Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 050 Date of Entry 09/17/2018 
FR Recognition Number 8-473
Standard
ASTM  F2885-17
Standard Specification for Metal Injection Molded Titanium-6Aluminum-4Vanadium Components for Surgical Implant Applications
Scope/Abstract
1.1 This specification covers chemical, mechanical, and metallurgical requirements for two types of metal injection molded (MIM) titanium-6aluminum-4vanadium components to be used in the manufacture of surgical implants.
1.2 The Type 1 MIM components covered by this specification may have been densified beyond their as-sintered density by post sinter processing.
Extent of Recognition
Partial recognition. The following part(s) of the standard is (are) not recognized:
Any reference to type 2 sintered material.
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.

This standard is recognized in part because it conflicts with another recognized standard. See ASTM F1472 below for type 2 sintered material.
Relevant FDA Guidance and/or Supportive Publications*
ASTM F1472-14 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium Alloy for Surgical Implant Applications (UNS R56400)

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 David M. Saylor
  FDA/OC/CDRH/OSEL/DBCMS/
  301-796-2626
  david.saylor@fda.hhs.gov
 Matthew Di Prima
  FDA/OC/CDRH/OSEL/DAM/
  301-796-2507
  matthew.diprima@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Materials