|
Class 2 Device Recall Signa Lx |
|
Date Initiated by Firm |
November 09, 2010 |
Date Posted |
February 24, 2011 |
Recall Status1 |
Terminated 3 on July 03, 2012 |
Recall Number |
Z-1424-2011 |
Recall Event ID |
57084 |
510(K)Number |
K962061
|
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product |
1) Signa Lx ( K962061 Signa Horizon Cx)
The Signa Horizon Cx System is indicated for use as a diagnostic imaging device to produce transverse, saggital, coronal and oblique images of the internal structures of the head or body. The images produced by the signa Horizon Cx System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
|
Code Information |
SERIAL 00000232068MR7 00000200618MR7 00000253500MR3 00000161854MR5 00000161128MR4 00000212453MR5 00000205557MR2 00000194277MR0 00000210629MR2 00000210789MR4 00000202408MR1 00000241094MR2 00000227605MR3 00000157686MR7 00000215316MR1 00000180421MR0 00000215313MR8 00000180899MR7 00000143624MR5 00000180897MR1 00000112550MR9 00000115704MR9 00000193709MR3 00000119503MR1 00000194626MR8 00000112027MR8 00000216508MR2 00000228081MR6 00000125783MR1 00000172839MR3 00000169151MR8 00000132803MR8 00000127413MR3 00000170580MR5 00000186751MR4 4351 00000235440MR5 00000170568MR0 00000220886MR6 00000198717MR1 00000156808MR8 00000205961MR6 00000226212MR9 00000241637MR8 00000155840MR2 00000181846MR7 00000217037MR1 00000148024MR3 00000226765MR6 00000159686MR5 00000222678MR5 00000178826MR4 00000183398MR7 00000218457MR0 00000159580MR0 00000164251MR1 00000235742MR4 00000170910MR4 00000121825MR4 00000239201MR7 00000165426MR8 00000124330MR2 00000231220MR5 00000123369MR1 00000208845MR8 00000166125MR5 00000142605MR5 00000215887MR1 00000199112MR4 00000167817MR6 00000151839MR8 00000180083MR8 00000154237MR2 00000232234MR5 00000223690MR9 00000224283MR2 00000214786MR6 00000214332MR9 00000221502MR8 00000238337MR0 00000235093MR2 00000151024MR7 00000126178MR3 00000168284MR8 00000212445MR1 00000233208MR8 00000162577MR1 00000214307MR1 00000137614MR4 00000233406MR8 00000172733MR8 00000116883MR0 00000108299MR9 00000166734MR4 00000194640MR9 00000200456MR2 00000235445MR4 4563 00000194635MR9 00000134508MR1 00000258182MR5 00000225022MR3 00000179498MR1 00000222291MR7 00000195819MR8 00000233015MR7 00000122281MR9 00000175057MR9 00000172833MR6 00000151957MR8 00000111157MR4 00000153576MR4 00000201909MR9 00000118962MR0 00000113891MR6 00000173562MR0 00000122780MR0 00000104691MR1 00000148234MR8 00000110539MR4 00000169157MR5 00000162596MR1 00000220883MR3 00000188412MR1 00000151995MR8 00000158407MR7 00000149036MR6 00000208841MR7 00000223493MR8 00000250829MR9 00000126170MR0 00000214783MR3 00000178814MR0 00000198053MR1 00000198726MR2 00000195806MR5 00000123365MR9 00000117426MR7 00000165790MR7 00000165893MR9 00000231850MR9 00000228263MR0 00000122352MR8 00000186754MR8 00000165431MR8 00000208812MR8 00000180436MR8 00002242271MR7 00000166444MR0 00000227939MR6 00000154876MR7 00000225881MR2 00000199734MR5 00000205542MR4 00000121256MR2 00000174692MR4 00000104728MR1 00000202048MR5 00000228542MR7 00000183397MR9 00000161850MR3 00000238787MR6 00000242501MR5 00000171554MR9 00000234778MR9 00000199096MR9 00000180905MR2 00000231332MR8 00000242819MR1 00000175034MR8 00000139719MR9 00000203407MR2 00000241159MR3 00000210640MR9 00000230634MR8 00000236233MR3 00000234874MR6 00000214768MR4 00000129567MR4 00000225341MR7 00000264785MR7 00000221898MR0 00000137858MR7 00000143615MR3 00000232057MR0 00000214355MR0 00000198054MR9 00000180086MR1 00000241160MR1 00000191917MR4 00000213656MR2 00000248919MR3 00000121831MR2 00000235094MR0 00000140903MR6 00000235722MR6 00000186261MR4 00000218181MR6 00000209588MR3 00000168170MR9 00000210326MR5 00000124046MR4 00000154116MR8 00000217380MR5 00000223029MR0 00000222899MR7 00000200446MR3 00000225893MR7 00000200615MR3 535668 00000135085MR9 00000193677MR2 00000211408MR0 00000149038MR2 00000210633MR4 00000235080MR9 00000169980MR0 00000214318MR8 00000248491MR3 00000211060MR9 00000237552MR5 00000186703MR5 00000140831MR9 00000188965MR8 00000219139MR3 00000229675MR4 00000135454MR7 00000142608MR9 00000206863MR3 00000163657MR0 00000228453MR7 00000124047MR2 00000126167MR6 00000200716MR9 00000174903MR5 00000213058MR1 00000200447MR1 00000177120MR3 00000108050MR6 00000149025MR9 00000133635MR3 00000257895MR3 00000113567MR2 00000168600MR5 00000217059MR5 00000219798MR6 00000234646MR8 00000214805MR4 00000217045MR4 00000188414MR7 00000170572MR2 00000181926MR7 00000221883MR2 00000236241MR6 00000198326MR1 00000175584MR2 00000107828MR6 00000238227MR3 00000184986MR8 00000212539MR1 00000138651MR5 00000171559MR8 00000202975MR9 00000136090MR8 00000219355MR5 00000009589GQ3 00000234249MR1 00000243480MR1 00000211062MR5 00000134300MR3 00000235863MR8 00000238662MR1 00000218880MR3 00000219145MR0 00000207222MR1 00000128859MR6 00000132797MR2 00000117782MR3 00000115702MR3 00000115446MR7 00000117235MR2 00000206390MR7 00000136087MR4 00000259460MR4 00000112346MR2 00000144306MR8 00000230461MR6 00000218733MR4 00000191050MR4 00000207843MR4 00000204226MR5 00000206857MR5 00000177768MR9 00000186746MR4 00000116879MR8 00000122905MR3 00000106749MR5 00000221117MR5 00000179827MR1 00000191918MR2 00000144302MR7 00000236232MR5 00000112355MR3 00000137855MR3 00000218081MR8 00000215512MR5 00000231219MR7 00000122345MR2 00000118123MR9 00000217027MR2 00000223638MR8 00000228633MR4 00000174096MR8 00000211071MR6 00000233224MR5 00000179225MR8 00000115017MR6 00000231346MR8 00000188411MR3 00000238113MR5 00000246547MR4 00000170567MR2 00000210075MR8 00000246546MR6 4209 00000112545MR9 00000243175MR7 00000191713MR7 00000130802MR2 00000119495MR0 00000112552MR5 00000186256MR4 00000242464MR6 00000214322MR0 00000199126MR4 00000248918MR5 00000137610MR2 00000211056MR7 00000200439MR8 00000200440MR6 00000231845MR9 00000232061MR2 00000184991MR8 00000185669MR9 00000170008MR7 00000238335MR4 00000180987MR0 00000241154MR4 00000218228MR5 00000179495MR7 00000143616MR1 00000159680MR8 00000236656MR5 00000156830MR2 00000247769MR3 00000209586MR7 00000172716MR3 00000210060MR0 00000188968MR2 00000220630MR8 00000174127MR1 00000154663MR9 00000185670MR7 00000180080MR4 00000172842MR7 00000210783MR7 00000171552MR3 