| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | IDENTITY, VERITY, SUSTAIN, VICTORY, ZEPHYR, & ACCENT FAMILIES OF IPG'S |
| Classification Name | pulse generator, permanent, implantable |
| Generic Name | tiered therapy implantable defibrillator |
| Applicant | ST. JUDE MEDICAL, INC. |
| PMA Number | P880086 |
| Supplement Number | S216 |
| Date Received | 02/22/2012 |
| Decision Date | 04/13/2012 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for implementing palladium-iridium (pd-ir) alloy wire as an alternate to the existing platinum-iridium (pt-ir) alloy feedthru wire in st. Jude medical high and low voltage pulse generators. |
|
|