  Class 2 Device Recall Terumo APS 1 see related information
Date Initiated by Firm October 12, 2004
Date Posted December 30, 2006
Recall Status1 Terminated 3 on February 01, 2007
Recall Number Z-0327-2007
Recall Event ID 36396
510(K)Number K022947  
Product Classification Perfusion System - Product Code DTQ
Product Terumo Advanced Perfusion System 1 Data Interface CDI 100; Catalog number 802558.
Code Information Serial numbers 00008 through 00011, 00013, 00020 through 00030, 00032 through 00034, 00036 through 00059 and 00061 through 00078.
Recalling Firm/
Manufacturer
Terumo Cardiovascular Systems Corp
6200 Jackson Rd
Ann Arbor MI 48103-9586
For Additional Information Contact
800-521-2818
Manufacturer Reason
for Recall
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
FDA Determined
Cause 2
Other
Action U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Quantity in Commerce 4740 for all affected products
Distribution Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = DTQ and Original Applicant = TERUMO CARDIOVASCULAR SYSTEMS CORP.
-
-