00000238521MR9 00000200741MR7 00000232590MR0 00000164479MR8 00000023067MR1 00000199354MR2 00000199746MR9 00000214224MR8 00000220885MR8 00000202483MR4 00000148006MR0 00000242496MR8 00000170909MR6 00000219807MR5 00000224270MR9 00000188964MR1 00000166954MR8 00000132806MR1 00000149024MR2 00000234867MR0 00000175035MR5 00000210798MR5 00000169450MR4 00000135342MR4 00000127053MR7 00000248334MR5 00000168586MR6 00000170991MR4 00000239401MR3 00000139717MR3 00000226228MR5 00000237994MR9 00000172664MR5 00000207855MR8 00000205981MR4 00000247024MR3 00000233218MR7 00000116499MR5 00000139119MR2 00000225885MR3 000000182739MR 00000115016MR8 00000127189MR9 00000132950MR7 00000133634MR6 00000170579MR7 00000128858MR8 00000194312MR5 00000224886MR2 00000213674MR5 00000205127MR4 00000246559MR9 00000185667MR3 00000241288MR0 00000206866MR6 00000196850MR2 00000180105MR9 00000233223MR7 00000187028MR6 00000237755MR4 00000224793MR0 00000186706MR8 00000208144MR6 00000174002MR6 00000179503MR8 00000132944MR0 00000154242MR2 00000222793MR2 00000170964MR1 00000228079MR0 00000191733MR5 00000174104MR0 00000200469MR5 00000139712MR4 00000165432MR6 00000138897MR4 00000142179MR1 00000142759MR0 00000179229MR0 00000239197MR7 00000127414MR1 00000179118MR5 00000180419MR4 00000161839MR6 00000187567MR3 00000119501MR5 00000132595MR0 00000249016MR7 00000239404MR7 00000198742MR9 00000215997MR8 00000194275MR4 00000183395MR3 00000130799MR0 00000158832MR6 00000177774MR7 00000227933MR9 00000152580MR7 00000229035MR1 00000209418MR3 00000221171MR2 00000122917MR8 00000200468MR7 00000169148MR4 00000174098MR4 00000185691MR3 00000130402MR1 00000148023MR5 00000167441MR5 00000122901MR2 00000164200MR8 00000112032MR8 00000142353MR2 00000167856MR4 00000233636MR0 00000142600MR6 00000116882MR2 00000130803MR0 535397 00000148029MR2 00000118128MR8 00000158689MR0 00000158693MR2 00000148004MR5 00000229412MR2 00000178823MR1 00000241293MR0 00000198720MR5 00000171436MR9 00000117229MR5 00000195076MR5 00000153276MR1 00000169150MR0 00000223636MR2 00000164345MR1 00000140607MR3 00000164192MR7 00000218301MR0 00000199358MR3 00000216709MR6 00000236236MR6 00000194279MR6 00000175572MR7 00000240233MR7 00000197817MR0 00000205976MR4 00000220420MR4 00000160058MR4 00000235604MR6 00000227608MR7 00000152415MR6 00000190307MR9 00000210123MR6 00000205123MR3 00000248333MR7 00000209583MR4 00000195094MR8 00000225091MR8 00000234246MR7 00000225342MR5 00000202243MR2 00000228626MR8 00000181841MR8 00000241286MR4 00000242135MR2 00000195079MR9 00000214333MR7 00000192330MR9 00000239204MR1 00000237113MR6 00000226231MR9 00000196464MR2 00000226749MR0 00000229409MR8 00000211913MR9 00000180989MR6 00000214222MR2 00000249404MR5 00000213650MR5 00000205969MR9 00000225888MR7 00000243264MR9 00000210637MR5 00000196472MR5 00000237762MR0 00000191722MR8 00000199637MR0 00000199459MR9 00000209420MR9 00000145113MR7 00000187897MR4 00000208784MR9 00000205966MR5 00000153282MR9 00000120988MR1 00000196473MR3 00000112935MR2 00000225464MR7 00000214317MR0 00000123368MR3 00000123372MR5 00000116887MR1 00000103317MR4 00000188420MR4 00000230632MR2 00000130805MR5 00000180082MR0 00000121823MR9 00000219646MR7 00000205964MR0 00000126638MR6 00000245317MR3 00000186267MR1 00000148027MR6 00000179500MR4 00000259303MR6 00000223552MR1 00000120517MR8 00000196451MR9 00000211084MR9 00000178168MR1 00000168285MR5 00000170577MR1 00000245319MR9 00000242499MR2 00000241765MR7 00000238414MR7 00000208840MR9 00000193244MR1 00000242807MR6 00000169174MR0 00000258392MR0 00000246772MR8 00000226686MR4 00000226213MR7 00000183411MR8 00000231832MR7 00000115716MR3 00000122774MR3 00000175021MR5 00000198724MR7 00000225087MR6 00000224268MR3 00000238336MR2 00000175425MR8 00000233209MR6 00000238519MR3 00000173558MR8 00000231331MR0 00000177773MR9 00000161127MR6 00000191743MR4 00000162590MR4 00000168965MR2 00000243675MR6 00000145682MR1 00000185668MR1 00000118120MR5 00000234871MR2 00000226515MR5 00000245214MR2 00000227747MR3 00000135086MR7 00000148261MR1 00000172163MR8 00000133307MR9 00000020535MR7 00000206606MR6 00000187570MR7 00000236506MR2 00000171592MR9 00000216501MR7 00000166122MR2 00000167438MR1 00000145686MR2 00000246868MR4 00000181911MR9 00000110460MR3 00000206394MR9 00000223260MR1 00000118960MR4 00000133745MR0 00000210796MR9 00000131062MR2 00000121257MR0 00000218452MR1 00000163656MR2 00000179122MR7 00000200437MR2 00000207840MR0 4269 00000204725MR6 00000223258MR5 00000234643MR5 00000180902MR9 00000240312MR9 3366 00000175574MR3 00000179390MR0 00000189562MR2 00000188544MR1 00000227246MR6 00000204694MR4 00000229966MR7 00000194623MR5 00000221115MR9 00000203717MR4 00000213689MR3 00000120521MR0 00000180417MR8 00000170978MR1 00000198314MR7 00000166709MR6 00000187580MR6 00000139127MR5 00000197816MR2 00000239668MR7 00000210423MR0 00000228543MR5 00000175798MR8 00000215521MR6 00000183401MR9 00000215503MR4 00000141667MR6 00000210055MR0 00000175786MR3 00000160299MR4 00000209437MR3 00000198324MR6 00000194627MR6 00000188966MR6 00000193372MR0 00000215884MR8 00000156723MR9 00000257937MR3 00000249634MR7 00000152584MR9 4429 00000167444MR9 00000131656MR1 00000199101MR7 00000127196MR4 00000229943MR6 00000219142MR7 00000242919MR9 00000141477MR0 00000165425MR0 535557 00000226223MR6 00000243676MR4 00000154667MR0 00000148232MR2 00000217980MR2 00000235868MR7 00000214320MR4 535722 00000199102MR5 00000179119MR3 00000238898MR1 00000211340MR5 00000217986MR9 00000212536MR7 00000116881MR4 00000228077MR4 00000166139MR6 00000166129MR7 00000104499MR9 00000159302MR9 00000193923MR0 00000218182MR4 00000156085MR3 00000112824MR8 00000172141MR4 00000147991MR4 00000179823MR0 00000179828MR9 00000237275MR3 00000147584MR7 00000183396MR1 00000113674MR6 00000205533MR3 00000170979MR9 00000214233MR9 00000229267MR0 00000164343MR6 00000173710MR5 00000192319MR2 00000127408MR3 00000164347MR7 00000179120MR1 00000179120MR1 00000104664MR8 00000225337MR5 00000214816MR1 00000177112MR0 00000169146MR8 00000245471MR8 00000236140MR0 00000204261MR2 4216 00000225875MR5 00000227612MR9 00000176698MR9 00000222665MR2 00000107705MR6 00000172851MR8 00000229945MR1 00000230627MR2 00000221416MR1 00000246556MR5 00000223525MR7 00000186271MR3 