  Class 2 Device Recall see related information
Date Initiated by Firm June 10, 2004
Date Posted July 27, 2004
Recall Status1 Terminated 3 on November 01, 2004
Recall Number Z-1165-04
Recall Event ID 29300
PMA Number P020018 
Product Classification System, Endovascular Graft, Aortic Aneurysm Treatment - Product Code MIH
Product Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded bifurcated main body graft - aortic graft diameter 22 mm - iliac leg diameters 11 mm - main body lengths cntralateral limb 88mm ipsilateral limb 133 mm - delivery system 18 FR; Catalog no. TFB-22-88.
Code Information Lots 1349235, F1548657, F1545296, F1552607, F1548658, F1552608, F1552609, F1552610, F1549743, F1549241, F1560040, F1560039, F1560041, 1349236 and 1349237.
Recalling Firm/
Manufacturer
Cook, Inc.
750 Daniels Way
Bloomington IN 47404-9120
For Additional Information Contact Customer Relations
800-457-4500
Manufacturer Reason
for Recall
Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
FDA Determined
Cause 2
Other
Action The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Quantity in Commerce 15
Distribution Nationwide, Canada, Denmark and Japan.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
PMA Database PMAs with Product Code = MIH and Original Applicant = COOK, INC.
-
-