| Trade Name | NUCLEUS 22 CHANNEL COCHLEAR IMPLANT SYSTEM |
| Classification Name | implant, cochlear |
| Generic Name | cochlear implant |
| Applicant | COCHLEAR CORP. |
| PMA Number | P840024 |
| Supplement Number | S059 |
| Date Received | 10/31/1995 |
| Decision Date | 12/11/1995 |
| Product Code | |
| Advisory Committee |
Ear Nose & Throat |
| Supplement Type | normal 180 day track |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval to make changes to the mini 22 cochlear implant integrated circuit (ic) test software (k15001) in order to add several test of improved quality assurance. One modification you have requested is to add a power supply leakage test to aid in ensuring that the bipolar tantalum power supply capacitor is not assembled in the reverse direction. The second modification you have requested is to implement an automatic evaluation of the data measured formt he integrated circuti prior to its final asembly in order to eliminate operator error. |