00000196442MR8 98120050 00000137037MR8 00000117776MR5 00000108694MR1 00000216723MR7 00000206881MR5 00000205128MR2 00000186262MR2 00000222173MR7 00000171974MR9 00000130791MR7 535833 00000133636MR1 00000133280MR8 00000128407MR4 00000140608MR1 00000240234MR5 00000218725MR0 00000147593MR8 00000143655MR9 00000125780MR7 00000145014MR7 00000207219MR7 00000199641MR2 535806 00000181829MR3 00000115090MR3 00000217975MR2 00000250673MR1 00000169142MR7 00000119500MR7 00000128854MR7 00000239390MR8 00000183403MR5 00000230623MR1 00000222289MR1 00000192338MR2 00000000D0C3VG 00000222383MR2 00000147988MR0 00000197836MR0 00000229958MR4 00000230258MR6 00000206125MR7 00000187592MR1 00000202460MR2 00000213588MR7 00000198327MR9 00000234648MR4 00000170960MR9 00000133273MR3 00000145099MR8 00000212956MR7 00000264790MR7 00000130406MR2 00000014113MR7 00000210430MR5 00000181839MR2 00000181913MR5 00000236234MR1 00000214775MR9 00000222667MR8 00000114140MR7 00000168629MR4 00000221174MR6 00000212958MR3 00000186701MR9 976 00000117554MR6 00000152599MR7 00000249153MR8 00000210782MR9 00000176675MR7 00000165972MR1 00000225884MR6 00000200436MR4 00000169551MR9 00000206594MR4 00000122779MR2 00000148028MR4 00000226767MR2 00000215333MR6 00000235278MR9 00000238899MR9 00000209438MR1 00000206855MR9 00000156082MR0 00000115440MR0 00000218443MR0 00000148629MR9 00000127416MR6 00000245782MR8 00000157705MR5 00000194292MR9 00000236066MR7 00000184981MR9 00000169156MR7 00000173707MR1 00000133315MR2 00000134509MR9 00000227094MR0 00000209444MR9 00000219639MR2 00000132798MR0 00000160055MR0 00000206377MR4 00000188980MR7 00000226956MR1 00000199360MR9 00000187639MR0 00000187577MR2 00000200719MR3 00000142607MR1 00000160290MR3 00000243468MR6 00000209597MR4 00000197813MR9 00000194284MR6 00000189837MR8 00000208813MR6 00000174696MR5 00000191746MR7 00000221895MR6 00000194286MR1 00000118447MR2 00000151983MR4 00000212377MR6 00000209439MR9 00000108695MR8 00000171551MR5 00000154886MR6 00000143668MR2 00000168599MR9 00000153590MR5 00000209412MR6 00000175422MR5 00000229707MR5 535707 00000185005MR6 00000250828MR1 00000000XXXXXX 00000168299MR6 00000171599MR4 00000228073MR3 00000180078MR8 00000158823MR5 00000169959MR4 00000226516MR3 00000133746MR8 00000117764MR1 00000111151MR7 00000218882MR9 00000230622MR3 00000137303MR4 00000214789MR0 00000116186MR8 00000223199MR1 00000187583MR0 00000180075MR4 00000140499MR5 00000235741MR6 00000117765MR8 00000142193MR2 00000127055MR2 00000154246MR3 00000255252MR9 00000233205MR4 00000136091MR6 00000214004MR4 00000167854MR9 00000235081MR7 535365 00000210078MR2 00000144359MR7 00000166442MR4 00000221178MR7 792 00000216719MR5 00000231844MR2 00000169529MR5 00000200442MR2 00000219165MR8 00000156692MR6 00000161881MR8 00000169140MR1 00000231339MR3 00000175418MR3 00000216605MR6 00000169969MR3 00000169546MR9 00000189510MR1 00000217431MR6 00000223268MR4 00000231500MR0 00000146386MR8 00000189566MR3 00000209414MR2 00000153561MR6 00000191036MR3 00000148626MR5 00000219651MR7 00000166953MR0 00000148633MR1 00000186259MR8 00000225889MR5 00000231851MR7 00000177118MR7 00000199456MR5 00000200735MR9 00000139122MR6 00000225873MR9 00000137297MR8 00000154259MR6 00000194305MR9 00000154109MR3 00000240392MR1 00000139124MR2 00000246254MR7 00000168288MR9 00000246773MR6 00000239649MR7 00000237273MR8 00000217652MR7 00000191042MR1 00000225344MR1 00000237126MR8 00000239645MR5 00000161131MR8 00000204255MR4 00000246867MR6 00000141661MR9 00000170585MR4 00000240314MR5 00000239388MR2 00000154905MR4 00000244540MR1 00000225874MR7 00000218445MR5 4513 00000212544MR1 00000237276MR1 00000223198MR3 00000162607MR6 00000151032MR0 00000219154MR2 00000128860MR4 00000225589MR1 00000166948MR0 00000151985MR9 00000193675MR6 00000236905MR6 00000205526MR7 00000246781MR9 00000153277MR9 00000253347MR9 00000229940MR2 00000141754MR2 00000169799MR4 00000137199MR6 00000118727MR7 00000225081MR9 00000185680MR6 00000153582MR2 00000236663MR1 00000141713MR8 00000201049MR4 00000122343MR7 00000199738MR6 00000191726MR9 00000117226MR1 00000231846MR7 00000242497MR6 00000211903MR0 00000243467MR8 00000147929MR4 00000225892MR9 00000199453MR2 00000151022MR1 00000120331MR4 00000132952MR3 00000117769MR0 00000249236MR1 00000231834MR3 00000234635MR1 000000175568MR 00000253348MR7 00000200473MR7 00000141481MR2 4358 00000233625MR3 00000205975MR6 00000217627MR9 00000187902MR2 00000168595MR7 00000209053MR8 00000183410MR0 00000223560MR4 00000219055MR1 00000191735MR0 00000214228MR9 00000237507MR9 00000227611MR1 00000231837MR6 00000212547MR4 00000158836MR7 00000194297MR8 00000164190MR1 00000201908MR1 00000210788MR6 00000170958MR3 00000208572MR8 00000170829MR6 00000131858MR3 00000226955MR3 00000258184MR1 535334 00000239200MR9 00000232992MR8 00000170586MR2 00000212390MR9 00000193234MR2 00000243475MR1 00000217029MR8 00000224269MR1 00000248497MR0 00000135763MR1 00000219351MR4 00000211303MR3 00000210787MR8 00000118956MR2 00000178133MR5 00000225897MR8 00000164342MR8 00000205963MR2 00000215305MR4 0000218281MR-4 00000207369MR0 00000169977MR6 00000209445MR6 00000180413MR7 00000165876MR4 00000193678MR0 00000159311MR0 00000248056MR4 00000207218MR9 00000172139MR8 00000228627MR6 00000234877MR9 00000218458MR8 00000186282MR0 00000235446MR2 0000000225091M 00000236229MR1 00000143665MR8 00000257920MR9 00000178132MR7 00000196471MR7 00000213586MR1 00000191031MR4 00000142816MR8 00000230880MR7 00000203348MR8 00000123370MR9 00000227609MR5 00000141752MR6 00000203129MR2 00000232058MR8 00000187560MR8 00000227243MR3 00000222165MR3 00000236562MR5 00000207837MR6 00000151037MR9 00000229413MR0 00000249402MR9 00000196443MR6 00000167452MR2 00000235859MR6 00000214316MR2 00000176684MR9 00000200460MR4 00000238105MR1 00000224284MR0 00000141881MR3 00000214325MR3 00000223864MR0 00000136093MR2 00000200438MR0 00000240315MR2 00000203724MR0 00000199099MR3 00000216731MR0 00000211066MR6 00000161883MR4 00000174118MR0 00000176584MR1 00000229026MR0 00000240235MR2 00000181812MR9 00000022022MR4 00000106093MR8 00000241085MR0 00000233221MR1 00000200710MR2 00000129571MR6 00000128221MR9 00000132599MR2 00000151956MR0 00000130405MR4 00000128413MR2 00000127193MR1 00000145016MR2 00000157431MR8 00000183554MR5 