839 5900963 5913409 5914290 5914524 5914527 5917773 5917979 5918561 5918563 5919301 5919305 5919476 5919477 5923551 6411196 5913166 5914289 5914522 5914526 5917805 5917919 5921153 5923549 5923557 5923679 5927800 5928323 5881906 5913552 5927797 5928324 5928325 5946921 5946922 5946924 6001590 6325187 6064801 6088835 6088838 6088903 6089355 6089356 6089357 6089360 6089361 6089862 6089863 6089864 6089865 6089866 6089867 6325193 3.5 mm Distributed OUS - 5643646 5690652 5700504 5700505 5643646B1 5700504B1 5810685 5810686 5810687 5810688 5810690 5810807 5748636 5748637 5748639 5748640 5748747 5751260 5751341 5751342 5751475 5751476 5751591 5751594 5759805 5748638 5748746 5748748 5759806 5759815 5810689 5810696 5810697 5810806 5810926 5810927 5817523 5831966 5831978 5864148 5864149 5864150 5864345 5864348 5864350 5864352 5866158 5866160 5866281 5866283 5866284 5878852 5878853 5878858 5878859 5878860 5878851 5880566 5878857 5880537 5880539 5880562 5880567 5880568 5880569 5881605 5881606 5881609 5881610 5883375 5935249 5927232 5927233 5927575 5927576 5927237 5927496 5935248 5935250 5927231 5927439 5927482 5927574 5935251 5935252 5935253 5935256 5935257 6016791 6016793 6016795 6016796 6016799 6016800 5663826 5663827 5664341 5664501 5664502 5675474 5675475 5675476 5675478 5675553 5675555 5675735 5691289 5704435 5704436 5704438 5705748 5705749 5705757 5706020 5706043 5664501B1 5704435B1 5705749B1 5706043B1 5752178 5752179 5754851 5754852 5769620 5769741 5769744 5769888 5769889 5770236 5770237 5770635 5770636 5769888B1 5755130 5770219 5780786 5781945 5782691 5794613 5754853 5755129 5755131 5769890 5770218 5770235 5770399 5770400 5770637 5779977 5780186 5780187 5781944 5781953 5781955 5781956 5782604 5782690 5779979 5780072 5780785 5781947 5782603 5782605 5782692 5806628 5806629 5806630 5807040 5817664 5817666 5817667 5818010 5818012 5818013 5817662 5817665 5817668 5817670 5818011 5845316 5845350 5845315 5845317 5845318 5845349 5855237 5858872 5858875 5858877 5858879 5859017 5859700 5859802 5859803 5859804 5859805 5855238 5855384 5858873 5858901 5859806 5862272 5878643 5878644 5883366 5883367 5883665 5884241 5884244 5859801 5878647 5883373 5883661 5883666 5883667 5884242 5884243 5895328 5895333 5895334 5895336 5904387 5904390 5904392 5913560 5913741 5913742 5913743 5913744 5913746 5928902 5928906 5928916 5928901 5928912 5928915 6053171 5963563 5981983 5981984 5981985 5982105 5988210 5963561 5963564 5963567 6008253 6008254 6008256 6010367 5913745 5963565 5963568 5963569 5982104 5987946 5987952 5996958 5996960 5997024 5997026 5997718 5987943 5996957 5996959 5997719 6002980 6002981 6003183 6003184 6003185 6010284 6010365 6010527 5997719B1 5963200 5982103 5987942 5987944 5987945 5988211 5988230 6008242 6009366 6010283 6010285 6010366 6010528 6018433 6018618 6009365 6009373 6014717 6018298 6018299 6018432 6018619 6023176 6023177 6023178 6023179 6023180 6023181 6027675 6027676 6027677 6027678 6027853 6027854 6027855 6027856 6027857 6049619 6049621 6049950 6053172 6053173 6053397 6053399 6743817 6744446 5680492 5680494 5680542 5680543 5647550 5647551 5647552 5647553 5647554 5647555 5647556 5647558 5655791 5655792 5655793 5655794 5655795 5655797 5657236 5657237 5657238 5657239 5657240 5657381 5657382 5680499 5647553B1 5647556B1 5655792B1 5655797B1 5604252 5713419 5739804 5744862 5745066 5745067 5739804B1 5745067B1 5664134 5713417 5713418 5730523 5730524 5741623 5744751 5749834 5785678 5786088 5786089 5739805 5739809 5741624 5741626 5744753 