• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 036 Date of Entry 07/09/2014 
FR Recognition Number 14-434
Standard
ASTM  F2228-13 (Reapproved 2021)
Standard Test Method for Non-Destructive Detection of Leaks in Medical Packaging Which Incorporates Porous Barrier Material by C02 Tracer Gas Method
Scope/Abstract
1.1 This non-destructive test method detects leaks in non-porous rigid thermoformed trays, as well as the seal between the porous lid and the tray. The test method detects channel leaks in packages as small as 100 μm (0.004 in.) diameter in the seal as well as 50 μm (0.002 in.) diameter pinholes, or equivalently d cracks in the tray, subject to trace gas concentration in the package, package design and manufacturing tolerances.
NOTE 1-This test method does not claim to challenge the porous (breathable) lidding material. Any defects that may exist in the porous portion of the package will not be detected by this test method.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
This standard is relevant for all medical devices where sterilization and packaging are components of the development and manufacture of the product.
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 Matthew Beckwith
  CDRH/OPEQ/OHTII/DHTIIC/
  240-402-3393
  Matthew.Beckwith@fda.hhs.gov
 Steven Turtil
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIB/
  301-796-6305
  steven.turtil@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Sterility
*These are provided as examples and others may be applicable.
-
-