00000184998MR3 00000172712MR2 00000217423MR3 00000254555MR6 00000232582MR7 00000237284MR5 00000169153MR4 00000235872MR9 00000215500MR0 00000161855MR2 00000214766MR8 00000198315MR4 00000199471MR4 00000148631MR5 00000248492MR1 00000172734MR6 00000238520MR1 00000124624MR8 00000170590MR4 00000122915MR2 00000128864MR6 00000222658MR7 00000156079MR6 00000110535MR2 00000194654MR0 00000236226MR7 00000219146MR8 00000205529MR1 00000210644MR1 00000141841MR7 00000110532MR9 00000209598MR2 00000214303MR0 00000179099MR7 00000196469MR1 00000131657MR9 00000159703MR8 00000112033MR6 00000198735MR3 00000236067MR5 00000200120MR4 00000156712MR2 00000166450MR7 00000225444MR9 00000235738MR2 00000129566MR6 00000198349MR3 00000146391MR8 00000199447MR4 00000225075MR1 00000118973MR7 00000249152MR0 00000172152MR1 00000207042MR3 00000126635MR2 00000210073MR3 00000199737MR8 00000236898MR3 00000207367MR4 00000128227MR6 00000184230MR1 00000197822MR0 00000167463MR9 00000151966MR9 00000122782MR6 00000223098MR5 00000240316MR0 00000223034MR0 00000155877MR4 00000159299MR7 00000249630MR5 00000171953MR3 00000201038MR7 00000226218MR6 00000139713MR2 535513 00000127568MR4 00000215309MR6 00000170913MR8 00000206599MR3 00000210071MR7 00000206396MR4 00000210371MR1 00000159324MR3 00000156616MR5 00000165456MR5 00000223257MR7 00000159307MR8 4318 00000206886MR4 00000171948MR3 00000228632MR6 00000258639MR4 00000237278MR7 00000229709MR1 00000230641MR3 00000163675MR2 00000234883MR7 00000237271MR2 00000165420MR1 00000210080MR8 00000219627MR7 00000172140MR6 00000151023MR9 00000143011MR5 00000228628MR4 00000223255MR1 00000142602MR2 00000235274MR8 00000140905MR1 00000255795MR7 00000253349MR5 00000181854MR1 00000218442MR2 00000181840MR0 00000193229MR2 00010149M00417 00000199454MR0 00000203763MR8 00000177111MR2 00000199370MR8 00000124625MR5 00000173554MR7 00000175055MR3 00000174706MR2 00000179496MR5 00000201047MR8 00000193713MR5 00000229944MR4 00000151954MR5 00000225590MR9 00000162576MR3 00000218225MR1 00000253342MR0 00000181834MR3 00000214818MR7 00000207357MR5 00000192321MR8 00000118126MR2 00000192336MR6 00000160054MR3 00000159579MR2 00000196446MR9 00000180079MR6 00000210451MR1 00000195084MR9 00000214358MR4 00000217988MR5 00000220419MR6 00000199732MR9 00000209417MR5 00000236235MR8 00000239883MR2 00000189259MR5 00000225594MR1 00000194639MR1 00000258183MR3 00000245477MR5 00000160063MR4 00000238781MR9 00000133064MR6 00000133313MR7 00000137036MR0 00000136184MR9 00000139126MR7 00000142194MR0 00000143672MR4 00000171439MR3 00000171598MR6 00000177772MR1 00000178139MR2 00000178847MR0 00000180973MR0 00000145019MR6 00000181850MR9 00000178824MR9 00000184969MR4 00000184973MR6 00000185651MR7 00000210117MR8 00000193228MR4 00000143661MR7 00000191047MR0 00000191232MR8 00000198177MR8 00000196448MR5 00000199107MR4 00000189262MR9 00000198174MR5 00000195802MR4 00000141514MR0 00000203709MR1 00000193666MR5 00000207210MR6 00000206867MR4 00000207211MR4 00000199355MR9 00000207234MR6 00000209094MR2 00000210786MR0 00000211425MR4 00000211422MR1 00000213053MR2 00000213652MR1 00000215330MR2 00000215334MR4 00000145860MR3 00000227604MR6 00000222283MR4 00000205138MR1 00000229942MR8 00000217041MR3 00000223873MR1 00000229963MR4 00000225343MR3 00000226219MR4 00000231511MR7 00000229265MR4 00000245315MR7 535774 00000230259MR4 00000218469MR5 00000228630MR0 00000228272MR1 00000237282MR9 00000238668MR8 00000232461MR4 00000236243MR2 00000216500MR9 00000245474MR2 00000246871MR8 00000246560MR7 00000246554MR0 00000242498MR4 00000156717MR1 00000226961MR1 00000191052MR0 00000172171MR1 00000236660MR7 00000236893MR4 00000229939MR4 00000187602MR8 00000143658MR3 00000147045MR9 00000203394MR2 00000164475MR6 00000145681MR3 00000138647MR3 00000152587MR2 00000164341MR0 00000154250MR5 00000157446MR6 00000158341MR8 00000158340MR0 00000159298MR9 00000159312MR8 00000151446MR2 00000159685MR7 00000165881MR4 00000161849MR5 00000162407MR1 00000162579MR7 00000164346MR9 00000166717MR9 00000141474MR7 00000158840MR9 00000151960MR2 00000171560MR6 00000167870MR5 00000168169MR1 00000169547MR7 00000168606MR2 00000165452MR4 00000169827MR3 00000172719MR7 00000171438MR5 000BASHZOSSER1 00000126636MR0 00000217443MR1 00000205534MR1 00000213221MR5 00000216175MR0 00000180418MR6 00000204710MR8 00000232986MR0 00000242820MR9 00000238427MR9 00000187559MR0 00000217631MR1 00000238891MR6 00000184978MR5 00000229044MR3 00000142703MR8 00000162357MR8 00000166736MR9 00000239881MR6 00000162587MR0 00000112929MR5 00000223553MR9 00000135345MR7 00000240401MR0 00000246561MR5 00000169559MR2 00000135457MR0 00000139720MR7 00000140602MR4 00000138705MR9 00000140609MR9 00000139715MR7 00000143664MR1 00000143670MR8 00000148753MR7 00000148761MR0 00000156619MR9 00000157433MR4 00000161122MR7 00000161129MR2 00000162405MR5 00000166720MR3 00000164490MR5 00000169149MR2 00000169542MR8 00000176588MR2 00000179089MR8 00000189428MR6 00000186710MR0 00000189839MR4 00000191730MR1 00000191217MR9 00000205547MR3 00000203395MR9 00000202477MR6 00000211410MR6 00000212454MR3 00000212537MR5 00000212378MR4 00000211908MR9 00000211415MR5 00000212455MR0 00000213085MR4 00000217024MR9 00000217033MR0 00000213051MR6 00000223105MR8 00000221170MR4 00000224834MR2 00000225468MR8 00000225076MR9 00000223867MR3 00000224276MR6 00000225079MR3 00000222169MR5 00000225465MR4 00000225598MR2 00000225466MR2 00000235848MR9 00000227595MR6 00000145017MR0 00000169554MR3 00000178818MR1 00000191051MR2 00000197810MR5 00000206860MR9 00000217035MR5 00000215514MR1 00000217031MR4 00000241761MR6 00000210065MR9 00000137197MR0 00000136183MR1 00000118446MR4 00000177116MR1 00000119782MR1 00000169548MR5 00000166128MR9 00000195798MR4 00000222381MR6 00000148627MR3 00000225595MR8 00000243103MR9 00000128869MR5 00000230625MR6 00000209045MR4 00000219184MR9 00000205962MR4 00000183400MR1 00000114144MR9 00000185657MR4 00000212851MR0 00000197843MR6 00000195081MR5 00000213655MR4 00000177122MR9 00000247025MR0 00000245210MR0 00000242817MR5 00000244538MR5 00000128871MR1 00000170839MR5 00000211423MR9 00000214343MR6 00000214130MR7 00000214793MR2 00000210810MR8 00000212341MR2 00000209057MR9 00000205453MR4 00000209453MR0 00000191744MR2 00000213219MR9 