| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | CUTTING BALLOON |
| Classification Name | catheters, transluminal coronary angioplasty, percutaneous |
| Generic Name | catheter, transluminal coronary angioplasty, percutaneous |
| Regulation Number | 870.5100 |
| Applicant | BOSTON SCIENTIFIC /INTERNATIONAL TECHNOLOGIES |
| PMA Number | P950020 |
| Supplement Number | S005 |
| Date Received | 05/22/2002 |
| Decision Date | 07/17/2002 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - other |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the 6 mm cutting balloon. The device, as modified, will be marketed under the trade name cutting balloon monorail/otw (6 mm) and is indicated for dilatation of stenosis in coronary arteries for the purpose of improving myocardial perfusion in those circumstances where a high pressure balloon resistant lesion is encountered. In addition, the target lesion should possess the following characteristics: discrete (<= 15 mm in length) or tubular (10 to 20 mm in length); with a reference vessel diameter ranging from 2. 0 mm to 4. 0 mm; readily accessible to the device; light to moderate tortuosity of proximal vessel segment, non-angulated lesion segment (<45 degrees), smooth angiographic contour; and absence of angiographically-visible thrombus and/or calcification. |
|
|