| Trade Name | MEDTRONIC SYNCHROMED INFUSION |
| Classification Name | pump, infusion, implanted, programmable |
| Generic Name | synchromed & isomed implantable infusion pumps |
| Applicant | MEDTRONIC NEUROLOGICAL |
| PMA Number | P860004 |
| Supplement Number | S081 |
| Date Received | 01/13/2006 |
| Decision Date | 03/22/2006 |
| Product Code | |
| Advisory Committee |
General Hospital |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
Yes
|
Approval Order Statement Approval for the following changes: 1) adding a sutureless pump connector to both models 8709 and 8731 catheters compatible with the synchromed and isomed implantable infusion pumps; 2) adding numbers (in centimeters) to the markings on the 8709 catheter body like those on the 8731 catheter; 3) adding a shorter (9. 3 cm) introducer needle to the model 8731 kit; 4) adding connector pins and strain relief sleeves to model 8731 catheter kit allowing the pump segment of the catheter to be trimmed; 5) create model 8578 as a revision kit for 8709 or 8709sc (sutureless connector) containing the sutureless connector segment and strain relief sleeves; 6) adding model 8598a for current 8598 kit (8731 spinal catheter segment revision kit) modified to contain the shorter (9. 3 cm) introducer needle; and 7) model 8596sc as the pump connector in the current 8596 kit (8731 pump catheter segment revision kit) was replaced with the sutureless pump connector. |