00000214311MR3 00000212456MR8 00000148754MR5 00000174097MR6 00000200108MR9 00000214327MR9 00000180116MR6 00000233013MR2 00000170996MR3 00000236669MR8 00000137202MR8 00000169553MR5 00000195794MR3 00000172158MR8 00000237556MR6 00000164491MR3 00000229706MR7 00000160062MR6 00000187564MR0 00000198559MR7 00000239643MR0 00000190311MR1 00000224271MR7 00000161755MR4 00000206391MR5 00000230266MR9 00000236897MR5 00000241764MR0 00000241284MR9 00000134301MR1 00000171809MR7 00000198321MR2 00000152588MR0 00000217056MR1 00000213044MR1 00000213649MR7 00000215510MR9 00000211409MR8 00000214792MR4 00000208573MR6 00000AS00343OW 0000017504ZMR1 00000156720MR5 00000110540MR2 00000206853MR4 00000162586MR2 00000199635MR4 00000162359MR4 00000172338MR6 00000189567MR1 00000217449MR8 00000207221MR3 00000216169MR3 00000225326MR8 3509 00000246869MR2 00000254642MR2 00000243171MR6 00000224292MR3 00000238334MR7 00000198747MR8 00000216170MR1 00000204699MR3 00000210084MR0 00000222669MR4 00000226518MR9 00000238110MR1 00000164778MR3 00000238901MR3 00000204249MR7 00000205530MR9 00000238235MR6 00000219655MR8 00000219655MR8 00000226759MR9 00000236567MR4 00000236568MR2 00000244006MR3 00000215888MR9 00000195796MR8 00000233206MR2 00000241292MR2 00000167431MR6 00000214773MR4 00000249632MR1 00000148795MR8 00000238424MR6 00000233630MR3 00000156521MR7 00000000MCD984 00000191225MR2 00000235448MR8 00000246552MR4 00000237124MR3 00000166719MR5 00000217058MR7 00000180975MR5 00000196449MR3 00000223933MR3 00000247770MR1 00000225896MRO 00000238338MR8 00000210066MR7 00000239669MR5 00000195807MR3 00000202360MR4 00000193692MR1 00000246549MR0 00000264789MR9 00000243364MR7 00000244907MR2 00000144317MR5 00000220426MR1 00000194274MR7 00000207371MR6 00000192335MR8 00000135340MR8 00000217428MR2 00000223201MR5 00000231833MR5 00000172720MR5 00000214223MR0 00000241755MR8 00000165424MR3 00000236239MR0 00000126175MR9 00000178811MR6 00000124332MR8 00000141560MR3 00000226958MR7 00000176693MR0 00000208564MR5 00000214141MR4 00000175032MR2 00000205129MR0 00000199110MR8 00000167440MR7 00000180959MR9 00000203733MR1 00000233629MR5 00000245948MR5 00000183393MR8 00000178820MR7 00000174094MR3 00000189515MR0 535348 00000216182MR6 00000241155MR1 00000197824MR6 00000221053MR2 00000193672MR3 00000115022MR6 00000243168MR2 00000225587MR5 00000216185MR9 00000168164MR2 00000134293MR0 00000227240MR9 00000243469MR4 00000217429MR0 00000164193MR5 00000237989MR5 00000162583MR9 00000227603MR8 00000225890MR3 00000227932MR1 00000236220MR0 00000248058MR0 00000193227MR6 00000114500MR2 35256 00000142867MR1 00000161139MR1 00000165457MR3 00000158359MR0 00000207825MR1 00000189516MR8 00000197819MR6 00000200119MR6 00000215302MR1 00000215878MR0 00000225592MR5 00000231526MR5 535580 00000242133MR7 00000202468MR5 00000166710MR4 00000154883MR3 00000219049MR4 00000110537MR8 00000236892MR6 00000215304MR7 00000173571MR1 00000111524MR5 00000180970MR6 00000193712MR7 00000125108MR1 00000229964MR2 00000190339MR2 00000195788MR5 00000223639MR6 00000208551MR9 00000242677MR3 00000177765MR5 00000229027MR8 00000194659MR9 00000231503MR4 00000239400MR5 00000208831MR8 00000233022MR3 00000114137MR3 00000181843MR4 00000212961MR7 00000120993MR1 00000112348MR8 00000117234MR5 00000219042MR9 00000210067MR5 00000245788MR5 00000245787MR7 00000187031MR0 00000235743MR2 00000179383MR5 00000189266MR0 00000153274MR6 00000178143MR4 00000221052MR4 00000223630MR5 00000179388MR4 00000245949MR3 00000126177MR5 00000209587MR5 00000211061MR7 0000000C004165 00000179096MR3 00000197815MR4 00000169993MR3 00000203359MR5 00000216179MR2 00000204698MR5 00000196674MR6 00000219799MR4 00000189840MR2 00000208839MR1 00000147583MR9 00000239393MR2 00000204241MR4 00000172721MR3 00000221175MR3 00000168282MR2 00000223259MR3 00000214000MR2 00000171562MR2 00000146387MR6 00000124052MR2 00000214809MR6 00000170007MR9 00000198348MR5 00000161135MR9 00000178825MR6 00000231523MR2 00000232985MR2 00000149023MR4 00000144080MR9 00000159831MR7 00000165895MR4 00000207836MR8 00000132013MR4 00000236659MR9 00000196436MR0 00000208814MR4 00000231333MR6 00000219793MR7 00000218448MR9 00000217634MR5 00000186255MR6 00000159713MR7 00000165901MR0 00000218290MR5 00000171600MR0 00000233026MR4 00000225593MR3 00000164191MR9 00000159300MR3 00000203360MR3 00000179102MR9 00000240403MR6 00000178131MR9 00000179502MR0 00000229410MR6 00000223265MR0 00000125372MR3 00000178827MR2 00000221167MR0 00000129562MR5 00000241758MR2 00000175566MR9 00000160050MR1 00000229030MR2 00000237547MR5 00000249512MR5 00000126641MR0 00000217625MR3 00000230631MR4 00000241092MR6 00000121829MR6 00000239386MR6 00000230260MR2 00000159687MR3 00000158333MR5 00000230868MR2 00000230639MR7 00000233210MR4 00000216502MR5 00000229408MR0 UNK2189500-300 00000236065MR9 00000186266MR3 00000125113MR1 UNK2189500-304 00000187576MR4 00000202182MR2 00000193233MR4 00000199359MR1 00000199457MR3 00000205548MR1 00000205528MR3 00000192326MR7 00000174099MR2 00000209411MR8 00000175039MR7 00000239387MR4 00000104683MR8 00000109617MR1 00000120989MR9 00000124626MR3 00000166947MR2 00000161840MR4 00000191032MR2 00000195797MR6 00000198058MR0 00000193715MR0 00000205560MR6 00000191728MR5 00000198332MR9 00000212376MR8 00000212447MR7 00000214321MR2 00000216163MR6 00000218196MR4 00000202407MR3 00000214314MR7 00000219640MR0 00000222664MR5 00000222663MR7 00000229029MR4 00000229028MR6 00000228636MR7 00000218085MR9 00000224799MR7 00000227598MR0 00000222392MR3 00000207625MR5 00000233600MR6 00000234647MR6 00000232065MR3 00000234248MR3 00000233411MR8 00000232062MR0 00000237119MR3 00000238514MR4 00000241161MR9 00000243167MR4 00000244605MR2 00000246244MR8 00000246245MR5 00000246557MR3 00000247022MR7 00000247018MR5 00000248489MR7 00000249403MR7 00000237747MR1 00000214126MR5 00000141877MR1 00000202053MR5 00000215297MR3 00000216172MR7 00000175782MR2 00000200731MR8 00000154900MR5 00000210791MR0 00000153638MR2 00000162349MR5 00000211077MR3 00000216496MR0 00000107169MR5 00000201067MR6 00000198730MR4 00000208571MR0 00000205531MR7 00000207372MR4 00000217648MR5 00000180420MR2 00000172177MR8 00000154239MR8 00000207236MR1 00000238895MR7 00000208880MR5 00000230257MR8 00000193243MR3 00000189379MR1 