 Class 1 Device Recall Shiley Single Cannula Cuffed Tracheostomy Tubesee related information
Date Initiated by FirmApril 13, 2010
Date PostedMay 06, 2010
Recall Status1 Terminated 3 on January 18, 2012
Recall NumberZ-1487-2010
Recall Event ID 55354
510(K)NumberK810106 
Product Classification Tube, tracheostomy (w/wo connector) - Product Code BTO
ProductShiley Single Cannula Cuffed Tracheostomy Tube, 10 SCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions.
Code Information Lot Numbers: 0907002007 through 0908001339
Recalling Firm/
Manufacturer
Nellcor Puritan Bennett Inc. (dba Covidien Ltd)
6135 Gunbarrel Ave
Boulder CO 80301-3214
For Additional Information Contact
303-305-2382
Manufacturer Reason
for Recall
Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
FDA Determined
Cause 2
Process design
ActionCovidien issued an "Urgent Medical Device Recall" notification dated April 13, 2010. Consignees were alerted to the affected device. If an affected product is already in use, the firm recommends that the tube be replaced as soon as the patient's physician determines it is clinically appropriate. All affected product is to be returned to the firm. For further Information, contact Covidien at 1-800-635-5267.
Quantity in Commerce230 units
DistributionWorldwide Distribution -- United States (including Guam and Puerto Rico), Australia, Belgium, Brazil, Canada, Chile, Colombia, Honduras, Japan, Mexico, Panama, Singapore and Uruguay.
Total Product Life CycleTPLC Device Report