5744754 5744874 5745979 5745980 5746002 5746004 5749831 5749832 5749833 5749836 5749838 5749942 5771749 5786091 5792320 5803638 5803794 5792317 5792318 5792319 5792561 5795509 5795573 5795574 5796479 5796695 5796479B1 5785676 5785677 5785732 5785734 5792562 5795505 5795507 5795575 5795833 5795835 5795838 5796198 5796199 5796200 5796368 5796480 5796609 5796611 5796696 5803637 5803639 5803789 5803798 5838441 5785677B1 5796609B1 5796367 5796369 5796481 5796610 5796693 5803636 5803640 5803782 5803786 5803790 5803793 5803795 5803796 5803797 5803799 5803800 5803841 5803843 5803844 5804283 5804284 5804285 5814242 5814152 5814153 5814154 5814160 5814241 5814363 5814364 5814366 5814377 5814379 5817521 5838436 5817298 5826891 5826892 5826893 5826894 5827001 5827002 5832088 5832089 5832094 5832095 5832282 5832284 5832091 5832097 5832283 5832285 5833766 5833768 5836743 5836745 5838537 5838863 5905167 5925475 5930375 5939835 5939838 5939863 5836741 5838861 5851849 5851850 5851851 5852006 5852011 5856283 5859464 5859466 5859468 5859470 5859471 5860714 5860715 5871176 5876593 5859469 5870443 5870445 5870447 5871175 5871178 5871256 5871454 5871455 5875212 5875213 5875214 5876598 5876600 5871178B1 5876602 5902558 5905135 5905164 5905168 5906112 5910057 5925366 5925367 5925368 5925926 5930373 5930378 5930379 5939833 5939839 5930373B1 5905163 5905165 5906114 5906115 5906116 5906117 5906119 5906120 5906162 5910056 5910058 5925477 5950717 5902941 5902942 5902943 5905140 5906113 5906161 5910059 5925369 5925474 5925925 5930376 5906113B1 5925927 5939837 5950608 5950610 5950716 5950718 5950720 5950865 5954289 5954296 5954297 5959207 5959208 5959209 5959542 5959552 5959201 5962345 5962346 5962349 5962352 5962354 5925924 5950715 5954300 5959210 5959217 5959545 5959546 5959551 5983565 5983566 5984527 5984528 5984530 5991233 5925476 5939832 5939840 5950609 5950611 5950612 5950719 5950866 5950867 5954288 5954295 5962350 5983567 5950868 5954290 5959547 5959548 5959549 5959550 5962348 5962351 5962353 5963575 5984226 5984227 5984229 5991238 5997801 5959543 5959544 5984532 6009055 6010665 6010785 6021625 6027242 6028299 6045882 5950607 5959202 5991237 5991832 6020722 6021635 6026815 6027088 6027243 6033903 5991235 5991833 6010666 6010670 6010784 6010789 6021401 6023372 6023374 6023403 6027086 6027244 6027245 6028301 6033799 6033976 6044316 6009051 6009052 6010782 6010787 6010788 6019019 6020655 6020656 6021622 6021623 6021624 6021626 6021636 6021637 6033793 5984228 5991234 5991834 6010667 6010669 6010783 6010790 5985012 5991830 5991831 5997796 5997797 5997798 5997799 5997800 6026816 6027085 6033975 6010664 6010786 6026814 6020658 6020659 6020661 6023371 6023373 6023375 6023404 6023405 6023406 6026817 6027087 6028300 6044311 6044313 6044314 6044317 6044318 6045887 6068722 6045815 6045820 6023376 6023402 6044312 6044315 6068724 6068725 5650177 5650179 5650222 5650223 5660345 5660346 5661620 5661722 5662477 5667689 5668545 5660345B1 5660346B1 5660347 5660359 5661721 5661723 5667692 5667694 5667942 5667945 5668459 5668543 5668544 5668688 5675517 5700896 5702186 5668543B1 5668688B1 5667690 5667943 5668240 5675512 5702187 5702188 5706382 5706384 5706387 5706448 5702187B1 5700894 5700895 5706056 5706057 5706058 5706059 5706386 5706388 5706389 5706449 5706450 5700895B1 5786224 5786232 5786233 5789305 5789307 5788996 5799897 5799978 5800207 5800217 5800443 5799894 5799979 5800206 5800208 5800216 5800218 5800219 5800441 5801481 5801482 5824012 5799895 5799977 