00000228640MR9 00000181813MR7 535371 535790 00000159695MR6 00000122778MR4 00000210765MR4 00000240396MR2 00000228080MR8 00000210642MR5 00000107163MR8 00000179504MR6 00000224286MR5 00000185655MR8 00000161863MR6 00000144982MR6 00000217976MR0 00000195789MR3 00000139718MR1 00000204704MR1 00000235266MR4 00000199636MR2 00000144992MR5 00000151642MR6 00000109631MR2 00000233635MR2 00000196669MR6 00000244002MR2 00000238900MR5 00000209080MR1 00000223102MR5 00000215489MR6 00000229710MR9 00000213672MR9 00000216016MR6 00000170953MR4 00000172440MR0 00000178819MR9 00000167437MR3 00000206393MR1 00000177771MR3 00000211079MR9 00000209599MR0 00000242924MR9 00000235855MR4 00000244535MR1 00000170987MR2 00000157367MR4 00000200736MR7 00000217042MR1 00000223269MR2 00000222391MR5 00000168596MR5 00000190303MR8 00000191035MR5 00000224278MR2 00000131054MR9 00000122358MR5 00000168602MR1 00000106090MR4 00000140368MR2 00000199445MR8 00000214308MR9 00000215540MR6 00000241087MR6 00000128223MR5 00000186707MR6 00000124862MR4 00000243174MR0 00000227089MR0 00000229961MR8 00000233027MR2 00000146385MR0 00000233415MR9 00000198317MR0 00000213818MR8 00000173563MR8 00000189845MR1 00000174006MR7 00000122280MR1 00000209442MR3 00000145999MR9 00000174705MR4 00000227936MR2 00000175575MR0 00000194290MR3 00000106092MR0 00000218447MR1 00000176583MR3 00000104500MR4 00000208565MR2 00000227594MR9 00000240390MR5 00000218889MR4 00000167857MR2 00000178828MR0 00000242818MR3 00000239663MR8 00000132093MR6 00000246551MR6 00000246239MR8 00000254379MR1 00000193676MR4 00000214803MR9 00000218890MR2 00000210628MR4 00000142187MR4 00000248052MR3 00000246241MR4 00000203134MR2 00000233216MR1 00000186708MR4 00000157430MR0 00000207361MR7 00000195836MR2 00000171550MR7 00000140604MR0 00000207634MR7 00000222168MR7 00000181820MR2 00000172145MR5 00000199466MR4 00000193673MR1 00000163660MR4 00000140500MR0 00000157432MR6 00000226768MR0 00000137041MR0 00000110458MR7 00000176667MR4 00000232988MR6 00000234708MR6 00000108300MR5 00000107832MR8 00000108297MR3 00000104663MR0 00000225083MR5 00000108051MR4 00000104690MR3 00000108682MR6 00000109306MR1 00000109312MR9 00000109623MR9 00000233643MR6 00000214823MR7 00000109620MR5 00000110461MR1 00000110459MR5 00000115024MR2 00000147990MR6 00000115021MR8 00000114497MR1 00000120324MR9 00000109627MR0 00000118450MR6 00000118971MR1 00000120979MR0 00000120986MR5 00000122904MR6 00000125784MR9 535709 00000119783MR9 00000124861MR6 00000125109MR9 00000124331MR0 00000125110MR7 00000120330MR6 00000125368MR1 00000126631MR1 00000118433MR2 00000126168MR4 00000110462MR9 00000128548MR5 00000128863MR8 00000129568MR2 00000131053MR1 00000120524MR4 00000131652MR0 00000131857MR5 00000132097MR7 00000132601MR6 00000130807MR1 00000133627MR0 00000134711MR1 00000137193MR9 00000134510MR7 00000137198MR8 00000129563MR3 00000137295MR2 00000134294MR8 00000128222MR7 00000137195MR4 00000137298MR6 00000136999MR0 00000137226MR7 00000137201MR0 00000138895MR8 00000138904MR8 00000128220MR1 00000139125MR9 00000147624MR1 00000132602MR4 00000152402MR4 00000151961MR0 00000157363MR3 00000156721MR3 00000151964MR4 00000152400MR8 00000146201MR9 00000156649MR6 00000151796MR0 00000159705MR3 00000156727MR0 00000140376MR5 00000158804MR5 00000158345MR9 00000154878MR3 00000159574MR3 00000158806MR0 00000156714MR8 00000157366MR6 00000159573MR5 00000160046MR9 00000161124MR3 00000161842MR0 00000162584MR7 00000162592MR0 00000162358MR6 00000164482MR2 00000164204MR0 00000166951MR4 00000167435MR7 00000166965MR4 00000165864MR0 00000167865MR5 00000169143MR5 00000158807MR8 00000156519MR1 00000170602MR7 00000170954MR2 00000171563MR0 00000171956MR6 00000161865MR1 00000159701MR2 00000174347MR5 00000169964MR4 00000174698MR1 00000161846MR1 00000176677MR3 00000176670MR8 00000178137MR6 00000136187MR2 00000176678MR1 00000178140MR0 00000171588MR7 00000207822MR8 00000164187MR7 00000228625MR0 00000176453MR9 00000196440MR2 00000177114MR6 00000172831MR0 00000183399MR5 00000187908MR9 00000191046MR2 00000190312MR9 00000178146MR7 00000176701MR1 00000177784MR6 00000179382MR7 00000179824MR8 00000179121MR9 00000180967MR2 00000180969MR8 00000181814MR5 00000181844MR2 00000183404MR3 00000180976MR3 00000181929MR1 00000166455MR6 00000185652MR5 00000195803MR2 00000198723MR9 00000186750MR6 00000195810MR7 00000186753MR0 00000197809MR7 00000186749MR8 00000198175MR2 00000184974MR4 00000192309MR3 00000187604MR4 00000200622MR9 00000193669MR9 00000187608MR5 00000174128MR9 00000189540MR8 00000190322MR8 00000191041MR3 00000191915MR8 00000193371MR2 00000192328MR3 00000195211MR8 00000196447MR7 00000197820MR4 00000195799MR2 00000195804MR0 00000198322MR0 00000199094MR4 00000198320MR4 00000198325MR3 00000203404MR9 00000203729MR9 00000223266MR8 00000223100MR9 00000196673MR8 00000202472MR7 00000203728MR1 00000202479MR2 00000203357MR9 00000202049MR3 00000202445MR3 00000198733MR8 00000195783MR6 00000195800MR8 00000203736MR4 00000203720MR8 00000202047MR7 00000205126MR6 00000202244MR0 00000204229MR9 00000203397MR5 00000202463MR6 00000205133MR2 00000203355MR3 00000199450MR8 00000204240MR6 00000204707MR4 00000202457MR8 00000205965MR7 00000205960MR8 00000203714MR1 00000203135MR9 00000203358MR7 00000204232MR3 00000207847MR5 00000206102MR6 00000207355MR9 00000204243MR0 00000191715MR2 00000205958MR2 00000208562MR9 00000189267MR8 00000207356MR7 00000203396MR7 00000209413MR4 00000208788MR0 00000207844MR2 00000192313MR5 00000210817MR3 00000216160MR2 00000215320MR3 00000216176MR8 00000208147MR9 00000208820MR1 00000207864MR0 00000210328MR1 00000214780MR9 00000207845MR9 00000207633MR9 00000211907MR1 00000213050MR8 00000214019MR2 00000212953MR4 00000213654MR7 00000215880MR6 00000216489MR5 00000214309MR7 00000216721MR1 00000217618MR8 00000206608MR2 00000220985MR6 00000221036MR7 00000225080MR1 00000215879MR8 00000207363MR3 00000215900MR2 00000216001MR8 00000217023MR1 00000217643MR6 00000221890MR7 00000220219MR0 00000219364MR7 00000208149MR5 00000216730MR2 00000216505MR8 00000218160MR0 00000219634MR3 00000219652MR5 00000221042MR5 00000221182MR9 00000221035MR9 00000219792MR9 00000218460MR4 00000222174MR5 00000218451MR3 00000222409MR5 00000222396MR4 00000216072MR9 00000216011MR7 00000222795MR7 