MEDTRONIC MED REL MEDTRONIC PUERTO RICO SOLETRA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number 7426 |
Device Problems
Inappropriate/Inadequate Shock/Stimulation (1574); Impedance Problem (2950)
|
Patient Problem
Electric Shock (2554)
|
Event Type
Injury
|
Event Description
|
It was reported the patient had turned the implantable neurostimulator (ins) off due to systemic shocking.The patient was having the ins replaced and during pre-operation impedances were measured to be slightly high, under 3000 ohms on electrodes 2, 3, and 4.Prior to this report all impedances were normal and no fractures were shown on x-ray.After replacing the ins, all impedances were normal.The patient status at the time of this report was alive with no injury.Follow up with the manufacturing representative indicated that there was no change in status and the device was still off.The patient had been referred to the healthcare professional that originally implanted and managed the ins.No outcome was reported regarding this event.Further follow-up is being conducted to obtain this information.If additional information is received, a follow up report will be sent.
|
|
Manufacturer Narrative
|
Concomitant products: product id 3387040, lot # v009174, implanted: (b)(6) 2007, product type lead; product id 7482a51, serial # (b)(4), implanted: (b)(6) 2007, product type extension; product id neu_unknown_prog, serial # unknown, product type programmer, physician.(b)(4).
|
|