| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | HIRESOLUTION BIONIC EAR SYSTEM |
| Classification Name | implant, cochlear |
| Generic Name | cochlear implant (pediatric) |
| Applicant | ADVANCED BIONICS CORP. |
| PMA Number | P960058 |
| Supplement Number | S036 |
| Date Received | 04/22/2004 |
| Decision Date | 08/04/2004 |
| Product Code | |
| Advisory Committee |
Ear Nose & Throat |
| Supplement Type | normal 180 day track |
| Supplement Reason | labeling change - other |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for labeling changes to indicate that the hiresolution bionic ear system is safe for mri at levels of 0. 3 tesla/12 mhz and 1. 5 tesla/64 mhz. |
|
|