| Trade Name | AFFINITY/INTEGRITY/VICTORY FAMILY OF PACEMAKERS |
| Classification Name | pulse generator, permanent, implantable |
| Generic Name | tiered therapy implantable defibrillator |
| Applicant | ST. JUDE MEDICAL, INC. |
| PMA Number | P880086 |
| Supplement Number | S156 |
| Date Received | 10/24/2007 |
| Decision Date | 12/12/2007 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - software |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for two corrective fixes to address a hardware issued that can leave the crt-d and icd systems listed above in a post-sense refractory state. For devices that are already implanted and/or already manufactured, changes will be made to the programmer software. During a patient follow-up visit, the programmer will interrogate the device and download firmware ram code to correct the anomalous condition if it occurs. For new devices, a modification has been made to the sio2 chip hardware to check the sense refractory state every two seconds and end it if appropriate. |