| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | SJM ICD'S PHOTON MICRO, EPIC, EPIC +, ATLAS, ATLAS + (ICDS) |
| Classification Name | pulse generator, permanent, implantable |
| Generic Name | transtelephonic follow-up/monitoring system |
| Applicant | ST. JUDE MEDICAL, INC. |
| PMA Number | P910023 |
| Supplement Number | S098 |
| Date Received | 02/02/2006 |
| Decision Date | 04/05/2006 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for minor modifications to the design of the high voltage capacitor, battery, feedthrus/headers and discrete components. Specifically, alteration of the battery pin configuration, modification of the feedthru wire tubing lengths and of the method used for applying insulation, change in the value of c36 capacitor from 10nf to 680pf, qualification of an alternate accelerometer substrate with integrated resistor capacitor, and the shifting of the anchor bridge to feedthru flange from 0. 17" to 0. 18". |
|
|