5800445 5800448 5824006 5824008 5824013 5826727 5826725 5826732 5828553 5828554 5832081 5832083 5832085 5839195 5839196 5839197 5847897 5847898 5847899 5847941 5847945 5848005 5848006 5848151 5851632 5839194 5847896 5848148 5851282 5851285 5851633 5848020 5851239 5851630 5851631 5863584 5863585 5863586 5867587 5867588 5871250 5880071 5848020B1 5863586B1 5848019 5880070 5880073 5880075 5880076 5880077 5890274 5890275 5890276 5890509 5890510 5890513 5890514 5890515 5902504 5902505 5880076B1 5931031 5931033 5931035 5984644 5984646 5941780 5942001 5942003 5953408 5953409 5953410 5982192 5984209 5984210 5984652 5942004 5962709 5963381 5963382 5963384 5963386 5984215 5931032 5931034 5942002 5947020 5947161 5962710 5963383 5975589 5982188 5982189 5984643 5984645 5984647 5987961 5995098 5995099 5995747 5962708 5975584 5975587 5975588 5984649 5984653 5987960 5947162 5975586 5982190 5982193 5984208 5984650 5984651 5984654 5988082 5994920 5995093 5995097 5995654 5995655 5995656 5995658 6002684 6002685 6002686 6002766 6002768 6002770 6002914 6002915 6003082 6006852 6006854 6007087 6007240 6007671 6007677 6007707 5995746 6002771 6003021 6006849 5982191 6006850 6007090 6007239 6007241 6007669 6007679 6018859 6006851 6006984 6006985 6006986 6007091 6007092 6007238 6018860 6019017 6006853 6006987 6007086 6007089 6007236 6007237 6007705 6018911 6021695 6021696 6026000 6026001 6025706 6025998 6025999 6025501 6025704 6025705 5651592 5651594 5651595 5651596 5651600 5654034 5654035 5654036 5654037 5654038 5654040 5654521 5654523 5654525 5661725 5661726 5661727 5662476 5654037B1 5662476B1 5661724 5693947 5693950 5651593 5693776 5693948 5721173 5745966 5786234 5786235 5786236 5745962 5745964 5745963 5745967 5761360 5763681 5763682 5770925 5770927 5770939 5771141 5771146 5771313 5771571 5771573 5770925B1 5771313B1 5770926 5771312 5799093 5799255 5799096 5799097 5799253 5799254 5799444 5799445 5814514 5814515 5814516 5814517 5814520 5814521 5814532 5814537 5832352 5814531 5814535 5832348 5832353 5836109 5848725 5848730 5848734 5853923 5853924 5853925 5853927 5871514 5848735 5871230 5871231 5871232 5871234 5871451 5871453 5871515 5871516 5886093 5886095 5886220 5886518 5886635 5886092 5886219 5902518 5902519 5902520 5902621 5937921 5871452 5871513 5886094 5886322 5886325 5886519 5886627 5886628 5886632 5898511 5898512 5938917 5942909 5942910 5943127 5956355 5956358 5938916 5942907 5942908 5942918 5942919 5942920 5943032 5943033 5943034 5943036 5943125 5943035 5943124 5943126 5956357 5959334 5959335 5959336 5959337 5963387 5963388 5963666 5963668 5967761 5956353 5956359 5956360 5963665 5963669 5963670 5963671 5967719 5967720 5967762 5967763 5967764 5967765 5967766 5967767 5967768 5967769 5967765B1 6002694 6002695 6018343 6028845 6031894 6031896 6031998 6032253 6002683 6031997 6032254 6055451 6002687 6002254 6002255 6002256 6015633 6018348 6054137 6054757 6055458 6055503 6057310 6057781 6031996 6054140 6054760 6055454 6058273 6058274 6018344 6028843 6028844 6031895 6032312 6032313 6054758 6057863 6058272 6054138 6054755 6058529 6054139 6055455 6057315 6057780 6057862 6089106 6089107 6089108 6089971 5756523 5756524 5756526 5756527 5756529 5756535 5756536 5769339 5769353 5769354 5769355 5769441 5756523B1 5756525 5769340 5779161 5779162 5779163 5786389 5798017 5798020 5798361 5798362 5786389B1 5784506 5784507 5784508 5784509 5798018 5798019 5812740 5812902 5812903 5814655 5814656 5814657 5814658 5820966 5821000 5821081 5821082 5821099 5821100 