| Trade Name | CONSULTA CRT-D, CONCERTO II CRT-D AND MAXIMO II CRT-D |
| Classification Name | defibrillator, automatic implantable cardioverter, with cardiac resynchronization (crt-d) |
| Generic Name | cardiac resynchronization therapy |
| Applicant | MEDTRONIC INC. |
| PMA Number | P010031 |
| Supplement Number | S141 |
| Date Received | 01/16/2009 |
| Decision Date | 03/11/2009 |
| Product Code | |
| Advisory Committee |
Cardiovascular |
| Supplement Type | real-time process |
| Supplement Reason | change design/components/specifications - component |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Approval for the following minor design changes: 1) two (2) ic design changes ¿ l310 and l354 2) two (2) alternate components ¿ diode change (currently: xd423 and proposed: xd445) ¿ insulated-gate bipolar transistor (igbt) change (currently: xd405 and proposed: xd447) 3) printed wire board (pwb) changes to accommodate the alternate (xd445 diode and xd447 igbt) components and manufacturability improvements. 4) four (4) ics (l303, l310, l354, and l409) to be transferred to an alternative ic fabrication vendor. |