| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | ST. JUDE MEDICAL EPIC II/EPIC II + AND ATLAS II/ATLAS II + ICDS AND CRT-D SYSTEMS |
| Classification Name | pulse generator, permanent, implantable |
| Generic Name | transtelephonic follow-up/monitoring system |
| Applicant | ST. JUDE MEDICAL, INC. |
| PMA Number | P910023 |
| Supplement Number | S126 |
| Date Received | 11/21/2006 |
| Decision Date | 12/13/2006 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | design change - minor |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the following modifications to the icd/crt-d device hybrid: 1) addition of a new c64 micro controller integrated circuit (ic); 2) changes to the c46 vss supply capacitor from 3. 3 uf to 33 uf; and 3) changes to diodes d17, d19 and d23 in the protection circuit. |
|
|