| Trade Name | SYNCHROMED II PROGRAMMABLE DRUG INFUSION SYSTEM |
| Classification Name | pump, infusion, implanted, programmable |
| Generic Name | implantable infusion pumps |
| Applicant | MEDTRONIC NEUROLOGICAL |
| PMA Number | P860004 |
| Supplement Number | S060 |
| Date Received | 03/24/2004 |
| Decision Date | 03/29/2004 |
| Product Code | |
| Advisory Committee |
General Hospital |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - material |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the following changes involving the motor coil for the synchromed ii implantable infusion pump: 1) change from a crimped to a heat-welded motor coil connector pin; and 2) change to a silicon varnish surface coating of the motor coil, the same coating as applied to the motor coil of the synchromed el implantable infusion pump. |