• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

TPLC - Total Product Life Cycle

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search show TPLC since Back to Search Results
Device catheter, assisted reproduction
Product CodeMQF
Regulation Number 884.6110
Device Class 2


Premarket Reviews
ManufacturerDecision
ALLWIN MEDICAL DEVICES, INC.
  SUBSTANTIALLY EQUIVALENT 2
CATHETER RESEARCH, INC.
  SUBSTANTIALLY EQUIVALENT 1
COOK INCORPORATED
  SUBSTANTIALLY EQUIVALENT 7
CROSSBAY MEDICAL INC.
  SUBSTANTIALLY EQUIVALENT 1
FEMASYS INC.
  SUBSTANTIALLY EQUIVALENT 1
FERTILIGENT, LTD
  SUBSTANTIALLY EQUIVALENT 2
KITAZATO CORPORATION
  SUBSTANTIALLY EQUIVALENT 3
KITAZATO MEDICAL CO., LTD.
  SUBSTANTIALLY EQUIVALENT 2
MEDGYN PRODUCTS, INC.
  SUBSTANTIALLY EQUIVALENT 1
ORIGIO A/S
  SUBSTANTIALLY EQUIVALENT 1
SHANGHAI HORIZON MEDICAL TECHNOLOGY CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1
WILLIAM A. COOK AUSTRALIA PTY LTD
  SUBSTANTIALLY EQUIVALENT 1

MDR Year MDR Reports MDR Events
2014 3 3
2015 2 2
2016 3 3
2017 10 10
2018 27 27
2019 16 16
2020 19 19
2021 13 13
2022 11 11
2023 12 12
2024 2 2

Device Problems MDRs with this Device Problem Events in those MDRs
Contamination /Decontamination Problem 29 29
Device Contamination with Chemical or Other Material 22 22
Device Contaminated During Manufacture or Shipping 22 22
Material Separation 11 11
Contamination 9 9
Unsealed Device Packaging 5 5
Device Operates Differently Than Expected 3 3
Difficult to Advance 3 3
Break 3 3
Nonstandard Device 3 3
Complete Blockage 2 2
Detachment of Device or Device Component 2 2
Material Too Rigid or Stiff 2 2
Packaging Problem 2 2
Appropriate Term/Code Not Available 1 1
Infusion or Flow Problem 1 1
Other (for use when an appropriate device code cannot be identified) 1 1
Defective Component 1 1
Bent 1 1
Deflation Problem 1 1
Entrapment of Device 1 1
Fluid/Blood Leak 1 1
Material Fragmentation 1 1
Mechanical Problem 1 1

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Consequences Or Impact To Patient 36 36
No Clinical Signs, Symptoms or Conditions 33 33
No Patient Involvement 25 25
No Known Impact Or Consequence To Patient 9 9
Foreign Body In Patient 6 6
Device Embedded In Tissue or Plaque 3 3
No Information 2 2
Insufficient Information 2 2
Blood Loss 2 2
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available 1 1
Unspecified Tissue Injury 1 1
Nonresorbable materials, unretrieved in body 1 1
Foreign body, removal of 1 1
Hemorrhage/Bleeding 1 1
Perforation 1 1

Recalls
Manufacturer Recall Class Date Posted
1 Cook Inc. II Nov-16-2019
2 Cook Inc. II Mar-26-2018
3 Smiths Medical ASD, Inc. II May-03-2013
4 Tenderneeds Fertility LLC II Aug-02-2021
5 Tenderneeds Fertility LLC II Apr-06-2021
-
-