00000221901MR2 00000208824MR3 00000221896MR4 00000226224MR4 00000225336MR7 00000235275MR5 00000233414MR2 00000232066MR1 00000249400MR3 00000224887MR0 00000222388MR1 00000222387MR3 00000222178MR6 00000221038MR3 00000219138MR5 00000222393MR1 00000216168MR5 00000224471MR3 00000219185MR6 00000225325MR0 00000225442MR3 00000229414MR8 00000221428MR6 00000225340MR9 00000224790MR6 00000222395MR6 00000227748MR1 00000224800MR3 00000229733MR1 00000229941MR0 00000231513MR3 00000229962MR6 00000235612MR9 00000230636MR3 00000232074MR5 00000229960MR0 00000219784MR6 00000218296MR2 00000221888MR1 00000229260MR5 00000227593MR1 00000233211MR2 00000220987MR2 00000233634MR5 00000234706MR0 00000235871MR1 00000234879MR5 00000237283MR7 00000235870MR3 00000236891MR8 00000239973MR1 00000231842MR6 00000236889MR2 00000239199MR3 00000248494MR7 00000228515MR3 00000236896MR7 00000217638MR6 00000232991MR0 00000213658MR8 00000221049MR0 00000247017MR7 00000249513MR3 00000243671MR5 00000179813MR1 00000200619MR5 00000178135MR0 00000238106MR9 00000233637MR8 00000232459MR8 00000217044MR7 00000233217MR9 00000232060MR4 00000234650MR0 00000232465MR5 00000237125MR0 00000235611MR1 00000236668MR0 00000233019MR9 00000235606MR1 00000235095MR7 00000236887MR6 00000237559MR0 00000235285MR4 00000223035MR7 00000237549MR1 00000235740MR8 00000234878MR7 00000235096MR5 00000237760MR4 00000237281MR1 00000235869MR5 00000246248MR9 00000238001MR2 00000237765MR3 00000242495MR0 00000233638MR6 00000242920MR7 00000237993MR1 00000238236MR4 00000228623MR5 00000237768MR7 00000238903MR9 00000246243MR0 00000239666MR1 00000236070MR9 00000236886MR8 00000239399MR9 00000239391MR6 00000241540MR4 00000241541MR2 00000243263MR1 00000242493MR5 00000242136MR0 00000241287MR2 00000242465MR3 00000242927MR2 00000243477MR7 00000244608MR6 00000243474MR4 00000245953MR5 00000242658MR3 00000244906MR4 00000248493MR9 00000244612MR8 00000246240MR6 00000242494MR3 00000213089MR6 00000232073MR7 00000104000MR5 00000180084MR6 00000248049MR9 00000230635MR5 00000245218MR3 00000176454MR7 00000130741MR2 00000245951MR9 00000195801MR6 00000196450MR1 00000162356MR0 00000151444MR7 00000236063MR4 00000189265MR2 00000211057MR5 00000191044MR7 00000221422MR9 00000133063MR8 00000113892MR4 00000178136MR8 00000160061MR8 00000189843MR6 00000234634MR4 00000170565MR6 00000234244MR2 00000221412MR0 00000233021MR5 00000207630MR5 00000233409MR2 00000184980MR1 00000218154MR3 00000200444MR8 00000237121MR9 00000229040MR1 UNK2189500-305 00000235284MR7 00000211906MR3 00000235605MR3 00000232983MR7 00000236219MR2 00000122279MR3 00000199098MR5 00000157693MR3 00000202180MR6 00000221411MR2 00000234711MR0 00000190306MR1 00000218313MR5 00000245217MR5 00000221429MR4 00000239877MR4 00000228624MR3 00000202481MR8 00000153580MR6 00000240394MR4 00000128109MR6 00000100073MR9 00000223346MR8 00000227242MR5 00000241539MR6 00000220417MR0 00000217616MR2 00000214132MR3 00000213090MR4 00000213647MR1 00000240398MR8 00000213218MR1 00000179101MR1 00000202461MR0 00000197812MR1 00000233023MR1 00000236068MR3 00000207362MR5 00000222179MR4 00000244003MR0 00000214335MR2 00000239397MR3 00000240274MR1 00000233214MR6 00000153565MR7 00000210074MR1 00000234637MR7 00000174694MR0 00000215539MR8 00000201912MR3 00000185656MR6 00000192322MR6 00000179091MR4 00000184968MR6 00000187581MR4 00000210818MR1 00000210110MR3 00000193682MR2 00000174902MR7 00000188408MR9 00000219641MR8 00000184979MR3 00000170588MR8 00000232586MR8 00000219365MR4 00000231342MR7 00000220625MR8 00000234636MR9 00000134297MR1 00000232064MR6 00000212556MR5 00000171821MR2 00000245784MR4 00000191063MR7 00000225588MR3 00000154889MR0 00000253343MR8 00000231336MR9 00000200107MR1 00000243170MR8 00000242921MR5 00000239885MR7 00000209463MR9 00000175417MR5 00000165877MR2 00000192332MR5 00000147042MR6 00000170981MR5 00000193705MR1 00000226244MR2 00000180076MR2 00000238666MR2 00000239205MR8 00000148039MR1 00000170956MR7 00000210124MR4 00000207821MR0 00000232067MR9 00000229965MR9 00000195790MR1 00000189543MR2 00000191030MR6 00000222801MR3 00000181833MR5 00000221891MR5 00000166714MR6 00000159304MR5 00000207626MR3 00000205970MR7 00000238894MR0 00000169555MR0 00000245785MR1 00000223559MR6 00000215998MR6 00000162582MR1 00000128549MR3 00000207231MR2 00000225339MR1 00000119699MR7 00000245472MR6 00000199104MR1 00000207853MR3 00000130410MR4 00000151958MR6 00000218291MR3 00000115086MR1 00000135456MR2 00000244005MR5 00000180964MR9 00000135089MR1 00000181852MR5 0000167881MR02 00000211080MR7 00000212959MR1 00000180961MR5 00000222666MR0 00000220624MR1 00000225330MR0 00000225093MR4 00000168286MR3 00000145859MR5 00000151798MR6 00000149032MR5 UNK21895000000 00000232460MR6 00000205559MR8 00000137607MR8 00000215542MR2 00000219133MR6 00000204708MR2 00000233212MR0 00000218161MR8 00000148030MR0 00000137605MR2 00000172717MR1 00000238784MR3 00000131648MR8 00000178144MR2 00000131860MR9 00000179116MR9 00000133306MR1 00000144925MR5 00000169141MR9 00000202415MR6 00000232984MR5 00000127584MR1 00000215766MR7 00000222384MR0 00000108055MR5 00000231852MR5 00000142184MR1 00000126172MR6 00000183402MR7 00000117232MR9 00000191056MR1 00000241632MR9 00000117427MR5 00000239405MR4 00000228516MR1 00000240391MR3 00000208559MR5 00000244910MR6 00000140601MR6 00000207617MR2 00000125105MR7 00000207819MR4 00000245473MR4 00000127583MR3 00000239396MR5 00000240318MR6 00000224297MR2 00000130804MR8 00000242676MR5 00000215331MR0 00000237756MR2 00000239664MR6 00000241157MR7 00000241757MR4 00000175416MR7 535714 00000110529MR5 00000219153MR4 00000118961MR2 00000172167MR9 00000174111MR5 00000191721MR0 00000225082MR7 00000242821MR7 00000213643MR0 00000112551MR7 00000218280MR6 00000226214MR5 00000231335MR1 00000115445MR9 00000238890MR8 00000203726MR5 00000237761MR2 00000225586MR7 00000222394MR9 00000142603MR0 00000259463MR8 00000260077MR3 00000231759MR2 00000210797MR7 00000193714MR3 00000175044MR7 00000234881MR1 00000214026MR7 00000180088MR7 00000234866MR2 00000168628MR6 00000202183MR0 00000200718MR5 00000201910MR7 00000234869MR6 00000241756MR6 00000254641MR4 00000198728MR8 00000200721MR9 00000202416MR4 00000236564MR1 00000235603MR8 00000117425MR9 00000210321MR6 00000215898MR8 