| Note: this medical device record is a supplement. The device description may have changed. Be sure to look at the original PMA to get an up-to-date view of this device. |
| |
| Trade Name | MEDTRONIC MINIMED CONTINUOUS GLUCOSE MONITORING SYSTEM |
| Classification Name | sensor, glucose, invasive |
| Applicant | MEDTRONIC INC. |
| PMA Number | P980022 |
| Supplement Number | S109 |
| Date Received | 02/08/2012 |
| Decision Date | 03/09/2012 |
| Product Code | |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-day notice |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
No
|
Approval Order Statement Implementation of a new cleaning system called isopropyl alcohol recirculating cleaning or iparc. The iparc system is used to clean the sensor base and needle assembly components of the sof-sensor glucose sensor (mmt-7002, mmt-7003). The sof-sensor is a component of the paradigm real-time system, paradigm real-time revel system, guardian real-time system, cgms ipro system and ipro2 professional continuous glucose monitoring (cgm) system. |
|
|