• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO SOLETRA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO SOLETRA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 7426
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Impedance Problem (2950)
Patient Problem Electric Shock (2554)
Event Type  Injury  
Event Description
It was reported the patient had turned the implantable neurostimulator (ins) off due to systemic shocking.The patient was having the ins replaced and during pre-operation impedances were measured to be slightly high, under 3000 ohms on electrodes 2, 3, and 4.Prior to this report all impedances were normal and no fractures were shown on x-ray.After replacing the ins, all impedances were normal.The patient status at the time of this report was alive with no injury.Follow up with the manufacturing representative indicated that there was no change in status and the device was still off.The patient had been referred to the healthcare professional that originally implanted and managed the ins.No outcome was reported regarding this event.Further follow-up is being conducted to obtain this information.If additional information is received, a follow up report will be sent.
 
Manufacturer Narrative
Concomitant products: product id 3387040, lot # v009174, implanted: (b)(6) 2007, product type lead; product id 7482a51, serial # (b)(4), implanted: (b)(6) 2007, product type extension; product id neu_unknown_prog, serial # unknown, product type programmer, physician.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLETRA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4301001
MDR Text Key5054407
Report Number3004209178-2014-22672
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/05/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2012
Device Model Number7426
Device Catalogue Number7426
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2014
Date Device Manufactured09/28/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-