New Search Back to Search Results
  Class 3 Device Recall see related information
Date Initiated by Firm March 24, 2005
Date Posted June 16, 2005
Recall Status1 Terminated 3 on January 23, 2012
Recall Number Z-0893-05
Recall Event ID 31586
510(K)Number K935673  
Product Classification Visual, Pregnancy Hcg, Prescription Use - Product Code JHI
Product AxSYM total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline Phosphatase Conjugate and one bottle of Specimen Diluent.
Code Information 2614Q100 (5/13/2005);  26278Q100, 26381Q100 (6/21/2005)
Recalling Firm/
Manufacturer
Abbott Health Products, Inc.
Hwy 2 Km 58.0
Barceloneta PR 00617
For Additional Information Contact Vana L. Smith, PhD
877-422-2688
Manufacturer Reason
for Recall
Abbott Diagnostics is investigation a recent increase in customer complaints for AxSYM calibration error code 1048 and controls out of range when using AxSYM Total B-hCG reagent lots. Testing has shown that protein aggregates are present in the specimen diluents of both IMx and the AxSYM Total B-hCG reagent kits. This is responsible for the calibration errors and shift in control values.
FDA Determined
Cause 2
Other
Action Abbot Health Products, Inc. informed and instructed their customers via Device Recall Letter to Identify if they have used or have inventory of the affected product. Discontinue use of and destroy any remaining inventory of reagent lots listed in the Device Recall letter according to their laboratory procedures. Record the quantity of the recalled reagents currently in inventory in their facility that will be destroyed on the attached reply form and fax the form back to Abbott 1-800-777-0051. (US only) Follow their laboratory''s procedure for communicating this issue to the health care providers they serve.
Quantity in Commerce 2496
Distribution Domestic , US Goverment and International ditribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JHI and Original Applicant = ABBOTT LABORATORIES
-
-