00000231218MR9 00000243177MR3 00000242824MR1 00000187901MR4 00000235619MR4 00000240275MR8 00000260054MR2 00000259184MR0 00000170908MR8 00000231222MR1 00000250674MR9 00000107168MR7 00000220632MR4 00000242678MR1 00000246776MR9 00000238332MR1 00000246548MR2 00000208146MR1 00000191229MR4 00000209452MR2 00000169152MR6 00000234632MR8 00000236061MR8 00000239878MR2 00000254558MR0 00000237120MR1 00000262056MR5 00000246255MR4 00000213048MR2 00000212964MR1 00000214776MR7 00000258640MR2 00000202406MR5 00000212444MR4 00000210814MR0 00000241631MR1 00000235857MR0 00000243472MR8 00000245944MR4 00000249017MR5 00000264157MR9 00000133741MR9 00000207841MR8 00000223631MR3 00000235861MR2 00000238516MR9 00000241158MR5 00000244613MR6 00000133739MR3 00000174695MR7 00000230879MR9 00000237557MR4 00000250509MR7 00000260078MR1 00000243481MR9 00000131855MR9 00000202045MR1 00000225331MR8 00000235447MR0 00000244912MR2 00000235099MR9 00000245794MR3 00000198731MR2 00000199744MR4 00000217628MR7 00000223637MR0 00000219362MR1 00000264784MR0 00000215890MR5 00000226509MR8 00000233624MR6 00000180795MR7 00000213687MR7 00000213657MR0 00000243674MR9 00000255251MR1 00000180937MR5 00000247021MR9 00000253190MR3 00000214778MR3 00000247774MR3 00000248335MR2 00000250827MR3 4134 00000132597MR6 00000231757MR6 00000214235MR4 00000193679MR8 00000216726MR0 00000140502MR6 00000239971MR5 00000153288MR6 00000191720MR2 00000202471MR9 00000198753MR6 00000237997MR2 00000235280MR5 00000212542MR5 00000208782MR3 00000189544MR0 00000175439MR9 00000120984MR0 00000232989MR4 00000239202MR5 00000229038MR5 00000213083MR9 00000222901MR1 00000221884MR0 00000131853MR4 00000218289MR7 00000156710MR6 00000205536MR6 00000242675MR7 00000125369MR9 00000205959MR0 00000214029MR1 00000211085MR6 00000118932MR3 00000106096MR1 00000107167MR9 00000191930MR7 00000166138MR8 00000190316MR0 00000224293MR1 00000246543MR3 00000223868MR1 00000221189MR4 00000226226MR9 00000156734MR6 00000158809MR4 00000165871MR5 00000154118MR4 00000223494MR6 00000211918MR8 00000148625MR7 00000175568MR5 00000116818MR6 00000187909MR7 000000191026MR 00000210455MR2 00000126176MR7 00000221033MR4 00000202060MR0 00000179093MR0 00000194633MR4 00000143622MR9 00000194293MR7 00000191224MR5 00000169552MR7 00000164467MR3 00000206017MR6 00000177141MR9 00000191924MR0 00000201060MR1 00000173553MR9 00000170952MR6 00000206602MR5 00000159684MR0 00000159322MR7 00000173763MR4 00000169512MR1 00000200717MR7 00000180422MR8 00000255801MR3 00000176673MR2 00000193691MR3 00000110543MR6 00000195787MR7 00000244911MR4 00000189511MR9 00000174126MR3 00000181836MR8 00000190314MR5 00000179818MR0 00000230629MR8 00000200123MR8 00000210444MR6 00000199739MR4 00000214337MR8 00000164470MR7 00000174906MR8 00000191932MR3 00000175787MR1 00000215875MR6 00000215506MR7 00000206392MR3 00000240400MR2 00000181937MR4 00000227239MR1 00000222400MR4 00000119504MR9 00000135096MR6 00000189269MR4 00000217642MR8 00000135764MR9 00000217985MR1 00000135761MR5 00000137609MR4 00000215509MR1 00000154662MR1 00000217610MR5 00000168172MR5 00000240313MR7 00000141478MR8 00000215886MR3 00000241763MR2 00000193725MR9 00000156121MR6 00000206854MR2 00000165786MR5 00000140839MR2 00000172840MR1 00000200626MR0 00000128415MR7 00000210063MR4 00000166441MR6 00000169960MR2 00000175789MR7 00000169979MR2 0000009737-003 00000223256MR9 00000170584MR7 00000119778MR9 00000237277MR9 00000169988MR3 00000170583MR9 00000217650MR1 00000115020MR0 00000199640MR4 00000179505MR3 00000114138MR1 00000168171MR7 00000119779MR7 00000141844MR1 00000221885MR7 00000213057MR3 00000213999MR6 00000175026MR4 00000132799MR8 00000111519MR5 00000113894MR0 00000166451MR5 00000176672MR4 00000225878MR8 00000239886MR5 00000174101MR6 00000189545MR7 00000209092MR6 00000181924MR2 00000230882MR3 00000121812MR2 00000113564MR9 00000199103MR3 00000159704MR6 00000147993MR0 00000194630MR0 00000218279MR8 00000239389MR0 00000147586MR2 00000199451MR6 00000204227MR3 00000141662MR7 00000171571MR3 00000216165MR1 00000218875MR3 00000180900MR3 00000169795MR2 00000193665MR7 00000169797MR8 00000243678MR0 |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety.
If the gradient cables are inadvertently swapped while servicing the gradients, and the Service Engineer
does not perform the geometry check as required in the service procedure, images m
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
On November 9, 2010, the firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated November 5, 2010 to all Consignees/Customers. The letter described the product, problem and action to be taken by the firm. GE instructed the customers to contact their local GEHC support Engineer, if they have any questions about the image orientation, or the procedures for geometry checks on thier system. Additionally, a GE Healthcare representative will install a device, which will reduce the potiential for the gradient cables to be swapped.
Contact your local GEHC Support Engineer if you have any questions concerning this notification or phone: 262-548-2731. |
Quantity in Commerce |
3105 |
Distribution |
Worldwide distribution: USA including DC and PR; and countries including: VIET NAM, VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKEY, TUNISIA, FORMER YUGOSLAV REPUBLIC OF MACEDONIA, THAILAND, TAIWAN, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, ROMANIA, REUNION, REPUBLIC OF KOREA, PORTUGAL POLAND, PHILIPPINES, PERU,PARAGUAY, PANAMA, PAKISTAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MEXICO, MAURITIUS, MARTINIQUE,MALTA, MALAYSIA, LUXEMBOURG,LITHUANIA, LIBYAN ARAB JAMAHIRIYA, LEBANON, LATVIA, KUWAIT, KENYA, KAZAKHSTAN, JORDAN, JAPAN, JAMAICA, ITALY, ISRAEL, ISLAMIC REPUBLIC OF IRAN, IRELAND, INDONESIA, INDIA, ICELAND, HUNGARY, HONG KONG, HONDURAS, GUYANA, GUATEMALA, GUADELOUPE, GREECE, GERMANY, GEORGIA, FRENCH POLYNESIA, FRANCE, FINLAND, EL SALVADOR, EGYPT, ECUADOR, DOMINICAN REPUBLIC, DENMARK, CECH REPUBLIC, CYPRUS, CROATIA, COLOMBIA, CHINA, CHILE, CANDA, BULGARIA, BRAZIL, BOSNIA AND HERZEGOVINA, BOLIVIA, BELGIUM, BELARUS, AZERBAIJAN, AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LNH and Original Applicant = GE MEDICAL SYSTEMS
|
|
|
|