• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC | Inspections
 


New Search Back to Search Results
Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device.
 
Trade NameATTAIN PERFORMA
Classification Namedrug eluting permanent left ventricular (lv) pacemaker electrode
Applicant MEDTRONIC INC.
PMA NumberP080006
Supplement NumberS078
Date Received04/03/2015
Decision Date04/28/2015
Product Code
OJX[ Registered Establishments with OJX ]
Advisory Committee Cardiovascular
Supplement Type30-day notice
Supplement Reason process change: manufacturing
Expedited Review Granted? No
Combination Product Yes
Approval Order Statement 
The following manufacturing changes: 1) the addition of an alternative testing system for electrode components; 2) use of factoryworks release 7. 9 manufacturing software; 3) a reduction in the exposure time during ethylene oxide half-cycle testing; 4) a reduction in the frequency of environmental monitoring testing in specific manufacturing areas; 5) an update to the maximum sterilization load profile temperature; and 6) use of factoryworks release 7. 11 manufacturing software.
-
-