| Trade Name | SYNCHROMED II INFUSION SYSTEM |
| Classification Name | pump, infusion, implanted, programmable |
| Applicant | MEDTRONIC INC. |
| PMA Number | P860004 |
| Supplement Number | S138 |
| Date Received | 08/10/2010 |
| Decision Date | 03/03/2011 |
| Product Code | |
| Advisory Committee |
General Hospital |
| Supplement Type | 135 review track for 30-day notice |
| Supplement Reason | process change: manufacturing |
| Expedited Review Granted? | No |
| Combination Product |
Yes
|
Approval Order Statement Approval of the following manufacturing process changes related to the baseline test of the implantable synchromed ii infusion pump, model 8637: 1) automation of the transfer of baseline test data; 2) change the duty cycle and lower the specification limit of the dynamic current test; and 3) an increase to the time to measure the micro awake current. |