| Trade Name | FACT(TM)AND ARC BALLOON CORNARY DILATATION CATHETERS |
| Classification Name | catheters, transluminal coronary angioplasty, percutaneous |
| Generic Name | ptca catheter with ultrasound |
| Regulation Number | 870.5100 |
| Applicant | CARDIOVASCULAR DYNAMICS, INC. |
| PMA Number | P910031 |
| Supplement Number | S011 |
| Date Received | 08/26/1996 |
| Decision Date | 04/09/1997 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | normal 180 day track |
| Supplement Reason | change design/components/specifications - other |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the following: 1) a change in the balloon design from a double layer of polyethylene with a polyethylene terephalate center region to a single layer of polyethylene terephalate; 2) a change in catheter shaft material from polyethylene to nylon; 3) a change in the internal shaft design from a two lumen side by side design to a two lumen coaxial design; 4) a change in the balloon/shaft bonding method from heat sealing to a urethane adhesive; 5) a change in the balloon lengths from 25 mm to 22 mm and from 20 mm to 18 mm; and 6) a change in the color of the catheter shaft from blue to teal. |