5821241 5821429 5835230 5835234 5835321 5835322 5835567 5835568 5835569 5835570 5887144 5903671 5903673 5911624 5869301 5869302 5869303 5869305 5887143 5887145 5887146 5887205 5887208 5887210 5887403 5887404 5887405 5896060 5896083 5903670 5903672 5911302 5911623 5930607 5935095 5947176 5923909 5947010 5947011 5947012 5947177 5947179 5947491 5947492 5947493 6010980 6011043 6328457 6010981 6011042 6049074 6049075 6049076 6049077 6049078 6049079 6056660 6056684 6057446 6057448 6057648 6057649 6057650 6057651 6057656 6057657 6057658 6056661 6056683 6057447 6305290 6305291 6305293 6305294 6317019 6317020 6317021 6317122 6317123 6317124 6317125 6317126 6328450 6328451 6328452 6328453 6328454 6328455 6328456 6750797 5639670 5639672 5643607 5672239 5672813 5672815 5672816 5676394 5677455 5677742 5643607B1 5672817 5676398 5676441 5676443 5676591 5677457 5677522 5677691 5677744 5672814 5709203 5709205 5709207 5709208 5739283 5739285 5739287 5819818 5819818B1 5790658 5802569 5802570 5802919 5802920 5803061 5803062 5803201 5803203 5808712 5808988 5808991 5809299 5809370 5811062 5811063 5811145 5811153 5811154 5814776 5832327 5802919B1 5802572 5803202 5808987 5808990 5808992 5809298 5809300 5809371 5809373 5809374 5809371B1 5809371B2 5811065 5814651 5814652 5814653 5814654 5814769 5814771 5814773 5815269 5815270 5815453 5815454 5815456 5815457 5828798 5814651B1 5828614 5828615 5828616 5844517 5903831 5927040 5927268 5862107 5875069 5875476 5875479 5875480 5875566 5875567 5877143 5877144 5877145 5885654 5895304 5877146 5885653 5895303 5895308 5902316 5903822 5903823 5903833 5904037 5905172 5905179 5930948 5931023 5875070 5875071 5875561 5902318 5902319 5905221 5905222 5917920 5927269 5927562 5902319B1 5886686 5886687 5902315 5904034 5885651 5902314 5927182 5927270 5927563 5927564 5930609 5930786 5930787 5930788 5930947 5930949 5931024 5902317 5905173 5927181 5927570 5930608 No 4.0 mm were distributed OUS No 5.0 mm were distributed OUS On August 5, 2004 an additional 38 lots were added to this recall: US Distribution only - 6536831 6538618 6539542 6539994 6540267 6540401 6540900 6540901 6540946 6541537 6543419 6543678 6543947 6544274 6544279 6544736 6544983 6544986 6545001 6545231 6545561 6547363 6547820 6547918 6548002 6548397 6548493 6548712 6548890 6548972 6548973 6549073 6549710 6549715 6549718 6550086 6550094 6550095 OUS - 6816206 6813839 6811839 6817267 6817117 |
Recalling Firm/ Manufacturer |
Boston Scientific Scimed 1 Scimed Pl Maple Grove MN 55311-1565
|
For Additional Information Contact |
Michelle Gudith 763-494-1194
|
Manufacturer Reason for Recall |
On a few occasions the balloon has failed to deflate within one minute after deployment of the stent.
|
FDA Determined Cause 2 |
Other |
Action |
July 2 and 16, 2004 recall letters were sent to user accounts, such as Hospitals and clinics of the need to recall TAXUS Express2 Paclitaxel-Eluting and Express 2 Coronary Stent Systems. Recall was due to characteristics in the delivery catheters that had the potential to impede balloon deflation during a coronary angioplasty procedure, a condition known as focal neckdown. An additional 38 lots were added to the recall on August 5, 2004, for the same reason. |
Quantity in Commerce |
205,983 units plus additional 3,000 units recalled on 8/5/04. |
Distribution |
Nationwide and Canada |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
PMA Database |
PMAs with Product Code = NIQ and Original Applicant = Boston Scientific Corp.
|