| Trade Name | BAUSCH & LOMB TECHNOLAS T217Z ZYOPTIX LASER SYSTEM |
| Classification Name | excimer laser system |
| Generic Name | excimer laser system |
| Applicant | TECHNOLAS GMBH PERFECT VISION |
| PMA Number | P990027 |
| Supplement Number | S010 |
| Date Received | 08/11/2008 |
| Decision Date | 05/18/2010 |
| Product Code | |
| Advisory Committee |
Ophthalmic |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - specifications |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the following changes: 1) removal of a redundant fluence test; 2) introduction of a new scanner system from gsi to the laser system: 3) extrapolation of the wavefront data in the diagnostic zywave software; 4) new zrs (zyoptix remote support) software to monitor the performance of the laser and diagnostic devices remotely via the internet; 5) modification of the zrs to include trulink, an enhancement to the zrs system; 6) introduction of a new op field illumination in the laser system; 7) change of the video chip in the ccd lenslet camera dmk4002-ir; and 8) change in the supplier of the temperature control unit in the laser system. |