|
Class 2 Device Recall GE Precision XR/a |
|
Date Initiated by Firm |
March 29, 2007 |
Date Posted |
September 30, 2008 |
Recall Status1 |
Terminated 3 on February 04, 2012 |
Recall Number |
Z-1713-2008 |
Recall Event ID |
49533 |
510(K)Number |
K993090
|
Product Classification |
X-Ray System - Product Code KPR
|
Product |
GE Healthcare Proteus XR/a Eclipse Collimator model # 2379827. |
Code Information |
42948HL0, 42949HL8, 43054HL6, 43055HL3, 43056HL1, 43057HL9, 43132HL0, 43133HL8, 43134HL6, 43135HL3, 43201HL3, 43202HL1, 43203HL9, 44605HL4, 44606HL2, 44607HL0, 44608HL8, 44609HL6, 44610HL4, 44681HL5, 45046HL0, 45047HL8, 45048HL6, 45128HL6, 45129HL4, 45130HL2, 45131HL0, 48119HL2, 48120HL0, 48121HL8, 48122HL6, 48123HL4, 48124HL2, 48125HL9, 48126HL7, 48127HL5, 48128HL3, 48164HL8, 48181HL2, 48274HL5, 48276HL0, 48277HL8, 48278HL6, 48680HL3, 48682HL9, 48683HL7, 48684HL5, 48685HL2, 48686HL0, 48687HL8, 48688HL6, 48689HL4, 48876HL7, 48877HL5, 48878HL3, 48995HL3, 48997HL1, 48998HL9, 49000HL3, 49208HL2, 49209HL0, 49210HL8, 49211HL6, 49391HL6, 49392HL4, 49393HL2, 49395HL7, 49494HL8, 49496HL3, 49497HL1, 49498HL9, 49499HL7, 49500HL2, 49501HL0, 49502HL8, 49503HL6, 49681HL0, 49682HL8, 49683HL6, 49684HL4, 49685HL1, 49686HL9, 49687HL7, 49688HL5, 49850HL1, 49851HL9, 49852HL7, 49853HL5, 49855HL0, 49856HL8, 49857HL6, 50027HL2, 50028HL0, 50029HL8, 50030HL6, 50031HL4, 50177HL5, 50178HL3, 50179HL1, 50180HL9, 50321HL9, 50322HL7, 50323HL5, 50324HL3, 50444HL9, 50445HL6, 50446HL4, 50447HL2, 50448HL0, 50449HL8, 50450HL6, 50451HL4, 50452HL2, 50453HL0, 50790HL5, 50791HL3, 50792HL1, 50793HL9, 50794HL9, 50795HL4, 50797HL0, 50798HL8, 50801HL0, 51143HL6, 51144HL4, 51145HL1, 51147HL7, 51486HL9, 51487HL7, 51488HL5, 52014HL8, and 70RD05001 |
Recalling Firm/ Manufacturer |
GE Healthcare 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
608-221-1551
|
Manufacturer Reason for Recall |
The actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.
|
FDA Determined Cause 2 |
Radiation Control for Health and Safety Act |
Action |
GE will be correcting the units by replacing all collimators that don't meet the 160-lux requirement. The field correction will be initiated and is scheduled to be completed by October 2008. |
Quantity in Commerce |
132 |
Distribution |
Worldwide Distribution --- USA including states of CA, DE, FL, GA, ID, IL, KS, KY, MD, MI, MO, MT, NC, NJ, NM, NY, OH, PA, SC, TX, UT, VA, and WV, and countries of Australia, Canada, China, France, Germany, Greece, Italy, Korea, Poland, Saudi Arabia, United Kingdom, and Venezuela. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPR and Original Applicant = GE MEDICAL SYSTEMS, INC.
|
|
|
|