| Trade Name | INSYNC MARQUIS/SENTRY/MAXIMO/CONCERTO CRT-ICDS |
| Classification Name | defibrillator, automatic implantable cardioverter, with cardiac resynchronization (crt-d) |
| Generic Name | cardiac resynchronization therapy |
| Applicant | MEDTRONIC VASCULAR |
| PMA Number | P010031 |
| Supplement Number | S069 |
| Date Received | 04/24/2007 |
| Decision Date | 06/25/2007 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the l325-03 and l325-04 integrated circuit (ic) to minimize the potential for a gate-oxide breakdown within the l325-01, l325-02, and l281-04 ics. |