• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Terumo APS 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Terumo APS 1 see related information
Date Initiated by Firm October 12, 2004
Date Posted December 30, 2006
Recall Status1 Terminated 3 on February 01, 2007
Recall Number Z-0333-2007
Recall Event ID 36396
510(K)Number K022947  
Product Classification Perfusion System - Product Code DTQ
Product Terumo Advanced Perfusion System 1 Flow Module; Catalog number 802018.
Code Information Serial numbers 00017, 00020, 00023 through 00026, 00029 through 00031, 00033, 00034, 00036 through 00048, 00051 through 00055, 00057 through 00078, 00080 through 00120, 00122 through 00134, 00136, 00138 through 00155, 00158 through 00169, 00171 through 00210, 00212 through 00243 and 00248.
Recalling Firm/
Manufacturer
Terumo Cardiovascular Systems Corp
6200 Jackson Rd
Ann Arbor MI 48103-9586
For Additional Information Contact
800-521-2818
Manufacturer Reason
for Recall
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
FDA Determined
Cause 2
Other
Action U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Quantity in Commerce 4740 for all affected products
Distribution Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = DTQ and Original Applicant = TERUMO CARDIOVASCULAR SYSTEMS CORP.
-
-