New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 055 Date of Entry 12/21/2020 
FR Recognition Number 8-538
Standard
ASTM  F2759-19
Standard Guide for Assessment of the Ultra High Molecular Weight Polyethylene (UHMWPE) Used in Orthopedic and Spinal Devices
Scope/Abstract
1.1 This guide covers general guidelines for the physical, chemical, biocompatibility, mechanical, and preclinical assessments of ultra-high molecular weight polyethylene (UHMWPE) in implantable orthopedic and spinal devices intended to replace a musculoskeletal joint. The UHMWPE components may include knee, hip, shoulder, elbow, ankle, total disc replacement, toe, finger, and wrist joint implant devices. This guide does not cover UHMWPE in fiber or tape forms.

1.2 This guide includes a description and rationale of assessments for the various UHMWPE types and processing conditions. Assessment testing based on physical, chemical, biocompatibility, mechanical, and preclinical analyses are briefly described and referenced. The user should refer to specific test methods for additional details.

1.3 This guide does not attempt to define all of the assessment methods associated with UHMWPE components in orthopedic and spinal devices.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
No Product Code information listed for 888.3xxx20, Polymeric components
Relevant FDA Guidance and/or Supportive Publications*
1. Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis, Issued April 2002.

2. Class II Special Controls Guidance: Shoulder Joint Metal/Polymer/Metal Nonconstrained or Semi-Constrained Porous-Coated Uncemented Prosthesis, Issued October 2000.

3. Class II Special Controls Guidance Documnt: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses, Issued January 2003.

4. Guidance Document for Testing Non-Articulating, 'Mechanically Locked', Modular Implant Components, Issued May 1995.

5. Guidance for Industry and FDA Staff Spinal System 510(k)s, Issued May 2004.

6. Guidance for Industry and/or FDA Reviewers/Staff - Guidance Document for the Preparation of IDEs for Spinal Systems, Issued January 2000.

7. Guidance for Industry and FDA Staff: Preparation and Review of Investigational Device Exemption Applications (IDEs) for Total Artificial Discs, Issued April 2008.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 Peter Allen
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA/
  301-796-6402
  peter.allen@fda.hhs.gov
 John Goode
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA/
  301-796-6407
  john.goode@fda.hhs.gov
 Jinrong (Jinny) Liu
  FDA/OC/CDRH/OPEQ/OHTII/DHTIIC/
  240-402-3160
  jinrong.liu@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
Materials
*These are provided as examples and others may be applicable.
-
-