| Trade Name | ANTENNA AND PORTABLE CHARGER |
| Classification Name | stimulator, spinal-cord, totally implanted for pain relief |
| Generic Name | implantable pulse generator system |
| Applicant | ST. JUDE MEDICAL |
| PMA Number | P010032 |
| Supplement Number | S054 |
| Date Received | 06/01/2012 |
| Decision Date | 07/12/2012 |
| Product Code | |
| Advisory Committee |
Neurology |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval to reverse the gender of the contacts of the plug and receptacle on the portable charger (models 3711 and 3715) i. E. , a receptacle for connecting the charging antenna will now have female pins for contacts, while the mating plug on the charging antenna will now have male contacts. Additionally, it was requested to change the pin-out connector from a 3 pin version to a 4 pin version. |