| Trade Name | ADAPTA, RELIA, SENSIA, & VERSA |
| Classification Name | pulse generator, permanent, implantable |
| Generic Name | implantable pulse generator |
| Applicant | MEDTRONIC INC. |
| PMA Number | P980035 |
| Supplement Number | S233 |
| Date Received | 06/29/2011 |
| Decision Date | 07/28/2011 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | 30-day notice |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Five changes to gen2 hybrid tester final test solution: 1) update to the force measure_ngt2 module; 2) update the pulse amplitude and pulse width test , pulse amplitude 7500 mv and amplitude 7500 mv test sequences; 3) widen the tolerance of the stability check criteria for the pulse amplitude and pulse width test ; 4) update the configuration settings for the pulse amplitude 7500 mv and unipolar bipolar switch test and the inter-chamber crosstalk test; and 5) update the electrically erasable programmable read only memory (eeprom) cyclic redundancy check (crc). |