Device dilator, urethral
Product CodeKOE
Regulation Number 876.5520
Device Class 2


Premarket Reviews
ManufacturerDecision
UROTRONIC INC
  SUBSTANTIALLY EQUIVALENT 1

MDR Year MDR Reports MDR Events
2019 20 20
2020 23 23
2021 19 19
2022 15 15
2023 20 20
2024 9 9

Device Problems MDRs with this Device Problem Events in those MDRs
Leak/Splash 12 12
Burst Container or Vessel 12 12
Material Puncture/Hole 12 12
Inflation Problem 9 9
Corroded 8 8
Material Deformation 7 7
Device Contaminated During Manufacture or Shipping 6 6
Device Contamination with Chemical or Other Material 6 6
Material Fragmentation 5 5
Break 5 5
Fluid/Blood Leak 5 5
Use of Device Problem 5 5
Contamination /Decontamination Problem 5 5
Tear, Rip or Hole in Device Packaging 3 3
Packaging Problem 3 3
Pressure Problem 3 3
Device Handling Problem 2 2
Unsealed Device Packaging 2 2
Output Problem 2 2
Unable to Obtain Readings 2 2
Material Twisted/Bent 2 2
Adverse Event Without Identified Device or Use Problem 2 2
Component Misassembled 1 1
Misassembled 1 1
Component Missing 1 1
Melted 1 1
Material Split, Cut or Torn 1 1
Therapeutic or Diagnostic Output Failure 1 1
Fracture 1 1
Gas/Air Leak 1 1
Difficult to Remove 1 1
Material Integrity Problem 1 1
Contamination 1 1
Moisture Damage 1 1
Material Rupture 1 1
Crack 1 1
Gel Leak 1 1
Noise, Audible 1 1
Material Separation 1 1
Decrease in Pressure 1 1
Physical Resistance/Sticking 1 1

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Clinical Signs, Symptoms or Conditions 59 59
No Consequences Or Impact To Patient 26 26
No Known Impact Or Consequence To Patient 20 20
No Patient Involvement 11 11
Foreign Body In Patient 2 2
Injury 1 1
Unspecified Mental, Emotional or Behavioural Problem 1 1
Genital Bleeding 1 1
Drug Resistant Bacterial Infection 1 1
Fever 1 1
Unspecified Infection 1 1
Muscle Weakness 1 1
Nausea 1 1
Pain 1 1
Perforation 1 1
Urinary Tract Infection 1 1
Chills 1 1

Recalls
Manufacturer Recall Class Date Posted
1 Cook Inc. II Feb-04-2020
-
-