Quick Links: Skip to main page content
Skip to Search
Skip to Topics Menu
Skip to Common Links
U.S. Food & Drug Administration
Follow FDA
En Español
Search FDA
Home
Food
Drugs
Medical Devices
Radiation-Emitting Products
Vaccines, Blood & Biologics
Animal & Veterinary
Cosmetics
Tobacco Products
TPLC - Total Product Life Cycle
FDA Home
medical devices
databases
-
510(k)
|
DeNovo
|
Registration & Listing
|
Adverse Events
|
Recalls
|
PMA
|
HDE
|
Classification
|
Standards
CFR Title 21
|
Radiation-Emitting Products
|
X-Ray Assembler
|
Medsun Reports
|
CLIA
|
TPLC
New Search
show TPLC since
2009
2010
2011
2012
2013
2014
2015
2016
2017
2018
2019
2020
2021
2022
2023
2024
Back to Search Results
Device
handpiece, contra- and right-angle attachment, dental
Product Code
EGS
Regulation Number
872.4200
Device Class
1
Premarket Reviews
Manufacturer
Decision
AJK ENGINEERING, INC.
SUBSTANTIALLY EQUIVALENT
1
KAVO DENTAL GMBH
SUBSTANTIALLY EQUIVALENT
1
LARES RESEARCH
SUBSTANTIALLY EQUIVALENT
1
MICRO-NX CO., LTD.
SUBSTANTIALLY EQUIVALENT
1
NAKANISHI INC.
SUBSTANTIALLY EQUIVALENT
4
SAESHIN PRECISION CO., LTD.
SUBSTANTIALLY EQUIVALENT
2
MDR Year
MDR Reports
MDR Events
2019
219
219
2020
191
191
2021
294
294
2022
279
279
2023
115
118
2024
13
13
Device Problems
MDRs with this Device Problem
Events in those MDRs
Detachment of Device or Device Component
554
555
Overheating of Device
422
422
Mechanical Problem
113
113
Temperature Problem
82
82
Output Problem
45
45
Dent in Material
44
44
Therapeutic or Diagnostic Output Failure
43
43
Defective Device
35
35
Physical Resistance/Sticking
32
32
Device Handling Problem
18
18
Loose or Intermittent Connection
15
15
Break
14
14
Failure to Align
11
11
Computer Operating System Problem
10
10
Mechanics Altered
9
9
Incorrect Measurement
8
8
Use of Device Problem
8
8
Device Slipped
7
7
Insufficient Information
7
9
Noise, Audible
6
6
Appropriate Term/Code Not Available
5
5
Data Problem
5
5
Retraction Problem
5
5
Material Separation
5
5
Improper or Incorrect Procedure or Method
5
5
Fitting Problem
3
3
Calibration Problem
3
3
Device Fell
3
3
Mechanical Jam
3
3
Unintended Movement
2
2
Unexpected Shutdown
2
2
Defective Component
2
2
Loss of Power
2
2
Energy Output Problem
2
2
Entrapment of Device
2
2
Unintended Ejection
1
1
Computer Software Problem
1
1
Failure to Cycle
1
1
Misconnection
1
1
Vibration
1
1
Component Missing
1
1
Device Displays Incorrect Message
1
1
Patient-Device Incompatibility
1
1
Connection Problem
1
1
Device Damaged by Another Device
1
1
Inappropriate or Unexpected Reset
1
1
Material Deformation
1
1
Intermittent Energy Output
1
1
Excessive Heating
1
1
Unintended Electrical Shock
1
1
Device Difficult to Maintain
1
1
Adverse Event Without Identified Device or Use Problem
1
1
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
No Clinical Signs, Symptoms or Conditions
511
514
Burn(s)
249
249
No Consequences Or Impact To Patient
169
169
Insufficient Information
82
82
Partial thickness (Second Degree) Burn
80
80
Burn, Thermal
79
79
Patient Problem/Medical Problem
66
66
Injury
62
62
Superficial (First Degree) Burn
43
43
No Known Impact Or Consequence To Patient
39
39
Foreign Body In Patient
15
15
No Information
12
12
Pain
9
9
Burning Sensation
4
4
Full thickness (Third Degree) Burn
3
3
Laceration(s)
3
3
Tooth Fracture
2
2
Needle Stick/Puncture
2
2
Skin Tears
2
2
Blister
2
2
Skin Burning Sensation
2
2
Unspecified Tissue Injury
1
1
Swelling/ Edema
1
1
No Code Available
1
1
Electric Shock
1
1
Numbness
1
1
Perforation
1
1
Scar Tissue
1
1
Ulceration
1
1
Recalls
Manufacturer
Recall Class
Date Posted
1
Young Dental Manufacturing I, LLC
II
Nov-09-2023
-
-