| Trade Name | EXPRESS2 MONORAIL AND OVER-THE-WIRE CORONARY STENT SYSTEMS |
| Classification Name | stent, coronary |
| Generic Name | intravascular stent with delivery system |
| Applicant | BOSTON SCIENTIFIC SCIMED, INC. |
| PMA Number | P020009 |
| Supplement Number | S015 |
| Date Received | 04/05/2004 |
| Decision Date | 05/05/2004 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Special Supplement |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval to implement the following: 1) the addition of an additional in-process inspection, 2) modification of a current in-process inspection, and 3) modification of a manufacturing process to address complaints related to failure of the delivery system balloon to properly deflate following stent deployment. |