| Trade Name | GALILEO III INTRAVASCULAR RADIOTHERAPY SYSTEM |
| Classification Name | intravascular radiation delivery system |
| Generic Name | percutaneous catheter, radionuclide brachytherapy source, remote controlled radionuclide applicator system |
| Applicant | GUIDANT CORP. |
| PMA Number | P000052 |
| Supplement Number | S014 |
| Date Received | 11/21/2002 |
| Decision Date | 12/17/2002 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Special Supplement |
| Supplement Reason | labeling change - instructions |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval to include a warning statement in the instructions for use for the galileo iii centering catheter advising the user not to exceed a pressure of 4 atm, as possible overexpansion of the inflation lumen could result. |