| Trade Name | ARCTIC FRONT CARDIAC CRYOABLATION CATHETER |
| Classification Name | catheter, percutaneous, cardiac ablation, for treatment of atrial fibrillation |
| Applicant | MEDTRONIC CRYOCATH LP |
| PMA Number | P100010 |
| Supplement Number | S015 |
| Date Received | 02/17/2012 |
| Decision Date | 04/12/2012 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | design change - minor |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for catheter changes to increase the uniformity of the distal balloon surface temperature (increased number of refrigerant ports on the injection tube coil, more distal position of the injection tube, increased refrigerant flow on the 28mm balloon), to add new visual marker on the catheter shaft and to implement additional minor changes to the design and labeling. |
|
|