| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | MEDSTREAM PROGRAMMABLE INFUSION SYSTEM |
| Classification Name | pump, infusion, implanted, programmable |
| Generic Name | pump, infusion, implanted, programmable |
| Applicant | CODMAN & SHURTLEFF, INC. |
| PMA Number | P890055 |
| Supplement Number | S040 |
| Date Received | 05/14/2012 |
| Decision Date | 06/25/2012 |
| Product Code | |
| Advisory Committee |
General Hospital |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - software |
| Expedited Review Granted? | No |
| Combination Product |
Yes
|
Approval Order Statement Approval for software modifications to the programmable infusion system. These modifications were data integrity check improvement, resolution of control unit keyboard, frame check sequence algorithm upgrade, software anomalies 015, 016, 017, 020 fix, and improvement of the bridge bolus, priming bolus and single bolus processes. |
|
|